Cheat sheet

CBSPD CSPDT Cheat Sheet

Practice this exam

Roles & Responsibilities

  • 13% of exam
  • Standards & Regulatory One-Way Flow Scope of Practice Recall & Records

Life Science

  • 9% of exam
  • Microbiology Bioburden Chain of Infection Spore Resistance

Decontamination & Disinfection

  • 19% of exam
  • Decontam PPE Cleaning Chemistries Spaulding Classification Disinfection Levels Cleaning Equipment

Preparation & Handling

  • 16% of exam
  • Packaging & Wrap Inspection & Assembly Peel Pouches Tray Weight

Sterilization

  • 18% of exam
  • Steam Sterilization Low-Temp Sterilization CI Types Biological Indicators Monitoring

Sterile Storage & Distribution

  • 11% of exam
  • Sterile Storage Event-Related Sterility Loaner Trays FIFO Rotation

Patient Care Equipment

  • 6% of exam
  • Powered Instruments Endoscopes Leak Test Warming Cabinets

Ethics

  • 8% of exam
  • Professional Conduct Documentation Follow IFU Report Defects

Quick Facts

  • Exam CSPDT
  • Body CBSPD
  • Questions 150 (125 scored)
  • Time 3 hours
  • Pass Scaled score 70
  • Fee $135
  • Valid 5 years
  • Provider PSI testing centers
  • Pass rate 48.1% (2025)

One-Way Flow RO

  • Decontam then Prep then Sterilize then Store
  • Dirty to clean Never backtrack Negative to positive pressure

Standards & Regulatory RO

  • ANSI/AAMI ST79 Steam sterilization
  • ANSI/AAMI ST91 Flexible endoscopes
  • ANSI/AAMI ST41/ST58 EO and low-temp
  • ANSI/AAMI ST108 Water quality
  • OSHA Worker safety, bloodborne
  • CDC Disinfection guideline
  • FDA Clears devices, sterilants
  • AORN Perioperative practices

Microbiology & Infection LI

  • Bioburden Microbial load present
  • Vegetative bacteria Easiest to kill
  • Spores Hardest to kill
  • Prions Extended special protocols
  • Chain of infection Source to susceptible host
  • SAL Sterility assurance 10⁻⁶
  • Pathogen Disease-causing microbe
  • Biofilm Shields attached microbes

Spaulding Levels DE

  • Critical sterilize, Semi HLD, Non low
  • Critical: sterilize Semi-critical: HLD Non-critical: low-level

Disinfection vs Sterilization DE

Disinfection

  • Reduces pathogens
  • May spare spores
  • HLD for semi-critical

Sterilization

  • Destroys all microbes
  • Includes spores
  • SAL 10⁻⁶
  • Most vs all microbes

Spaulding to Reprocessing DE

  • Enters sterile tissue → Sterilize ( Critical )
  • Enters vascular system → Sterilize ( Critical )
  • Surgical implant → Sterilize + BI ( Critical )
  • Touches mucous membrane → High-level disinfect ( Semi-critical )
  • Flexible endoscope → HLD or sterilize ( Semi-critical )
  • Contacts intact skin → Low-level disinfect ( Non-critical )
  • BP cuff, stethoscope → Intermediate/low ( Non-critical )
  • Heat-sensitive critical → Low-temp sterilize ( EO/H2O2 )

Decontam PPE & Area DE

  • Fluid-resistant gown Protects torso and arms
  • Gloves General-purpose utility gloves
  • Face shield/mask Splash protection
  • Hair + shoe covers Contain contaminants
  • Decontam temp 60-65°F (16-18°C)
  • Decontam humidity 30-60% RH
  • Air exchanges ≥10 per hour
  • Room pressure Negative pressure

Critical vs Semi-critical DE

Critical

  • Sterile tissue/blood
  • Must sterilize
  • Implants, instruments

Semi-critical

  • Mucous membranes
  • HLD minimum
  • Endoscopes, scopes
  • Sterilize vs HLD

Cleaning Chemistries DE

  • Enzymatic detergent Breaks down bioburden
  • Enzymes Protease, lipase, amylase
  • Neutral pH ~7, protects instruments
  • Keep instruments moist Prevent dried soil
  • Point-of-use care Pre-clean at OR
  • Utility water Wash and rinse
  • Critical water Final rinse
  • Rinse residue Removes detergent film

