Study guide

11.4 Point-of-Care and Waived Hematology Tests

AMT Certified Medical Laboratory Assistant CMLA

Track your progress as you read

Finished sections get a check mark.

1 Introduction

1.1 Current Exam Facts 1.2 Blueprint Weights and Study Strategy

2 Chapter 2: Safety Terminology and Infection Control

2.1 Safety, Infection, and OSHA Terminology 2.2 Chain of Infection and Precautions 2.3 Isolation, Aseptic Technique, and Disease Prevention

3 Chapter 3: OSHA, HazCom, and Public-Health Reporting

3.1 OSHA Bloodborne Pathogens and Needlestick Safety 3.2 Hazard Communication, SDS, and GHS 3.3 Mandated Safety Rules and Public-Health Reporting

4 Chapter 4: Quality Assessment and Quality Control

4.1 Quality Control and Assessment Terminology 4.2 QC vs QA, Internal vs External QC, and Proficiency Testing 4.3 Daily QC, Temperature Logs, and Corrective Action 4.4 Instrument Maintenance and Calibration in CMLA Scope

5 Chapter 5: Waived-Test Integrity and Confirmatory Testing

5.1 Off-Label Waived Testing and High-Complexity Default 5.2 Following Manufacturer Instructions for Waived Tests 5.3 Confirmatory and Supplemental Testing

6 Chapter 6: Phlebotomy Foundations

6.1 Phlebotomy Terminology and Equipment 6.2 Serum, Plasma, Whole Blood, and Anticoagulation 6.3 Patient Identification and Container Identification

7 Chapter 7: Venipuncture, Order of Draw, and Specimen Integrity

7.1 Patient Preparation and Physiologic Effects 7.2 Vein Selection and Venipuncture Technique 7.3 Order of Draw and Tube Additives 7.4 Short Draws, Hemolysis, and Plasma-on-Cells Limits

8 Chapter 8: Capillary Collection, Processing, and Legal Specimens

8.1 Capillary Puncture Including Infants 8.2 Post-Care, Labeling, Transport, Processing, and Send-Outs 8.3 Specimens Used as Legal Evidence

9 Chapter 9: Non-Blood Collection and Preanalytical Errors

9.1 Urine Collection Types and Patient Instruction 9.2 Urine Suitability, Storage, Preservatives, and Rejection 9.3 Other Specimens and Time/Temperature Requirements 9.4 Shipping, Slide Prep, Preanalytical Errors, and Complications

10 Chapter 10: Clinical Chemistry

10.1 Clinical Chemistry Terminology and Test Families 10.2 Chemistry Instrumentation, Cleaning, and Maintenance 10.3 Point-of-Care and Waived Chemistry Testing 10.4 Toxicology Within CMLA Scope

11 Chapter 11: Hematology and Coagulation

11.1 Hematology Terminology, Blood Functions, and Cell Types 11.2 Hematology Instrumentation, QC, and Calibration 11.3 Coagulation: PT, PTT, INR, Heparin, and Warfarin 11.4 Point-of-Care and Waived Hematology Tests

12 Chapter 12: Immunology, Serology, and Immunohematology

12.1 Antigen-Antibody Principles 12.2 Blood Types, Rh, and Immunohematology Basics 12.3 Waived Immunology, Serology, and Immunohematology Procedures

13 Chapter 13: Microbiology

13.1 Microbiology Terminology and Specimen Appropriateness 13.2 Aseptic Processing, Storage, and Transport of Micro Specimens 13.3 Media, Cultures, and Procedure Matching 13.4 Stains, KOH, Wet Preps, Reagents, and Controls

14 Chapter 14: Urinalysis and Microscopy

14.1 Urinalysis Terminology 14.2 Waived Urinalysis Procedures and Specimen Handling 14.3 Microscopy Parts, Magnification, Wet Prep, and Cleaning

15 Chapter 15: Basic Anatomy and Physiology

15.1 Anatomy Relevant to Laboratory Testing 15.2 Physiology Relevant to Laboratory Testing

16 Chapter 16: Legal, Ethical, and HIPAA Foundations

16.1 Consent, ABN, Test Ordering, and PPM Restrictions 16.2 Patient's Bill of Rights, Refusal, and Medicolegal Risk 16.3 Professional Appearance, Conduct, and Social Media 16.4 HIPAA Confidentiality and Record Release

17 Chapter 17: Clerical Skills, Communication, and Critical Values

17.1 Reports, Abnormal Results, and Critical-Value Notification 17.2 HIT Data, Inventory, Expirations, and Billing/Coding 17.3 Patient Instruction, Communication, Telephone, and Time Management

11.4 Point-of-Care and Waived Hematology Tests

Key Takeaways Explain This Topic AI

CMLA performs FDA-waived/POC hematology: waived hemoglobin, assigned spun microhematocrit, waived ESR, and cartridge devices such as i-STAT-type systems only when that cartridge is waived and on the laboratory menu.

