Study guide
Signal Processing: Demodulation, Rejection & Pre/Post-Processing
ARDMS SPI Sonography Principles & Instrumentation
Track your progress as you read
Finished sections get a check mark.
1 Introduction & the SPI Exam
What the SPI Is & Why Physics Matters The V24.1 Content Outline & How to Study It Math, Units & Decibels You Must Know
2 Properties of Ultrasound Waves
Sound as a Mechanical Longitudinal Wave Period, Frequency & Wavelength Propagation Speed & the Wavelength Equation Amplitude, Power & Intensity
3 Pulsed Ultrasound
Continuous vs Pulsed Wave; PRF & PRP Pulse Duration, Spatial Pulse Length & Duty Factor Bandwidth, Q-Factor & Damping
4 Interaction of Sound with Tissue
Attenuation: Absorption, Reflection & Scattering The Attenuation Coefficient & Half-Value Layer Acoustic Impedance & Reflection Refraction, Snell's Law & the Range Equation
5 Transducers & Sound Beams
Construction & the Piezoelectric Effect The Sound Beam: Near/Far Field & Focusing Array Transducers Transducer Selection, Frequency Trade-offs & Nonimaging Probes
6 Resolution
Axial Resolution Lateral & Elevational Resolution Temporal Resolution & Frame Rate
7 Pulse-Echo Instrumentation & the Imaging Chain
Beam Former, Pulser & Output Power The Receiver: Amplification & Time-Gain Compensation Dynamic Range & Compression Signal Processing: Demodulation, Rejection & Pre/Post-Processing Scan Converter, Display & Image Storage
8 Display Modes & Image-Optimization Features
A-Mode, B-Mode & M-Mode Harmonic Imaging Spatial & Frequency Compounding, Panoramic Imaging 3D/4D Imaging & Contrast Agents
9 Doppler Principles & Instrumentation
The Doppler Effect, Shift & Equation The Doppler Angle & Cosine Continuous-Wave Doppler Pulsed-Wave Doppler, Sample Volume & Nyquist Limit Aliasing & How to Eliminate It Spectral Doppler Analysis & Waveforms Color & Power Doppler Hemodynamics & Doppler Measurements
10 Artifacts
Propagation Artifacts Attenuation & Beam-Width Artifacts Doppler & Color Artifacts
11 Bioeffects, Safety & Quality Assurance
Bioeffects Mechanisms: Thermal & Cavitation Thermal & Mechanical Index & ALARA Quality Assurance & Tissue-Mimicking Phantoms Statistical Parameters: Sensitivity, Specificity & Predictive Value Patient Care, Ergonomics, Infection Control & Workflow
Signal Processing: Demodulation, Rejection & Pre/Post-Processing
Key Takeaways Explain This Topic AI
Pre-processing functions are applied to the signal before it is stored in scan-converter memory and cannot be changed afterward; post-processing functions are applied after storage and can be adjusted on a frozen image.
Write zoom re-scans the region of interest at higher line density before storage, genuinely improving resolution within the zoomed area.
Read zoom magnifies already-stored pixels after the fact and does not add new resolution or information.
Persistence (frame averaging) blends consecutive frames to reduce noise and speckle but reduces temporal resolution for fast-moving structures.
Frequency compounding combines images acquired using multiple frequency bands to reduce speckle noise.
Last updated: July 2026
Overview: Pre-Processing vs. Post-Processing
Before a returning echo signal becomes a pixel on the screen, it passes through a series of signal-processing steps. SPI groups these into two broad categories:
Category When it happens Can it be changed after the image is frozen/stored?
Pre-processing Applied to the signal before it is stored in the scan converter's image memory No — pre-processing choices are "baked in" to the stored image data
Post-processing Applied to the image data after it is stored in the scan converter's image memory Yes — post-processing can be adjusted on a frozen or previously acquired image, including on a stored/reviewed image
Nearly every function in this section can be classified as pre- or post-processing, and SPI regularly tests which category a given function belongs to.
