Study guide
14.2 Waived Urinalysis Procedures and Specimen Handling
AMT Certified Medical Laboratory Assistant CMLA
Track your progress as you read
Finished sections get a check mark.
1 Introduction
1.1 Current Exam Facts 1.2 Blueprint Weights and Study Strategy
2 Chapter 2: Safety Terminology and Infection Control
2.1 Safety, Infection, and OSHA Terminology 2.2 Chain of Infection and Precautions 2.3 Isolation, Aseptic Technique, and Disease Prevention
3 Chapter 3: OSHA, HazCom, and Public-Health Reporting
3.1 OSHA Bloodborne Pathogens and Needlestick Safety 3.2 Hazard Communication, SDS, and GHS 3.3 Mandated Safety Rules and Public-Health Reporting
4 Chapter 4: Quality Assessment and Quality Control
4.1 Quality Control and Assessment Terminology 4.2 QC vs QA, Internal vs External QC, and Proficiency Testing 4.3 Daily QC, Temperature Logs, and Corrective Action 4.4 Instrument Maintenance and Calibration in CMLA Scope
5 Chapter 5: Waived-Test Integrity and Confirmatory Testing
5.1 Off-Label Waived Testing and High-Complexity Default 5.2 Following Manufacturer Instructions for Waived Tests 5.3 Confirmatory and Supplemental Testing
6 Chapter 6: Phlebotomy Foundations
6.1 Phlebotomy Terminology and Equipment 6.2 Serum, Plasma, Whole Blood, and Anticoagulation 6.3 Patient Identification and Container Identification
7 Chapter 7: Venipuncture, Order of Draw, and Specimen Integrity
7.1 Patient Preparation and Physiologic Effects 7.2 Vein Selection and Venipuncture Technique 7.3 Order of Draw and Tube Additives 7.4 Short Draws, Hemolysis, and Plasma-on-Cells Limits
8 Chapter 8: Capillary Collection, Processing, and Legal Specimens
8.1 Capillary Puncture Including Infants 8.2 Post-Care, Labeling, Transport, Processing, and Send-Outs 8.3 Specimens Used as Legal Evidence
9 Chapter 9: Non-Blood Collection and Preanalytical Errors
9.1 Urine Collection Types and Patient Instruction 9.2 Urine Suitability, Storage, Preservatives, and Rejection 9.3 Other Specimens and Time/Temperature Requirements 9.4 Shipping, Slide Prep, Preanalytical Errors, and Complications
10 Chapter 10: Clinical Chemistry
10.1 Clinical Chemistry Terminology and Test Families 10.2 Chemistry Instrumentation, Cleaning, and Maintenance 10.3 Point-of-Care and Waived Chemistry Testing 10.4 Toxicology Within CMLA Scope
11 Chapter 11: Hematology and Coagulation
11.1 Hematology Terminology, Blood Functions, and Cell Types 11.2 Hematology Instrumentation, QC, and Calibration 11.3 Coagulation: PT, PTT, INR, Heparin, and Warfarin 11.4 Point-of-Care and Waived Hematology Tests
12 Chapter 12: Immunology, Serology, and Immunohematology
12.1 Antigen-Antibody Principles 12.2 Blood Types, Rh, and Immunohematology Basics 12.3 Waived Immunology, Serology, and Immunohematology Procedures
13 Chapter 13: Microbiology
13.1 Microbiology Terminology and Specimen Appropriateness 13.2 Aseptic Processing, Storage, and Transport of Micro Specimens 13.3 Media, Cultures, and Procedure Matching 13.4 Stains, KOH, Wet Preps, Reagents, and Controls
14 Chapter 14: Urinalysis and Microscopy
14.1 Urinalysis Terminology 14.2 Waived Urinalysis Procedures and Specimen Handling 14.3 Microscopy Parts, Magnification, Wet Prep, and Cleaning
15 Chapter 15: Basic Anatomy and Physiology
15.1 Anatomy Relevant to Laboratory Testing 15.2 Physiology Relevant to Laboratory Testing
16 Chapter 16: Legal, Ethical, and HIPAA Foundations
16.1 Consent, ABN, Test Ordering, and PPM Restrictions 16.2 Patient's Bill of Rights, Refusal, and Medicolegal Risk 16.3 Professional Appearance, Conduct, and Social Media 16.4 HIPAA Confidentiality and Record Release
17 Chapter 17: Clerical Skills, Communication, and Critical Values
17.1 Reports, Abnormal Results, and Critical-Value Notification 17.2 HIT Data, Inventory, Expirations, and Billing/Coding 17.3 Patient Instruction, Communication, Telephone, and Time Management
14.2 Waived Urinalysis Procedures and Specimen Handling
Key Takeaways Explain This Topic AI
Unpreserved routine UA is examined promptly—commonly within 2 hours of voiding—or refrigerated; mix before dipping and bring refrigerated cups toward room temperature per SOP.
