ARRT Nuclear Medicine Technology (N) Cheat Sheet
Patient Care
- 12% of exam
- Informed Consent HIPAA Standard Precautions Allergic Reactions
Safety
- 12.5% of exam
- Radiation Protection Cardinals NRC Dose Limits Posting & Area Classifications Survey & Dosimetry Instruments
Image Production
- 16.5% of exam
- Collimator Types Gamma Camera QC PET/CT & SPECT Terms Health IT & Informatics
Procedures
- 59% of exam
- Radionuclide Table Cardiac Procedures Endocrine & Oncology GI & GU Procedures Other Imaging Generator & Radiopharmacy QC
Quick Facts
- Exam ARRT NMT (N)
- Credential Nuclear Medicine Technology
- Time 230 min (250 total)
- Pass Scaled score 75
- Fee $225
- Questions 230 (200 scored)
- Provider Pearson VUE
- Blueprint Jan 1 2022
Patient Care Essentials PA
- Informed consent Physician explains; tech witnesses
- HIPAA minimum necessary Access only what's needed
- Chain of infection 6 links; break one
- Standard precautions All patients; PPE by exposure
- Early allergy signs Urticaria, pruritus, flushing
- Metformin and FDG PET Hold ~48h before scan
Cardinal Principles SA
- Time, Distance, Shielding = ALARA
- Time: minimize duration Distance: inverse square law Shielding: match radiation type ALARA: as low as achievable
Radiation Area vs High Radiation Area SA
Radiation Area
- Over 5 mrem/hr
- At 30 cm
- Standard posting required
High Radiation Area
- Over 100 mrem/hr
- At 30 cm
- Stricter access control
- 20x higher threshold
Radiation Protection Cardinals SA
- Time Minimize exposure duration
- Distance Inverse square law falloff
- Shielding Match material to radiation
- ALARA As low as reasonably achievable
- Deterministic effects Threshold dose; severity increases
- Stochastic effects No threshold; only probability increases
NRC Dose Limits SA
- Public -> Occupational -> Eye -> Skin
- Public: 100 mrem/yr Occupational: 5 rem/yr Eye: 15 rem/yr Skin: 50 rem/yr
Photoelectric Effect vs Compton Scattering SA
Photoelectric
- Low-energy photon absorbed
- Ejects inner-shell electron
- No scatter produced
Compton
- Higher-energy photon interaction
- Ejects outer-shell electron
- Main source of scatter
- Absorption vs scatter
NRC Dose Limits (10 CFR 20) SA
- Occupational whole-body 5 rem (50 mSv)/year
- Public dose limit 100 mrem (1 mSv)/year
- Declared pregnant worker 500 mrem per gestation
- Lens of eye 15 rem (150 mSv)/year
- Skin extremities 50 rem (0.5 Sv)/year
- Minor worker (under 18) 10% of adult limits
GM Counter vs Ionization Chamber SA
GM Counter
- Highly sensitive
- For low-level contamination
- Saturates at high dose
Ion Chamber
- Less sensitive
- Accurate at high dose
- For area, spill surveys
- Low-level vs high dose-rate
Posting & Area Classifications SA
- Radiation area Over 5 mrem/hr at 30cm
- High radiation area Over 100 mrem/hr at 30cm
- Very high radiation area Over 500 rad/hr at 1m
- Restricted area Access controlled for protection
- Controlled area Occupational dose is managed
- Airborne radioactivity area Airborne concentration limits exceeded
Survey & Dosimetry Instruments SA
- GM counter Low-level contamination; saturates high-dose
- Ion chamber Higher dose rates; less sensitive
- Well counter In vitro samples; high sensitivity
- Thyroid probe In vivo uptake; collimated NaI
- Pocket dosimeter Real-time direct-reading personal dose
- TLD/OSL badge Cumulative personnel dose record
SPECT vs Planar IM
SPECT
- 3D tomographic imaging
- Multiple angle views
- Better contrast resolution
Planar
- 2D single view
- Faster acquisition time
- Whole-body screening
- Depth vs speed
Image Artifact Troubleshooting IM
- Cold or hot spot → Check PMT/crystal ( Recalibrate camera )
- Nonuniform flood, edge packing → Reset energy window ( Off-peak photopeak )
- Blurred edges, streaking → Repeat acquisition ( Patient motion )
- Star-pattern artifact → High-energy collimator ( Septal penetration )
- Ring artifact on SPECT → Update uniformity map ( Detector nonuniformity )
- Missing outer body (PET/CT) → Extend CT field ( Truncation artifact )