The heterophile mononucleosis (mono) test is immunology (Chapter 12), not this hematology section.

Clotted, underfilled, or overfilled EDTA invalidates many hematology results; recollect rather than reporting a printed number from a bad tube.

Lipemia and hemolysis can bias photometric hemoglobin; mix settled tubes before sampling, and escalate interferents per the IFU and SOP.

Off-label changes such as extra incubation or dilution of a waived hemoglobin or ESR kit default the method to high complexity, which a CMLA must not independently perform.

Last updated: August 2026

CMLA III.6.D is one competency stated as practice: perform point-of-care and waived hematological tests . Terminology, cell functions, instrument QC, and PT/PTT/INR pairing were the earlier lettered areas. This section is the bench: which hematology methods a Certified Medical Laboratory Assistant (CMLA) actually runs, which look-alikes belong in other chapters, and which specimen problems (clotted ethylenediaminetetraacetic acid (EDTA) , overfill/underfill, lipemia , hemolysis ) make a printed number unreportable.

Quick Answer: Run Food and Drug Administration (FDA)–waived hemoglobin, assigned spun microhematocrit , waived erythrocyte sedimentation rate (ESR) , and i-STAT-type cartridges only when that cartridge is waived and on the menu. Follow the instructions for use (IFU) . Reject clotted or misfilled EDTA. Watch lipemia and hemolysis on photometric hemoglobin. The mono test is not this section (Chapter 12). Do not modify a waived kit.

What counts as waived hematology on CMLA

Clinical Laboratory Improvement Amendments (CLIA) waived tests are simple methods with insignificant risk of an erroneous result when the IFU is followed ( 42 CFR 493.15 ). The original federal waived list includes spun microhematocrit , hemoglobin by copper sulfate , and nonautomated ESR , among other non-hematology tests. FDA has additionally waived specified hemoglobin photometers and some ESR and cartridge systems. Your legal menu is whatever is waived (or otherwise assigned) on the laboratory's CLIA certificate plus the current IFU—not a brand you saw in another clinic.

Method CMLA performance Classic integrity rules

Waived hemoglobin photometer (cuvette or strip) Yes, if trained and on the menu Mix the specimen; correct volume; no clots; no off-label dilution or extra incubation

Spun microhematocrit Yes when the laboratory uses this assigned method Seal, balance, spin the timed IFU cycle; read the red-cell column, not the buffy coat; no air bubbles

Waived ESR kit/device Yes if that kit is waived and assigned Level, still, timed; no clots, vibration, or delay beyond IFU

Cartridge blood analyzer (i-STAT-type) Only the cartridges that are FDA-waived and listed on the laboratory menu Fill the cartridge as IFU shows; no air; no off-menu cartridge

Automated CBC counter and smear differential No independent CMLA performance Deliver mixed unclotted EDTA to MLT/MLS personnel

Heterophile mono test Not this section Immunology/serology, Chapter 12

Fecal occult blood appears on the federal waived list but is not a hematology cell-count method; skip it here (chemistry/urinalysis-adjacent waived testing is covered elsewhere). Copper-sulfate hemoglobin is historically waived; many sites have replaced it with photometers. If your SOP still uses copper sulfate, follow that IFU exactly—do not invent a specific-gravity cutoff AMT did not print.

i-STAT-type devices are a CMLA trap because the analyzer is not automatically waived . Some cartridges are FDA-waived (for example certain chemistry/hematocrit combinations cleared as waived); other cartridges on the same handle are moderate or high complexity. You perform only the waived cartridge that is on the menu, with that cartridge's IFU. Grabbing a nonwaived hematology or coagulation cartridge because the device is on the counter is practicing beyond the certificate.

How you actually perform the common waived tests

Waived hemoglobin. Confirm patient identification, mix well-mixed capillary or venous whole blood exactly as the IFU specifies , fill the cuvette or strip without bubbles, and read in the labeled time window. Apply the lot calibration (code chip or factory cuvette). Run required QC first (Section 11.2). Capillary first-drop technique follows Chapter 8: wipe the first drop when the IFU/SOP says to, do not milk the finger, and do not use a clotted globule. A hemoglobin photometer result is not a CBC. If the provider ordered a CBC, the waived hemoglobin is not a silent substitute unless the SOP says so.

