Study guide
Patient Identification, Gowning & Metal Removal
ARRT Imaging Assistant Magnetic Resonance I.A.(MR
Track your progress as you read
Finished sections get a check mark.
1 Chapter 1: Patient Care — Ethics, Legal Standards & Professional Practice
ARRT Standards of Ethics & Professional Scope Legal Concepts, Patient Rights & Confidentiality Consent, Advance Directives & HIPAA Patient Communication, Education & Anxiety Management
2 Chapter 2: Patient Care — Preparation, Transfer & Vital Signs
Patient Identification, Gowning & Metal Removal Safe Patient Handling, Ergonomics & Transfers Lateral Transfers, Positioning Aids & Fall Prevention Vital Signs & Physiological Monitoring Medical Equipment, Sedation Monitoring & Physiological Gating
3 Chapter 3: Patient Care — Infection Control & Medical Emergencies
CDC Standard Precautions & Hand Hygiene PPE, Sharps & Exposure Prevention Transmission-Based Precautions & Environmental Disinfection Contrast Media Fundamentals & Patient Safety Administration Routes, Venipuncture & Injector Preparation Acute Medical Emergencies & Adverse Contrast Reactions
4 Chapter 4: MR Safety — Environment, Personnel & Access Control
The Four MR Safety Zones & Facility Architecture Shielding & Facility Engineering Controls MR Personnel Classifications & Supervisory Roles MR Safety Leadership & Assistant Responsibilities Ferromagnetic Screening Workflows & Detection Systems Ferromagnetic Detection & Its Limitations
5 Chapter 5: MR Safety — Implants, Devices & Bioeffects
ASTM F2503 Medical Device Labeling & Compatibility Interpreting MR Conditional Parameters Screening High-Risk Implants & Foreign Bodies Static Magnetic Field & Mechanical Hazards RF Fields, Gradient Fields & Patient Bioeffects Switched Gradients, Hearing Protection & Patient Symptoms
6 Chapter 6: MR Safety — Emergencies, Cryogens & Quality Culture
Cryogen Hazards & Magnet Quench Protocols Zone IV Emergency Response & Fire Safety Safety Event Reporting, Root-Cause Analysis & Quality Control Mobile, Point-of-Care & Remote MR Safety
7 Chapter 7: Positioning & Landmarking — Neurological Imaging
Brain MRI Anatomy, Landmarking & Positioning Brain Coil Setup, Landmarking & Imaging Planes Internal Auditory Canals (IAC), Pituitary & Orbit MRI Cervical Spine MRI Anatomy & Positioning Thoracic & Lumbar/Sacral Spine MRI Positioning TMJ & Soft-Tissue Neck Positioning
8 Chapter 8: Positioning & Landmarking — Body & Torso Imaging
Breast MRI Anatomy, Positioning & Coil Setup Chest, Thoracic & Brachial Plexus MRI Abdomen, Liver & Pancreas MRI Pelvis, MRCP & Specialized Enterography Studies
9 Chapter 9: Positioning & Landmarking — Upper Extremity MSK
Shoulder MRI Anatomy & Positioning Elbow & Forearm MRI Positioning Wrist & Hand MRI Positioning
10 Chapter 10: Positioning & Landmarking — Lower Extremity MSK
Hip & Pelvic Girdle MSK Positioning Hip and Pelvis: Coil Coverage & Supported Setup Knee MRI Anatomy & Positioning Knee Coil Selection, Position & Landmarking Ankle, Foot & Toe Positioning Upper & Lower Long-Bone Positioning
11 Chapter 11: Clinical Practice Integration & Exam Strategy
ARRT I.A.(MR) Exam Blueprint & Safety Passing Threshold Clinical Competency Tracking & Verification Protocols Clinical Information Systems, Handoffs & Downtime
Patient Identification, Gowning & Metal Removal
Key Takeaways Explain This Topic AI
Use two approved patient identifiers; room location and appointment time are not identifiers.
Ask patients to state identifiers when able and reconcile the order, wristband and records.
Facility-provided MR attire reduces hidden conductive-fabric and pocket-item risks.
