Cheat sheet
ARRT Radiography Cheat Sheet
Patient Care
- 16-17% of exam
- Vitals & Assessment Contrast Media Infection & Emergencies Contrast Picker
Safety
- 25% of exam
- Radiation Units NCRP Dose Limits Protection & ALARA Radiobiology & Fluoro
Image Production
- 25-26% of exam
- Exposure Factors Image Quality Digital Imaging QC & Artifacts Exposure Fix Picker
Procedures
- 33% of exam
- Chest & Thorax Spine & Skull Extremities & Pelvis Abdomen & Contrast Projection Picker
Quick Facts
- Exam ARRT Radiography
- Scored 200 (+30 pilot)
- Time 230 min
- Pass Scaled 75
- Fee $225
- Format Mostly MCQ (Pearson VUE)
- Domains 4 categories
- Blueprint Jan 2022
Barium vs Water-Soluble PA
Barium
- Intact GI tract
- Better coating
- Not if perforation
Water-soluble
- Suspected perforation
- Iodinated, absorbable
- Safe in peritoneum
- Intact vs perforation
Contrast Picker PA
- Suspected GI perforation → Water-soluble iodinated
- Impaired renal function → Check eGFR, hydrate
- Prior contrast reaction → Premedicate steroids
- Joint evaluation → Arthrography, iodinated
- Uterus and tubes → Hysterosalpingography
Vitals & Assessment PA
- Pulse 60-100 bpm adult
- Respirations 12-20 per minute
- SpO2 95-100% normal
- Hypoxemia Below 90% SpO2
- Blood pressure About 120/80 mmHg
- Nasal cannula 1-6 L/min low-flow
Contrast Media PA
- LOCM Nonionic, fewer reactions
- HOCM Ionic, more reactions
- Barium Intact GI tract only
- Water-soluble If perforation suspected
- eGFR/creatinine Screen before iodinated
- Shellfish myth Not a contraindication
Infection & Emergencies PA
- Hand hygiene Most effective control
- Airborne N95 + negative-pressure room
- Droplet Surgical mask, close range
- Contact Gown and gloves
- PPE order Gown, mask, goggles, gloves
- Epinephrine First for anaphylaxis
- Vasovagal Supine, legs elevated
Cardinal Rules SA
- Time - Distance - Shielding
- Time: less beam-on Distance: inverse square Shielding: lead barrier Distance is most effective
Gray vs Sievert SA
Gray (Gy)
- Absorbed dose
- Energy per kg
- Tissue dose
Sievert (Sv)
- Equivalent dose
- Weighted by risk
- Stochastic reporting
- Absorbed vs biological
Radiation Units SA
- Gray Absorbed dose Gy
- Sievert Equivalent/effective dose Sv
- Roentgen Exposure (traditional unit; SI = C/kg)
- Becquerel One disintegration per second
- Coulomb/kg SI exposure unit
- LET / RBE Biological damage factors
Dose Limits SA
- Whole 50, Lens 150, Skin 500
- 50 mSv: whole body 150 mSv: eye lens 500 mSv: skin/extremities 1 mSv: public
NCRP Dose Limits SA
- Occupational 50 mSv per year
- Cumulative 10 mSv times age
- Public 1 mSv per year
- Fetus 5 mSv over gestation
- Monthly fetus 0.5 mSv after declaration
- Lens of eye 150 mSv per year
- Skin/extremities 500 mSv per year
Protection & ALARA SA
- ALARA As Low As Reasonably Achievable
- Cardinal rules Time, distance, shielding
- Inverse square Double distance, quarter dose
- Filtration 2.5 mm Al above 70kVp
- Collimation Less scatter, lower dose
- Protective apron 0.25-0.5 mm Pb equivalent
- Distance Most effective protection
Radiobiology & Fluoro SA
- Stochastic No threshold, probability
- Deterministic Threshold, tissue reactions
- Bremsstrahlung Braking radiation
- Characteristic Shell-transition x-rays
- Photoelectric Absorption, more contrast
- Compton Scatter, patient dose
- Last-image hold Freezes frame, cuts dose
15% kVp Rule IM
- +15% kVp = double exposure
- +15% kVp: 2x exposure Halve mAs to hold Lower dose, lower contrast
mAs vs kVp IM
mAs
- Receptor exposure
- Density/brightness
- Directly proportional
kVp
- Beam penetration
- Contrast scale
- More scatter
- Exposure vs contrast
Exposure Fix Picker IM
- Quantum mottle noise → Increase mAs ( More photons )
- Low contrast, scatter fog → Add grid, collimate
- Need lower patient dose → +15% kVp, halve mAs
- Motion blur → Shorter exposure time
- Too much magnification → Raise SID, reduce OID
- Overexposed, high DI → Reduce mAs
Exposure Factors IM
- mAs Controls receptor exposure
- kVp Controls penetration/contrast
- 15% rule +15% kVp doubles exposure
- SID Affects exposure and magnification
- OID Increases magnification, air gap
- AEC Auto-terminates exposure
- Focal spot Small = better detail
EI vs DI IM
EI
- Receptor exposure
- Detector-specific
- Technique feedback
DI
- Deviation from target
- Zero is ideal
- +3 ≈ 2× target receptor exposure
- Amount vs deviation