Read each reagent-strip pad at the IFU time; runover (urine cascading from pad to pad) contaminates adjacent reactions and can invalidate the screen.
Dipstick pads screen glucose, protein, blood, leukocyte esterase, nitrite, ketones, pH, specific gravity, urobilinogen, and bilirubin—they are chemical analyses, not a culture report.
Confirmatory tablet or precipitation tests (Ictotest, sulfosalicylic acid, historically Clinitest) are used only when the current IFU and SOP still list them; many laboratories have retired Clinitest.
Qualitative urine hCG is a waived immunoassay (Chapter 12), not a Chemstrip pad; leftover patient urine is not quality control.
Last updated: August 2026
CMLA III.9.B tests whether you can prepare, store, dispose of, and properly transport UA specimens and perform urinalysis tests within the laboratory assistant’s scope . Scope is unmodified waived reagent-strip UA (visual or a waived reader) plus the handling that keeps the result valid. Slide preparation is started here and finished in Section 14.3. Qualitative urine human chorionic gonadotropin (hCG) is commonly waived; it is an immunoassay (Chapter 12), not a dipstick pad—follow that cassette’s IFU and do not treat a pregnancy device as a Chemstrip.
Quick Answer: Mix, label, and test unpreserved UA promptly (commonly within 2 hours ) or refrigerate. Bring cold specimens toward room temperature before dipping. Read each pad at the IFU time. Prevent runover . Do not report a strip as culture. Tablet confirmations (Ictotest, SSA, historically Clinitest) follow the current IFU; many laboratories have retired Clinitest. Dispose of urine as a biohazard under Standard Precautions.
Handle the specimen before you dip
Identity. Two identifiers match the order (CLSI PRE01). An unlabeled bathroom cup is a rejection, not a dip (Chapter 9).
Suitability. Match the cup to the test. A boric-acid C&S tube is often the wrong aliquot for reagent strips because the preservative can distort pH and other pads. Fecal contamination, leakage, insufficient volume, and delayed unrefrigerated urine are rejections.
Mix. Cells and casts settle. Invert gently. Do not shake a foam that you then call protein.
Temperature. If the cup was refrigerated, let it warm toward room temperature per SOP before chemical testing. Cold urine can dull some pad reactions and dump amorphous crystals into the sediment. Warming to dissolve amorphous urates or phosphates is an SOP step, not a reason to incubate the cup on a heater block you invented.
Timing. Unpreserved routine UA: examine promptly, commonly within 2 hours , or refrigerate (typically 2–8°C). Delay grows bacteria (false nitrite, false cloudiness), consumes glucose, lyses cells and casts, fades bilirubin and urobilinogen, and raises pH as urea becomes ammonia.
Transport. Leak-proof lid, biohazard bag, upright, labeled, no delay on a warm counter. Send-outs follow CLSI PRE04/PRE05 and the receiving laboratory’s catalog. Do not ship an open cup in a staff coat pocket.
Dispose. Treat urine as potentially infectious. Empty per SOP into the designated sink or waste, then discard the cup in biohazard waste. Used strips are wet with urine—biohazard, not office trash. Do not pour leftover UA down a public restroom because it is “only urine,” and do not store leftover cups in the staff refrigerator beside lunch.
Waived chemical UA: the pads
A typical 10-parameter reagent strip (follow the brand in your hand ) includes the pads below. Visual reading uses the chart on the bottle; a waived automated reader times and reads for you if that system is cleared as waived and you use it unmodified.
- Pad Detects CMLA meaning High-yield limit
- Glucose Glucose (oxidase/peroxidase) Glycosuria screen Specific for glucose; not galactose
Protein Mainly albumin (protein-error of indicators) Proteinuria screen May miss light chains; highly alkaline urine can false-positive
Blood Heme peroxidase activity Intact RBCs, free hemoglobin, myoglobin Microscope needed to confirm intact hematuria
Leukocyte esterase Esterase in neutrophils Supports pyuria Lymphocytes have little esterase; vaginal contamination can positive
Nitrite Nitrate → nitrite (Griess) Suggests nitrate-reducing bacteria Needs bladder time; Enterococcus often negative; not a culture
- Ketones Nitroprusside (mainly acetoacetate) Ketosis screen Poor for β-hydroxybutyrate
- pH Double indicator Acid vs alkaline urine Delay and urea-splitting bacteria alkalinize
- Specific gravity Ionic polyelectrolyte Concentration estimate Not a refractometer
Urobilinogen Ehrlich-type / diazonium Normal small amount Increased in hemolysis or hepatocellular disease; light-sensitive
Bilirubin Diazonium Conjugated bilirubin Light-sensitive; confirm per SOP (Ictotest)
Physical (macroscopic) observations still count. A red, blood-pad-negative specimen may be beets, porphyrins, or dye. A cloudy, nitrite-negative specimen may be crystals, mucus, or organisms that do not reduce nitrate. Phenazopyridine (pyridium) turns urine orange and can mask pad colors—do not force a read the IFU says is unreadable.
