8.1 Capillary Puncture Including Infants
AMT Certified Medical Laboratory Assistant CMLA
Track your progress as you read
Finished sections get a check mark.
1 Introduction
1.1 Current Exam Facts 1.2 Blueprint Weights and Study Strategy
2 Chapter 2: Safety Terminology and Infection Control
2.1 Safety, Infection, and OSHA Terminology 2.2 Chain of Infection and Precautions 2.3 Isolation, Aseptic Technique, and Disease Prevention
3 Chapter 3: OSHA, HazCom, and Public-Health Reporting
3.1 OSHA Bloodborne Pathogens and Needlestick Safety 3.2 Hazard Communication, SDS, and GHS 3.3 Mandated Safety Rules and Public-Health Reporting
4 Chapter 4: Quality Assessment and Quality Control
4.1 Quality Control and Assessment Terminology 4.2 QC vs QA, Internal vs External QC, and Proficiency Testing 4.3 Daily QC, Temperature Logs, and Corrective Action 4.4 Instrument Maintenance and Calibration in CMLA Scope
5 Chapter 5: Waived-Test Integrity and Confirmatory Testing
5.1 Off-Label Waived Testing and High-Complexity Default 5.2 Following Manufacturer Instructions for Waived Tests 5.3 Confirmatory and Supplemental Testing
6 Chapter 6: Phlebotomy Foundations
6.1 Phlebotomy Terminology and Equipment 6.2 Serum, Plasma, Whole Blood, and Anticoagulation 6.3 Patient Identification and Container Identification
7 Chapter 7: Venipuncture, Order of Draw, and Specimen Integrity
7.1 Patient Preparation and Physiologic Effects 7.2 Vein Selection and Venipuncture Technique 7.3 Order of Draw and Tube Additives 7.4 Short Draws, Hemolysis, and Plasma-on-Cells Limits
8 Chapter 8: Capillary Collection, Processing, and Legal Specimens
8.1 Capillary Puncture Including Infants 8.2 Post-Care, Labeling, Transport, Processing, and Send-Outs 8.3 Specimens Used as Legal Evidence
9 Chapter 9: Non-Blood Collection and Preanalytical Errors
9.1 Urine Collection Types and Patient Instruction 9.2 Urine Suitability, Storage, Preservatives, and Rejection 9.3 Other Specimens and Time/Temperature Requirements 9.4 Shipping, Slide Prep, Preanalytical Errors, and Complications
10 Chapter 10: Clinical Chemistry
10.1 Clinical Chemistry Terminology and Test Families 10.2 Chemistry Instrumentation, Cleaning, and Maintenance 10.3 Point-of-Care and Waived Chemistry Testing 10.4 Toxicology Within CMLA Scope
11 Chapter 11: Hematology and Coagulation
11.1 Hematology Terminology, Blood Functions, and Cell Types 11.2 Hematology Instrumentation, QC, and Calibration 11.3 Coagulation: PT, PTT, INR, Heparin, and Warfarin 11.4 Point-of-Care and Waived Hematology Tests
12 Chapter 12: Immunology, Serology, and Immunohematology
12.1 Antigen-Antibody Principles 12.2 Blood Types, Rh, and Immunohematology Basics 12.3 Waived Immunology, Serology, and Immunohematology Procedures
13 Chapter 13: Microbiology
13.1 Microbiology Terminology and Specimen Appropriateness 13.2 Aseptic Processing, Storage, and Transport of Micro Specimens 13.3 Media, Cultures, and Procedure Matching 13.4 Stains, KOH, Wet Preps, Reagents, and Controls
14 Chapter 14: Urinalysis and Microscopy
14.1 Urinalysis Terminology 14.2 Waived Urinalysis Procedures and Specimen Handling 14.3 Microscopy Parts, Magnification, Wet Prep, and Cleaning
15 Chapter 15: Basic Anatomy and Physiology
15.1 Anatomy Relevant to Laboratory Testing 15.2 Physiology Relevant to Laboratory Testing
16 Chapter 16: Legal, Ethical, and HIPAA Foundations
16.1 Consent, ABN, Test Ordering, and PPM Restrictions 16.2 Patient's Bill of Rights, Refusal, and Medicolegal Risk 16.3 Professional Appearance, Conduct, and Social Media 16.4 HIPAA Confidentiality and Record Release
17 Chapter 17: Clerical Skills, Communication, and Critical Values
17.1 Reports, Abnormal Results, and Critical-Value Notification 17.2 HIT Data, Inventory, Expirations, and Billing/Coding 17.3 Patient Instruction, Communication, Telephone, and Time Management
8.1 Capillary Puncture Including Infants
Key Takeaways Explain This Topic AI
CLSI GP42, Collection of Capillary Blood Specimens, 7th edition (2020), is the capillary standard on AMT CMLA-REF-2025-1; PRE02 is venous collection, not skin puncture.