Check the exact patch or wearable-device labeling; medication interruption and replacement need clinical authorization.
Last updated: October 2026
Patient Identification, Gowning & Metal Removal
Important
Patient safety in magnetic resonance imaging begins long before the scanner table moves into the bore. The Imaging Assistant serves as the frontline clinical barrier against catastrophic projectile incidents, radiofrequency burns, and misidentified diagnostic exams. Rigorous adherence to two-identifier verification and complete metal elimination protocols is mandatory for every patient.
The Joint Commission Two-Identifier Mandate & Active Verification
The Joint Commission National Patient Safety Goals (NPSG) establish that healthcare practitioners must use at least two independent patient identifiers prior to administering care, performing diagnostic imaging, or transferring patients. In the MRI department, misidentifying a patient carries dual catastrophic risks: performing an examination on the incorrect anatomical region or wrong individual, and exposing an unscreened patient with contraindicated ferromagnetic implants to the static magnetic field ( B 0 B_0 B 0 ).
Acceptable vs. Prohibited Identifiers
Standard acceptable identifiers include:
- Full legal name (first, middle, and last name as recorded in the legal medical record)
- Date of birth (DOB) (month, day, and year)
- Medical record number (MRN) (unique institutional electronic health record identifier)
- Assigned barcode wristband identification
Warning
Never use the patient's room number, bed assignment, physical location in the waiting room, or scheduled appointment time as an identifier. Physical locations are fluid and easily swapped during hospital patient transfers, making them a frequent root cause of sentinel identification errors.
Active Communication Methodology
Patient identification must always be conducted using active verbal questioning . An Imaging Assistant must never use passive questioning, such as asking, "Are you Mr. Robert Jones?" or "Is your birthday July 14th?"
Patients who are anxious, sedated, hearing-impaired, or cognitively compromised frequently nod or answer "yes" to any authoritative inquiry. Instead, the assistant must directly state:
"Please state your full legal name and date of birth for me."
While the patient speaks, the assistant must simultaneously cross-reference the spoken response against two tangible sources:
The patient's identification wristband securely fastened to their wrist or ankle.
The electronic health record (EHR) order requisition and MRI safety screening worksheet.
For non-verbal, unconscious, pediatric, or confused patients, identity must be verified by reconciling the patient wristband with an accompanying parent, legal guardian, or designated transport nurse who can explicitly confirm the individual's identity.
Gowning Standards & Invisible Hazards in Zone IV
Under the facility's gowning policy, patients change out of unverified personal clothing into facility-approved and laundered hospital attire, typically 100% cotton or poly-cotton scrubs with fabric ties or non-metallic plastic fasteners.
The Lethal Fallacy of "Safe" Street Clothes
Patients frequently object to gowning, claiming their street clothes contain no metal because they lack external metallic zippers, belts, or buttons. However, street clothes harbor critical hidden hazards:
Antimicrobial and Anti-Odor Athletic Apparel: High-performance modern sportswear (such as athletic shirts, yoga leggings, compression shorts, and running socks) frequently integrates microscopic filaments of silver ( A g + Ag^+ A g + ) or copper . These metallic nanoparticles are woven directly into the synthetic fabric matrix to destroy odor-causing bacteria. Under the intense oscillating radiofrequency (RF) transmit fields of the MRI scanner, these metallic microfibers act as electrical conductors. Induced eddy currents circulate through the fabric, generating rapid resistive heating that produces devastating second- and third-degree skin burns.
Invisible Fasteners: Hidden brass grommets, concealed magnetic snaps, metallic underwires in brassieres, metallic thread embroidery, and metallic foil branding logos heat rapidly and cause skin contact burns.
Ferromagnetic Rivets and Zippers: Denim jeans contain heavy steel or brass rivets that distort the local magnetic field and produce severe susceptibility artifacts, in addition to carrying rotational torque risks.
Caution
Underwear policy must be strictly enforced. Underwire bras must be removed without exception due to RF heating and severe local magnetic distortion. Patients must wear facility-provided disposable undergarments or go without undergarments if their personal clothing contains elastane-metallic blends.
