4.4 Scan Preparation: Patient Data Entry, Artifact Removal, and Infection Control
ARRT Bone Densitometry Exam
Track your progress as you read
Finished sections get a check mark.
1 Chapter 1: Exam Orientation & the ARRT Blueprint
1.1 About the ARRT Bone Densitometry Examination 1.2 Eligibility, Structured Education, and Clinical Experience
2 Chapter 2: Osteoporosis, Bone Biology & Risk Factors
2.1 Osteoporosis: Definitions, Types, and WHO Diagnostic Criteria 2.2 Bone Functions, Structural Anatomy, and Bone Types 2.3 Bone Cells, the Remodeling Cycle, and Calcium Homeostasis 2.4 Risk Factors, Secondary Causes, and Bone-Compromising Medications
3 Chapter 3: Indications, Prevention & Treatment
3.1 The Bone Mass Measurement Act: Covered Indications and Coding 3.2 Prevention: Nutrition, Exercise, Smoking Cessation, and Fall Prevention 3.3 Osteoporosis Drug Therapy: Antiresorptive, Anabolic, and Hormonal Agents
4 Chapter 4: Patient Preparation, Support & Radiation Safety
4.1 Patient Support, Disabilities, Ergonomics, and Transfer Techniques 4.2 Patient Instructions: Explaining the Exam, Dose, Duration, and Motion Control 4.3 Patient History, Contraindications, and Pregnancy Screening 4.4 Scan Preparation: Patient Data Entry, Artifact Removal, and Infection Control 4.5 Radiation Safety in DXA: ALARA, Distance, and Dose Levels
5 Chapter 5: DXA Physics & Instrumentation
5.1 Physical Principles of Dual-Energy X-ray Absorptiometry 5.2 Beam Geometries: Pencil Beam and Fan Beam Systems 5.3 Fan-Beam Geometry: Source–Object–Detector Distances, Magnification, and Centering 5.4 The Scan Analysis Algorithm: Soft Tissue, Bone Edge Detection, BMC, Area, and BMD
6 Chapter 6: Measuring & Reporting Bone Mineral Density
6.1 Statistics for Densitometry: Mean, Standard Deviation, and Coefficient of Variation 6.2 T-Scores, Z-Scores, and Reference Database Selection 6.3 Clinical Indications for BMD Testing and the FRAX Tool 6.4 Vertebral Fracture Assessment and Trabecular Bone Score 6.5 Whole Body DXA, Body Composition, and Pediatric Scanning
7 Chapter 7: Quality Control, Precision & Equipment Management
7.1 Accuracy, Precision, and the Sources of Precision Error 7.2 Daily Quality Control, Phantoms, and Control Charts 7.3 In Vivo Precision Studies and RMS-SD Calculation 7.4 Determining and Applying the Least Significant Change 7.5 Maintenance, Repair, Relocation, Cross-Calibration, and Software Upgrades
8 Chapter 8: DXA Informatics & Data Management
8.1 DXA Informatics, DICOM, PACS, and Data Management
9 Chapter 9: Lumbar Spine DXA
9.1 Lumbar Spine Anatomy, Landmarks, and Regions of Interest 9.2 Lumbar Spine Scan Acquisition and ROI Edge Placement 9.3 Lumbar Spine: Compensating for Body Habitus, Variant Anatomy, Pathology, and Low Bone Density 9.4 Lumbar Spine Positioning: The Leg Block, Centering, and Common Challenges 9.5 Lumbar Spine Problems: Poor Bone Edge Detection and Obscured Intervertebral Spaces 9.6 Lumbar Spine: Variant Anatomy, Fractures, Degenerative Pathology, Calcification, and Surgery 9.7 Lumbar Spine Analysis: Vertebral Exclusion and Reporting Rules 9.8 Lumbar Spine Follow-Up Scans and Rate of Change
10 Chapter 10: Proximal Femur DXA
10.1 Proximal Femur Anatomy, Landmarks, and Regions of Interest 10.2 Femur Selection: Right, Left, and Dual Femur Scanning 10.3 Proximal Femur Scan Parameters and Compensation 10.4 Proximal Femur Positioning: Neck Rotation, Shaft Placement, and Positioning Aids 10.5 Proximal Femur Problems: Ischium, Short Neck, Arthritis, Fracture, and Hardware 10.6 Femoral ROI Analysis: Femoral Neck, Total Hip, and Ward's Area 10.7 Proximal Femur Follow-Up Scans and Rate of Change
11 Chapter 11: Forearm DXA
11.1 Forearm Anatomy, Landmarks, and Adjacent Structures 11.2 Forearm Indications and Side Selection 11.3 Forearm Length Measurement, Scan Parameters, and Positioning 11.4 Forearm Regions of Interest: One-Third, Mid, Ultradistal, and Total 11.5 Forearm Problems: Edge Detection, Artifacts, Low Bone Density, and Joint Disease 11.6 Scan Analysis: Region Placement, Window Level, and Reported Values 11.7 Forearm Follow-Up Scans and Rate of Change
