Cheat sheet

ARDMS SPI Cheat Sheet

Practice this exam

Perform Ultrasound Examinations

  • 23% of exam
  • Wave Basics Decibel Math Attenuation Impedance + Range Pulsed Parameters

Manage Ultrasound Transducers

  • 7% of exam
  • Construction Beam + Arrays Transducer Types Damping

Optimize Sonographic Images

  • 26% of exam
  • Resolution Gain + TGC Processing Display Modes Artifacts

Apply Doppler Concepts

  • 34% of exam
  • Doppler Equation CW vs PW Spectral + Color Hemodynamics Doppler Artifacts

Provide Clinical Safety & QA

  • 10% of exam
  • Bioeffects + ALARA QA Phantoms Stats + Safety ALARA

Quick Facts

  • Exam SPI
  • Body ARDMS / Inteleos
  • Questions ~110 MCQ
  • Time 2 hours
  • Passing Score 555 / 700 scaled
  • Fee $275
  • Outline V24.1 (task-based)
  • Biggest Domain Doppler, 34%
  • Prerequisite Physics course, grade C+
  • Validity Window 5 yrs with specialty

Decibel Doubling Rule PE

  • +3 dB doubles; -3 dB halves intensity
  • +3dB: ×2 -3dB: ÷2 +10dB: ×10 +20dB: ×100

Wave Basics & Speed PE

  • Period (T) T = 1/frequency
  • Frequency (f) f = 1/period
  • Propagation speed eq c = f × λ
  • Soft tissue speed c = 1540 m/s
  • Speed alt unit 1.54 mm/µs
  • λ shortcut (mm) λ = 1.54/f(MHz)
  • 5 MHz period 0.2 µs
  • 5 MHz wavelength 0.308 mm
  • Diagnostic freq range 2–15 MHz
  • Ultrasound threshold Above 20 kHz
  • Speed order (slow→fast) Air, fat, tissue, bone
  • Speed depends on Stiffness up, density down

Decibel (dB) Math PE

  • dB formula 10 × log(I2/I1)
  • +3 dB Doubles intensity
  • 3 dB Halves intensity
  • +10 dB ×10 intensity
  • 10 dB ÷10 intensity
  • +20 dB ×100 intensity
  • 6 dB Quarter of intensity
  • Amplitude → power Power ∝ amplitude²
  • 2× amplitude 4× power (quadruples)

Attenuation & Loss PE

  • Attenuation coefficient 0.5 dB/cm/MHz soft tissue
  • Total attenuation (dB) 0.5 × freq × depth
  • HVL / DOP Depth at −3 dB
  • Absorption Dominant loss, converts heat
  • Reflection Large smooth boundary echo
  • Scattering Small or rough surfaces
  • Attenuation rises with Higher frequency, greater depth
  • Frequency selection rule Highest freq that penetrates

Impedance, Refraction & Range PE

  • Acoustic impedance (Z) Z = ρ × c
  • Impedance units Rayls
  • IRC ((Z2−Z1)/(Z2+Z1))²
  • ITC 1 − IRC
  • Bigger Z mismatch Stronger reflected echo
  • Specular reflection Angle-dependent, large surface
  • Rayleigh scattering Angle-independent, small surfaces
  • Refraction requires Oblique angle + speed change
  • Snell's Law sinθt/sinθi = c2/c1
  • Range equation Depth = (c × t)/2
  • Range eq factor of 2 Round-trip travel time

Pulsed Ultrasound Parameters PE

  • PRP PRP = 1/PRF
  • Imaging PRF range 1–10 kHz
  • Round-trip per 1 cm ~13 microseconds
  • Pulse duration (PD) # cycles × period
  • Spatial pulse length # cycles × wavelength
  • Duty factor PD/PRP = PD × PRF
  • Imaging duty factor 0.1%–1%
  • CW duty factor 100%
  • PD & SPL control Fixed by transducer only
  • Bandwidth Range of freqs in pulse
  • Q-factor Frequency ÷ bandwidth
  • Imaging Q-factor Low, wide bandwidth
  • CW Q-factor High, narrow bandwidth
  • Damping effect Shortens pulse, lowers sensitivity

Transducer Construction TR

  • PZT crystal Active piezoelectric ceramic element
  • Curie point Polarization temperature threshold
  • Overheat past Curie Permanently destroys piezoelectric effect
  • Piezoelectric effect Reciprocal: voltage ↔ vibration
  • Resonant frequency c(PZT) ÷ (2 × thickness)
  • Thinner element Higher resonant frequency
  • Matching layer ¼ wavelength, steps impedance
  • Backing material Dampens pulse, shortens SPL

