Cheat sheet
ARDMS SPI Cheat Sheet
Perform Ultrasound Examinations
- 23% of exam
- Wave Basics Decibel Math Attenuation Impedance + Range Pulsed Parameters
Manage Ultrasound Transducers
- 7% of exam
- Construction Beam + Arrays Transducer Types Damping
Optimize Sonographic Images
- 26% of exam
- Resolution Gain + TGC Processing Display Modes Artifacts
Apply Doppler Concepts
- 34% of exam
- Doppler Equation CW vs PW Spectral + Color Hemodynamics Doppler Artifacts
Provide Clinical Safety & QA
- 10% of exam
- Bioeffects + ALARA QA Phantoms Stats + Safety ALARA
Quick Facts
- Exam SPI
- Body ARDMS / Inteleos
- Questions ~110 MCQ
- Time 2 hours
- Passing Score 555 / 700 scaled
- Fee $275
- Outline V24.1 (task-based)
- Biggest Domain Doppler, 34%
- Prerequisite Physics course, grade C+
- Validity Window 5 yrs with specialty
Decibel Doubling Rule PE
- +3 dB doubles; -3 dB halves intensity
- +3dB: ×2 -3dB: ÷2 +10dB: ×10 +20dB: ×100
Wave Basics & Speed PE
- Period (T) T = 1/frequency
- Frequency (f) f = 1/period
- Propagation speed eq c = f × λ
- Soft tissue speed c = 1540 m/s
- Speed alt unit 1.54 mm/µs
- λ shortcut (mm) λ = 1.54/f(MHz)
- 5 MHz period 0.2 µs
- 5 MHz wavelength 0.308 mm
- Diagnostic freq range 2–15 MHz
- Ultrasound threshold Above 20 kHz
- Speed order (slow→fast) Air, fat, tissue, bone
- Speed depends on Stiffness up, density down
Decibel (dB) Math PE
- dB formula 10 × log(I2/I1)
- +3 dB Doubles intensity
- 3 dB Halves intensity
- +10 dB ×10 intensity
- 10 dB ÷10 intensity
- +20 dB ×100 intensity
- 6 dB Quarter of intensity
- Amplitude → power Power ∝ amplitude²
- 2× amplitude 4× power (quadruples)
Attenuation & Loss PE
- Attenuation coefficient 0.5 dB/cm/MHz soft tissue
- Total attenuation (dB) 0.5 × freq × depth
- HVL / DOP Depth at −3 dB
- Absorption Dominant loss, converts heat
- Reflection Large smooth boundary echo
- Scattering Small or rough surfaces
- Attenuation rises with Higher frequency, greater depth
- Frequency selection rule Highest freq that penetrates
Impedance, Refraction & Range PE
- Acoustic impedance (Z) Z = ρ × c
- Impedance units Rayls
- IRC ((Z2−Z1)/(Z2+Z1))²
- ITC 1 − IRC
- Bigger Z mismatch Stronger reflected echo
- Specular reflection Angle-dependent, large surface
- Rayleigh scattering Angle-independent, small surfaces
- Refraction requires Oblique angle + speed change
- Snell's Law sinθt/sinθi = c2/c1
- Range equation Depth = (c × t)/2
- Range eq factor of 2 Round-trip travel time
Pulsed Ultrasound Parameters PE
- PRP PRP = 1/PRF
- Imaging PRF range 1–10 kHz
- Round-trip per 1 cm ~13 microseconds
- Pulse duration (PD) # cycles × period
- Spatial pulse length # cycles × wavelength
- Duty factor PD/PRP = PD × PRF
- Imaging duty factor 0.1%–1%
- CW duty factor 100%
- PD & SPL control Fixed by transducer only
- Bandwidth Range of freqs in pulse
- Q-factor Frequency ÷ bandwidth
- Imaging Q-factor Low, wide bandwidth
- CW Q-factor High, narrow bandwidth
- Damping effect Shortens pulse, lowers sensitivity
Transducer Construction TR
- PZT crystal Active piezoelectric ceramic element
- Curie point Polarization temperature threshold
- Overheat past Curie Permanently destroys piezoelectric effect
- Piezoelectric effect Reciprocal: voltage ↔ vibration
- Resonant frequency c(PZT) ÷ (2 × thickness)
- Thinner element Higher resonant frequency
- Matching layer ¼ wavelength, steps impedance
- Backing material Dampens pulse, shortens SPL
Beam Geometry & Arrays TR
- Near zone length (NZL) D² / (4 × λ)
- NZL alt form Radius² ÷ wavelength
- Larger diameter/freq Lengthens near zone
- Beam waist End of near zone
- Focusing effect Narrows beam at set depth
- Linear/curved array Sequential firing, rectangular/trapezoidal image
- Phased array Full aperture, electronic steer
