Cheat sheet
CBSPD CSPDT Cheat Sheet
Roles & Responsibilities
- 13% of exam
- Standards & Regulatory One-Way Flow Scope of Practice Recall & Records
Life Science
- 9% of exam
- Microbiology Bioburden Chain of Infection Spore Resistance
Decontamination & Disinfection
- 19% of exam
- Decontam PPE Cleaning Chemistries Spaulding Classification Disinfection Levels Cleaning Equipment
Preparation & Handling
- 16% of exam
- Packaging & Wrap Inspection & Assembly Peel Pouches Tray Weight
Sterilization
- 18% of exam
- Steam Sterilization Low-Temp Sterilization CI Types Biological Indicators Monitoring
Sterile Storage & Distribution
- 11% of exam
- Sterile Storage Event-Related Sterility Loaner Trays FIFO Rotation
Patient Care Equipment
- 6% of exam
- Powered Instruments Endoscopes Leak Test Warming Cabinets
Ethics
- 8% of exam
- Professional Conduct Documentation Follow IFU Report Defects
Quick Facts
- Exam CSPDT
- Body CBSPD
- Questions 150 (125 scored)
- Time 3 hours
- Pass Scaled score 70
- Fee $135
- Valid 5 years
- Provider PSI testing centers
- Pass rate 48.1% (2025)
One-Way Flow RO
- Decontam then Prep then Sterilize then Store
- Dirty to clean Never backtrack Negative to positive pressure
Standards & Regulatory RO
- ANSI/AAMI ST79 Steam sterilization
- ANSI/AAMI ST91 Flexible endoscopes
- ANSI/AAMI ST41/ST58 EO and low-temp
- ANSI/AAMI ST108 Water quality
- OSHA Worker safety, bloodborne
- CDC Disinfection guideline
- FDA Clears devices, sterilants
- AORN Perioperative practices
Microbiology & Infection LI
- Bioburden Microbial load present
- Vegetative bacteria Easiest to kill
- Spores Hardest to kill
- Prions Extended special protocols
- Chain of infection Source to susceptible host
- SAL Sterility assurance 10⁻⁶
- Pathogen Disease-causing microbe
- Biofilm Shields attached microbes
Spaulding Levels DE
- Critical sterilize, Semi HLD, Non low
- Critical: sterilize Semi-critical: HLD Non-critical: low-level
Disinfection vs Sterilization DE
Disinfection
- Reduces pathogens
- May spare spores
- HLD for semi-critical
Sterilization
- Destroys all microbes
- Includes spores
- SAL 10⁻⁶
- Most vs all microbes
Spaulding to Reprocessing DE
- Enters sterile tissue → Sterilize ( Critical )
- Enters vascular system → Sterilize ( Critical )
- Surgical implant → Sterilize + BI ( Critical )
- Touches mucous membrane → High-level disinfect ( Semi-critical )
- Flexible endoscope → HLD or sterilize ( Semi-critical )
- Contacts intact skin → Low-level disinfect ( Non-critical )
- BP cuff, stethoscope → Intermediate/low ( Non-critical )
- Heat-sensitive critical → Low-temp sterilize ( EO/H2O2 )
Decontam PPE & Area DE
- Fluid-resistant gown Protects torso and arms
- Gloves General-purpose utility gloves
- Face shield/mask Splash protection
- Hair + shoe covers Contain contaminants
- Decontam temp 60-65°F (16-18°C)
- Decontam humidity 30-60% RH
- Air exchanges ≥10 per hour
- Room pressure Negative pressure
Critical vs Semi-critical DE
Critical
- Sterile tissue/blood
- Must sterilize
- Implants, instruments
Semi-critical
- Mucous membranes
- HLD minimum
- Endoscopes, scopes
- Sterilize vs HLD
Cleaning Chemistries DE
- Enzymatic detergent Breaks down bioburden
- Enzymes Protease, lipase, amylase
- Neutral pH ~7, protects instruments
- Keep instruments moist Prevent dried soil
- Point-of-use care Pre-clean at OR
- Utility water Wash and rinse
- Critical water Final rinse
- Rinse residue Removes detergent film
Cleaning vs Disinfection DE
Cleaning
- Removes soil
- Physical + detergent
- Must come first
Disinfection
- Kills microbes
- Chemical or thermal
- After cleaning
- Remove vs kill
Cleaning Equipment DE
