Study guide

Amplitude, Power & Intensity

ARDMS SPI Sonography Principles & Instrumentation

Track your progress as you read

Finished sections get a check mark.

1 Introduction & the SPI Exam

What the SPI Is & Why Physics Matters The V24.1 Content Outline & How to Study It Math, Units & Decibels You Must Know

2 Properties of Ultrasound Waves

Sound as a Mechanical Longitudinal Wave Period, Frequency & Wavelength Propagation Speed & the Wavelength Equation Amplitude, Power & Intensity

3 Pulsed Ultrasound

Continuous vs Pulsed Wave; PRF & PRP Pulse Duration, Spatial Pulse Length & Duty Factor Bandwidth, Q-Factor & Damping

4 Interaction of Sound with Tissue

Attenuation: Absorption, Reflection & Scattering The Attenuation Coefficient & Half-Value Layer Acoustic Impedance & Reflection Refraction, Snell's Law & the Range Equation

5 Transducers & Sound Beams

Construction & the Piezoelectric Effect The Sound Beam: Near/Far Field & Focusing Array Transducers Transducer Selection, Frequency Trade-offs & Nonimaging Probes

6 Resolution

Axial Resolution Lateral & Elevational Resolution Temporal Resolution & Frame Rate

7 Pulse-Echo Instrumentation & the Imaging Chain

Beam Former, Pulser & Output Power The Receiver: Amplification & Time-Gain Compensation Dynamic Range & Compression Signal Processing: Demodulation, Rejection & Pre/Post-Processing Scan Converter, Display & Image Storage

8 Display Modes & Image-Optimization Features

A-Mode, B-Mode & M-Mode Harmonic Imaging Spatial & Frequency Compounding, Panoramic Imaging 3D/4D Imaging & Contrast Agents

9 Doppler Principles & Instrumentation

The Doppler Effect, Shift & Equation The Doppler Angle & Cosine Continuous-Wave Doppler Pulsed-Wave Doppler, Sample Volume & Nyquist Limit Aliasing & How to Eliminate It Spectral Doppler Analysis & Waveforms Color & Power Doppler Hemodynamics & Doppler Measurements

10 Artifacts

Propagation Artifacts Attenuation & Beam-Width Artifacts Doppler & Color Artifacts

11 Bioeffects, Safety & Quality Assurance

Bioeffects Mechanisms: Thermal & Cavitation Thermal & Mechanical Index & ALARA Quality Assurance & Tissue-Mimicking Phantoms Statistical Parameters: Sensitivity, Specificity & Predictive Value Patient Care, Ergonomics, Infection Control & Workflow

Amplitude, Power & Intensity

Key Takeaways Explain This Topic AI

Amplitude is the magnitude of variation of an acoustic variable (such as pressure) from its baseline value.

Power is proportional to the square of amplitude; doubling amplitude quadruples power.

Intensity is defined as power divided by the beam's cross-sectional area, expressed in watts per square centimeter (W/cm²).

Intensity naming conventions describe spatial and temporal averaging or peak conditions, such as SPTA (spatial peak, temporal average) and SPPA (spatial peak, pulse average).

SPTA intensity is most closely linked to thermal bioeffects because it averages energy over the entire pulse repetition period, including the quiet listening time.

Last updated: July 2026

Amplitude: The Size of the Variation

Amplitude describes the strength, or magnitude, of the variation in an acoustic variable — most commonly pressure — from its normal, undisturbed (baseline) value to its maximum value. A "loud" or "strong" pulse has a large pressure amplitude (a big swing between the peak of a compression and the trough of a rarefaction); a "quiet" or "weak" pulse has a small amplitude. Amplitude is controlled at the machine by output power (via the pulser/beam former, covered in Chapter 7) and is entirely independent of frequency — you can have a high-amplitude, low-frequency pulse or a low-amplitude, high-frequency pulse. Because pressure amplitude is the most clinically relevant of the four acoustic variables, it is commonly expressed in pascals (Pa) or megapascals (MPa) when describing peak rarefactional pressure for safety-index calculations (covered in Chapter 11), but for conceptual purposes it is simplest to think of amplitude as "how big the pressure swing is" above and below the undisturbed baseline value.

