Study guide

9.3 Patient-Related Artifacts

ARRT Mammography Postprimary

Track your progress as you read

Finished sections get a check mark.

1 Introduction & Exam Overview

1.1 Current Exam Facts 1.2 Postprimary Eligibility, Structured Education, and Clinical Requirements 1.3 Scoring, Retakes, Fees, and Credential Maintenance 1.4 2025 Content Outline, Domain Weights, and Study Strategy

2 Patient Communication

2.1 Pre-Exam Instructions 2.2 Explaining the Exam, Rapport, and Support 2.3 Patient Education: ACS/ACR Guidelines, BSE, CBE, Dose, and Modalities

3 Results Communication and Breast Cancer Risk

3.1 Process for Receiving Results and Additional Imaging 3.2 Epidemiology and Inherent Risk Factors 3.3 Social Risk Factors, Access, and Hormone Use 3.4 Signs and Symptoms of Breast Disease

4 History, Priors, and Treatment Context

4.1 Medical History, Physical Findings, and Gender Affirmation 4.2 Previous Mammograms 4.3 Surgical, Nonsurgical, and Reconstruction Options

5 Mammography Unit Design

5.1 kVp Range, Tube, Filtration, Window, and Focal Spot 5.2 Compression Paddles, Grids, and System Geometry

6 Digital Acquisition, Display, and Informatics

6.1 FFDM-DR/2D, DBT/3D, and Synthesized Imaging 6.2 Image Receptors and Acquisition vs Interpretation Monitors 6.3 HIS/RIS/EMR, PACS, and CAD

7 Accreditation, MQSA, EQUIP, and Density Reporting

7.1 Accreditation Bodies, FDA/MQSA Personnel, Records, Audits, and Policies 7.2 EQUIP: QA Oversight, Clinical Image Quality, and Corrective Action 7.3 Breast Density Reporting Requirements

8 Quality Control Testing

8.1 Mammographer QC: Phantom Image, Compression Thickness, Visual Checklist 8.2 Mammographer QC: Monitors, Repeat Analysis, Viewing Conditions, Compression Force, Detector Calibration 8.3 Physicist Tests: MEE, Collimation, Spatial Resolution/MTF, SNR/CNR, AEC, AGD, kVp, HVL 8.4 Physicist Tests: Artifacts, Phantom Quality, Illuminance, Tech QC Program Review, Compression Application/Paddle Alignment, Workstation QC

9 Technique and Image Evaluation

9.1 Technical Factors: kVp, mAs, AEC vs Manual, Target/Filter, Focal Spot, Grid, Mag, Labeling 9.2 EQUIP Image-Quality Criteria 9.3 Patient-Related Artifacts

10 Localization and Anatomy

10.1 Clock Position, Quadrants, and Triangulation 10.2 External Anatomy 10.3 Internal Anatomy: Fascia, Glandular Tissue, and the TDLU 10.4 Cooper Ligaments, Pectoral Muscle, Vascular/Lymphatic Systems, and Cytology

11 BI-RADS Lexicon

11.1 Architectural Distortion, Asymmetry, and Mass Features 11.2 Calcification Morphology and Distribution 11.3 Assessment Categories, Breast Density, and Recommendations

12 Breast Pathology

12.1 Benign Pathology and Mammographic Appearance 12.2 Lesions with Upgrade Potential 12.3 DCIS, IDC, and Invasive Lobular Carcinoma 12.4 Inflammatory Carcinoma, Paget Disease, Sarcoma, Lymphoma, and Metastases

13 Standard Screening Views

13.1 Craniocaudal (CC) Positioning 13.2 Mediolateral Oblique (MLO) Positioning

14 Additional Views I

14.1 Mediolateral (ML) and Lateromedial (LM) 14.2 XCCL, XCCM, Cleavage (CV), and Axillary Tail (AT) 14.3 Tangential (TAN) and Rolled Views (RL, RM, RS, RI)

15 Additional Views II

15.1 Implant Displaced (ID), Nipple in Profile, and Anterior Compression 15.2 Spot Compression and Magnification

16 Patient Variance

16.1 Chest Wall Variations, Kyphosis/Lordosis, Protruding Abdomen, and Extreme Breast Size 16.2 Irradiated Breast, Reduction Mammoplasty, and Postsurgical Breast 16.3 Male Patients and Lactating Breast 16.4 Breast Augmentation and Implanted Devices

17 Imaging Examinations

17.1 Screening vs Diagnostic Mammography 17.2 Breast Ultrasound and Breast MRI

18 Interventional Procedures

18.1 History, Consent, Time-Out, Setup, and Postprocedure Instructions 18.2 Stereotactic and Ultrasound-Guided Biopsy 18.3 Needle Localization, Specimen Imaging, and Clip Placement 18.4 Handling and Disposing of Biohazardous Materials

9.3 Patient-Related Artifacts

Key Takeaways Explain This Topic AI

Removable causes—deodorant, clothing, hair, eyeglasses, skin folds, and extra body parts—that overlay parenchyma or mimic calcifications are technical repeats.

Deodorant and powder appear as high-density axillary specks that mimic microcalcifications; wipe the skin and repeat that view.

Pacemakers, ports, and implants are not wiped away: fix AEC with manual technique if hardware covers the sensor, include as much breast as possible, and leave interpretation to the radiologist.

A classic intramammary lymph node is a reniform upper-outer density with a fatty hilum—it is anatomy, not an artifact, and does not warrant a technical repeat.

Motion blur of trabeculae and calcifications is the most common patient-related FFDM artifact and warrants a shorter, better-immobilized repeat when the patient can cooperate.