Beam Geometry & Arrays TR

  • Near zone length (NZL) D² / (4 × λ)
  • NZL alt form Radius² ÷ wavelength
  • Larger diameter/freq Lengthens near zone
  • Beam waist End of near zone
  • Focusing effect Narrows beam at set depth
  • Linear/curved array Sequential firing, rectangular/trapezoidal image
  • Phased array Full aperture, electronic steer
  • Annular array Ring elements, mechanical steer
  • 2D matrix array Rows + columns, real-time 3D/4D
  • CW pencil probe 2 crystals, no image
  • CW limitation Range ambiguity, no depth

Axial Resolution Synonyms (LARRD) OP

  • Longitudinal, Axial, Range, Radial, Depth resolution
  • L: Longitudinal A: Axial R: Range R: Radial D: Depth

Axial vs Lateral Resolution OP

Axial

  • = SPL / 2
  • Along beam axis
  • Set by pulse length

Lateral

  • = beam width
  • Across beam axis
  • Best at focus
  • Depth direction vs side-to-side

Artifact Cause Identification OP

  • Parallel bands, decreasing brightness → Reverberation ( 2 strong reflectors )
  • Duplicate structure past diaphragm → Mirror image
  • Anechoic band deep to mass → Acoustic shadowing
  • Bright band deep to cyst → Posterior enhancement
  • Thin shadow at cyst edge → Refraction artifact
  • Structure at wrong depth → Speed displacement ( Non-1540 m/s path )

Resolution Types & Formulas OP

  • Axial resolution SPL ÷ 2
  • Axial resolution value ~0.5 mm typical
  • Axial improves with Higher freq, more damping
  • LARRD Axial resolution synonyms mnemonic
  • Axial NOT affected by PRF, depth, gain, power
  • Lateral resolution = beam width
  • Lateral best at Transmit focal zone
  • LATA Lateral resolution synonyms mnemonic
  • Multiple focal zones Better lateral, lower frame rate
  • Elevational resolution Set by element height
  • 1.5D array benefit Electronic elevational focusing
  • Temporal resolution Controlled by frame rate
  • Frame rate decreases with Depth, width, lines, foci up
  • M-mode temporal res Highest, single scan line

Lateral Resolution Synonyms (LATA) OP

  • Lateral, Angular, Transverse, Azimuthal resolution
  • L: Lateral A: Angular T: Transverse A: Azimuthal

Gain vs Output Power OP

Gain

  • Post-echo amplification
  • No bioeffect change
  • Adjust freely

Output Power

  • Pre-echo, patient dose
  • Raises bioeffect risk
  • ALARA: minimize first
  • Receiver control vs transmitter control

Gain, TGC & Dynamic Range OP

  • Pulser Sets transmit pulse amplitude
  • Output power up Raises intensity + bioeffect risk
  • Apodization Tapers aperture, cuts side lobes
  • Overall gain Amplifies all depths equally
  • Gain vs bioeffects Gain does not raise risk
  • TGC / DGC Corrects attenuation by depth
  • Only patient-dose control Output power, not gain/TGC
  • Dynamic range dB, largest to smallest signal
  • Narrow dynamic range More contrast, fewer shades
  • Wide dynamic range Less contrast, more shades

Reverberation vs Mirror Image OP

Reverberation

  • Equally spaced bands
  • 2 parallel reflectors
  • Decreasing brightness

Mirror Image

  • Single duplicate structure
  • 1 strong reflector
  • Beyond true structure
  • Repeating vs single duplicate

Signal Processing & Storage OP

  • Pre-processing Before storage, not adjustable later
  • Post-processing After storage, adjustable on freeze
  • Write zoom Rescans, truly improves resolution
  • Read zoom Magnifies pixels, no new data
  • Persistence Frame averaging, lowers temporal res
  • Gray shades formula 2^(bits per pixel)
  • Standard bit depth 8 bits = 256 shades
  • DICOM Medical image file standard
  • PACS Stores, distributes DICOM images

Shadowing vs Enhancement OP

Shadowing

  • Anechoic band below
  • Strong attenuator
  • TGC assumption too high

Enhancement

  • Bright band below
  • Weak attenuator
  • TGC assumption too low
  • High vs low attenuation

Display Modes & Harmonics OP

  • A-mode Amplitude spikes, ophthalmic use
  • B-mode Brightness dots, standard 2D
  • M-mode Position vs time tracing
  • M-mode use Rapid repetitive motion, valves
  • Harmonic frequency ~2× fundamental transmit frequency
  • Harmonics origin Generated in tissue, not interface
  • Harmonic imaging benefit Less clutter, better lateral res
  • Contrast harmonic imaging Exploits microbubble nonlinear signal
  • Spatial compounding Multi-angle average, less speckle
  • Frequency compounding Multi-band average, less speckle
  • Panoramic imaging Stitched sweep, not real-time
  • 3D acquisition Wobbling 1D or matrix array
  • 4D imaging Real-time 3D, live volumes
  • Contrast microbubbles IV gas bubbles, capillary-sized
  • Contrast + high MI Can rupture microbubbles

