Study guide
5.1 Ohm’s Law, AC/DC Circuits, and Resistance
Registered Nerve Conduction Studies Technologist R.NCS.T.
Track your progress as you read
Finished sections get a check mark.
1 Exam Overview
1.1 Current Exam Facts 1.2 Eligibility, Documentation, and Application 1.3 Scoring, Recertification, and Test Day
2 Membrane Physiology and Action Potentials
2.1 Resting Membrane Potential and Cell Membrane Physiology 2.2 Action Potential Phases, All-or-None Response, and Refractory Periods 2.3 Saltatory Conduction, Myelin, and Nodes of Ranvier
3 Peripheral Nerve Structure and Cell Bodies
3.1 Fascicles and Nerve Connective Tissues 3.2 Peripheral Nerve Fiber Types 3.3 Motor and Sensory Cell-Body Locations
4 Muscle Physiology, Roots, and Plexuses
4.1 Skeletal Muscle Structure: Myofibrils, Sarcomeres, and Bands 4.2 Excitation-Contraction Coupling 4.3 Spinal Roots, Dorsal and Ventral Roots, and Rami 4.4 Brachial and Lumbosacral Plexus Organization
5 Electronics, Instrumentation, and Stimulation
5.1 Ohm’s Law, AC/DC Circuits, and Resistance 5.2 Amplifiers, Filters, Averaging, and Display Settings 5.3 Stimulators, Polarity, and Stimulation Pitfalls
6 Waveform Parameters, Conduction Velocity, and Pediatric NCS
6.1 Latency, Amplitude, Duration, Area, and Rise Time 6.2 Conduction Velocity Calculation and Normal Values 6.3 Pediatric Nerve Conduction Studies
7 Lesion Patterns and Wallerian Degeneration
7.1 Preganglionic versus Postganglionic Lesions 7.2 Focal Lesions, Conduction Block, Demyelination, and Temporal Dispersion 7.3 Axonal Loss, Wallerian Degeneration, and Nerve-Injury Grades
8 Late Responses and the Blink Reflex
8.1 F Waves: Origin, Technique, and Clinical Use 8.2 H Reflexes: Origin, Technique, and Clinical Use 8.3 A Waves and the Blink Reflex
9 NMJ Physiology and Repetitive Nerve Stimulation
9.1 Neuromuscular Junction Anatomy and Physiology 9.2 Repetitive Nerve Stimulation: Technical Setup 9.3 Post-Activation Facilitation, Exhaustion, and Pseudofacilitation
10 Neuromuscular Junction Disorders
10.1 Myasthenia Gravis (Postsynaptic) 10.2 Lambert-Eaton Myasthenic Syndrome 10.3 Botulism and Presynaptic Toxin Patterns
11 Upper-Extremity Nerve Conduction Protocols
11.1 Median Motor, Sensory, AIN, and Palmar Cutaneous Studies 11.2 Ulnar Motor, Sensory, Dorsal Ulnar Cutaneous, and Medial Antebrachial Studies 11.3 Radial Motor, Posterior Interosseous, and Superficial Radial Studies
12 Upper-Extremity Entrapments and Mononeuropathies
12.1 Median Neuropathy at the Wrist (Carpal Tunnel) 12.2 Ulnar Neuropathy at the Elbow and Guyon’s Canal 12.3 Radial Neuropathy (Saturday Night Palsy) and Related UE Mononeuropathies
13 Lower-Extremity Nerve Conduction Protocols
13.1 Common, Deep, and Superficial Fibular (Peroneal) Studies 13.2 Tibial Motor and Medial/Lateral Plantar Studies 13.3 Sural, Saphenous, and Lateral Femoral Cutaneous Studies
14 Lower-Extremity Neuropathies and Proximal Nerves
14.1 Fibular Neuropathy at the Fibular Head versus L5 Radiculopathy 14.2 Tarsal Tunnel, Tibial Neuropathy, and S1 Localization 14.3 Femoral, Sciatic, and Lateral Femoral Cutaneous Neuropathies
15 Cranial Nerves, Other Nerves, and Plexus Lesions
15.1 Cranial Nerves with Facial, Trigeminal, and Accessory Studies 15.2 Phrenic, Suprascapular, Long Thoracic, and Axillary Nerves 15.3 Plexus Lesions: Erb’s Palsy, Neurogenic TOS, and Parsonage-Turner
16 Root, Motor Neuron, Polyneuropathy, and Muscle Disorders
16.1 Radiculopathy, Root Avulsion, and Sensory Neuronopathy 16.2 Motor Neuron Disease and Multifocal Motor Neuropathy 16.3 Polyneuropathies, Mononeuropathy Multiplex, and Muscle Disorders
17 Temperature, Age, Height, and Martin-Gruber Anastomosis
17.1 Temperature Effects and Warming Methods 17.2 Age and Height Effects on Nerve Conduction 17.3 Martin-Gruber Anastomosis Patterns
18 Anomalous Innervation, Technical Factors, and Safety