Infant heelsticks use the medial or lateral plantar surface of the heel; never the posterior curvature, the arch, or an infant's fingers (osteomyelitis of the calcaneus).
Depth is controlled by choosing the correct single-use safety lancet, not by pressing harder; typical infant heel devices are designed so the blade does not reach bone.
Wipe away the first drop, then fill microtainers in common CLSI/practice order: EDTA first for hematology to limit clots, then other additive tubes, then serum last.
Newborn screening cards need completely filled circles, horizontal air-drying, and no stacking while wet; capillary glucose may run higher than venous, and hemolysis risk is higher.
Last updated: August 2026
CMLA Work Area II lists capillary work as two competencies: perform capillary punctures (II.3.B.9) and perform capillary punctures on infants (II.3.B.10). The supporting standard on AMT CMLA-REF-2025-1 is CLSI GP42 , Collection of Capillary Blood Specimens , 7th edition (2020). GP42 is not PRE02 . PRE02 is venous collection; GP42 is skin puncture. Reciting the venous order of draw does not make a correct heelstick.
Quick Answer: Older children and adults: palmar distal 3rd or 4th finger. Infants: medial or lateral plantar heel—never the posterior curvature, never the arch, never an infant finger. Use a device-controlled safety lancet so you do not hit bone. Warm, wipe the first drop, fill EDTA microtainers first, and air-dry newborn screening cards without stacking them wet.
When capillary is the right collection
Capillary puncture (skin puncture, fingerstick, or heelstick) samples mixed arteriolar, venous, and interstitial blood from a dermal incision. Choose it when the method is validated on capillary blood, the volume is small, venous access is poor or restricted (neonates, burns, oncology, repeated glucose checks), or the order is a newborn screening (NBS) filter-paper card. Do not use capillary as a shortcut for tests that require venous or arterial blood: routine coagulation (PT/aPTT), blood cultures, most erythrocyte sedimentation rates, and chemistry panels that exceed microcollection volume.
Capillary blood is not "tiny venous blood." Glucose may run slightly higher on capillary than venous samples, especially after a meal. Hemolysis is more common because of squeezing and small-bore collection, so potassium, lactate dehydrogenase, and aspartate aminotransferase can read falsely high. Do not interchange a fingerstick result with a venous reference interval unless the method and the laboratory directory say you may.
Adult and pediatric finger sites
For patients with adequate finger tissue (classically children older than about one year who are walking, and adults), puncture the palmar surface of the distal segment of the third (middle) or fourth (ring) finger . Stay slightly off the pad center. Align the incision across (perpendicular to) the fingerprint ridges so blood forms a round drop instead of running along the grooves.
Do not use the thumb (pulse, thicker callus), the index finger (callus, used constantly), the fifth (pinky) finger (thin tissue, bone closer to skin), a swollen, cyanotic, scarred, burned, or infected finger, the same hole repeatedly, or the earlobe (not a CLSI-recommended diagnostic site).
Clean with 70% isopropyl alcohol and air-dry . Residual alcohol hemolyzes cells and stings. Povidone-iodine left on the skin can contaminate capillary samples and falsely elevate potassium, phosphorus, and uric acid—do not substitute it for alcohol on a routine fingerstick unless a legal-collection SOP says otherwise (Section 8.3).
Infant heel sites — the high-stakes map
Infants, especially neonates, do not have enough tissue on the fingers. Finger punctures in this group risk osteomyelitis (bone infection) and bone injury. The approved site is the medial or lateral plantar surface of the heel —the sides of the plantar heel, not the back and not the bottom center.