Quality control. Run the manufacturer’s UA controls at the frequency the IFU and SOP require. Do not use leftover patient urine as homemade QC (that modification defaults the method to high complexity—Chapter 5). Check strip-bottle dating, cap the bottle immediately (pads are humidity-sensitive), and never combine leftover strips from two bottles.
Timing of reads and runover
The IFU assigns seconds , not “when it looks done.” Leukocyte esterase is often the longest visual pad (commonly about 2 minutes on classic Multistix-style inserts); glucose, bilirubin, and several others are shorter. If you use a waived reader, you use its timing, calibration, and QC as cleared. Reading after the stop time is a modification . So is diluting the urine so a viscous specimen “flows nicer.”
Runover is excess urine flooding from one pad onto the next so chemicals mix. Acid from an adjacent pad can distort protein; leftover bleach or strong oxidizers on the container can false-positive blood or glucose. Technique:
Dip briefly to wet all pads; do not soak.
Drag the edge along the container rim.
Blot the edge of the strip on absorbent paper.
Hold the strip level (or seat it in the reader) so urine does not cascade down the pads.
Do not lay a dripping strip face-down on the color chart.
Compare each pad at its time; do not read the whole strip at 30 seconds because you are busy.
Confirmatory tablet and precipitation tests — follow the IFU
Some positives still need a second method. These are confirmatory tests in AMT language. Use them only when the current IFU and SOP still list them.
Ictotest (tablet diazonium method) is a classic, more sensitive and specific confirmation for urine bilirubin than the strip pad. Follow the current tablet IFU for number of drops, mat, and time.
Sulfosalicylic acid (SSA) precipitation can detect proteins the albumin-biased strip misses (including some globulins and Bence Jones proteins). Interpret turbidity against the SOP chart. SSA is not a dipstick, and boric-acid aliquots are often the wrong matrix.
Clinitest (copper-reduction tablet) historically detected reducing substances , including glucose and galactose —the reason pediatric laboratories used it when the glucose oxidase pad was negative but a reducing sugar might still be present. Many laboratories have retired Clinitest : tablets are caustic, the reaction boils, and commercial availability has collapsed. If your SOP still lists it, follow that IFU and the chemical-hygiene rules. If it does not, do not improvise Benedict’s solution at the bench. A CMLA does not independently invent a galactosemia workup.
Acetest tablets historically confirmed ketones. Same rule: current IFU or not at all.
Do not substitute a second dip for a required confirmatory tablet. Do not use an expired Clinitest “because we still have a few.” Off-label confirmatory recipes default to high complexity (Chapter 5).
What waived UA is not
Do not report a dipstick as culture. Leukocyte esterase and nitrite support a UTI screen. They do not identify the organism, do not give a colony count, and do not generate a susceptibility panel. C&S is a microbiology order with a clean-catch (or catheter) specimen. A positive strip may prompt the provider to add culture; it does not let you type an organism and a drug list from a pad.
Waived urine hCG remains Chapter 12: qualitative cassette, labeled specimen type, read window, control line. First-morning urine is preferred for hCG because it is more concentrated; that preference does not turn the chemistry strip into a pregnancy test.
Automated nonwaived microscopy analyzers and manual sediment identification are moderate-complexity work. Perform them only when the laboratory’s CLIA certificate, SOP, and your competency assessment say you may (Section 14.3). A Certificate of Waiver site that only dips strips does not become a microscopy laboratory because you found a scope in a closet.
In practice
You receive a labeled random cup, 45 minutes old, for waived UA. You mix, dip without runover, time the pads with the IFU or waived reader, record color and clarity, and note leukocyte esterase moderate and nitrite positive. You do not write “positive C&S.” You do not add Clinitest because the glucose pad is negative in an adult with no SOP for reducing substances. You dispose of the strip and leftover urine as biohazard. If culture is also ordered, you follow the sterile-cup or boric-acid pathway from Chapter 9 instead of dipping that preservative tube.
- Exam traps
- Reading every pad at 30 seconds, or waiting “until the color looks real.”
Allowing runover, then trusting protein and pH.
Dipping a refrigerated unmixed cup, or a boric-acid C&S aliquot.
Reporting nitrite or leukocyte esterase as C&S.
Using leftover patient urine as QC, or expired Clinitest as a pediatric screen.
Calling a urine hCG cassette a tenth dipstick pad.
Keep the method waived by matching specimen, strip, clock, and IFU—and keep culture in the microbiology column.