12 Chapter 12: Cross-Site Artifacts & Clinical Integration
12.1 Recognition, Impact, and Management of DXA Artifacts 12.2 Clinical Interpretation, Serial Comparison, and Reporting
4.4 Scan Preparation: Patient Data Entry, Artifact Removal, and Infection Control
Key Takeaways Explain This Topic AI
Demographic entry is not clerical: date of birth, sex, and ethnicity select the reference database used to compute T-scores and Z-scores, so an error changes the reported diagnosis without changing the measured BMD.
Height and weight must be measured at the visit rather than reported, because height loss is a clinical finding and weight drives both scan mode selection and FRAX.
Artifact-producing objects include zippers, snaps, buttons, underwire, belts, coins, jewelry, navel piercings, and metallic thread over the lower back, abdomen, and hips.
Non-removable artifacts such as surgical hardware, pacemakers, and umbilical piercings that cannot be removed must be documented rather than ignored, so the interpreting physician and future technologists know.
Table pads, leg blocks, hip positioners, and straps are disinfected between every patient with an EPA-registered product used for its full wet contact time.
Last updated: September 2026
4.4 Scan Preparation: Patient Data Entry, Artifact Removal, and Infection Control
Quick Answer: Three preparation tasks each have the power to invalidate an otherwise perfect acquisition. Data entry selects the reference database that converts BMD into a T-score and Z-score. Artifact removal determines whether the edge-detection algorithm sees bone or metal. Infection control protects a population that is old, often immunosuppressed by glucocorticoids, and frequently moving between healthcare facilities.
Patient Identification
Before anything else, confirm identity using two patient identifiers — typically full name and date of birth — against the requisition and the scanner worklist. Bone densitometry has an unusual exposure here: studies are compared across years, so a misidentified scan does not merely produce one wrong report. It contaminates a longitudinal record, and the error may not surface until a follow-up comparison produces an impossible change.
Verify that the order matches the patient and the sites . An order for an axial study does not authorize a forearm substitute; that requires contacting the ordering provider.
Entry of Patient Data
Every field entered before acquisition has a downstream consequence.
Field Why it matters
Date of birth Determines age, which selects the Z-score age-matched reference. An error shifts the Z-score without changing BMD
Sex Selects the reference database. ISCD recommends a uniform Caucasian (non-race-adjusted) female NHANES III reference for femoral neck and total hip T-scores in both sexes in the United States
Ethnicity Affects reference selection on some systems and some sites; enter what the facility protocol and the manufacturer's database require, consistently across serial studies
Height Measure it at the visit. Historical height loss is itself a clinical indicator of vertebral fracture and may trigger VFA
- Weight Drives acquisition mode selection, affects soft-tissue baseline, and is a FRAX input
- Referring provider, indication, prior study date Support coverage, reporting, and valid serial comparison
Two data-entry habits prevent most reference-database errors. First, measure height and weight rather than accepting reported values — patients systematically report their younger height, and the difference is diagnostically interesting. Second, check the demographics carried forward on a follow-up study ; an incorrect date of birth entered years ago silently propagates.