- Photopenic rim at edges → Re-register CT/PET ( Attenuation misalignment )
- Diffuse extra-osseous uptake → Check free pertechnetate ( Poor labeling )
Collimator Types IM
- Parallel-hole General purpose; no magnification
- Pinhole Magnifies small superficial organs
- Converging (fan-beam) Magnifies; smaller field size
- Diverging Minifies; wider field coverage
- LEHR Best resolution; longer scan time
- ME/HE collimators For I-131, higher-energy isotopes
Gamma Camera QC IM
- Uniformity (flood) Daily; extrinsic or intrinsic
- Center of rotation Weekly; SPECT cameras only
- Resolution/linearity Monthly; bar phantom test
- Multi-window registration Quarterly; dual-isotope alignment check
- Energy resolution Acceptance testing; annual verification
PET/CT & SPECT Terms IM
- Attenuation correction CT density corrects photon loss
- SUV Standardized uptake value; tumor metric
- Coincidence detection Two 511 keV photons simultaneously
- Iterative reconstruction OSEM; better noise than FBP
- Filtered back projection (FBP) Older method; streak artifacts
- CTAC artifact Metal, contrast, or motion misregistration
- Matrix size Pixel grid; affects resolution
Health IT & Informatics IM
- DICOM Transmits/stores medical images
- HL7 Exchanges orders and reports
- PACS Archives and displays images
- RIS Scheduling, orders, and reporting
Half-Life Order PR
- F-18 -> Tc-99m -> I-123 -> Mo-99
- F-18: 110 min Tc-99m: 6 hours I-123: 13 hours Mo-99: 66 hours
Constancy vs Accuracy PR
Constancy
- Daily check
- Cs-137 reference source
- Detects day-to-day drift
Accuracy
- Install + annual check
- NIST-traceable source
- Confirms true activity
- Drift check vs true value
Radiopharmaceutical Selection by Indication PR
- Thyroid uptake or Graves → I-123 ( Oral capsule )
- Thyroid ablation or therapy → I-131 ( High-dose therapy )
- Myocardial perfusion imaging → Tc-99m sestamibi/tetrofosmin ( Or Tl-201 )
- Bone metastases survey → Tc-99m MDP ( 3-phase for infection )
- Acute cholecystitis workup → Tc-99m mebrofenin (HIDA) ( Nonvis at 60 min )
- Parathyroid adenoma localization → Tc-99m sestamibi dual-phase ( Delayed washout )
- Neuroendocrine tumor imaging → Ga-68 DOTATATE ( Replaced OctreoScan )
- Pheochromocytoma or neuroblastoma → I-123 MIBG ( Block thyroid first )
- GFR calculation needed → Tc-99m DTPA ( Glomerular filtration )
- Poor renal function → Tc-99m MAG3 ( Tubular secretion )
Radionuclide Table PR
- Tc-99m 6 h, 140 keV IT decay; workhorse agent
- Mo-99 66 h, generator parent β⁻ decay to Tc-99m
- I-123 13 h, 159 keV EC decay; diagnostic thyroid
- I-131 8 d, 364 keV β⁻/γ; thyroid ablation
- F-18 110 min, 511 keV β+ decay; PET FDG
- Ga-67 78 h, 93/184/300 keV EC decay; infection, tumor
- In-111 2.8 d, 171/245 keV EC decay; tumor, infection
- Tl-201 73 h, 69-83/167 keV EC decay; cardiac perfusion
- Xe-133 5.24 d, 81 keV β⁻/γ; lung ventilation
- Lu-177 6.65 d, 208 keV β⁻ decay; PRRT, PSMA
- Y-90 64 h, pure beta No primary gamma photon
Dose Calibrator QC PR
- C-A-L-G: Constancy, Accuracy, Linearity, Geometry
- Constancy: daily check Accuracy: install, annual Linearity: install, quarterly Geometry: install only
Linearity vs Geometry PR
Linearity
- Across activity range
- Install + quarterly
- Catches nonlinear response
Geometry
- Different volumes, containers
- Install only
- Catches volume-dependent error
- Activity range vs container size
Pharmacologic Stress Agent Picker PR
- Standard vasodilator stress needed → Regadenoson ( Fixed dose )
- Reactive airway disease → Regadenoson preferred ( A2A selective )
- AV block, no pacemaker → Avoid vasodilators ( Use dobutamine )
- Recent caffeine or theophylline → Hold before vasodilator ( Antagonizes receptors )
- Cannot exercise, no contraindication → Adenosine or dipyridamole ( Alternative vasodilators )
- Vasodilator contraindicated → Dobutamine ( Inotrope, chronotrope )
Generator & Radiopharmacy QC PR
- Mo-99 breakthrough ≤0.15 µCi/mCi Tc-99m
- Al breakthrough ≤10 µg/mL eluate
- Stannous ion Reduces Tc for labeling