Spun microhematocrit. Fill the capillary tube from mixed EDTA or from a capillary collection to the IFU mark, seal one end with clay (not your glove), wipe the outside, place the sealed end against the gasket, balance the rotor, close the lid, and spin for the manufacturer's time and speed. Read the packed red-cell column as a percentage of total blood column. Do not include the buffy coat (white cells and platelets) in the hematocrit. Air bubbles, unsealed tubes, an unbalanced head, and reading before the cells are packed invalidate the result. Duplicate tubes are used when the SOP requires them; a large mismatch means recollect or repeat per SOP, not averaging a bubble tube with a good tube.

Waived ESR. Mix the assigned anticoagulated blood, fill to the mark without bubbles, place the tube vertical in a rack that is not bumped, and time the IFU interval. Classic teaching is that delays, clots, vibration, direct sunlight, and a tilted tube falsely change how far red cells fall. Follow the kit; do not move a nonautomated ESR tube halfway through to a busier bench.

Specimen problems: clots, fill, lipemia, hemolysis

Waived methods do not forgive bad specimens. A number on the screen is not automatically reportable.

Problem What it does CMLA action

Clotted EDTA Cells and platelets trapped in fibrin; counts and many whole-blood hemoglobin/hematocrit results invalid Reject and recollect; do not pick out the clot

Underfilled EDTA Excess EDTA shrinks red cells; hematocrit and mean cell volume can fall; morphology distorts Recollect to the fill line; do not top off from another tube

Overfilled EDTA Additive may be insufficient; microclots form Recollect; do not shake harder and hope

Unmixed / settled Sampling the plasma layer under-reads hemoglobin/hematocrit; sampling the packed layer over-reads Invert again per IFU immediately before testing

Hemolysis Free hemoglobin and turbidity; photometric hemoglobin can be biased; many methods flag or invalidate Do not report against the IFU; recollect a nonhemolyzed specimen if possible

Lipemia (and marked icterus) Turbidity (and pigment) can falsely raise some spectrophotometric hemoglobin readings Follow IFU interference limits; notify qualified personnel; do not report a known invalid

Microhematocrit air bubble or broken seal Column height is not a true packed-cell fraction Repeat with a new tube; do not estimate past the bubble

Hemolysis from a traumatic stick, a squeezed finger, or a shaken tube is a collection error you prevent. If the waived photometer still prints a number, the IFU—not the waiting provider—decides whether that number may leave the bench. Lipemia is milky plasma from high triglycerides (or recent fat load). You may see it in a spun hematocrit tube above the cells. Some hemoglobin methods read the turbidity as extra hemoglobin. Escalate; do not mentally subtract a guess.

Capillary versus venous hemoglobin and hematocrit can differ slightly. Do not relabel a fingerstick as a venous CBC specimen. Warm the site, wipe as required, and fill without air.

What this section is not

Mono test: heterophile agglutination is immunology (Chapter 12). Mention it only to avoid running it as a hematology CBC substitute.

High-complexity differential or core-lab CBC: still MLT/MLS work even if you can power the analyzer on.

Off-label tweaks: extra incubation of a faint hemoglobin cuvette, saline dilution so a thick sample will fill, an expired ESR kit, or an unapproved specimen type defaults the method to high complexity (Chapter 5). A CMLA must not independently perform that high-complexity testing. A control line or a plausible number does not restore waived status.

Scenario

Clinic orders a waived hemoglobin and an ESR. The EDTA tube is underfilled and the plasma layer looks milky. You do not squeeze extra drops from a previously poked finger into the short EDTA tube, and you do not run the lipemic sample because the photometer will print something. You recollect a proper EDTA, mix it, run in-control QC on the hemoglobin photometer, and set up the waived ESR on an unclotted, well-mixed specimen. A nurse hands you an i-STAT cartridge that is not on the waived menu and asks you to incubate a faint cuvette longer. You decline both: off-menu cartridge and extra incubation are not CMLA waived hematology.

Exam traps

Treating every cartridge on a handheld analyzer as waived.

Putting the mono test in this section.

Reporting hemoglobin from clotted, underfilled, hemolyzed, or frankly lipemic specimens against the IFU.

Including buffy coat in a spun hematocrit, or spinning an unbalanced centrifuge.

Diluting or over-incubating a waived kit to make it work.

Perform the waived method exactly. Protect the specimen. Leave mono, differentials, and nonwaived cartridges off this bench.

Loading diagram... Waived Hematology Path Versus Tests That Do Not Belong Here