- Annular array Ring elements, mechanical steer
- 2D matrix array Rows + columns, real-time 3D/4D
- CW pencil probe 2 crystals, no image
- CW limitation Range ambiguity, no depth
Axial Resolution Synonyms (LARRD) OP
- Longitudinal, Axial, Range, Radial, Depth resolution
- L: Longitudinal A: Axial R: Range R: Radial D: Depth
Axial vs Lateral Resolution OP
Axial
- = SPL / 2
- Along beam axis
- Set by pulse length
Lateral
- = beam width
- Across beam axis
- Best at focus
- Depth direction vs side-to-side
Artifact Cause Identification OP
- Parallel bands, decreasing brightness → Reverberation ( 2 strong reflectors )
- Duplicate structure past diaphragm → Mirror image
- Anechoic band deep to mass → Acoustic shadowing
- Bright band deep to cyst → Posterior enhancement
- Thin shadow at cyst edge → Refraction artifact
- Structure at wrong depth → Speed displacement ( Non-1540 m/s path )
Resolution Types & Formulas OP
- Axial resolution SPL ÷ 2
- Axial resolution value ~0.5 mm typical
- Axial improves with Higher freq, more damping
- LARRD Axial resolution synonyms mnemonic
- Axial NOT affected by PRF, depth, gain, power
- Lateral resolution = beam width
- Lateral best at Transmit focal zone
- LATA Lateral resolution synonyms mnemonic
- Multiple focal zones Better lateral, lower frame rate
- Elevational resolution Set by element height
- 1.5D array benefit Electronic elevational focusing
- Temporal resolution Controlled by frame rate
- Frame rate decreases with Depth, width, lines, foci up
- M-mode temporal res Highest, single scan line
Lateral Resolution Synonyms (LATA) OP
- Lateral, Angular, Transverse, Azimuthal resolution
- L: Lateral A: Angular T: Transverse A: Azimuthal
Gain vs Output Power OP
Gain
- Post-echo amplification
- No bioeffect change
- Adjust freely
Output Power
- Pre-echo, patient dose
- Raises bioeffect risk
- ALARA: minimize first
- Receiver control vs transmitter control
Gain, TGC & Dynamic Range OP
- Pulser Sets transmit pulse amplitude
- Output power up Raises intensity + bioeffect risk
- Apodization Tapers aperture, cuts side lobes
- Overall gain Amplifies all depths equally
- Gain vs bioeffects Gain does not raise risk
- TGC / DGC Corrects attenuation by depth
- Only patient-dose control Output power, not gain/TGC
- Dynamic range dB, largest to smallest signal
- Narrow dynamic range More contrast, fewer shades
- Wide dynamic range Less contrast, more shades
Reverberation vs Mirror Image OP
Reverberation
- Equally spaced bands
- 2 parallel reflectors
- Decreasing brightness
Mirror Image
- Single duplicate structure
- 1 strong reflector
- Beyond true structure
- Repeating vs single duplicate
Signal Processing & Storage OP
- Pre-processing Before storage, not adjustable later
- Post-processing After storage, adjustable on freeze
- Write zoom Rescans, truly improves resolution
- Read zoom Magnifies pixels, no new data
- Persistence Frame averaging, lowers temporal res
- Gray shades formula 2^(bits per pixel)
- Standard bit depth 8 bits = 256 shades
- DICOM Medical image file standard
- PACS Stores, distributes DICOM images
Shadowing vs Enhancement OP
Shadowing
- Anechoic band below
- Strong attenuator
- TGC assumption too high
Enhancement
- Bright band below
- Weak attenuator
- TGC assumption too low
- High vs low attenuation
Display Modes & Harmonics OP
- A-mode Amplitude spikes, ophthalmic use
- B-mode Brightness dots, standard 2D
- M-mode Position vs time tracing
- M-mode use Rapid repetitive motion, valves
- Harmonic frequency ~2× fundamental transmit frequency
- Harmonics origin Generated in tissue, not interface
- Harmonic imaging benefit Less clutter, better lateral res
- Contrast harmonic imaging Exploits microbubble nonlinear signal
