9.3 Patient-Related Artifacts — Beam Hardening, Motion, Metal & Edge Gradient
ARRT Computed Tomography CT
Track your progress as you read
Finished sections get a check mark.
1 Introduction & Exam Overview
1.1 ARRT CT Exam Facts, Format & Scoring 1.2 How to Study & Test-Taking Strategy for the CT Exam
2 Patient Assessment, Communication & Safe Handling
2.1 Pre-Scan Assessment — Scheduling, History, Lab Values & Communication 2.2 Patient Education, Consent, Ergonomics & Positioning for the Scan 2.3 Patient Monitoring, Accessory Devices, Medications & Infection Control
3 Contrast Media & Injection Procedures
3.1 Contrast Media Types, Properties & Special Considerations 3.2 Venipuncture, Administration Routes & Power Injection Techniques 3.3 Contrast Timing, Reactions & Postprocedure Care
4 CT Radiation Physics
4.1 X-ray Production & Target Interactions 4.2 X-ray Beam Characteristics & Geometry 4.3 Photon Interactions with Matter & Tissue Attenuation
5 Radiation Protection & Dose Management
5.1 Minimizing Patient Dose — Technical Parameters & Modulation 5.2 Shielding & Patient-Specific Dose Considerations (Pediatric/Adult/Pregnancy) 5.3 Dose Measurements, Reporting & Personnel Protection
6 CT Scanner Components & Imaging Parameters
6.1 CT Unit Components — Gantry, Detectors & Data Acquisition System 6.2 Core Imaging Parameters — kVp, mAs, Pitch & Collimation 6.3 Slice Thickness, Scan Planes & Field of View
7 Data Acquisition Modes & Image Reconstruction
7.1 CT Acquisition Modes — Sequential, Helical & Volumetric Scanning 7.2 Cine/Shuttle Acquisition & Dual-Energy CT 7.3 Reconstruction Algorithms — Filtered Backprojection vs. Iterative Reconstruction 7.4 Reconstruction Parameters — Slice Thickness, Interval & Interpolation
8 Postprocessing, Image Display & Informatics
8.1 Postprocessing — MPR, 3D Rendering & Quantitative Analysis 8.2 Image Display Fundamentals — Pixels, Matrix & Field of View 8.3 Windowing & Measurement Tools — W/L, W/W & Region of Interest 8.4 CT Informatics — PACS, DICOM, Networking & Downtime Procedures
9 Image Quality & Artifacts
9.1 Image Quality Metrics — Resolution, Noise & Uniformity 9.2 CT Number, QA/QC & Accreditation 9.3 Patient-Related Artifacts — Beam Hardening, Motion, Metal & Edge Gradient 9.4 Equipment Artifacts — Rings, Streaks, Tube Arcing & Cone Beam
10 Head CT Procedures
10.1 Brain & Cranium CT — Routine and Perfusion Imaging 10.2 Temporal Bone, Orbit & Sinus CT 10.3 Maxillofacial, Mandible & TMJ CT
11 Spine & Musculoskeletal CT
11.1 Spine CT — Cervical, Thoracic & Lumbosacral Protocols 11.2 CT Myelography & Postmyelogram Imaging 11.3 Extremity CT — Upper & Lower Extremity, Postarthrogram 11.4 Shoulder, Bony Pelvis & Hip CT
12 Neck & Chest CT
12.1 Neck CT — Larynx/Airway & Soft Tissue 12.2 Cardiac CT — Coronary Angiography, Calcium Scoring, Gating & TAVR 12.3 Chest CT — Mediastinum, Airway, Chest Wall & Esophagram 12.4 HRCT, Lung Nodules & Low-Dose Lung Cancer Screening
13 Abdominal CT Procedures
13.1 Multiphase Liver, Pancreas & Adrenal CT 13.2 Renal CT — Multiphase Kidney, Urogram & Stone Protocol 13.3 GI Tract CT & Spleen Imaging
14 Pelvic CT Procedures
14.1 Pelvis CT — Bladder, Cystogram & Reproductive Organs 14.2 CT Colonography & Colorectal Imaging
15 Vascular, Interventional & Trauma CT
15.1 CT Angiography & Vascular Procedures — CTA, CTV, Runoff, Dissection & PE 15.2 CT-Guided Interventional Procedures — Biopsy, Drainage & Aspiration 15.3 CT Trauma Protocols & Surgical Planning
9.3 Patient-Related Artifacts — Beam Hardening, Motion, Metal & Edge Gradient
Key Takeaways Explain This Topic AI
Beam hardening occurs because lower-energy photons are absorbed preferentially, producing dark interpetrous streaks or a 'cupping' pattern in large uniform structures.
