9.4 Equipment Artifacts — Rings, Streaks, Tube Arcing & Cone Beam

ARRT Computed Tomography CT

Track your progress as you read

Finished sections get a check mark.

1 Introduction & Exam Overview

1.1 ARRT CT Exam Facts, Format & Scoring 1.2 How to Study & Test-Taking Strategy for the CT Exam

2 Patient Assessment, Communication & Safe Handling

2.1 Pre-Scan Assessment — Scheduling, History, Lab Values & Communication 2.2 Patient Education, Consent, Ergonomics & Positioning for the Scan 2.3 Patient Monitoring, Accessory Devices, Medications & Infection Control

3 Contrast Media & Injection Procedures

3.1 Contrast Media Types, Properties & Special Considerations 3.2 Venipuncture, Administration Routes & Power Injection Techniques 3.3 Contrast Timing, Reactions & Postprocedure Care

4 CT Radiation Physics

4.1 X-ray Production & Target Interactions 4.2 X-ray Beam Characteristics & Geometry 4.3 Photon Interactions with Matter & Tissue Attenuation

5 Radiation Protection & Dose Management

5.1 Minimizing Patient Dose — Technical Parameters & Modulation 5.2 Shielding & Patient-Specific Dose Considerations (Pediatric/Adult/Pregnancy) 5.3 Dose Measurements, Reporting & Personnel Protection

6 CT Scanner Components & Imaging Parameters

6.1 CT Unit Components — Gantry, Detectors & Data Acquisition System 6.2 Core Imaging Parameters — kVp, mAs, Pitch & Collimation 6.3 Slice Thickness, Scan Planes & Field of View

7 Data Acquisition Modes & Image Reconstruction

7.1 CT Acquisition Modes — Sequential, Helical & Volumetric Scanning 7.2 Cine/Shuttle Acquisition & Dual-Energy CT 7.3 Reconstruction Algorithms — Filtered Backprojection vs. Iterative Reconstruction 7.4 Reconstruction Parameters — Slice Thickness, Interval & Interpolation

8 Postprocessing, Image Display & Informatics

8.1 Postprocessing — MPR, 3D Rendering & Quantitative Analysis 8.2 Image Display Fundamentals — Pixels, Matrix & Field of View 8.3 Windowing & Measurement Tools — W/L, W/W & Region of Interest 8.4 CT Informatics — PACS, DICOM, Networking & Downtime Procedures

9 Image Quality & Artifacts

9.1 Image Quality Metrics — Resolution, Noise & Uniformity 9.2 CT Number, QA/QC & Accreditation 9.3 Patient-Related Artifacts — Beam Hardening, Motion, Metal & Edge Gradient 9.4 Equipment Artifacts — Rings, Streaks, Tube Arcing & Cone Beam

10 Head CT Procedures

10.1 Brain & Cranium CT — Routine and Perfusion Imaging 10.2 Temporal Bone, Orbit & Sinus CT 10.3 Maxillofacial, Mandible & TMJ CT

11 Spine & Musculoskeletal CT

11.1 Spine CT — Cervical, Thoracic & Lumbosacral Protocols 11.2 CT Myelography & Postmyelogram Imaging 11.3 Extremity CT — Upper & Lower Extremity, Postarthrogram 11.4 Shoulder, Bony Pelvis & Hip CT

12 Neck & Chest CT

12.1 Neck CT — Larynx/Airway & Soft Tissue 12.2 Cardiac CT — Coronary Angiography, Calcium Scoring, Gating & TAVR 12.3 Chest CT — Mediastinum, Airway, Chest Wall & Esophagram 12.4 HRCT, Lung Nodules & Low-Dose Lung Cancer Screening

13 Abdominal CT Procedures

13.1 Multiphase Liver, Pancreas & Adrenal CT 13.2 Renal CT — Multiphase Kidney, Urogram & Stone Protocol 13.3 GI Tract CT & Spleen Imaging

14 Pelvic CT Procedures

14.1 Pelvis CT — Bladder, Cystogram & Reproductive Organs 14.2 CT Colonography & Colorectal Imaging

15 Vascular, Interventional & Trauma CT

15.1 CT Angiography & Vascular Procedures — CTA, CTV, Runoff, Dissection & PE 15.2 CT-Guided Interventional Procedures — Biopsy, Drainage & Aspiration 15.3 CT Trauma Protocols & Surgical Planning

9.4 Equipment Artifacts — Rings, Streaks, Tube Arcing & Cone Beam

Key Takeaways Explain This Topic AI

A ring artifact is a complete circular pattern centered on the axis of rotation caused by a single miscalibrated detector element; it is fixed by detector recalibration, not by adjusting patient technique.

Equipment-related streak artifact reflects data inconsistency between detector channels or projection views and reproduces across different patients, distinguishing it from patient-caused streaking.

Tube arcing is a sudden electrical discharge inside the x-ray tube from aging, contamination, or wear, requiring service or tube replacement, not technique adjustment.

Cone beam artifact appears in wide-detector-array scanners where outer detector rows are struck at an oblique angle, most pronounced in peripheral slices of wide-volume acquisitions.

The key distinguishing test for any equipment artifact is reproducibility across different patients and studies on the same scanner — patient-caused artifacts do not repeat this way.

