ARRT CT Cheat Sheet
Patient Care
- 13% of exam
- Contrast Media Contrast Safety Reactions Injection Technique
Safety
- 13% of exam
- Radiation Physics Dose Metrics Dose Reduction Personnel Protection
Image Production
- 32% of exam
- CT Components Acquisition Parameters Reconstruction Artifacts Image Quality
Procedures
- 43% of exam
- Head & Spine Neck & Chest Abdomen & Pelvis Vascular & Interventional
Quick Facts
- Exam ARRT CT
- Pathway Postprimary
- Total Questions 195 (165 scored)
- Test Time 195 minutes
- Pass Score Scaled 75
- Pass Rate 74.7% (2025)
- Fee $225 ARRT pathway
- Content Spec Eff. Sep 1 2026
- Provider Pearson VUE
Contrast Reaction Ladder PA
- Mild -> Moderate -> Severe
- Mild: hives, itching Moderate: bronchospasm, hypotension Severe: anaphylactoid, arrest
Bolus Tracking vs Test Bolus PA
Bolus tracking
- Monitors live ROI
- Auto-triggers diagnostic scan
- Used during real injection
Test bolus
- Small preview injection
- Measures time to peak
- Given before diagnostic bolus
- Live trigger vs preview
Contrast Protocol Picker PA
- Suspected pulmonary embolism → CTPA bolus tracking ( Pulmonary artery ROI )
- Suspected acute stroke → Noncontrast head CT ( Rule out bleed first )
- Suspected renal colic → Noncontrast stone protocol ( Avoid contrast mimicry )
- Characterize liver lesion → Triple-phase liver CT ( Arterial, portal, delayed )
- Suspected interstitial lung disease → Prone HRCT ( Fibrosis vs atelectasis )
- High lung-cancer risk → Low-dose screening CT ( Lung-RADS scoring )
Contrast Media Types PA
- Nonionic Standard low-osmolar contrast
- Ionic Higher osmolality, older type
- Barium sulfate GI tract contrast
- Water-soluble IV or oral iodinated
- Osmolarity Reaction-risk factor
- Air Colonography bowel distension
Contrast Safety Checks PA
- eGFR Renal function check
- Metformin Hold if renal impaired
- Prior reaction Triggers premedication protocol
- Pregnancy Risk-benefit review
- Lactation Feeding usually continues
- hCG Pregnancy screening lab
Contrast Reaction Levels PA
- Mild Hives, itching, nausea
- Moderate Bronchospasm, hypotension
- Severe Anaphylactoid, cardiac arrest
- Nephrotoxicity Delayed kidney complication
- Extravasation Contrast leaks into tissue
- Treatment Steroids, antihistamine, epinephrine
Injection & Bolus Technique PA
- Power injector Controlled flow rate
- Single-phase One constant flow rate
- Multiphase Variable flow rate
- Bolus tracking Live ROI auto-trigger
- Test bolus Small timing preview
- Scan delay Fixed post-injection wait
Dose Metric Order SA
- CTDIvol -> DLP -> SSDE
- CTDIvol: per location DLP: total exam dose SSDE: size-adjusted estimate
CTDIvol vs DLP SA
CTDIvol
- Dose at one location
- Measured in mGy
- Based on standard phantom
DLP
- Total dose across exam
- Measured in mGy·cm
- CTDIvol times scan length
- Point dose vs total dose
Dose Reduction Tool Picker SA
- Variable patient thickness → Tube current modulation ( Auto mA adjustment )
- Noisy low-dose images → Iterative reconstruction ( Reduces image noise )
- Need faster coverage → Pitch above one ( Lower relative dose )
- Cardiac motion control → ECG gating ( Prospective preferred )
- Track cumulative exam dose → CTDIvol and DLP ( Location and total )
- Adjust dose for size → SSDE calculation ( Uses patient diameter )
Radiation Physics Basics SA
- Bremsstrahlung Braking radiation, most x-rays
- Characteristic Electron shell interaction
- Photoelectric Full photon absorption
- Compton Partial scatter interaction
- Inverse square law Intensity drops with distance
- Attenuation Depends on tissue density