- Spatial compounding Multi-angle average, less speckle
- Frequency compounding Multi-band average, less speckle
- Panoramic imaging Stitched sweep, not real-time
- 3D acquisition Wobbling 1D or matrix array
- 4D imaging Real-time 3D, live volumes
- Contrast microbubbles IV gas bubbles, capillary-sized
- Contrast + high MI Can rupture microbubbles
Propagation & Attenuation Artifacts OP
- Reverberation Equal parallel bands, decreasing brightness
- Mirror image Duplicate past strong reflector
- Speed displacement Wrong depth, non-1540 path
- Range ambiguity PRF too high for depth
- Acoustic shadowing Anechoic band, strong attenuator
- Posterior enhancement Bright band, weak attenuator
- Refraction edge shadow Thin shadow at curved edge
- Grating lobes Full-strength, element spacing ≥λ
- Side lobes Lower-intensity, any transducer
- Beam-width artifact Worst away from focal zone
- Slice-thickness artifact Elevational plane partial volume
- Beam-geometry artifact fix Technique change, not gain/TGC
Color Doppler Convention (BART) DO
- Blue Away, Red Toward the transducer
- B: Blue A: Away R: Red T: Toward
PW vs CW Doppler DO
PW
- Sample gate depth
- Has Nyquist limit
- Can alias
CW
- No depth gate
- No Nyquist limit
- Never aliases
- Depth localization vs velocity range
PW vs CW Doppler Selection DO
- Need high velocity, stenosis → CW Doppler ( No upper limit )
- Need depth localization → PW Doppler ( Sample gate )
- Suspect PW aliasing → Raise PRF, scale ( First fix )
- Aliasing at max PRF → Switch to CW ( No Nyquist limit )
- Flow mostly one direction → Shift baseline ( Reallocates display )
- Need exact depth of jet → PW, not CW ( CW = range ambiguity )
Doppler Equation & Angle DO
- Doppler shift equation fD = 2×ft×v×cosθ/c
- fD definition fD = fr − ft
- Flow toward transducer Positive Doppler shift
- Flow away from transducer Negative Doppler shift
- c in Doppler eq 1540 m/s, assumed constant
- cos 0° = 1, max shift
- cos 60° = 0.5, half shift
- cos 90° = 0, no shift
- Max angle rule 60° or less for velocity
- Angle correction cursor Align to vessel/flow direction
Color vs Power Doppler DO
Color
- Shows direction + velocity
- Can alias
- Mean frequency shift
Power
- Shows presence only
- Never aliases
- Signal amplitude
- Velocity map vs sensitivity map
Aliasing Fix Sequence DO
- Aliasing on PW/color → Increase PRF, scale
- Scale already maxed → Lower transmit frequency
- Still aliasing → Increase Doppler angle
- Flow one-directional → Shift baseline
- Exceeds all PW limits → Switch to CW Doppler
CW, PW & Nyquist Limit DO
- CW Doppler crystals 2 crystals, transmit + receive
- CW aliasing Never aliases, no PRF
- CW velocity range No upper limit
- CW depth limitation Range ambiguity, no localization
- PW Doppler crystal 1 crystal, sample gate
- Nyquist limit PRF / 2
- Aliasing trigger True shift exceeds PRF/2
- Deeper sample volume Lower max PRF, more aliasing
- Sample gate size Sensitivity vs depth accuracy
- Fix aliasing (best) Raise PRF, scale
- Fix aliasing (other) Shift baseline, lower freq
- Fix aliasing (switch mode) Use CW Doppler
Spectral, Color & Power Doppler DO
- Spectral display FFT of Doppler signal
- FFT output Velocity vs time
- Spectral broadening Fills window, turbulence or technical
- Wall filter Removes low-freq wall motion
- Wall filter too high Erases real slow flow
- Velocity scale control = PRF setting
- Color Doppler method Autocorrelation, mean shift
- BART convention Toward = red, away = blue
- Power Doppler Amplitude, not direction/velocity
- Power Doppler aliasing Never aliases
- Power Doppler strength More sensitive, low flow
- Tissue Doppler (TDI) Low-velocity, high-amplitude tissue motion
Hemodynamics & Measurements DO