Partial volume averaging occurs when a voxel contains a mixture of tissue types, most problematic with thick slices, small structures, or oblique orientation.
Metallic artifact combines severe photon starvation with beam hardening; metal artifact reduction (MAR) algorithms and dual-energy virtual monochromatic imaging reduce but do not fully eliminate it.
Motion artifact is corrected with faster acquisition, breath-hold instructions, and ECG gating for cardiac motion — not by adjusting kVp or mAs, which affect noise, not motion.
Out-of-field (truncation) artifact appears when part of the patient extends beyond the scan field of view and is fixed by proper positioning or extended-FOV reconstruction, not by treating it as a scanner malfunction.
Last updated: July 2026
Why This Topic Matters
Artifact Recognition and Reduction (2.C) is one of the most heavily tested lettered items inside the Image Evaluation and Archiving subcategory, and artifact-identification scenarios are a hallmark ARRT CT question format: a description of an image appearance is given, and the candidate must name the artifact, its cause, and its fix. This section covers the six sub-items that originate from the patient — the anatomy or positioning being scanned — as distinct from the equipment-based artifacts covered in the next section. Recognizing which category an artifact belongs to matters clinically too: patient-related artifacts are addressed with technique or positioning changes, while equipment artifacts require service intervention.
Beam Hardening (or Cupping)
A clinical CT x-ray beam is polychromatic — it contains a spectrum of photon energies, not a single energy. As this beam passes through tissue, lower-energy photons are absorbed preferentially (the photoelectric effect is strongly energy-dependent), so the beam's average energy increases, or "hardens," the farther it travels through the patient. This produces two classic appearances:
Dark streaks or bands between two dense structures — the textbook example is a dark band between the petrous bones on a posterior fossa head CT (sometimes called an interpetrous or Hounsfield artifact).
Cupping artifact in a large, relatively homogeneous structure (a full-size body phantom, or a large patient's abdomen), where the periphery appears artifactually brighter/higher in HU than the center, because central rays traverse more attenuating tissue and harden more.
Correction: software beam-hardening correction algorithms applied at reconstruction, physical prefiltration of the beam (flattening the energy spectrum before it reaches the patient), higher kVp (reduces the relative severity), and dual-energy CT, which can generate monochromatic images that avoid the beam-hardening effect entirely.
Partial Volume Averaging
Partial volume averaging occurs when a single voxel contains a mixture of two or more tissue types with different attenuation coefficients. The reconstructed CT number for that voxel is a weighted average of the tissues inside it — not an accurate representation of either one. It is most problematic when:
Slices are thick relative to the structure being imaged.
The structure of interest is small .
The structure runs obliquely to the scan plane, increasing the length of overlap within a voxel.
A classic clinical trap: a small, truly low-density renal cyst adjacent to the collecting system can appear falsely denser on a thick-slice scan because the voxel averages in higher-density adjacent tissue — potentially masking a stone or a lesion that thinner slices would have revealed clearly. Correction: thinner acquisition and reconstruction slices, isotropic voxel reconstruction, and orienting the scan plane perpendicular to the structure of interest when clinically feasible.