Photon Interaction Pair SA
- Photoelectric absorbs, Compton scatters
- Photoelectric: full absorption Compton: partial scatter Both shape image contrast
CTDIvol vs SSDE SA
CTDIvol
- Uses standard phantom size
- Same for all patients
- Console-displayed value
SSDE
- Uses patient's actual size
- Adjusted per individual
- More accurate estimate
- Standard vs personalized dose
Dose Measurement Units SA
- CTDIvol Dose at scan location mGy
- DLP Total dose across exam mGy·cm
- SSDE Size-adjusted dose estimate
- Absorbed dose Measured in mGy
- Effective dose Measured in mSv
- SI units Primary CT dose units
Dose Reduction Techniques SA
- Tube current modulation Auto-adjusts mA
- Iterative reconstruction Lower noise, lower dose
- Pitch above 1 Faster, lower dose
- Collimation Matches beam to anatomy
- Shielding Lead apron protection
- Gating Times cardiac dose
HU Reference Points IM
- Air -1000 | Water 0 | Bone +1000
- Air: -1000 HU Water: 0 HU Fat: -50 to -100 Bone: +1000 HU+
kVp vs mAs IM
kVp
- Controls beam energy
- Affects penetration and contrast
- Raise for larger patients
mAs
- Controls photon quantity
- Mainly affects image noise
- Raise mainly reduces noise
- Energy vs quantity
Artifact Fix Picker IM
- Streaks between dense bones → Beam-hardening correction ( Or raise kVp )
- Metal implant present → MAR algorithm ( Metal artifact reduction )
- Patient movement blur → Faster scan or gating ( Not a kVp fix )
- Circular pattern on image → Detector recalibration ( Equipment issue, not patient )
- Small lesion looks blurred → Thinner slice thickness ( Trades noise for detail )
CT Unit Components IM
- Gantry Houses tube and detectors
- Slip rings Enable continuous rotation
- DAS Digitizes detector signal
- Generator Powers the x-ray tube
- Detectors Convert photons to signal
Pitch Rule IM
- Below 1 overlap, above 1 spacing
- Below 1: higher dose At 1: contiguous data Above 1: faster, lower dose
Pitch Below vs Above 1 IM
Pitch below 1
- Overlapping helical data
- Higher patient dose
- Better spatial resolution
Pitch above 1
- Wider helical spacing
- Lower relative dose
- Faster table coverage
- Overlap vs speed
Acquisition Parameters IM
- kVp Beam energy, penetration
- mAs Photon quantity, noise
- Pitch Table feed per rotation
- Collimation Total beam width
- SFOV Scan field of view
- Slice thickness Z-axis resolution
Beam Hardening vs Metal Artifact IM
Beam hardening
- Caused by dense tissue
- Dark streaks or cupping
- Corrected with algorithms or kVp
Metal artifact
- Caused by metal implants
- Bright and dark streaking
- Corrected with MAR algorithms
- Tissue cause vs metal cause
Reconstruction & Postprocessing IM
- Filtered back projection Traditional reconstruction method
- Iterative Reduces noise per dose
- Reconstruction interval Overlap spacing, no dose
- MPR Multiplanar reformatted images
- MIP Max intensity projection
- Isotropic voxels Equal x, y, z size
Motion vs Ring Artifact IM
Motion artifact
- Caused by patient movement
- Blur or streak pattern
- Fixed by speed or gating
Ring artifact
- Caused by detector fault
- Circular pattern on image
- Fixed by recalibration
- Patient cause vs equipment cause
Artifact Recognition IM
- Beam hardening Dark streaks, cupping
- Metal artifact Bright and dark streaking
- Motion Blur from patient movement
- Ring artifact Detector miscalibration circle
- Partial volume Averaged mixed-tissue voxel
Reconstruction Interval vs Slice Thickness IM
Reconstruction interval
- Spacing between displayed images
- Can overlap freely
- Adds no extra dose
Slice thickness
- Set at acquisition time