Demodulation and Rejection
Demodulation converts the raw, high-frequency radiofrequency (RF) echo signal into a usable video signal. It has several sub-steps, including amplification (Section 7.2), compression (Section 7.3), and rectification (converting the RF waveform, which has both positive and negative voltage swings, into an all-positive envelope that represents echo amplitude over time). Demodulation happens early, before the signal reaches the scan converter, and is therefore a pre-processing function.
Rejection (sometimes called suppression) is a receiver function that eliminates low-amplitude signals below a chosen threshold — typically electronic noise or very weak scatter — so they do not appear on the display as unwanted low-level gray "haze." Raising rejection cleans up the image by removing faint noise, but rejection set too high can also discard genuine weak echoes; it is generally classified as pre-processing since it acts on the signal before scan conversion.
Read Zoom vs. Write Zoom
Zoom magnifies a region of interest, but SPI distinguishes two very different mechanisms:
Write zoom (also called "true" or "research" zoom) is applied before the scan converter writes the image into memory: the sonographer selects a region of interest and the system re-scans that region using a higher scan-line density, actually acquiring more data points to fill the zoomed region. Because it changes what data is written to memory, write zoom is a pre-processing function, and it genuinely improves detail/resolution within the zoomed region.
Read zoom (also called "pan" zoom) is applied after the image is already stored in scan-converter memory: the system simply magnifies the existing stored pixels, enlarging them on the display. Because it only reads and re-displays already-stored data, read zoom is a post-processing function, and it does not add new information or improve resolution — it only makes existing pixels bigger (similar to zooming in on a digital photo, which can produce a blocky/pixelated look at high magnification).
Gray-Scale Maps
A gray-scale map (also called a gray-scale curve or transfer curve) is the function that assigns a specific shade of gray to each possible echo-amplitude value after dynamic range compression (Section 7.3). Different preset gray-scale maps emphasize different parts of the amplitude range — for example, one map might favor contrast in low-amplitude (weak echo) regions, while another favors contrast in high-amplitude (strong echo) regions. Because gray-scale maps are typically adjustable on a stored/frozen image, they are classified as a post-processing function.
Persistence (Frame Averaging)
Persistence , also called frame averaging or temporal compounding , is a post-processing technique that blends several consecutive image frames together before displaying the composite. Because random noise varies frame-to-frame while real anatomic structures remain consistent across frames, averaging frames together smooths out noise and speckle, producing a less "grainy" image. The trade-off is temporal resolution: high persistence blurs rapid motion (because it blends frames spanning a longer time interval), so persistence should be reduced when imaging fast-moving structures such as a beating fetal heart or valve leaflets.
Edge Enhancement
Edge enhancement is a processing technique that increases the amplitude difference at boundaries between adjacent structures with different echogenicity, making borders/edges appear sharper and more distinct. It works by exaggerating amplitude transitions rather than adding new anatomic information, improving the subjective visibility of a boundary such as an organ capsule or a mass margin.
Frequency Compounding
Frequency compounding transmits and/or processes multiple different frequency bands (rather than a single frequency) from the same location and combines the resulting images into a single composite frame. Because each frequency band produces slightly different speckle and interference patterns, averaging across bands reduces speckle noise and improves overall image uniformity — a mechanism distinct from spatial compounding (Chapter 8), which instead combines images acquired from multiple beam-steering angles .
Summary Table: Classifying Each Function
- Function Pre- or Post-Processing Primary Effect
- Demodulation/rectification Pre Converts RF signal to usable video envelope
- Rejection Pre Removes low-level noise below a threshold
- Write zoom Pre Re-scans region at higher line density — true resolution gain
- Read zoom Post Magnifies stored pixels — no new resolution
- Gray-scale map Post Reassigns amplitude-to-gray-shade mapping
- Persistence (frame averaging) Post Smooths noise/speckle; reduces temporal resolution
- Edge enhancement Post Sharpens borders between structures
- Frequency compounding Pre (acquisition-based) Reduces speckle by combining frequency bands
Mastering the pre-/post-processing split, and especially the read-zoom-vs-write-zoom distinction, directly covers Domain 3 tasks 3.E (magnification), 3.M (gray scale, frequency compounding, persistence), and 3.N (edge enhancement).