Study guide

9.4 Lumbar Spine Positioning: The Leg Block, Centering, and Common Challenges

ARRT Bone Densitometry Exam

Track your progress as you read

Finished sections get a check mark.

1 Chapter 1: Exam Orientation & the ARRT Blueprint

1.1 About the ARRT Bone Densitometry Examination 1.2 Eligibility, Structured Education, and Clinical Experience

2 Chapter 2: Osteoporosis, Bone Biology & Risk Factors

2.1 Osteoporosis: Definitions, Types, and WHO Diagnostic Criteria 2.2 Bone Functions, Structural Anatomy, and Bone Types 2.3 Bone Cells, the Remodeling Cycle, and Calcium Homeostasis 2.4 Risk Factors, Secondary Causes, and Bone-Compromising Medications

3 Chapter 3: Indications, Prevention & Treatment

3.1 The Bone Mass Measurement Act: Covered Indications and Coding 3.2 Prevention: Nutrition, Exercise, Smoking Cessation, and Fall Prevention 3.3 Osteoporosis Drug Therapy: Antiresorptive, Anabolic, and Hormonal Agents

4 Chapter 4: Patient Preparation, Support & Radiation Safety

4.1 Patient Support, Disabilities, Ergonomics, and Transfer Techniques 4.2 Patient Instructions: Explaining the Exam, Dose, Duration, and Motion Control 4.3 Patient History, Contraindications, and Pregnancy Screening 4.4 Scan Preparation: Patient Data Entry, Artifact Removal, and Infection Control 4.5 Radiation Safety in DXA: ALARA, Distance, and Dose Levels

5 Chapter 5: DXA Physics & Instrumentation

5.1 Physical Principles of Dual-Energy X-ray Absorptiometry 5.2 Beam Geometries: Pencil Beam and Fan Beam Systems 5.3 Fan-Beam Geometry: Source–Object–Detector Distances, Magnification, and Centering 5.4 The Scan Analysis Algorithm: Soft Tissue, Bone Edge Detection, BMC, Area, and BMD

6 Chapter 6: Measuring & Reporting Bone Mineral Density

6.1 Statistics for Densitometry: Mean, Standard Deviation, and Coefficient of Variation 6.2 T-Scores, Z-Scores, and Reference Database Selection 6.3 Clinical Indications for BMD Testing and the FRAX Tool 6.4 Vertebral Fracture Assessment and Trabecular Bone Score 6.5 Whole Body DXA, Body Composition, and Pediatric Scanning

7 Chapter 7: Quality Control, Precision & Equipment Management

7.1 Accuracy, Precision, and the Sources of Precision Error 7.2 Daily Quality Control, Phantoms, and Control Charts 7.3 In Vivo Precision Studies and RMS-SD Calculation 7.4 Determining and Applying the Least Significant Change 7.5 Maintenance, Repair, Relocation, Cross-Calibration, and Software Upgrades

8 Chapter 8: DXA Informatics & Data Management

8.1 DXA Informatics, DICOM, PACS, and Data Management

9 Chapter 9: Lumbar Spine DXA

9.1 Lumbar Spine Anatomy, Landmarks, and Regions of Interest 9.2 Lumbar Spine Scan Acquisition and ROI Edge Placement 9.3 Lumbar Spine: Compensating for Body Habitus, Variant Anatomy, Pathology, and Low Bone Density 9.4 Lumbar Spine Positioning: The Leg Block, Centering, and Common Challenges 9.5 Lumbar Spine Problems: Poor Bone Edge Detection and Obscured Intervertebral Spaces 9.6 Lumbar Spine: Variant Anatomy, Fractures, Degenerative Pathology, Calcification, and Surgery 9.7 Lumbar Spine Analysis: Vertebral Exclusion and Reporting Rules 9.8 Lumbar Spine Follow-Up Scans and Rate of Change

10 Chapter 10: Proximal Femur DXA

10.1 Proximal Femur Anatomy, Landmarks, and Regions of Interest 10.2 Femur Selection: Right, Left, and Dual Femur Scanning 10.3 Proximal Femur Scan Parameters and Compensation 10.4 Proximal Femur Positioning: Neck Rotation, Shaft Placement, and Positioning Aids 10.5 Proximal Femur Problems: Ischium, Short Neck, Arthritis, Fracture, and Hardware 10.6 Femoral ROI Analysis: Femoral Neck, Total Hip, and Ward's Area 10.7 Proximal Femur Follow-Up Scans and Rate of Change

11 Chapter 11: Forearm DXA

11.1 Forearm Anatomy, Landmarks, and Adjacent Structures 11.2 Forearm Indications and Side Selection 11.3 Forearm Length Measurement, Scan Parameters, and Positioning 11.4 Forearm Regions of Interest: One-Third, Mid, Ultradistal, and Total 11.5 Forearm Problems: Edge Detection, Artifacts, Low Bone Density, and Joint Disease 11.6 Scan Analysis: Region Placement, Window Level, and Reported Values 11.7 Forearm Follow-Up Scans and Rate of Change

12 Chapter 12: Cross-Site Artifacts & Clinical Integration

12.1 Recognition, Impact, and Management of DXA Artifacts 12.2 Clinical Interpretation, Serial Comparison, and Reporting

9.4 Lumbar Spine Positioning: The Leg Block, Centering, and Common Challenges

Key Takeaways Explain This Topic AI

The leg block or positioning cushion flexes the hips and knees to flatten lumbar lordosis so intervertebral disc spaces open and lie parallel to the beam.

The spine must appear straight and centered in the scan field with approximately equal soft tissue on each side.

The arms are positioned out of the scan field, typically across the chest or at the sides, so they do not contaminate the soft-tissue baseline.