Power: Amplitude Squared

Power is the rate at which acoustic energy is transferred through the medium, measured in watts (W) or milliwatts (mW). Power is not simply proportional to amplitude — it follows a square-law relationship :

  • Formula Meaning
  • Power ∝ Amplitude² Power is proportional to the square of amplitude

This square-law relationship has a large practical consequence: small changes in amplitude produce disproportionately large changes in power. Doubling amplitude does not double power — it quadruples it (2² = 4). Tripling amplitude produces nine times the power (3² = 9). This is the same square-law logic used throughout physics (kinetic energy proportional to velocity squared, for example) and it reappears on the SPI exam specifically in the amplitude/power relationship.

Intensity: Power Spread Over Area

Intensity takes power a step further by accounting for how tightly that power is focused. It is defined as:

  • Formula Meaning
  • Intensity = Power / Area Expressed in watts per square centimeter (W/cm²)

For a fixed amount of power, concentrating it into a smaller cross-sectional area (a narrower beam) produces a higher intensity , and spreading the same power over a larger area produces a lower intensity . Because power itself is proportional to amplitude², intensity is also proportional to amplitude² for a fixed beam area — meaning a small increase in output amplitude at the machine can meaningfully raise the intensity delivered to the patient.

Naming the Many Faces of Intensity

Ultrasound intensity is not a single fixed number — it varies from moment to moment (as pulses turn on and off) and from location to location within the beam (highest at the beam's center, lower toward the edges). To describe intensity precisely, ultrasound physics uses a two-part naming convention:

Spatial term — describes intensity across the beam's cross-section: peak (the single highest point in the beam) or average (averaged across the whole beam area)

Temporal term — describes intensity over time: peak (during the pulse only), pulse average (averaged across just the pulse-on duration), or average (averaged across the entire pulse repetition period, including the "silent" listening time between pulses)

Abbreviation Full name What it represents

SPTP Spatial Peak, Temporal Peak The single highest instantaneous intensity anywhere in the beam — the largest value of all

SPPA Spatial Peak, Pulse Average Highest beam-center intensity, averaged only across the "on" portion of each pulse

SATA Spatial Average, Temporal Average Intensity averaged across the whole beam and across the entire pulse cycle (on and off time)

SPTA Spatial Peak, Temporal Average Highest beam-center intensity, averaged across the entire pulse repetition period (on and off time)

Because diagnostic ultrasound uses a very low duty factor (the pulse is "on" only a tiny fraction of the time — detailed in Chapter 3), the time-averaged values are dramatically smaller than the peak values: SATA < SPTA < SPPA < SPTP , from lowest to highest.

Why It Matters: The Link to Bioeffects

Intensity values are not just abstract labels — they are the basis for regulatory safety limits and bioeffect risk assessment (fully covered in Chapter 11). SPTA is the value most closely tied to thermal bioeffects , because it factors in the entire pulse repetition period, including the quiet listening time when tissue has a chance to cool — making it the best available proxy for cumulative heating. Higher amplitude leads to higher power, which leads to higher intensity, which leads to greater potential for tissue heating and, at very high pressures, cavitation. This chain of relationships (amplitude to power to intensity to bioeffect risk) is exactly why output power controls, TGC/gain, and machine safety indices are treated as physically distinct concepts on the SPI exam: gain amplifies received echoes electronically with no patient bioeffect cost, while output power changes the amplitude actually transmitted into the patient and does carry bioeffect implications.

SPI Exam Tip

Expect a direct amplitude/power squaring calculation ("if amplitude triples, power increases by what factor?") and at least one item asking you to identify which intensity abbreviation is used for a described averaging scenario, or which one is most relevant to thermal bioeffect limits (SPTA).