Cleaning vs Disinfection DE

Cleaning

  • Removes soil
  • Physical + detergent
  • Must come first

Disinfection

  • Kills microbes
  • Chemical or thermal
  • After cleaning
  • Remove vs kill

Cleaning Equipment DE

  • Manual cleaning Scrub below water line
  • Pre-clean first Before ultrasonic
  • Ultrasonic Cavitation removes fine soil
  • Cavitation Bubbles implode on soil
  • Washer-disinfector Automated wash + thermal
  • Thermal rinse ≈180°F (82°C)
  • Cart washer Case carts, containers
  • Lumen flushing Connect irrigation ports

Spaulding Classification DE

  • Critical Enters sterile tissue/vascular
  • Critical example Implants, surgical instruments
  • Critical process Sterilization required
  • Semi-critical Contacts mucous membranes
  • Semi-critical example Endoscopes, laryngoscopes
  • Semi-critical process High-level disinfection minimum
  • Non-critical Contacts intact skin
  • Non-critical process Low/intermediate disinfection

Disinfection Levels & Agents DE

  • High-level (HLD) Kills all but spores
  • OPA 0.55%, ~12 min
  • Glutaraldehyde ≥2.4%, ≥20 min
  • Peracetic acid Sporicidal liquid
  • Hydrogen peroxide 7.5% HLD
  • Intermediate-level Kills TB, not spores
  • Low-level Vegetative bacteria only
  • Alcohol 70-90%, not sporicidal

Packaging & Wrap PR

  • Sequential wrap Two layers, two steps
  • Simultaneous wrap Bonded double layer
  • Envelope fold Diagonal technique
  • Square fold Straight technique
  • Peel pouch Single instrument, paper-plastic
  • Pouch stacking Paper faces paper
  • Rigid container Reusable, filters/valves
  • Tamper seal Shows unopened

Inspection & Assembly PR

  • Inspect cleanliness Under magnification/light
  • Check function Sharpness, alignment
  • Box locks open Allow sterilant contact
  • Lubricate Instrument milk, water-soluble
  • Tray weight ≤25 lbs total
  • Count sheet Verify tray contents
  • Tip guards Protect sharp tips
  • Absorbent liner Wicks moisture

Steam Temps ST

  • Gravity 250, Prevac 270-275, BD 273
  • Gravity 121°C/250°F Prevac 132°C/270°F 135°C/275°F Bowie-Dick 134°C

CI vs BI ST

Chemical (CI)

  • Chemical color change
  • Confirms conditions
  • Instant read

Biological (BI)

  • Live spores
  • Proves lethality
  • Needs incubation
  • Conditions vs kill

Sterilization Method Picker ST

  • Heat/moisture tolerant → Steam ( Preferred )
  • Dense metal, no lumen → Gravity steam ( Longer cycle )
  • Porous or lumened load → Prevacuum steam ( Air removal )
  • Heat-sensitive, no lumen → H2O2 gas plasma ( Low-temp )
  • Long lumens → EO ( Aeration needed )
  • Oils or powders → Dry heat ( No moisture )
  • Immersible, heat-sensitive → Liquid chemical ( Peracetic acid )
  • Need fast turnover → IUSS ( Not for implants )

Steam Sterilization ST

  • Gravity displacement Steam pushes air down
  • Prevacuum Vacuum removes air first
  • SFPP Steam-flush pressure-pulse
  • Moist heat action Coagulates proteins
  • Wrapped gravity 121°C, 30 min
  • Prevacuum 132°C 4 minutes exposure
  • Prevacuum 135°C 3 minutes exposure
  • Drying phase Prevents wet packs

BI Organisms ST

  • Steam = Geo, EO = atrophaeus
  • Steam/H2O2/ozone: G. stearothermophilus EO/dry heat: B. atrophaeus

Gravity vs Prevacuum ST

Gravity

  • Steam displaces air
  • No vacuum
  • 121°C 30 min

Prevacuum

  • Vacuum removes air
  • Faster penetration
  • 132°C 4 min
  • Passive vs dynamic air

Indicator & Monitoring Picker ST

  • New/repaired sterilizer → 3 BI cycles ( Qualification )
  • Every prevacuum day → Bowie-Dick ( Empty chamber )
  • Confirm exposure outside → Type 1 CI ( Process )
  • Inside every package → Type 4-6 CI ( Internal )
  • Implant load → BI + Type 5 ( Quarantine result )
  • Routine load monitor → BI, daily preferred ( Weekly minimum )
  • Positive BI → Recall load ( Quarantine )
  • Wet pack found → Do not use ( Investigate )

Low-Temp Sterilization ST

  • Ethylene oxide (EO) Heat/moisture-sensitive items
  • EO parameters Gas, temp, humidity, time
  • EO aeration Removes toxic residual
  • H2O2 gas plasma ≈50°C, lumen limits
  • Vaporized H2O2 Oxidizes cell components
  • Ozone Oxygen + water
  • Liquid chemical Peracetic acid immersion
  • Dry heat Oils, powders, 160-190°C