Last updated: July 2026

Why This Topic Matters

The final sub-item under Artifact Recognition and Reduction (2.C.7) is equipment artifacts , broken into four named types: rings, streaks, tube arcing, and cone beam . These differ fundamentally from the patient-related artifacts covered in the previous section because their cause lives inside the scanner hardware, not the patient's anatomy or position. The ARRT CT exam tests whether a technologist can tell the difference, because the correct response is different: patient-related artifacts are addressed with technique or positioning changes made by the technologist, while equipment artifacts require notifying service engineering or a medical physicist — repeating the scan with different patient technique will not fix a hardware problem.

Ring Artifact

A ring artifact is caused by a single detector element (or a small group of elements) that has drifted out of calibration or is malfunctioning. To understand why this produces a ring specifically, recall the geometry of a rotate-rotate (third-generation) CT system: the x-ray tube and detector array rotate together around the patient, and a given detector element samples the same radial distance from the isocenter at every projection angle throughout the full rotation. If that one element has a consistent measurement error, the reconstruction places that error at the same radial position on every projection — which reconstructs as a complete (or partial) circular ring, centered on the axis of rotation , rather than a random or localized artifact.

Appearance: one or more sharp, complete or near-complete circular/annular patterns, typically centered on the scan isocenter, and — critically — visible across images from multiple different patients scanned on that unit.

Correction: this is a hardware issue. It is resolved through detector recalibration (an air/reference calibration procedure performed by service or physics), not by changing the current patient's scan parameters. A technologist who recognizes a ring artifact should notify service/physics rather than attempt to "fix" it by rescanning with different technique.

Streak Artifact (Equipment-Related)

While streaking can come from many patient-related sources (motion, metal, beam hardening — covered in the previous section), equipment-related streak artifact specifically arises from data inconsistency between adjacent detector channels or projection views — for example, from detector malfunction, electronic noise, or a calibration error somewhere in the acquisition chain.

Appearance: linear streaking patterns that do not correlate with any specific patient anatomy or dense object, and that reproduce regardless of who or what is being scanned.

Correction: system calibration, detector service, or a software/firmware update, depending on the root cause identified by engineering. The key diagnostic clue for the technologist is reproducibility : a streak pattern that appears consistently across different patients and different studies on the same scanner points to equipment, not the patient in front of you. Reproducible artifact patterns should be documented and reported.

Tube Arcing

Tube arcing is a sudden, unwanted electrical discharge (an arc) occurring inside the x-ray tube. Common causes include tube aging, internal contamination or outgassing within the tube envelope, or wear of the filament or anode. Beyond the image artifact itself, tube arcing can also present clinically as an aborted exposure , an error message, or an irregular exposure — not just a visual artifact on the reconstructed image.

Appearance: severe, often transient streak artifacts coinciding with the discharge event.

Correction: tube arcing is an equipment malfunction requiring biomedical or service engineering evaluation, and may require tube reseasoning or outright tube replacement. It is not correctable through technologist technique adjustment. Recurrent arcing is both an image-quality and a safety concern significant enough that the scanner should be taken out of clinical service until evaluated.

Cone Beam Artifact

As multidetector CT (MDCT) scanners have expanded to wide-area detector arrays — modern systems range from roughly 64-detector-row up to 320-detector-row configurations, providing several centimeters of z-axis (longitudinal) coverage per single rotation — the x-ray beam geometry is no longer a simple flat, two-dimensional fan. Instead, it forms a three-dimensional cone . The detector rows located farthest from the central plane of that cone are struck by x-rays at an increasingly oblique angle relative to the patient's long axis, rather than perpendicular to it, which the simpler fan-beam reconstruction model does not fully account for.

Appearance: shading, streaking, or generally reduced image quality that is most pronounced in the outer/peripheral detector rows — that is, the edges of a wide-volume acquisition — and that becomes more significant as the total number of detector rows and z-axis coverage increases.

Correction: dedicated cone-beam reconstruction algorithms (such as Feldkamp-type or iterative cone-beam correction methods) built into modern reconstruction software specifically to account for the oblique ray geometry, rather than assuming a simplified fan-beam model.

Equipment Artifact Mechanical Cause Appearance Correction

Ring Miscalibrated/malfunctioning detector element Complete circular ring centered on isocenter, across multiple patients Detector recalibration by service/physics

Streak (equipment) Data inconsistency between detector channels Non-anatomic linear streaking, reproducible across patients System calibration, detector/firmware service

Tube arcing Electrical discharge from tube aging/contamination/wear Severe transient streaks; may accompany aborted exposure Biomedical/service evaluation, tube reseasoning or replacement

Cone beam Oblique ray angle at outer detector rows in wide-array scanners Shading/streaking in peripheral slices of wide-volume acquisitions Dedicated cone-beam reconstruction algorithms

Exam Scenario

A technologist performs a wide-volume 320-detector-row cardiac CT and notices a subtle shading artifact confined to the outermost slices of the acquired volume, while the central slices are essentially artifact-free. This pattern — worse at the periphery of a wide-coverage acquisition, absent centrally — is characteristic of a cone beam artifact tied to the oblique ray geometry striking the outer detector rows, and is addressed through the scanner's cone-beam reconstruction algorithm rather than by repeating the scan with different mAs.

Contrast that with a second scenario: a technologist notices the same complete circular artifact appearing on head CT images from several different patients scanned on the same unit over the course of a week. Because the pattern is identical regardless of which patient is scanned, this is a ring artifact from a miscalibrated detector element, and the correct response is to report it to service/physics for detector recalibration — not to keep scanning additional patients on the assumption the artifact is patient-related.