Study guide

13.1 Craniocaudal (CC) Positioning

ARRT Mammography Postprimary

Track your progress as you read

Finished sections get a check mark.

1 Introduction & Exam Overview

1.1 Current Exam Facts 1.2 Postprimary Eligibility, Structured Education, and Clinical Requirements 1.3 Scoring, Retakes, Fees, and Credential Maintenance 1.4 2025 Content Outline, Domain Weights, and Study Strategy

2 Patient Communication

2.1 Pre-Exam Instructions 2.2 Explaining the Exam, Rapport, and Support 2.3 Patient Education: ACS/ACR Guidelines, BSE, CBE, Dose, and Modalities

3 Results Communication and Breast Cancer Risk

3.1 Process for Receiving Results and Additional Imaging 3.2 Epidemiology and Inherent Risk Factors 3.3 Social Risk Factors, Access, and Hormone Use 3.4 Signs and Symptoms of Breast Disease

4 History, Priors, and Treatment Context

4.1 Medical History, Physical Findings, and Gender Affirmation 4.2 Previous Mammograms 4.3 Surgical, Nonsurgical, and Reconstruction Options

5 Mammography Unit Design

5.1 kVp Range, Tube, Filtration, Window, and Focal Spot 5.2 Compression Paddles, Grids, and System Geometry

6 Digital Acquisition, Display, and Informatics

6.1 FFDM-DR/2D, DBT/3D, and Synthesized Imaging 6.2 Image Receptors and Acquisition vs Interpretation Monitors 6.3 HIS/RIS/EMR, PACS, and CAD

7 Accreditation, MQSA, EQUIP, and Density Reporting

7.1 Accreditation Bodies, FDA/MQSA Personnel, Records, Audits, and Policies 7.2 EQUIP: QA Oversight, Clinical Image Quality, and Corrective Action 7.3 Breast Density Reporting Requirements

8 Quality Control Testing

8.1 Mammographer QC: Phantom Image, Compression Thickness, Visual Checklist 8.2 Mammographer QC: Monitors, Repeat Analysis, Viewing Conditions, Compression Force, Detector Calibration 8.3 Physicist Tests: MEE, Collimation, Spatial Resolution/MTF, SNR/CNR, AEC, AGD, kVp, HVL 8.4 Physicist Tests: Artifacts, Phantom Quality, Illuminance, Tech QC Program Review, Compression Application/Paddle Alignment, Workstation QC

9 Technique and Image Evaluation

9.1 Technical Factors: kVp, mAs, AEC vs Manual, Target/Filter, Focal Spot, Grid, Mag, Labeling 9.2 EQUIP Image-Quality Criteria 9.3 Patient-Related Artifacts

10 Localization and Anatomy

10.1 Clock Position, Quadrants, and Triangulation 10.2 External Anatomy 10.3 Internal Anatomy: Fascia, Glandular Tissue, and the TDLU 10.4 Cooper Ligaments, Pectoral Muscle, Vascular/Lymphatic Systems, and Cytology

11 BI-RADS Lexicon

11.1 Architectural Distortion, Asymmetry, and Mass Features 11.2 Calcification Morphology and Distribution 11.3 Assessment Categories, Breast Density, and Recommendations

12 Breast Pathology

12.1 Benign Pathology and Mammographic Appearance 12.2 Lesions with Upgrade Potential 12.3 DCIS, IDC, and Invasive Lobular Carcinoma 12.4 Inflammatory Carcinoma, Paget Disease, Sarcoma, Lymphoma, and Metastases

13 Standard Screening Views

13.1 Craniocaudal (CC) Positioning 13.2 Mediolateral Oblique (MLO) Positioning

14 Additional Views I

14.1 Mediolateral (ML) and Lateromedial (LM) 14.2 XCCL, XCCM, Cleavage (CV), and Axillary Tail (AT) 14.3 Tangential (TAN) and Rolled Views (RL, RM, RS, RI)

15 Additional Views II

15.1 Implant Displaced (ID), Nipple in Profile, and Anterior Compression 15.2 Spot Compression and Magnification

16 Patient Variance

16.1 Chest Wall Variations, Kyphosis/Lordosis, Protruding Abdomen, and Extreme Breast Size 16.2 Irradiated Breast, Reduction Mammoplasty, and Postsurgical Breast 16.3 Male Patients and Lactating Breast 16.4 Breast Augmentation and Implanted Devices

17 Imaging Examinations

17.1 Screening vs Diagnostic Mammography 17.2 Breast Ultrasound and Breast MRI

18 Interventional Procedures

18.1 History, Consent, Time-Out, Setup, and Postprocedure Instructions 18.2 Stereotactic and Ultrasound-Guided Biopsy 18.3 Needle Localization, Specimen Imaging, and Clip Placement 18.4 Handling and Disposing of Biohazardous Materials

13.1 Craniocaudal (CC) Positioning

Key Takeaways Explain This Topic AI

Set the C-arm at 0° and raise the detector to the elevated inframammary fold; seeing the IMF on the finished CC image is not a CC criterion.