Propagation & Attenuation Artifacts OP

  • Reverberation Equal parallel bands, decreasing brightness
  • Mirror image Duplicate past strong reflector
  • Speed displacement Wrong depth, non-1540 path
  • Range ambiguity PRF too high for depth
  • Acoustic shadowing Anechoic band, strong attenuator
  • Posterior enhancement Bright band, weak attenuator
  • Refraction edge shadow Thin shadow at curved edge
  • Grating lobes Full-strength, element spacing ≥λ
  • Side lobes Lower-intensity, any transducer
  • Beam-width artifact Worst away from focal zone
  • Slice-thickness artifact Elevational plane partial volume
  • Beam-geometry artifact fix Technique change, not gain/TGC

Color Doppler Convention (BART) DO

  • Blue Away, Red Toward the transducer
  • B: Blue A: Away R: Red T: Toward

PW vs CW Doppler DO

PW

  • Sample gate depth
  • Has Nyquist limit
  • Can alias

CW

  • No depth gate
  • No Nyquist limit
  • Never aliases
  • Depth localization vs velocity range

PW vs CW Doppler Selection DO

  • Need high velocity, stenosis → CW Doppler ( No upper limit )
  • Need depth localization → PW Doppler ( Sample gate )
  • Suspect PW aliasing → Raise PRF, scale ( First fix )
  • Aliasing at max PRF → Switch to CW ( No Nyquist limit )
  • Flow mostly one direction → Shift baseline ( Reallocates display )
  • Need exact depth of jet → PW, not CW ( CW = range ambiguity )

Doppler Equation & Angle DO

  • Doppler shift equation fD = 2×ft×v×cosθ/c
  • fD definition fD = fr − ft
  • Flow toward transducer Positive Doppler shift
  • Flow away from transducer Negative Doppler shift
  • c in Doppler eq 1540 m/s, assumed constant
  • cos 0° = 1, max shift
  • cos 60° = 0.5, half shift
  • cos 90° = 0, no shift
  • Max angle rule 60° or less for velocity
  • Angle correction cursor Align to vessel/flow direction

Color vs Power Doppler DO

Color

  • Shows direction + velocity
  • Can alias
  • Mean frequency shift

Power

  • Shows presence only
  • Never aliases
  • Signal amplitude
  • Velocity map vs sensitivity map

Aliasing Fix Sequence DO

  • Aliasing on PW/color → Increase PRF, scale
  • Scale already maxed → Lower transmit frequency
  • Still aliasing → Increase Doppler angle
  • Flow one-directional → Shift baseline
  • Exceeds all PW limits → Switch to CW Doppler

CW, PW & Nyquist Limit DO

  • CW Doppler crystals 2 crystals, transmit + receive
  • CW aliasing Never aliases, no PRF
  • CW velocity range No upper limit
  • CW depth limitation Range ambiguity, no localization
  • PW Doppler crystal 1 crystal, sample gate
  • Nyquist limit PRF / 2
  • Aliasing trigger True shift exceeds PRF/2
  • Deeper sample volume Lower max PRF, more aliasing
  • Sample gate size Sensitivity vs depth accuracy
  • Fix aliasing (best) Raise PRF, scale
  • Fix aliasing (other) Shift baseline, lower freq
  • Fix aliasing (switch mode) Use CW Doppler

Spectral, Color & Power Doppler DO

  • Spectral display FFT of Doppler signal
  • FFT output Velocity vs time
  • Spectral broadening Fills window, turbulence or technical
  • Wall filter Removes low-freq wall motion
  • Wall filter too high Erases real slow flow
  • Velocity scale control = PRF setting
  • Color Doppler method Autocorrelation, mean shift
  • BART convention Toward = red, away = blue
  • Power Doppler Amplitude, not direction/velocity
  • Power Doppler aliasing Never aliases
  • Power Doppler strength More sensitive, low flow
  • Tissue Doppler (TDI) Low-velocity, high-amplitude tissue motion

Hemodynamics & Measurements DO

  • Laminar flow Orderly, narrow spectral window
  • Turbulent flow Disorganized, fills spectral window
  • Parabolic profile Fast center, zero at wall
  • Plug flow Uniform velocity, large vessels
  • Bernoulli equation ΔP = 4v²
  • Resistive Index (RI) (PSV − EDV) / PSV
  • Pulsatility Index (PI) (PSV − EDV) / mean
  • Velocity from shift v = fD×c/(2×ft×cosθ)

