Study guide
5.1 Tube Additives & Color Selection
AMCA PTC Phlebotomy Technician Exam
Track your progress as you read
Finished sections get a check mark.
1 Chapter 1: Introduction & Exam Overview
1.1 Current Exam Facts & Logistics 1.2 Exam Blueprint & Study Strategy
2 Chapter 2: Professionalism & Regulatory Compliance
2.1 Professional Conduct, Diversity & Communication 2.2 Scope of Practice & AMCA Code of Ethics 2.3 Patient Rights, Consent & Documentation Basics 2.4 Patient Privacy & HIPAA Compliance 2.5 CLIA, OSHA Oversight & Incident Reporting 2.6 Continuing Education & Recertification
3 Chapter 3: Infection Control & Safety
3.1 OSHA Bloodborne Pathogens Standard & Exposure Control 3.2 Hand Hygiene & Respiratory Etiquette 3.3 Personal Protective Equipment & Standard Precautions 3.4 Transmission-Based Isolation Precautions 3.5 Cleaning, Disinfection & Sharps Disposal 3.6 Emergency Care, First Aid & CPR Response 3.7 Workplace Safety: CDC, OSHA & NFPA
4 Chapter 4: Patient Identification, Assessment & Site Preparation
4.1 Requisition Review & Professional Introduction 4.2 Two-Identifier Rule & Misidentification Protocol 4.3 Collection Compliance & Patient History Assessment 4.4 Patient Positioning for Safe Collection 4.5 Venipuncture Site Selection & Anatomy 4.6 Aseptic Site Preparation 4.7 Tourniquet Use & CLSI Guidelines
5 Chapter 5: Equipment Selection & Maintenance
5.1 Tube Additives & Color Selection 5.2 Analytical Interference Factors 5.3 Antiseptic Agent Selection 5.4 Venipuncture Equipment Selection 5.5 Capillary & Non-Blood Collection Equipment 5.6 Equipment Quality Control & Inspection
6 Chapter 6: Collection of Blood & Non-Blood Specimens
6.1 Venipuncture Procedure & Needle Technique 6.2 CLSI Order of Draw 6.3 Prioritizing Timed, STAT & Fasting Specimens 6.4 Chain of Custody Collections 6.5 Capillary Puncture Methods 6.6 Collection Complications & Response 6.7 Special Patient Circumstances 6.8 Blood Culture Collection 6.9 Special Collections: GTT, TDM & Blood Bank 6.10 Point-of-Care Testing 6.11 Blood Volume & Tube Fill Requirements 6.12 Non-Blood Specimen Collection 6.13 Post-Puncture Patient Care 6.14 Recording Collection Details 6.15 Measuring & Recording Vital Signs
7 Chapter 7: Specimen Processing & Transportation
7.1 Specimen Labeling & Inversion 7.2 Specimen Quality Assessment & Rejection 7.3 Centrifugation & Aliquoting 7.4 Preparing Non-Blood Specimens for the Lab 7.5 Special Transport, Delivery & Corrective Action
5.1 Tube Additives & Color Selection
Key Takeaways Explain This Topic AI
Evacuated collection tubes are manufactured with pre-measured negative pressure designed to draw exact blood volumes for accurate blood-to-additive ratios.
Expired collection tubes must never be used because vacuum loss causes underfilled tubes, and additive breakdown compromises laboratory test accuracy.
Anticoagulants (EDTA, Sodium Citrate, Heparin, Potassium Oxalate) preserve whole blood or plasma, whereas clot activators accelerate serum separation.
Thixotropic gel separators (SST/PST) form an impenetrable physical barrier between blood cells and liquid serum/plasma upon centrifugation.
Pre-collection quality control mandates inspecting needle sterile seals, checking bevels for hooks or barbs, verifying tube expiration dates, and ensuring tourniquet hygiene.
Last updated: July 2026
5.1 Tube Additives & Color Selection
The integrity of laboratory diagnostic results depends entirely upon the quality of the blood specimen collected. Evacuated collection tubes are engineered to precise physical and chemical specifications. A phlebotomist must understand how vacuum dynamics work, how chemical additives interact with biological components, and how pre-collection equipment inspection prevents pre-analytical errors. This section outlines essential concepts tested on the AMCA Phlebotomy Technician Certification (PTC) exam.
Vacuum Tube Dynamics & Expiration Dates
Evacuated tubes are manufactured with a pre-measured negative pressure (vacuum) . When the posterior needle pierces the rubber stopper of an evacuated tube inserted into a holder, the pressure differential between the patient's vein (positive pressure) and the inside of the tube (negative pressure) automatically pulls blood into the tube.
The Critical Importance of Draw Volume
Each evacuated tube is calibrated to pull an exact volume of blood (e.g., 2.7 mL, 4.0 mL, 6.0 mL). The pre-measured vacuum is designed to stop drawing blood precisely when the proper fill volume is reached, ensuring the correct blood-to-additive ratio .
Underfilled Tubes ("Short Draws") : Occur when a tube loses vacuum or venipuncture is interrupted prematurely. Underfilling alters the chemical concentration of additives, leading to critical pre-analytical testing errors.
Overfilled Tubes : Extremely rare with evacuated systems, but can occur if blood is forcibly injected via a syringe without a proper transfer device.
Impact of Underfilled Collection Tubes
Tube Type & Additive Target Ratio / Concentration Result of Short Draw (Underfilling) Diagnostic Impact
Light Blue (Sodium Citrate) Strict 9:1 Blood-to-Anticoagulant ratio. Excess free sodium citrate binds excess calcium in testing reagents. Falsely prolonged PT/INR and aPTT coagulation test results.
Lavender / Pink (EDTA) 1.5 – 2.0 mg EDTA per mL of blood. Excess EDTA concentration causes red blood cells to shrink (crenation). Altered cell morphology, false low Hematocrit (HCT), incorrect RBC indices.
Green (Heparin) 12 – 30 USP units per mL of blood. Excess heparin alters plasma electrolyte concentration and enzyme activity. False elevations or depressions in STAT chemistry analytes.
Gray (Sodium Fluoride / Oxalate) 2 mg Fluoride / 3 mg Oxalate per mL. Excess oxalate causes hemolysis and altered cell membrane permeability. Invalid blood glucose and blood alcohol concentration values.
Tube Expiration Date Standards
Every collection tube box and individual tube label bears a manufacturer expiration date . The AMCA exam strictly enforces the rule that expired tubes must NEVER be used under any circumstances .
Consequences of Using Expired Tubes
Loss of Vacuum : Over time, tube stoppers slowly leak micro-amounts of air, resulting in diminished vacuum pressure and incomplete sample filling (short draws).
Chemical Additive Degradation : Liquid anticoagulants (such as sodium citrate) can evaporate through plastic walls, changing their concentration. Powdered additives (such as EDTA or heparin) can crystallize or degrade, losing their ability to prevent clotting.
Separator Gel Failure : Thixotropic gel in SST or PST tubes can lose stability over time, preventing complete barrier formation during centrifugation.