The scan field must include enough anatomy to identify levels, showing T12 with its ribs at the top and part of the iliac crest at the bottom.

In scoliosis the patient is centered on the curve rather than the table midline, with the field widened to include the full lateral excursion, and the modification documented.

Last updated: September 2026

9.4 Lumbar Spine Positioning: The Leg Block, Centering, and Common Challenges

Quick Answer: Supine, midline centered, arms out of the field , and hips and knees flexed over a leg block to flatten lumbar lordosis. The spine should appear straight and centered with roughly equal soft tissue on each side, and the field should show T12 with ribs at the top and part of the iliac crest at the bottom so levels can be identified. Every deviation forced by anatomy is documented and reproduced.

What the Leg Block Actually Does

In a supine patient lying flat, the lumbar spine arches forward in physiologic lordosis. The vertebral bodies tilt relative to the tabletop, the intervertebral disc spaces close anteriorly, and the beam passes obliquely through each body.

Flexing the hips and knees over a leg block (also called a positioning cushion, spine block, or leg elevator) rotates the pelvis posteriorly and flattens the lordotic curve. The consequences are all favorable:

Disc spaces open and lie perpendicular to the beam , so intervertebral markers can be placed accurately and vertebral levels separate cleanly.

Vertebral bodies present squarely , reducing projected overlap between adjacent bodies.

Edge detection improves because bone margins are better defined.

Reproducibility improves , because the block imposes a standardized position rather than depending on how the patient happened to lie.

Typical hip and knee flexion is in the range of 60 to 90 degrees, with the manufacturer's block sized to produce it. What matters more than the exact angle is that the same block, at the same setting, is used at every visit — which means documenting the block and the setting.

The Standard Position, Step by Step

Patient supine , head on a thin radiolucent pad or directly on the table per protocol.

Longitudinal alignment: the patient's midsagittal plane aligned to the table centerline, verified with the laser or light localizer.

Leg block placed under both lower legs so hips and knees flex symmetrically. Both legs must rest on the block — asymmetric leg elevation tilts the pelvis and rotates the spine.

Arms out of the scan field: across the chest or at the sides away from the trunk, per manufacturer instruction. Arms lying within the baseline region contaminate the soft-tissue estimate.

Shoulders and hips square to the table; no rotation. Check by comparing the two anterior superior iliac spines for equal height.

Scan field set to begin below the level of the iliac crest and extend above T12, so the acquisition includes part of L5 or the crest inferiorly and T12 with ribs superiorly.

Instruct : breathe quietly and normally, remain still, do not talk.

Verify the localizer image before analysis : spine straight, centered, equal soft tissue bilaterally, correct levels included, no artifacts.

What a Correctly Positioned Spine Looks Like

  • Criterion Correct appearance
  • Alignment Spine straight and vertical in the image, not curved or angled
  • Centering Spine in the middle of the field with approximately equal soft tissue left and right
  • Rotation Spinous processes appear midline within the vertebral bodies; pedicles symmetric
  • Level coverage T12 with ribs visible at top; iliac crest or L5 visible at bottom
  • Disc spaces Open and clearly demarcated, allowing accurate marker placement
  • Field content No arms, no external artifacts, soft tissue present on both sides

If the spine appears angled or curved in a patient without scoliosis, the usual cause is patient rotation or asymmetric leg elevation — reposition rather than accept it.

  • Common Challenges
  • Scoliosis

A scoliotic spine cannot be made straight. The goals shift to capturing the whole spine and being reproducible.

Center on the curve , not the table midline, so the entire spine stays within the field with soft tissue on both sides.

Widen the scan field enough to include the full lateral excursion; a truncated spine is unanalyzable.

Expect rotational deformity. Scoliosis includes vertebral rotation, so bodies project asymmetrically, which degrades edge detection and precision.

Document the centering used , in enough detail to repeat it.

Recognize when the spine is not usable. Severe scoliosis with rotation and secondary degenerative change may make the spine uninterpretable; the hip then governs, and the report should say why.

Kyphosis

Severe thoracic kyphosis prevents the upper back and head from resting on the table, leaving the patient arched and unstable.

Support the head and upper thoracic region with radiolucent padding , positioned outside the measured lumbar field .

Do not place padding under the lumbar spine itself; on fan-beam systems that changes object height and therefore projected area.

Record the padding used and its placement.

Hip or Knee Contracture and Joint Replacement

Use the largest hip and knee flexion the patient tolerates , even if it is far short of the standard block position.

If the block cannot be used at all, support the legs with sponges and record the achieved position: "leg block not tolerated; knees supported on small sponge, hips flexed approximately 25 degrees."

Reproduce that exact arrangement at follow-up. A patient scanned flat at baseline and over a full block at follow-up will show a BMD change that is purely geometric.

Obesity

Verify table weight and field-width limits first.

Displace a pendulous panniculus symmetrically where possible so the soft-tissue baseline is not lopsided.

Use a slower acquisition mode for photon statistics.

Confirm the arms are clear of the trunk, which is harder in large patients and more important because the baseline region is already compromised.

Pain, Tremor, and Inability to Cooperate

Position gently and efficiently; a patient held in a painful position will move.

Shorten table time by acquiring only ordered sites.

Use the manufacturer's faster mode if motion risk outweighs noise, and document that choice.

Consider whether the study should be deferred rather than repeated three times.

The Positioning–Precision Link

Positioning is the single largest contributor to in vivo precision error, and the lumbar spine is where standardization pays most. Three habits carry most of the benefit: use the same block at the same setting, center the same way, and record any deviation the moment you make it rather than at the end of the day.