18.1 Riche-Cannieu Anastomosis and Accessory Fibular Nerve 18.2 Technical Factors and Troubleshooting Abnormal Studies 18.3 Infection Control, Electrical Safety, HIPAA, and OSHA
5.1 Ohm’s Law, AC/DC Circuits, and Resistance
Key Takeaways Explain This Topic AI
Ohm’s law is V = I × R; 20 mA through 5 kΩ requires 100 V from the stimulator
High electrode-skin impedance increases noise and stimulus artifact and can prevent a constant-current stimulator from delivering the current shown on the panel
Alternating current (AC) line interference is 60 Hz in North America; direct current (DC) does not reverse polarity
Series resistances add; two equal parallel resistors present half the resistance of one branch
- Tissue injury tracks current through the body (I = V / R), not open-circuit voltage alone
- Last updated: September 2026
Nerve conduction study (NCS) instruments are voltmeters paired with current sources. The biologic signal you record is a small voltage . The pulse you deliver is a brief current . Domain II of the American Association of Electrodiagnostic Technologists (AAET) Registered Nerve Conduction Study Technologist (R.NCS.T.) outline — Basic Electronics, Instrumentation and Stimulation — is only about 5% of the exam by weight, but it decides whether every other domain’s waveforms are trustworthy. This ExamFinch section is independent teaching for Ohm’s law, alternating current (AC) versus direct current (DC), resistors, and electrode-skin impedance. It does not claim AAET review, partnership, or official status.
Ohm’s law: V = I × R
Ohm’s law states that voltage (V) equals current (I) times resistance (R) :
V = I × R
The same relationship answers three laboratory questions when you rearrange it:
- Quantity you need Formula Laboratory use
- Voltage V = I × R What voltage the stimulator must produce to push a chosen current through skin
- Current I = V / R What current actually flows if voltage is held fixed and resistance changes
- Resistance R = V / I What opposition the electrode-skin path is presenting
Units on the instrument:
Voltage in volts (V). Displayed waveforms use millivolts (mV) for compound muscle action potentials (CMAPs) and microvolts (µV) for sensory nerve action potentials (SNAPs). 1 millivolt = 1,000 microvolts.
Current in amperes (A). Stimulator panels use milliamperes (mA). 1 mA = 0.001 A.
Resistance in ohms (Ω). Electrode checks use kilohms (kΩ). 1 kΩ = 1,000 Ω.
A convenient unit identity: mA × kΩ = V . The milli- and kilo- prefixes cancel, so 20 milliamperes through 5 kilohms is 100 volts without converting to amperes first. Use the identity as a sanity check, not as a substitute for knowing V = I × R in base units.
Worked example with lab-relevant numbers
A constant-current stimulator is set to 20 mA (0.020 A). Electrode-skin resistance under the stimulating cathode is 5 kΩ (5,000 Ω).
V = I × R = 0.020 × 5,000 = 100 V
Unit trick: 20 mA × 5 kΩ = 100 V .
If the same 20 mA is requested but the skin is poorly prepared and resistance rises to 10 kΩ :
V = 0.020 × 10,000 = 200 V
High impedance is therefore not only a “noisy screen” problem. The stimulator must generate more voltage to keep current constant. That extra voltage drop at the skin-electrode interface enlarges stimulus artifact — the shock-related deflection that can overlap or bury the response — and increases noise pickup. If the stimulator reaches its compliance voltage limit and cannot supply the needed voltage, delivered current falls below the number on the panel. The nerve is then under-stimulated even though the intensity control still reads 20 mA.
A second pass with a smaller sensory-range current: 8 mA through 2 kΩ needs only 16 V . The same 8 mA through 20 kΩ needs 160 V . Cleaning oil, light abrasion where the protocol allows, fresh gel or paste, and fully seated electrodes drop R, so the same clinical current requires less voltage and leaves a cleaner baseline.