A demographic error is uniquely dangerous because the scan looks fine. BMD in g/cm² is a physical measurement and is unaffected. Only the derived scores move — and those are what the report says.
Removal of Artifact-Producing Objects
The scanner cannot distinguish a snap from a bone edge; it sees attenuation. Anything dense in the beam path is either included in the bone map or excluded as an artifact, and either way the measurement is wrong.
Change the patient into a gown unless clothing is verifiably free of metal over the scan field. It is faster than debating whether a particular garment qualifies.
Category Examples
Clothing hardware Zippers, snaps, buttons, rivets, grommets, hooks, metallic thread, sequins, reflective prints
- Undergarments Underwire bras, hook-and-eye closures, bodyshapers with metal or dense elastic panels
- Accessories Belts, belt buckles, keys, coins, phones, wallets in back pockets
- Jewelry Long necklaces that fall over the spine field, navel and abdominal piercings, body chains
Medical and personal Ostomy appliances with metal clips, incontinence products with metallic-backed liners, transdermal patches with metallic backing, some compression garment closures
Ingested or injected Recently taken calcium tablets, barium, iodinated contrast, radiopharmaceuticals
The scan field is wider than patients assume. For the lumbar spine it extends from roughly the iliac crest to above T12; for the femur it includes the pelvis and proximal shaft. A necklace resting on the sternum is out of field; the same necklace on a supine patient often is not.
Non-Removable Artifacts
Some artifacts cannot be removed, and the correct response is documentation, not silence :
Surgical hardware: spinal fusion rods, pedicle screws, interbody cages, hip arthroplasty components, plates and screws
Implanted devices: pacemakers and defibrillators, intrathecal pumps, spinal cord stimulators, IVC filters, aortic stents and endografts
- Radiopaque contraceptive devices in the pelvis
- Piercings that cannot be removed by the patient
- Vascular calcification and osteophytes — pathologic rather than foreign, but equally non-removable
For these, the responses are: exclude the affected vertebra or region where exclusion rules permit; select an alternative site where the order and protocol allow; record precisely what was present and where , so the interpreting physician can judge validity and the next technologist can reproduce the same analysis; and photograph or annotate the scan if the system supports it.
Documenting an artifact that was accommodated is what makes a follow-up study comparable. An unrecorded excluded vertebra is a future serial-comparison error.
Contraindication Confirmation
Before acquisition, confirm the screening answers obtained during history-taking: no recent barium study, no recent iodinated contrast, no recent nuclear medicine radiopharmaceutical, no oral calcium within the facility's specified interval, and pregnancy status resolved for patients of childbearing potential. These are pre-scan checks precisely because performing the study and discovering the contradiction afterwards wastes the appointment and the dose.
Infection Control
The DXA population skews old, and many patients are on glucocorticoids, biologic therapy, or chemotherapy. The scan table is a large shared horizontal surface that every patient contacts directly.
Between every patient, disinfect the table pad, leg elevation block, hip positioners and foot braces, forearm positioning tray, restraining straps, and any handrails or grab points.
Use an EPA-registered hospital disinfectant compatible with the manufacturer's materials — some products degrade table upholstery and radiolucent foam — and observe the full wet contact time stated on the label, commonly 1 to 3 minutes. A surface wiped and immediately dried has not been disinfected.
Perform hand hygiene before and after every patient contact, after glove removal, and after handling positioning accessories. Alcohol-based hand rub is appropriate unless hands are visibly soiled or the patient has a spore-forming organism such as C. difficile , in which case soap and water is required.
Change table paper or linens between patients.
Follow transmission-based precautions communicated by the referring unit. Patients on contact precautions may require dedicated equipment handling, gowning, and terminal cleaning after the study.
Handle sharps and body fluids per facility policy, and clean spills immediately using the facility's spill protocol rather than a routine wipe.
Keep the workstation clean too. Keyboard, mouse, and trackball are touched with the same hands that positioned the patient and are frequently the least-cleaned surfaces in the room.