Last updated: August 2026

The 2025 ARRT Mammography content specifications added patient-related artifacts under technique and image evaluation. Equipment ghosts and software processing lines are quality-control problems. This section is what the patient brings into the field—and whether that appearance is a technical repeat or something the radiologist must interpret. Patient-care chapters already cover wiping deodorant before the first exposure. Here the question is: what does it look like on the image, and do you repeat?

Motion

Motion is the most common patient-related artifact on FFDM. Appearance: loss of sharp trabeculae, a smeared skin line, calcifications that look larger and less distinct than on the other view. On DBT, blur is often clearer on projection frames than on reconstructed slices. On contrast-enhanced mammography, motion between low- and high-energy frames can produce banding.

If blur overlays parenchyma, repeat with more effective immobilization and a shorter exposure. Do not send a smeared MLO because “she was shaking.” If the patient cannot hold still after a good-faith attempt, document the limitation and still provide the sharpest achievable image. Motion is an EQUIP sharpness fail and a patient-related artifact at the same time; the action is still a better-immobilized retake when cooperation is possible.

Clothing and External Objects

Buttons, embroidery, wrinkles in a gown, jewelry, and eyeglasses are high-attenuation or textured shadows. Glasses commonly clip the superior CC or anterior MLO. Hair hanging over the superior breast makes streaky, wavy lucent or dense lines that can mimic architectural distortion.

Remove the object, smooth hair off the field, and repeat . These are never “leave for the radiologist” findings when they overlie breast tissue. A necklace under the gown is not a calcification cluster.

Anatomy: Skin Folds and Other Body Parts

Skin folds appear as curvilinear lucent lines or dark air-gap bands where skin is doubled. A fold across the posterior breast can hide a mass. Other body parts—chin on the superior CC, abdomen or opposite breast on the MLO, a hand or shoulder in the axillary tail—add extra density and can fool AEC.

If the fold or extra part overlays tissue you need, lift, taut the skin, and repeat . A tiny fold in far axillary skin that does not cover parenchyma may be documented rather than endlessly repeated; that is judgment, not laziness. Geriatric loose skin folds easily—that is technique, not “the patient failed.”

Deodorant, Antiperspirant, and Powder

Patients should arrive without deodorant; that instruction lives in patient care. On the image , talc and antiperspirant leave high-density, specklike particles, typically in the axilla and skin folds, that mimic microcalcifications .

If you see axillary specks that sit on the skin line or in a fold, wipe the skin and repeat . Do not annotate “deodorant?” and send the study as if the radiologist should guess. If specks persist after cleaning and sit in a pattern of dermal calcifications (lucent centers, known skin calcifications), leave that differential for the IP—you already did the technical step.

Implanted Devices

Pacemakers and infusion ports overlay the MLO more often than the CC. They are dense, structured, and often have a processing halo on digital images. They can sit on the AEC cell and blow the exposure.

Reposition to include as much breast as possible without dragging the generator into the middle of the glandular tissue if the device can be avoided.

Switch to manual if the device remains over the sensor.

Do not repeat in order to make the device disappear. Document its presence. The radiologist interprets tissue around hardware; you are responsible for exposure and for not excluding posterior breast while “avoiding” the device.

Tissue expanders, sternal wires, and injection-related foreign material follow the same rule: they are inside the patient, not dirt on the skin.

Intramammary Lymph Nodes Versus Artifact

A prominent intramammary lymph node (IMLN) is usually an upper-outer-quadrant, reniform, circumscribed density with a fatty hilum , often projected over the pectoral margin on the MLO. It is not an artifact. Do not wipe the skin, do not repeat to “get rid of the mass,” and do not mark it as a mole unless there is a corresponding skin lesion.

If the density is not circumscribed, has no hilum, is high density, or is new compared with priors, it is a finding—still not a technical repeat. Leave it for the radiologist.

Tattoos, seborrheic keratoses, and moles can mimic masses or calcifications; those you mark and, if they look like surface specks, correlate with the skin.

Appearance → Cause → Action

Appearance on the image Likely cause Action

Directional blur of trabeculae, skin, or calcifications Patient motion; long exposure Repeat with immobilization / shorter time; document if unavoidable

High-density specks in axilla or inframammary fold, often on the skin Deodorant, antiperspirant, powder Wipe skin; repeat that view

  • Streaky lines over the superior breast Hair Move hair; repeat
  • Geometric dense frames at the superior edge Eyeglasses Remove glasses; repeat
  • Curvilinear lucent line or dark band Skin fold / air gap Smooth and retake if it overlays parenchyma
  • Extra soft-tissue mass at chest wall or IMF Abdomen, chin, opposite breast, hand Reposition; repeat
  • Fabric weave, button, necklace Clothing / jewelry Remove; repeat

High-attenuation generator or catheter reservoir on MLO Pacemaker or port Manual technique if over AEC; include tissue; do not treat as dirt; leave for the IP

  • Reniform upper-outer density with fatty hilum Intramammary node Not an artifact; no technical repeat
  • Raised skin lesion mimicking a mass Mole, wart, or keratoma BB/mole marker; document

If you can take the cause off the patient or out of the field in the room, and it hides or mimics disease, it is a technical repeat . If the cause is inside the patient (node, implant, pacemaker) or a documented positioning limit, do not manufacture a repeat that cannot change the physics. The radiologist owns interpretation; you own whether the image is a fair test of the tissue.

Loading diagram... Patient-related finding: technical repeat versus leave for the radiologist