- Manual cleaning Scrub below water line
- Pre-clean first Before ultrasonic
- Ultrasonic Cavitation removes fine soil
- Cavitation Bubbles implode on soil
- Washer-disinfector Automated wash + thermal
- Thermal rinse ≈180°F (82°C)
- Cart washer Case carts, containers
- Lumen flushing Connect irrigation ports
Spaulding Classification DE
- Critical Enters sterile tissue/vascular
- Critical example Implants, surgical instruments
- Critical process Sterilization required
- Semi-critical Contacts mucous membranes
- Semi-critical example Endoscopes, laryngoscopes
- Semi-critical process High-level disinfection minimum
- Non-critical Contacts intact skin
- Non-critical process Low/intermediate disinfection
Disinfection Levels & Agents DE
- High-level (HLD) Kills all but spores
- OPA 0.55%, ~12 min
- Glutaraldehyde ≥2.4%, ≥20 min
- Peracetic acid Sporicidal liquid
- Hydrogen peroxide 7.5% HLD
- Intermediate-level Kills TB, not spores
- Low-level Vegetative bacteria only
- Alcohol 70-90%, not sporicidal
Packaging & Wrap PR
- Sequential wrap Two layers, two steps
- Simultaneous wrap Bonded double layer
- Envelope fold Diagonal technique
- Square fold Straight technique
- Peel pouch Single instrument, paper-plastic
- Pouch stacking Paper faces paper
- Rigid container Reusable, filters/valves
- Tamper seal Shows unopened
Inspection & Assembly PR
- Inspect cleanliness Under magnification/light
- Check function Sharpness, alignment
- Box locks open Allow sterilant contact
- Lubricate Instrument milk, water-soluble
- Tray weight ≤25 lbs total
- Count sheet Verify tray contents
- Tip guards Protect sharp tips
- Absorbent liner Wicks moisture
Steam Temps ST
- Gravity 250, Prevac 270-275, BD 273
- Gravity 121°C/250°F Prevac 132°C/270°F 135°C/275°F Bowie-Dick 134°C
CI vs BI ST
Chemical (CI)
- Chemical color change
- Confirms conditions
- Instant read
Biological (BI)
- Live spores
- Proves lethality
- Needs incubation
- Conditions vs kill
Sterilization Method Picker ST
- Heat/moisture tolerant → Steam ( Preferred )
- Dense metal, no lumen → Gravity steam ( Longer cycle )
- Porous or lumened load → Prevacuum steam ( Air removal )
- Heat-sensitive, no lumen → H2O2 gas plasma ( Low-temp )
- Long lumens → EO ( Aeration needed )
- Oils or powders → Dry heat ( No moisture )
- Immersible, heat-sensitive → Liquid chemical ( Peracetic acid )
- Need fast turnover → IUSS ( Not for implants )
Steam Sterilization ST
- Gravity displacement Steam pushes air down
- Prevacuum Vacuum removes air first
- SFPP Steam-flush pressure-pulse
- Moist heat action Coagulates proteins
- Wrapped gravity 121°C, 30 min
- Prevacuum 132°C 4 minutes exposure
- Prevacuum 135°C 3 minutes exposure
- Drying phase Prevents wet packs
BI Organisms ST
- Steam = Geo, EO = atrophaeus
- Steam/H2O2/ozone: G. stearothermophilus EO/dry heat: B. atrophaeus
Gravity vs Prevacuum ST
Gravity
- Steam displaces air
- No vacuum
- 121°C 30 min
Prevacuum
- Vacuum removes air
- Faster penetration
- 132°C 4 min
- Passive vs dynamic air
Indicator & Monitoring Picker ST
- New/repaired sterilizer → 3 BI cycles ( Qualification )
- Every prevacuum day → Bowie-Dick ( Empty chamber )
- Confirm exposure outside → Type 1 CI ( Process )
- Inside every package → Type 4-6 CI ( Internal )
- Implant load → BI + Type 5 ( Quarantine result )
- Routine load monitor → BI, daily preferred ( Weekly minimum )
- Positive BI → Recall load ( Quarantine )
- Wet pack found → Do not use ( Investigate )
Low-Temp Sterilization ST
- Ethylene oxide (EO) Heat/moisture-sensitive items
- EO parameters Gas, temp, humidity, time
- EO aeration Removes toxic residual