Doppler & Color Artifacts DO

  • Aliasing cause Shift exceeds Nyquist, PRF/2
  • Twinkling artifact Red/blue mosaic, renal calculus
  • Blooming artifact Color overflows vessel wall
  • Blooming cause Excess color gain/power
  • Flash artifact Motion overwhelms wall filter
  • Artifactual broadening Oversized sample gate, not disease

ALARA Priority Order SA

  • Gain first, minimize time, output power last
  • 1. Increase gain 2. Limit dwell time 3. Lower output power 4. Freeze when idle

TI vs MI SA

TI

  • Thermal risk index
  • TIS/TIB/TIC variants
  • Tied to power/dwell

MI

  • Cavitation risk index
  • Pr / √freq
  • FDA limit 1.9
  • Heat risk vs cavitation risk

Bioeffects, TI, MI & ALARA SA

  • Bioeffect mechanisms Thermal + mechanical, cavitation
  • Absorption up with Higher frequency, longer dwell
  • Bone absorption Highest, hot spot at interface
  • Inertial cavitation Violent collapse, free radicals
  • Thermal Index (TI) Power used ÷ 1°C-rise power
  • TI variants TIS, TIB, TIC
  • Mechanical Index (MI) Peak rarefaction pressure/√freq
  • FDA output limit ISPTA ≤ 720 mW/cm²
  • FDA ophthalmic limit ISPTA ≤ 50 mW/cm²
  • FDA MI limit MI ≤ 1.9
  • ALARA priority Gain over output power
  • ODS standard Real-time on-screen TI/MI

Sensitivity vs Specificity SA

Sensitivity

  • TP / (TP+FN)
  • True-positive detection
  • Rules out disease

Specificity

  • TN / (TN+FP)
  • True-negative detection
  • Rules in disease
  • SnNout vs SpPin

QA & Tissue Phantoms SA

  • Phantom speed 1540 m/s, matches assumption
  • Core QA tests DOP, axial/lateral res, dead zone
  • More core tests Distance accuracy, uniformity
  • Depth of penetration Max depth, weak echoes
  • Transducer integrity check Detects dead/dropped elements
  • Dropped element sign Dark stripe artifact
  • QA documentation Dated, retained every check

Stats, Infection & Patient Safety SA

  • Sensitivity TP / (TP + FN)
  • Specificity TN / (TN + FP)
  • PPV TP / (TP + FP)
  • NPV TN / (TN + FN)
  • Accuracy (TP+TN) / total
  • Sens/spec vs PPV/NPV Intrinsic vs prevalence-dependent
  • Spaulding: critical devices Require sterilization
  • Spaulding: semicritical High-level disinfection, TV/TR probes
  • Spaulding: noncritical Low-level disinfection
  • Patient ID check 2 identifiers before scanning
  • Contrast safety Screen contraindications, monitor reaction

Common Traps

Frequency vs Resolution

  • Higher freq = better res ≠ Higher freq = less penetration

Gain vs Output Power

  • Gain: no patient dose ≠ Output power: raises patient dose

Axial vs Lateral Resolution

  • Axial = pulse length ≠ Lateral = beam width

PW vs CW Aliasing

  • PW can alias ≠ CW never aliases

TI vs MI

  • TI = thermal risk ≠ MI = cavitation risk

Sensitivity vs Specificity

  • Sensitivity finds true positives ≠ Specificity finds true negatives

Legacy vs V24.1 Outline

  • Legacy: 5 topics, retired ≠ V24.1: 5 tasks, current

Reverberation vs Mirror Image

  • Reverberation: repeating bands ≠ Mirror: single duplicate

Write Zoom vs Read Zoom

  • Write zoom: adds resolution ≠ Read zoom: just magnifies

Last Minute

  • 1. Doppler 34%: biggest domain
  • 2. c = 1540 m/s in tissue
  • 3. Wavelength(mm) = 1.54 / f(MHz)
  • 4. Axial resolution = SPL / 2
  • 5. Nyquist limit = PRF / 2
  • 6. Doppler shift = 2ft·v·cosθ/c
  • 7. Attenuation = 0.5 dB/cm/MHz
  • 8. MI limit ≤ 1.9, FDA
  • 9. ISPTA limit ≤ 720 mW/cm²
  • 10. 60° = max Doppler angle
  • 11. Fix aliasing: raise PRF first
  • 12. ΔP = 4v², Bernoulli equation

ARDMS SPI Sonography Principles and Instrumentation

  • Practice Questions 124 questions Study Guide 46 topics Cheat Sheet 211 facts Flashcards 50 cards Blogs 2 blogs Free CE Podcast