Posteromedial cleavage/sternal tissue is the CC non-negotiable, because that tissue is the tissue the MLO most often excludes.

ACR 1999 and the ACR Learning Network require the CC posterior nipple line (PNL) to fall within 1 cm of the same-side MLO PNL.

The nipple must be in profile on at least one view of each breast; pectoralis appears on only about 30–40% of correctly positioned CC images and is not required.

Lateral exaggeration that throws the nipple off midline and drops medial tissue fails the CC; add an XCCL instead of ruining the standard view.

Last updated: August 2026

Why the CC is a screening essential

A screening mammogram is four images: right and left craniocaudal (CC) plus right and left mediolateral oblique (MLO) . ARRT Procedures 2.A Views tests those projections as a pair, not as two interchangeable pictures of the same tissue. The CC is the superior-to-inferior projection: the C-arm sits at 0° , the image receptor (IR) is horizontal under the breast, and compression is applied from above. Its clinical job is to show posteromedial tissue the MLO is anatomically poor at including, and to give a second plane so a finding can be placed medial or lateral to the nipple.

The positioning standard ARRT still uses is the Clinical Image Evaluation chapter of the ACR 1999 Mammography Quality Control Manual . ACR accreditation still scores positioning from that chapter (the first of eight clinical-image parameters). The ACR Learning Network Mammography Positioning Improvement Collaborative later converted the same ideas into a meets / does not meet table for quality improvement — not the accreditation score sheet, but the language departments use to fail an image.

  • Set the C-arm, then set height from a lifted fold
  • Detector height vs seeing the IMF

Rotate the C-arm to 0° before the patient steps in so the detector is horizontal and the beam is vertical. Height is not a remembered number on the column. Lift the breast and elevate the inframammary fold (IMF) with your hand, then bring the detector up to that raised crease. If the detector is too low, the breast slumps, posterior tissue falls off the chest-wall edge, and the nipple points toward the floor. If the detector is too high, the IMF is over-elevated and you lose inferior and posterior tissue — a classic items-style trap.

Hold two facts at once:

You use the elevated IMF as a height landmark.

You do not need to see the IMF on the finished CC. An open fold is an MLO criterion. Chasing a visible IMF on the CC usually means the detector is in the wrong place or you have turned the patient into a mock-oblique.

Stance, two-handed lift, and the dropped shoulder

The patient stands slightly off the unit, both feet facing the machine, weight even, hips back, torso leaning in. Shoulders are relaxed — relaxed is not dropped. A dropped ipsilateral shoulder drags posterior tissue off the detector and can project the shoulder, chin, or upper arm over the breast. ACR Learning Network treats "other body parts projected over breast" (contralateral breast, hand, shoulder, chin) as a major fail with no partial credit.

Stand on the medial side of the breast you are imaging. Turn the patient's head toward you (away from the side being examined) so the mandible and ipsilateral shoulder clear the field. With both hands, pull the breast up and away from the chest wall onto the detector and anchor it. Do not let go until the paddle has fully compressed. The same early release that creates camel-nose sagging on the MLO also lets posterior CC tissue sneak back toward the chest wall.

The nipple should point straight at the back center of the detector. That alignment is how you know the breast is not rolled and not exaggerated.

  • Medial (cleavage / sternal) tissue is the CC's non-negotiable
  • Cleavage and sternal contact

The medial breast is relatively fixed at the sternum. It cannot be stretched the way lateral tissue can, so on the MLO it is the tissue most likely to be excluded. Therefore the CC must include posteromedial, cleavage, and sternal-contact tissue. Drape the contralateral breast over the corner of the detector so the sternum meets the IR. Skin-to-skin cleavage on the image is the proof you reached midline. If medial tissue is missing on the CC, it is often missing from the entire screening study.

Do not trade medial tissue for a photogenic lateral margin. If far-posterior lateral glandular tissue is still off the detector after a correctly centered CC, add a dedicated exaggerated craniocaudal lateral (XCCL) . Do not rotate a "standard" CC until the nipple sits in the medial third of the image.