- H2O2 gas plasma ≈50°C, lumen limits
- Vaporized H2O2 Oxidizes cell components
- Ozone Oxygen + water
- Liquid chemical Peracetic acid immersion
- Dry heat Oils, powders, 160-190°C
EO vs H2O2 Plasma ST
Ethylene oxide
- Penetrates long lumens
- Toxic residual
- Needs aeration
H2O2 plasma
- No toxic residue
- Lumen limits
- Fast cycle
- Lumens vs no residue
Sterilization Monitoring ST
- Mechanical/physical Gauges, printouts, time/temp
- Chemical indicator Confirms conditions met
- Biological indicator Proves spore kill
- Bowie-Dick Daily air-removal test
- Every load External + internal CI
- Implant load BI + rapid readout
- Qualification 3 consecutive BI runs
- Positive BI Recall and quarantine
Chemical Indicator Types ST
- Type 1 Process/exposure indicator
- Type 2 Bowie-Dick test
- Type 3 Single-variable indicator
- Type 4 Multi-variable indicator
- Type 5 Integrating indicator
- Type 6 Emulating indicator
- Internal CI Inside every pack
- Type 5/6 Track all parameters
Biological Indicators ST
- Steam BI G. stearothermophilus
- H2O2/plasma BI G. stearothermophilus
- Ozone BI G. stearothermophilus
- EO BI B. atrophaeus
- Dry heat BI B. atrophaeus
- Spore form Most resistant challenge
- PCD/challenge pack Holds the BI
- Control BI Confirms viable spores
Event vs Time Sterility DI
Event-related
- Sterile until compromised
- Integrity-based
- Current standard
Time-related
- Expires by date
- Older practice
- Not integrity-based
- Integrity vs calendar
Sterile Storage Conditions DI
- Storage temp ≤75°F (24°C)
- Storage humidity ≤70% RH
- Air exchanges ≥4 per hour
- Room pressure Positive pressure
- From floor 8-10 inches
- From ceiling 18 inches
- From outside wall 2 inches
- Access control Limited traffic, closed
Distribution & Loaners DI
- Event-related sterility Integrity, not date
- FIFO First in, first out
- Handle minimally Prevent package damage
- Inspect before use No holes or moisture
- Loaner trays Arrive early, decontaminate
- Loaner IFU Vendor provides instructions
- Case cart Transport to OR
- Compromised package Considered unsterile
Patient Care Equipment EQ
- Powered instruments Never immerse motors
- Air/battery lines Wipe, do not soak
- Endoscope channels Brush and flush all
- Leak test Before flexible reprocessing
- Warming cabinet Blankets ≤130°F
- Suction tubing Single-use preferred
Ethics & Conduct ET
- Never skip steps Patient safety first
- Follow IFU Manufacturer instructions
- Document loads Traceability and recall
- Report defects Remove from service
- Confidentiality Protect patient data
- Speak up Stop unsafe processing
Common Traps
Indicator vs kill
- CI confirms conditions ≠ BI proves lethality
Cleaning vs sterilizing
- Cleaning removes soil ≠ Sterilizing kills spores
HLD vs sterilization
- HLD spares some spores ≠ Sterilization kills all
Gravity vs prevac time
- Gravity 30 min ≠ Prevac 4 min
Steam vs EO spore
- Steam: G. stearothermophilus ≠ EO: B. atrophaeus
Event vs date sterility
- Event-related integrity ≠ Not calendar expiry
Bowie-Dick purpose
- Tests air removal ≠ Not sterility or kill
Last Minute
- 1. Clean before disinfect or sterilize
- 2. Critical items must be sterilized
- 3. Semi-critical needs high-level disinfection
- 4. Gravity: 121°C for 30 min
- 5. Prevacuum: 132°C for 4 min
- 6. Prevacuum: 135°C for 3 min
- 7. Bowie-Dick tests air removal
- 8. CI confirms; BI proves kill
- 9. Steam BI: G. stearothermophilus
- 10. EO BI: B. atrophaeus
- 11. Implant loads: BI first
- 12. Storage ≤75°F, ≤70% humidity
- 13. Sterility is event-related
- 14. Decontam: negative pressure, 10 ACH
CSPDT Certified Sterile Processing and Distribution Technician