- Laminar flow Orderly, narrow spectral window
- Turbulent flow Disorganized, fills spectral window
- Parabolic profile Fast center, zero at wall
- Plug flow Uniform velocity, large vessels
- Bernoulli equation ΔP = 4v²
- Resistive Index (RI) (PSV − EDV) / PSV
- Pulsatility Index (PI) (PSV − EDV) / mean
- Velocity from shift v = fD×c/(2×ft×cosθ)
Doppler & Color Artifacts DO
- Aliasing cause Shift exceeds Nyquist, PRF/2
- Twinkling artifact Red/blue mosaic, renal calculus
- Blooming artifact Color overflows vessel wall
- Blooming cause Excess color gain/power
- Flash artifact Motion overwhelms wall filter
- Artifactual broadening Oversized sample gate, not disease
ALARA Priority Order SA
- Gain first, minimize time, output power last
- 1. Increase gain 2. Limit dwell time 3. Lower output power 4. Freeze when idle
TI vs MI SA
TI
- Thermal risk index
- TIS/TIB/TIC variants
- Tied to power/dwell
MI
- Cavitation risk index
- Pr / √freq
- FDA limit 1.9
- Heat risk vs cavitation risk
Bioeffects, TI, MI & ALARA SA
- Bioeffect mechanisms Thermal + mechanical, cavitation
- Absorption up with Higher frequency, longer dwell
- Bone absorption Highest, hot spot at interface
- Inertial cavitation Violent collapse, free radicals
- Thermal Index (TI) Power used ÷ 1°C-rise power
- TI variants TIS, TIB, TIC
- Mechanical Index (MI) Peak rarefaction pressure/√freq
- FDA output limit ISPTA ≤ 720 mW/cm²
- FDA ophthalmic limit ISPTA ≤ 50 mW/cm²
- FDA MI limit MI ≤ 1.9
- ALARA priority Gain over output power
- ODS standard Real-time on-screen TI/MI
Sensitivity vs Specificity SA
Sensitivity
- TP / (TP+FN)
- True-positive detection
- Rules out disease
Specificity
- TN / (TN+FP)
- True-negative detection
- Rules in disease
- SnNout vs SpPin
QA & Tissue Phantoms SA
- Phantom speed 1540 m/s, matches assumption
- Core QA tests DOP, axial/lateral res, dead zone
- More core tests Distance accuracy, uniformity
- Depth of penetration Max depth, weak echoes
- Transducer integrity check Detects dead/dropped elements
- Dropped element sign Dark stripe artifact
- QA documentation Dated, retained every check
Stats, Infection & Patient Safety SA
- Sensitivity TP / (TP + FN)
- Specificity TN / (TN + FP)
- PPV TP / (TP + FP)
- NPV TN / (TN + FN)
- Accuracy (TP+TN) / total
- Sens/spec vs PPV/NPV Intrinsic vs prevalence-dependent
- Spaulding: critical devices Require sterilization
- Spaulding: semicritical High-level disinfection, TV/TR probes
- Spaulding: noncritical Low-level disinfection
- Patient ID check 2 identifiers before scanning
- Contrast safety Screen contraindications, monitor reaction
Common Traps
Frequency vs Resolution
- Higher freq = better res ≠ Higher freq = less penetration
Gain vs Output Power
- Gain: no patient dose ≠ Output power: raises patient dose
Axial vs Lateral Resolution
- Axial = pulse length ≠ Lateral = beam width
PW vs CW Aliasing
- PW can alias ≠ CW never aliases
TI vs MI
- TI = thermal risk ≠ MI = cavitation risk
Sensitivity vs Specificity
- Sensitivity finds true positives ≠ Specificity finds true negatives
Legacy vs V24.1 Outline
- Legacy: 5 topics, retired ≠ V24.1: 5 tasks, current
Reverberation vs Mirror Image
- Reverberation: repeating bands ≠ Mirror: single duplicate
Write Zoom vs Read Zoom
- Write zoom: adds resolution ≠ Read zoom: just magnifies
Last Minute
- 1. Doppler 34%: biggest domain
- 2. c = 1540 m/s in tissue
- 3. Wavelength(mm) = 1.54 / f(MHz)
- 4. Axial resolution = SPL / 2
- 5. Nyquist limit = PRF / 2
- 6. Doppler shift = 2ft·v·cosθ/c
- 7. Attenuation = 0.5 dB/cm/MHz
- 8. MI limit ≤ 1.9, FDA
- 9. ISPTA limit ≤ 720 mW/cm²
- 10. 60° = max Doppler angle
- 11. Fix aliasing: raise PRF first
- 12. ΔP = 4v², Bernoulli equation
ARDMS SPI Sonography Principles and Instrumentation