Motion Artifact
Motion artifact results from any patient movement during data acquisition — voluntary (shifting position, coughing, talking) or involuntary (respiratory motion, cardiac pulsation, peristalsis, vascular pulsation). It appears as blurring, streaking, or a double-edge (ghosting) pattern , most visible at high-contrast interfaces such as the diaphragm or aorta. Correction: shorter scan time (faster gantry rotation, wider detector coverage per rotation), clear breath-hold instructions and patient communication, immobilization devices, and — specifically for cardiac motion — ECG gating (prospective or retrospective), which synchronizes acquisition or reconstruction to a defined phase of the cardiac cycle. Adjusting kVp or mAs does not correct motion; those parameters govern noise and dose, not temporal blur.
Metallic Artifact
High-density, high-atomic-number objects — dental amalgam, hip or joint prostheses, surgical clips, pacemaker leads, spinal hardware — cause severe photon starvation (nearly all photons absorbed or scattered before reaching the detector) combined with intense beam hardening. The result is bright and dark streaks radiating outward from the metal, frequently obscuring adjacent anatomy. Correction: dedicated metal artifact reduction (MAR) algorithms (iterative or projection-interpolation based), dual-energy virtual monochromatic imaging at higher keV levels, adjusted technique (higher kVp/mAs when clinically appropriate), and thinner slices. MAR reduces but does not fully eliminate metallic artifact — the exam expects "reduces," not "eliminates," as the correct framing.
Edge Gradient Artifact
Edge gradient artifact arises from undersampling at sharp, high-contrast interfaces — for example, bone-air or bone-soft tissue boundaries at the skull base or petrous ridge — where the system's spatial and contrast resolution cannot fully resolve the abrupt attenuation change. It appears as fine streak lines emanating from the high-contrast edge. Correction: thinner slices, higher-resolution reconstruction algorithms, and adequate sampling (an appropriately small pitch where clinically warranted).
Patient Positioning (Out-of-Field) / Truncation Artifact
When part of the patient's body extends beyond the scan field of view (SFOV) — common with large or bariatric patients, or when a patient's arms remain at their sides during an abdominal/pelvic scan — the reconstruction algorithm's assumption of complete data within the field is violated. Missing peripheral data produces a bright rim or streak artifact at the image edge (truncation artifact) , and can also distort CT numbers even in the clinically important central anatomy. Correction: proper patient positioning and centering within the SFOV before scanning (for example, raising the arms above the head for abdominal/pelvic CT whenever possible), selecting an appropriately large SFOV, or applying extended-FOV/truncation-correction reconstruction algorithms when repositioning is not possible.
Artifact Cause Appearance Correction
Beam hardening/cupping Preferential absorption of low-energy photons Dark interpetrous streaks; peripheral-brighter cupping Correction software, prefiltration, higher kVp, dual-energy
Partial volume averaging Voxel mixes multiple tissue types Inaccurate averaged CT number Thinner slices, perpendicular scan plane
Motion Voluntary/involuntary patient movement Blurring, streaking, ghosting Faster acquisition, breath-hold, ECG gating
Metallic Photon starvation + beam hardening from dense metal Bright/dark streaks radiating from metal MAR algorithms, dual-energy, adjusted technique
Edge gradient Undersampling at sharp high-contrast interfaces Fine streaks at bone-air/bone-tissue edges Thinner slices, higher-resolution reconstruction
Out-of-field/truncation Patient extends beyond SFOV Bright rim/streak at image periphery Proper positioning, larger SFOV, extended-FOV algorithm
Exam Scenario
A bariatric patient is scanned for an abdominal/pelvic CT but cannot raise both arms above the head due to shoulder mobility limitations, so the arms remain at the patient's sides within the gantry. The resulting image shows streak artifact concentrated at the lateral image periphery, while the central abdominal anatomy is comparatively less affected. The technologist should recognize this as an out-of-field (truncation) artifact related to the patient's arms extending beyond the SFOV — not a scanner malfunction — and should consider selecting a larger SFOV or an extended-FOV reconstruction algorithm rather than assuming the equipment needs service.