Image Quality IM
- Contrast Density difference (kVp)
- Spatial resolution Recorded detail/sharpness
- Distortion Size or shape misrepresentation
- Magnification Enlarged by OID
- Quantum mottle Noise from low mAs
- SNR Signal-to-noise ratio
Window Width vs Level IM
Window width
- Displayed contrast
- Gray range shown
- Post-processing only
Window level
- Displayed brightness
- Midpoint value
- Post-processing only
- Contrast vs brightness
Digital Imaging IM
- EI Receptor exposure indicator
- DI Deviation from target (0 ideal)
- Dose creep DI +3 ≈ 2× receptor exposure
- DQE Efficiency of signal capture
- Direct DR a-Se converts to charge
- Indirect DR Scintillator makes light first
QC & Artifacts IM
- Window width Displayed contrast
- Window level Displayed brightness
- kVp accuracy Within about 5%
- Reproducibility Output varies under 5%
- Flat-field Corrects dead pixels
- Grid cutoff Primary-beam absorption loss
- Moire CR grid-line aliasing
Oblique Spine (posterior obliques) PR
- Posterior obliques: Cervical up, Lumbar down
- Cervical RPO/LPO foramina: farthest (up) Lumbar RPO/LPO zygapophyseal: nearest (down) Scotty dog = lumbar
Fluid vs Air Decubitus PR
Fluid (effusion)
- Affected side down
- Fluid layers dependent
- Horizontal beam
Air (pneumothorax)
- Affected side up
- Air rises nondependent
- Horizontal beam
- Fluid down, air up
Projection Picker PR
- Suspect pleural effusion → Lateral decub, side down ( Fluid layers dependent )
- Suspect pneumothorax → Lateral decub, side up ( Air rises nondependent )
- Free air, cannot stand → Left lateral decubitus ( Air over liver )
- Lung apices clear → AP lordotic (Lindblom)
- Scaphoid detail → Stecher, ulnar deviation
- Carpal tunnel → Gaynor-Hart
- Open glenohumeral joint → Grashey method
- Suspected hip fracture → Danelius-Miller lateral
Chest & Thorax PR
- PA chest SID 72 inches (180 cm)
- CR centering T7, inferior scapular angle
- Full inspiration 10 posterior ribs
- Fluid decub Affected side down
- Air decub Affected side up
- AP lordotic Apices free of clavicles
- SC joints PA or shallow oblique
Decubitus Rule PR
- Fluid down, Air up
- Effusion: affected side down Pneumothorax: affected side up
Frog-leg vs Danelius-Miller PR
Frog-leg (Cleaves)
- Routine lateral hip
- Requires abduction
- Not for fracture
Danelius-Miller
- Trauma lateral hip
- Horizontal beam
- No limb movement
- Routine vs trauma
Abdomen & Contrast PR
- KUB centering Iliac crests
- Upright abdomen 2 inches above crests
- Free air Left lateral decubitus
- Erect wait 5 minutes before exposure
- Esophagram/UGI RAO position
- Small bowel Images 15-30 minutes
- IVU centering Iliac crests
Spine & Skull PR
- Open-mouth Dens, C1-C2
- Cervical posterior oblique Foramina farthest (upside)
- Lumbar posterior oblique Zygapophyseal joints (Scotty dog)
- Lateral lumbar Centered at L4
- Towne 30 caudad, occipital bone
- Waters Maxillary sinuses, orbital floors
- Caldwell 15 caudad, frontal sinuses
Extremities & Pelvis PR
- PA oblique hand 45 external rotation
- Stecher Scaphoid, ulnar deviation
- Gaynor-Hart Carpal canal/tunnel
- Lateral elbow 90 degrees flexion
- Grashey Open glenohumeral joint
- AP mortise ankle 15-20 internal rotation
- Calcaneus axial 40 cephalad angle
- AP pelvis 15-20 internal rotation
Trauma / Mobile / OR PR
- Trauma C-spine Cross-table lateral first
- Must include C7-T1 visualized
- Trauma rule CR perpendicular, 2 projections 90
- Mobile chest Upright as tolerated
- C-arm Tube under table
- Markers In field, at exposure
Common Traps
Gray vs Sievert
- Gray = absorbed dose ≠ Sievert = equivalent dose
mAs vs kVp
- mAs = exposure ≠ kVp = contrast
Shellfish myth
- Shellfish allergy irrelevant ≠ Prior reaction matters
Barium danger
- Barium if intact ≠ Water-soluble if perforation
Decub direction
- Fluid: side down ≠ Air: side up
EI is not dose
- EI = receptor exposure ≠ Not patient dose
Distance beats shielding
- Distance most effective ≠ Shielding not first
Last Minute
- 1. Procedures 33%, biggest domain
- 2. Pass = scaled score 75
- 3. 200 scored, 30 pilot
- 4. 230 minutes total test time
- 5. PA chest = 72-inch SID
- 6. Distance beats time and shielding
- 7. Occupational limit 50 mSv yearly
- 8. mAs = exposure; kVp = contrast
- 9. Perforation = water-soluble contrast
- 10. Fluid decub down, air up
- 11. Epinephrine first for anaphylaxis
- 12. Gray = absorbed; Sievert = equivalent