Lateral tissue without excessive exaggeration

Once medial contact is locked, pull lateral posterior tissue onto the detector as the paddle comes down. Keep the paddle flush with the chest wall so retroglandular fat is not left behind. ACR Learning Network scores CC exaggeration as a minor criterion: nipple midline with no exaggeration meets ; a slight medial or lateral offset with no posterior tissue lost partially meets ; nipple off midline and lost medial or lateral posterior tissue does not meet .

Pectoralis on the CC is a bonus, not a requirement. The 1999 manual notes that even with correct positioning the pec appears on only about 30–40% of CC images (the Learning Network rounds this to about 30%). Chasing pec on every CC is how technologists rotate laterally, lose cleavage, and still fail the posterior nipple line. Posterior tissue on the CC is judged by the retroglandular fat plane and by PNL, not by whether a strip of muscle happened to come along.

  • Posterior nipple line: the 1 cm rule
  • How to measure PNL

The posterior nipple line (PNL) is the only numeric check that the CC included comparable depth to the MLO. Measure from the base of the nipple (from the skin line if the nipple is rolled) back to the pectoralis at 90°. If no pec is visible, measure to the posterior edge of the image. On the MLO, draw from the nipple perpendicular to the anterior margin of the pec, or to the posterior image edge if the muscle is short.

ACR 1999 and the Learning Network agree: PNL on the CC must be within 1 cm of PNL on the same-side MLO. The usual failure is a CC that is more than 1 cm shorter than the MLO, meaning posterior tissue was left on the chest wall. Fix it with a higher-quality lift, correct detector height, sternal contact, and a shoulder that is not dropped. PNL is a Learning Network major criterion: fail it on either breast and the exam fails that collaborative's "all 14 majors" bar.

A matching PNL does not prove you pulled as far as anatomy allows. If both views are equally shallow, both can pass the 1 cm rule and still miss posterior tissue. Always look for retroglandular fat and compare with prior depth.

Nipple in profile, motion, and what the CC is not

The nipple must be in profile on at least one view of each breast (CC or MLO). Ideal is both views; the scored minimum is one per breast. A buried nipple superimposes retroareolar ducts and can hide a retroareolar mass or calcification. If forcing profile on the CC would drop posterior tissue, keep the posterior tissue and obtain a dedicated nipple-in-profile image.

Motion is a major fail with no partial credit. On the CC it usually means the patient breathed, the shoulder crept down, or you released the breast before the paddle locked. Ask for a still breath-hold — not a deep inspiratory chest-lift that pulls tissue out of the field. Left-to-right compression on the same projection should be comparable; a large mismatch is a clue that one side was under-compressed.

CC critique: meets vs does not meet

Use this table when you hang a CC. "Meets" language follows ACR 1999 plus the Learning Network major/minor split. IMF is listed so you remember it is out of scope for this view.

  • Criterion Meets Does not meet
  • C-arm / projection 0° vertical beam, detector horizontal Oblique C-arm used as a "CC"

Detector height At the elevated IMF; breast lifted onto IR Too low (slump, short PNL) or too high (lost inferior/posterior tissue)

Medial tissue Cleavage/sternal contact; posteromedial tissue on the image Medial cutoff; sternum never reached the detector

Exaggeration (minor) Nipple midline; no tissue lost Nipple off midline with lost medial or lateral posterior tissue

Posterior tissue / PNL (major) Retroglandular fat visible; CC PNL within 1 cm of MLO PNL No posterior plane; CC PNL >1 cm shorter than MLO

Pectoralis on CC Optional; seen in ~30–40% of correct CCs Treating absent pec as an automatic fail

Nipple in profile (major, per breast) In profile on at least the CC or the MLO of that breast Buried on both views of that breast

Motion (major) Sharp Cooper ligaments and calcifications Any motion blur

Cutoff / overlay (major) Entire CC breast on detector; no chin, hand, shoulder, or opposite breast over tissue Lateral or medial cutoff; body part over breast

  • Skin folds (minor) None, or tiny folds that do not hide tissue Large folds obscuring glandular tissue
  • Open IMF Not a CC goal Failing a CC because the fold is not on the image
  • High-yield CC errors

Lateral exaggeration that loses medial tissue — nipple migrates medially, cleavage disappears, and a medial cancer can sit off the film.

Dropped shoulder — posterior tissue walks off; shoulder or chin may overlay the breast.

Motion — blur of ligaments and calcifications; always a major fail.

Detector too high or too low — over-elevating the IMF loses inferior/posterior tissue; a low detector lets the breast slump.

Letting go before lock — posterior tissue retracts and PNL shortens.

Treating IMF visualization as a CC goal — use the raised fold only to set height.

If lateral glandular tissue is still missing after a midline CC that already meets PNL and medial criteria, the next image is an XCCL, not another rotated "CC."