External Metal Screening & Head-to-Toe Removal Protocol
Once the patient is gowned, the Imaging Assistant conducts a meticulous physical inventory to strip all external ferromagnetic and conductive items. Ferromagnetic items become lethal airborne projectiles when accelerated by the static magnetic field gradient, while conductive non-ferromagnetic items create RF burn risks and obscure diagnostic anatomy.
Anatomical Region / Item Category Specific Items to Remove Primary Mechanism of Hazard Required Clinical Action
Cranial & Facial Items Eyeglasses, bobby pins, barrettes, wig clips, metallic hair weaves Projectile hazard from steel clips; massive susceptibility artifact on brain and spine MRI Complete removal; inspect scalp for hidden metal hairpins; place in locker
Auditory & Dental Hearing aids, cochlear sound processors, removable dentures, partial plates with metal clasps Hearing aid internal electronics destroyed by B 0 B_0 B 0 ; dislodgement of dental appliances; severe local image distortion Remove hearing aids before Zone III; store dentures in moist, labeled container
Torso & Body Adornments Body piercings (navel, nipple, dermal anchors), wire undergarments, metallic necklaces RF inductive loop burns; RF antenna effect; translational pulling forces Remove all body jewelry; if piercing is non-removable, consult technologist for MR Conditional material verification
Extremities & Electronics Smartwatches, fitness trackers, rings, bracelets, cell phones, pagers Destruction of electronic microchips; battery thermal runaway; RF skin burns Remove all wristwear; notify the technologist if a ring cannot be removed; do not improvise cutting or clearance
Pockets & Wallets Keys, pocketknives, coins, credit cards, magnetic transit passes, lighters Lethal projectile hazard; static field degausses and erases magnetic strips on cards Meticulous physical check; empty all pockets into locker before entering Zone III
Transdermal Medication Patches & Radiofrequency Heating
Transdermal medication patches deliver continuous systemic drug dosages through the dermis. Some transdermal patch formulations incorporate an aluminized metallic backing or thin metallic foil layer within their laminated structure. This metallic foil acts as an impermeable moisture barrier to maintain steady drug diffusion.
The Burn Mechanism
During an MRI pulse sequence, the scanner's RF body coil transmits high-frequency electromagnetic pulses (typically 64 MHz at 1.5 Tesla, 128 MHz at 3.0 Tesla). When these RF waves encounter the conductive metallic foil of a transdermal patch, they induce circulating electrical eddy currents. Because the metallic layer is microscopic in thickness, its electrical resistance is high, causing immediate and intense resistive heating. The heat is trapped against the patient's skin, resulting in full-thickness contact burns and blister formation.
Product-Specific Patch Review
Patch composition varies by manufacturer and formulation. Identify the medication, brand, strength and application time, then check the current labeling. A drug name alone cannot establish whether its patch contains metal. Some patches require removal before MRI to avoid burns; the treating team must arrange interruption, disposal and replacement without changing the dose inadvertently.
Clinical Management Protocol for Patches
Visual Skin Inspection: During the gowning process, the Imaging Assistant must visually inspect the patient's upper arms, back, chest, and flanks for transdermal patches.
Technologist and Nurse Notification: The assistant must never independently peel off or discard a prescription medication patch without explicit authorization from the supervising MRI technologist and the patient's registered nurse or physician.
Medication disposal: Follow the product labeling and institutional pharmaceutical-waste process. Do not independently flush medication or discard a controlled-drug patch into ordinary trash. Document medication changes through the authorized clinician.
Post-Exam Replacement Plan: Ensure that the nursing team is alerted so a fresh patch can be applied immediately following the completion of the imaging exam.
Patient Belongings & Zone II Locker Management
All removed personal belongings must be secured in designated patient lockers located strictly within Zone II , outside the controlled-access doors of Zone III.
Lockers must be secured with physical brass keys, magnetic keycards, or digital electronic keypads. Under no circumstances should locker keys made of ferromagnetic steel be handed to the patient to carry into Zone IV. Belongings management forms must record the locker number, ensuring patient possessions remain protected and preventing unauthorized metal from crossing the Zone III boundary.