- Labeling efficiency Depends on pH, time, temp
- Radiochemical purity (RCP) TLC measures unbound activity
- Elution Saline extracts Tc-99m only
Thyroid Iodine Choice PR
- I-123 diagnoses; I-131 treats
- I-123: 13h imaging I-131: 8d therapy Low-iodine diet: before I-131 TSH stimulation required
Free vs Hydrolyzed-Reduced Tc PR
Free Pertechnetate
- Unbound Tc-99m
- Thyroid, stomach, salivary uptake
- Signals poor labeling
Hydrolyzed-Reduced Tc
- Under-reduced by stannous ion
- Liver, spleen uptake
- Colloid-like behavior
- Both are labeling failures
Dose Calibrator QC PR
- Constancy Daily, Cs-137 source
- Accuracy Install + annual, NIST source
- Linearity Install + quarterly, activity range
- Geometry Install only, varied containers
- Attenuators (liners) Match isotope energy range
Stress vs Rest Imaging PR
Stress
- Pharmacologic or exercise
- Reveals reversible ischemia
- Usually acquired first
Rest
- Baseline perfusion state
- Confirms fixed defects
- Redistribution comparison point
- Compare for reversibility
Cardiac Procedures PR
- Reversible defect Stress only; ischemic, viable
- Fixed defect Stress and rest; scarred tissue
- MUGA ECG-gated; ejection fraction calc
- Sestamibi/tetrofosmin Mitochondrial uptake; no redistribution
- Vasodilator stress Regadenoson, adenosine, or dipyridamole
- Dobutamine Inotrope; used if contraindicated
MAG3 vs DTPA PR
MAG3
- Tubular secretion
- High extraction efficiency
- Preferred, poor function
DTPA
- Glomerular filtration
- Calculates GFR
- Normal renal function
- Secretion vs filtration
Endocrine & Oncology Procedures PR
- RAIU % uptake at set intervals
- Parathyroid scan Dual-phase sestamibi; adenoma persists
- Low-iodine diet Before I-131; maximizes uptake
- MIBG Norepinephrine analog; neuroendocrine tumors
- Ga-68 DOTATATE Somatostatin receptor; replaces OctreoScan
- FDG prep Fast 4-6h; check glucose
In-111 Pentetreotide vs Ga-68 DOTATATE PR
In-111 Pentetreotide
- Older SPECT agent
- OctreoScan brand name
Ga-68 DOTATATE
- Newer PET agent
- Higher resolution imaging
- Largely replaced by PET
GI & GU Procedures PR
- Gastric emptying Solid meal; 4-hour protocol
- Meckel scan Pertechnetate; ectopic gastric mucosa
- HIDA scan No gallbladder by 60 min
- GI bleed scan Tagged RBCs; detects slow bleeds
- MAG3 Tubular secretion; poor renal function
- DTPA renal Glomerular filtration; calculates GFR
Other Imaging Procedures PR
- 3-phase bone scan Flow, blood pool, delayed
- Brain death study No flow; hollow skull sign
- MAA lung perfusion Microembolizes capillaries; reduce dose caution
- Xe-133 ventilation True gas; closed delivery system
- DTPA aerosol Nebulized; uneven obstructed deposition
- Sentinel node mapping Sulfur colloid; first-draining node
- Cisternography/shunt patency CSF flow; confirms patency, leaks
Common Traps
Constancy vs Accuracy
- Constancy checks daily drift ≠ Accuracy confirms true activity value
Radiation Area vs High Radiation Area
- 5 mrem/hr threshold ≠ 100 mrem/hr threshold
Free vs Bound Pertechnetate
- Free causes thyroid, stomach uptake ≠ Bound stays on kit ligand
Reversible vs Fixed Defect
- Reversible resolves at rest ≠ Fixed persists at rest too
I-123 vs I-131 Use
- I-123 is for diagnosis ≠ I-131 is for therapy
GM Counter vs Ion Chamber
- GM for low-level survey work ≠ Chamber for high dose rates
Deterministic vs Stochastic Effects
- Deterministic has a dose threshold ≠ Stochastic has no dose threshold
Last Minute
- 1. Tc-99m: 6h, 140 keV
- 2. F-18: 110 min, 511 keV
- 3. I-123 diagnoses; I-131 treats
- 4. Occupational 5 rem; public 100 mrem
- 5. Radiation Area = 5 mrem/hr
- 6. Medical event: 20% + threshold rem
- 7. Mo-99 breakthrough ≤0.15 µCi/mCi
- 8. Constancy daily; accuracy install + annual
- 9. MAG3 = poor function; DTPA = GFR
- 10. SPECT 180° arc for cardiac
- 11. Free pertechnetate = poor labeling
- 12. Y-90 = pure beta, no gamma
ARRT Nuclear Medicine
- Practice Questions 270 questions Study Guide 36 topics Cheat Sheet 99 facts Flashcards 50 cards Videos 1 video Blogs 1 blog Free CE Podcast