- Affects raw resolution
- Affects noise per slice
- Display spacing vs raw data
Image Quality Metrics IM
- Spatial resolution Fine structural detail
- Contrast resolution Density difference detection
- Temporal resolution Freezes motion in time
- Noise Grainy image appearance
- HU Hounsfield unit scale
- Window width Displayed contrast range
Prospective vs Retrospective Gating PR
Prospective gating
- Triggers at one phase
- Lower radiation dose
- Less arrhythmia tolerance
Retrospective gating
- Covers full cardiac cycle
- Higher radiation dose
- Allows functional data
- Triggered vs continuous
Procedure Protocol Picker PR
- Complex intra-articular fracture → Thin-slice CT with MPR ( Not first-line imaging )
- Suspected aortic dissection → CTA root to femorals ( Full coverage required )
- Peripheral arterial disease symptoms → Runoff CT angiography ( Bolus-timed table speed )
- MRI unsafe with hardware → CT myelography ( Intrathecal contrast injection )
- Deep bone or soft-tissue biopsy → CT-guided biopsy ( Real-time needle guidance )
Head, Spine & MSK PR
- Noncontrast head CT Rules out hemorrhage first
- Temporal bone CT Thin-slice bone kernel
- C-spine trauma Thin axial, reformats
- Postmyelogram Follows intrathecal injection
- Postarthrogram Follows joint contrast injection
CTA vs CTV PR
CTA
- Arterial-phase timing
- Evaluates arteries
- Earlier peak enhancement
CTV
- Venous-phase timing
- Evaluates veins
- Later peak enhancement
- Artery phase vs vein phase
Neck & Chest Procedures PR
- CTPA Bolus-tracked pulmonary timing
- HRCT Thin-slice, often prone
- Coronary calcium score Prospective gated, noncontrast
- Low-dose lung screening Lung-RADS classification
- Cardiac CTA ECG-gated coronary study
Abdomen & Pelvis Procedures PR
- Multiphase liver Arterial, portal, delayed
- Renal stone protocol Noncontrast, low-dose
- Urogram Excretory-phase kidney timing
- CT colonography Insufflation, two positions
- Appendicitis CT IV contrast standard
Vascular & Interventional PR
- Aortic CTA Root through femoral arteries
- Runoff CTA Bolus-timed leg coverage
- CT-guided biopsy Real-time needle path
- CT fluoroscopy Near real-time guidance
- Percutaneous drainage Catheter to fluid collection
Common Traps
CTDIvol vs DLP
- CTDIvol measures one location ≠ DLP measures total exam dose
Bolus Tracking vs Test Bolus
- Tracking triggers the live scan ≠ Test bolus previews timing only
Beam Hardening vs Metal Artifact
- Hardening comes from dense tissue ≠ Metal artifact comes from implants
Pitch Below vs Above One
- Below one raises patient dose ≠ Above one lowers patient dose
Prospective vs Retrospective Gating
- Prospective triggers a single phase ≠ Retrospective covers the full cycle
kVp vs mAs Effects
- kVp mainly changes penetration ≠ mAs mainly changes image noise
CTA vs CTV Timing
- CTA times the arterial phase ≠ CTV times the venous phase
Last Minute
- 1. 195 total = 165 scored + 30 pilot
- 2. Test time is 195 minutes
- 3. Pass score = scaled 75 of 99
- 4. Procedures = 43% of scored exam
- 5. Image Production = 32% of exam
- 6. Patient Care ties Safety at 13%
- 7. CTDIvol = dose per scan location
- 8. DLP = total dose for exam
- 9. Pitch above one lowers dose
- 10. Noncontrast head CT excludes hemorrhage first
- 11. Renal stone CT skips IV contrast
- 12. Three attempts allowed within three years
- 13. Exam fee is $225 standard
- 14. Supporting credential required before applying
ARRT Computed Tomography
- Practice Questions 200 questions Study Guide 48 topics Cheat Sheet 90 facts Flashcards 50 cards Videos 2 videos Blogs 3 blogs Free CE Podcast