Never puncture the posterior curvature of the heel (the calcaneus is closest to the skin there), the arch of the foot (nerves, tendons, cartilage), fingers or toes of a neonate, previous punctures, bruises, or edematous skin. If flow is poor, warm again; do not "just try the back of the heel."
Imagine a heel-print. The safe zones are the outer and inner edges of that print, still on the plantar surface—not the Achilles-adjacent curve.
Depth, devices, and osteomyelitis
Depth is controlled by choosing the correct single-use safety lancet , not by pressing harder. Infant heel lancets make a shallow, device-limited incision so the blade does not reach the calcaneus. Typical teaching associated with GP42 keeps infant heel depth at or below 2.0 mm ; premature and low-birth-weight infants need still shallower, device-specific lancets. Adult finger devices are not interchangeable with neonatal heel devices.
Osteomyelitis of the calcaneus is the named complication of puncturing too deep or in the wrong place. Related injuries include cartilage damage and calcaneal osteochondritis. If blood does not flow, you do not "go deeper" with a longer blade. You rewarm, lower the heel below the heart, and, if needed, select a new approved site with a new device. Activate the safety feature; used lancets go in a sharps container. OSHA treats them as contaminated sharps.
Warming, first drop, and milking
Warm the site about 3–5 minutes with a commercial heel warmer or a warm (not scalding) cloth. The goal is increased arterial flow ( arterialization ), not a burn. Manufacturer warmers are built not to exceed a safe skin temperature (often about 42°C); a microwave-hot washcloth is not a substitute.
Hold the finger or heel dependent. Puncture, then wipe away the first drop . That drop is mixed with tissue fluid and leftover alcohol. Collect later drops into microcollection tubes (microtainers) , into capillary tubes when allowed, or onto filter paper. Some waived glucose meters specify a different first-drop rule in the instructions for use (IFU); unless that IFU says otherwise, wipe.
Do not milk the site with a pumping squeeze. Gentle, intermittent pressure is not wringing the finger. Excess squeezing hemolyzes red cells and dilutes the specimen with interstitial fluid—both of which change glucose, potassium, and cell counts.
Microtainer order of fill (common CLSI/practice)
Capillary additive order is not the venous order of draw. Platelets in capillary blood aggregate quickly, so hematology cannot wait while you fill a serum tube. Common CLSI GP42 / laboratory practice —label it as such on CMLA; the facility IFU still wins—is:
- Fill order Container Why it is there
- 1 (if ordered) Capillary blood gas Gases change with delay and air exposure
- 2 EDTA microtainer Hematology (CBC, smear) before clots and platelet clumps form
- 3 Other additive (heparin, oxalate/fluoride) Chemistry or glucose preservative after EDTA
- 4 Serum microtainer Clot last; clotting does not harm a serum tube
Mix additive microtainers by inversion immediately. Underfilling EDTA still changes morphology and counts. Do not scrape dried blood off the skin into a tube.
Newborn screening filter paper
State NBS programs use dried-blood-spot cards. Fill each printed circle completely so blood soaks through from one side. Do not layer extra drops on a partially dried circle, do not paint with a capillary tube unless the program IFU allows a named applicator, and do not contaminate circles with lotion, powder, or alcohol-wet gloves.
Air-dry the card horizontally at ambient temperature—commonly at least 3 hours —away from sunlight and heat. Do not stack wet cards ; stacking smears spots and can transfer blood between patients. Do not dry with a hair dryer or incubator. Once dry, sleeve the card per the state kit and ship as the program specifies. Timing (often after 24 hours of age, with repeat rules for early collection) is a state protocol , not a number you invent.
Capillary versus venous results
Glucose: capillary may be higher than venous, especially postprandial.
Hemolysis risk: higher with skin puncture and squeezing.
Hematocrit/hemoglobin: often slightly higher in capillary samples.
Potassium and some enzymes: falsely high if hemolyzed.
Volume: you will run out of blood; plan the tubes, EDTA first.
Exam traps
Infant fingers are not small adult fingers.
Posterior heel and arch are forbidden, not backup sites.
Venous order of draw is the wrong answer for microtainers.
The first drop is discarded, not the best drop.
Wet NBS cards are never stacked to save space.
When CMLA shows a heel or a microtainer rack, name the site, the device, the first tube, and the bone you are trying not to hit.