EO vs H2O2 Plasma ST

Ethylene oxide

  • Penetrates long lumens
  • Toxic residual
  • Needs aeration

H2O2 plasma

  • No toxic residue
  • Lumen limits
  • Fast cycle
  • Lumens vs no residue

Sterilization Monitoring ST

  • Mechanical/physical Gauges, printouts, time/temp
  • Chemical indicator Confirms conditions met
  • Biological indicator Proves spore kill
  • Bowie-Dick Daily air-removal test
  • Every load External + internal CI
  • Implant load BI + rapid readout
  • Qualification 3 consecutive BI runs
  • Positive BI Recall and quarantine

Chemical Indicator Types ST

  • Type 1 Process/exposure indicator
  • Type 2 Bowie-Dick test
  • Type 3 Single-variable indicator
  • Type 4 Multi-variable indicator
  • Type 5 Integrating indicator
  • Type 6 Emulating indicator
  • Internal CI Inside every pack
  • Type 5/6 Track all parameters

Biological Indicators ST

  • Steam BI G. stearothermophilus
  • H2O2/plasma BI G. stearothermophilus
  • Ozone BI G. stearothermophilus
  • EO BI B. atrophaeus
  • Dry heat BI B. atrophaeus
  • Spore form Most resistant challenge
  • PCD/challenge pack Holds the BI
  • Control BI Confirms viable spores

Event vs Time Sterility DI

Event-related

  • Sterile until compromised
  • Integrity-based
  • Current standard

Time-related

  • Expires by date
  • Older practice
  • Not integrity-based
  • Integrity vs calendar

Sterile Storage Conditions DI

  • Storage temp ≤75°F (24°C)
  • Storage humidity ≤70% RH
  • Air exchanges ≥4 per hour
  • Room pressure Positive pressure
  • From floor 8-10 inches
  • From ceiling 18 inches
  • From outside wall 2 inches
  • Access control Limited traffic, closed

Distribution & Loaners DI

  • Event-related sterility Integrity, not date
  • FIFO First in, first out
  • Handle minimally Prevent package damage
  • Inspect before use No holes or moisture
  • Loaner trays Arrive early, decontaminate
  • Loaner IFU Vendor provides instructions
  • Case cart Transport to OR
  • Compromised package Considered unsterile

Patient Care Equipment EQ

  • Powered instruments Never immerse motors
  • Air/battery lines Wipe, do not soak
  • Endoscope channels Brush and flush all
  • Leak test Before flexible reprocessing
  • Warming cabinet Blankets ≤130°F
  • Suction tubing Single-use preferred

Ethics & Conduct ET

  • Never skip steps Patient safety first
  • Follow IFU Manufacturer instructions
  • Document loads Traceability and recall
  • Report defects Remove from service
  • Confidentiality Protect patient data
  • Speak up Stop unsafe processing

Common Traps

Indicator vs kill

  • CI confirms conditions ≠ BI proves lethality

Cleaning vs sterilizing

  • Cleaning removes soil ≠ Sterilizing kills spores

HLD vs sterilization

  • HLD spares some spores ≠ Sterilization kills all

Gravity vs prevac time

  • Gravity 30 min ≠ Prevac 4 min

Steam vs EO spore

  • Steam: G. stearothermophilus ≠ EO: B. atrophaeus

Event vs date sterility

  • Event-related integrity ≠ Not calendar expiry

Bowie-Dick purpose

  • Tests air removal ≠ Not sterility or kill

Last Minute

  • 1. Clean before disinfect or sterilize
  • 2. Critical items must be sterilized
  • 3. Semi-critical needs high-level disinfection
  • 4. Gravity: 121°C for 30 min
  • 5. Prevacuum: 132°C for 4 min
  • 6. Prevacuum: 135°C for 3 min
  • 7. Bowie-Dick tests air removal
  • 8. CI confirms; BI proves kill
  • 9. Steam BI: G. stearothermophilus
  • 10. EO BI: B. atrophaeus
  • 11. Implant loads: BI first
  • 12. Storage ≤75°F, ≤70% humidity
  • 13. Sterility is event-related
  • 14. Decontam: negative pressure, 10 ACH

CSPDT Certified Sterile Processing and Distribution Technician

  • Practice Questions 150 questions Study Guide 34 topics Cheat Sheet 140 facts Flashcards 50 cards Shorts 1 short Videos 1 video Blogs 1 blog Free CE Podcast