Why high impedance increases noise conceptually: thermal (Johnson) noise grows with resistance, and a high-impedance electrode is a better antenna for ambient electric fields. Why it increases artifact : the stimulator’s larger voltage pulse couples more easily into the recording leads, and a high-impedance recording contact converts more of that coupled current into a displayed voltage (again V = I × R at the recording interface).
Alternating current (AC) versus direct current (DC)
Direct current (DC) does not reverse polarity with time. A battery is the classroom example. Slow electrode polarization and some baseline offsets behave like DC problems: the trace wanders or sits off-center until the high-pass filter (Section 5.2) removes the slow drift.
Alternating current (AC) reverses polarity at a regular frequency. North American line power is 60 hertz (Hz) AC (50 Hz in many other regions). The patient, cable loops, and the technologist act as antennas. The interference is a thick, regular ripple — 60 cycles in one second, or one cycle about every 16.7 milliseconds . Differential amplification (Section 5.2) is the main defense. A 60 Hz notch filter is a last resort because it can ring and distort CMAPs and SNAPs.
NCS stimulators deliver brief rectangular pulses of current through tissue. That pulse is not wall AC injected into the nerve. Do not confuse “we stimulate electrically” with “we are applying 60 Hz line current.” The pulse has a polarity (cathode versus anode, Section 5.3) and a duration measured in tenths of a millisecond, not a 60 Hz sine wave.
Resistors in a simple DC circuit obey Ohm’s law with a single number. Skin and electrodes in NCS are closer to impedances : they oppose pulsed and AC current with resistance plus capacitance . Machines that offer an “impedance check” are reporting that opposition, usually in kilohms. For exam teaching, treat a high kilohm reading as high opposition at that contact, whether the stem says resistance or impedance.
Resistors and electrode-skin impedance
A resistor opposes current. Skin, electrode paste, hair, oil, and the metal-electrolyte interface all act as resistors. Electrode-skin impedance is the opposition at the recording or stimulating contact.
Low, matched impedances let the tiny biologic voltage reach the amplifier with less noise and less artifact.
High impedance increases thermal noise, increases pickup of ambient AC, and, at stimulating electrodes, increases the voltage needed for a given current.
Mismatched impedances between the active recording electrode and the reference convert common-mode 60 Hz into a differential voltage the amplifier cannot fully cancel. Two electrodes at 2 kΩ each behave better than one at 2 kΩ and one at 20 kΩ, even if the average looks “acceptable.”
Practical reductions: clean oil, gentle abrasion when the protocol allows, fresh conductive gel, full skin contact, quiet grouped leads, and an impedance check when the instrument provides one. Raising display gain does not fix high impedance; it only magnifies the noise.
Many laboratories teach a working target in the low kilohm range with closely matched contacts. Exact numeric cutoffs vary by machine and protocol; the exam-relevant idea is low and equal , not a single unpublished official threshold.
Series versus parallel at a basic level
Series: resistances add. Current is the same through each element; voltages add. Two 5 kΩ contacts stacked in series present 10 kΩ . Dried paste plus a loose clip is a series stack. If 10 mA flows through that 10 kΩ series path, the total voltage drop is 100 V, split across the elements.
Parallel: voltage across each branch is the same; currents add. Reciprocals add: 1/R_total = 1/R1 + 1/R2. Two equal 10 kΩ paths in parallel present 5 kΩ . A well-gelled larger contact can behave like parallel paths and lower net impedance.
Recording montages are not a textbook series circuit, but extra interfaces (hair, dried gel, corrosion on a clip) still stack like series resistance. Lowering those extras is how you lower R in Ohm’s law without changing the physiology you came to measure.
Quick numeric drill:
2 kΩ in series with 3 kΩ → 5 kΩ total.
4 kΩ in parallel with 4 kΩ → 2 kΩ total.
2 kΩ in parallel with 2 kΩ in parallel with 2 kΩ → 0.67 kΩ total (three equal branches).
Safety preview: current, not voltage, injures
Tissue injury and cardiac risk scale with current through the patient , not with the largest voltage printed on a power-supply sticker. Ohm’s law is the reason: I = V / R . Wet skin (low R) lets more current flow for the same voltage. A high open-circuit voltage that cannot drive current through a very high resistance is not the same hazard as a modest voltage that drives tens of milliamperes across the thorax.
Keep stimulators connected only when you intend to stimulate. Do not assume “it is only a few volts” if a current path exists through the patient. Macroshock, leakage current, isolated power, and grounding rules belong in the electrical-safety chapter; this section only plants the electronics fact those rules rest on: current injures . Read the later safety chapter before you treat this preview as complete.