Study guide
3.10 Parameter Tradeoffs: Resolution and Scan Time
ARRT MRI Exam
Track your progress as you read
Finished sections get a check mark.
1 Chapter 1: Patient Care
1.1 Patient Interactions and Ethical/Legal Aspects 1.2 Patient Assessment and Monitoring 1.3 Communication and Safety Checks 1.4 Infection Control and Aseptic Technique 1.5 Contrast Administration and Safety
2 Chapter 2: Safety
2.1 ACR Safety Zones and Personnel 2.2 Static Magnetic Field (B0) Hazards 2.3 Radiofrequency (B1) Field Hazards 2.4 Time-Varying Gradient Field Hazards 2.5 Acoustic Noise and Patient Protection 2.6 Patient Screening Workflows 2.7 Implant and Device Safety 2.8 Emergency Response and Quenches 2.9 Specialized Thermal and Dress Safety
3 Chapter 3: Image Production
3.1 Fundamental Physics and Nuclear Magnetism 3.2 Tissue Characteristics and Relaxation 3.3 Spatial Localization and Gradients 3.4 Spin Echo Pulse Sequences 3.5 Gradient Echo Pulse Sequences 3.6 Inversion Recovery Sequences 3.7 Advanced Sequence Families 3.8 k-Space and Data Acquisition 3.9 Parameter Tradeoffs: SNR and CNR 3.10 Parameter Tradeoffs: Resolution and Scan Time 3.11 Artifact Recognition and Management 3.12 Coil Selection and Positioning 3.13 Quality Assurance and Equipment Operation
4 Chapter 4: Procedures
4.1 Brain and Head/Neck Protocols 4.2 Spine Protocols 4.3 Abdomen Protocols 4.4 Pelvis Protocols 4.5 Musculoskeletal Joints 4.6 Musculoskeletal Soft Tissue 4.7 Vascular, Breast, and Specialized Procedures
3.10 Parameter Tradeoffs: Resolution and Scan Time
Key Takeaways Explain This Topic AI
Acquisition Time (Ta) for a standard 2D spin echo sequence is calculated using the formula Ta = TR × Ny × NEX, where Ny is the number of phase encoding steps.
Fast Spin Echo (FSE) sequences accelerate scan time by the Echo Train Length (ETL) or Turbo Factor, yielding the scan time formula Ta = (TR × Ny × NEX) / ETL.
Spatial Resolution is determined by voxel size; smaller voxels yield higher spatial resolution but result in lower SNR and potentially longer scan times.
Parallel Imaging (SENSE/GRAPPA) uses multi-channel coils to skip phase encoding steps, reducing scan time by the acceleration factor (R) but decreasing SNR by a factor of g × √R (where g is the geometry factor).
Last updated: July 2026
Parameter Tradeoffs: Resolution and Scan Time
In MRI, optimizing the balance between spatial resolution and scan time is a critical task for the technologist. High spatial resolution is required to resolve small anatomical details, but it typically demands smaller voxel sizes. Smaller voxels contain fewer spins, resulting in lower SNR, which technologists often compensate for by using parameters that increase the total acquisition time.
Acquisition Time Calculations
The scan time for an MRI sequence depends on the pulse sequence family and the selected scan parameters. Technologists must master the mathematical formulas for calculating acquisition times for the ARRT registry exam.
1. 2D Spin Echo (SE) and Gradient Echo (GRE) Sequence
For standard 2D sequences, the acquisition time ( T a T_a T a ) is determined by: T a = TR × N y × NEX T_a = \text{TR} \times N_y \times \text{NEX} T a = TR × N y × NEX
Where:
TR (Repetition Time): The time between successive RF excitation pulses (in seconds).
N y N_y N y (Phase Matrix): The number of phase encoding steps.
NEX (Number of Excitations): The number of times each phase line is collected.
Worked Example: A T1-weighted spin echo sequence is programmed with a TR of 500 ms, a phase matrix of 256, and a NEX of 2. T a = 0.5 s × 256 × 2 = 256 seconds = 4 minutes and 16 seconds T_a = 0.5\text{ s} \times 256 \times 2 = 256\text{ seconds} = 4\text{ minutes and 16 seconds} T a = 0.5 s × 256 × 2 = 256 seconds = 4 minutes and 16 seconds
2. Fast Spin Echo (FSE) Sequence
FSE sequences collect multiple phase encoding lines during each TR interval by applying a series of 180-degree refocusing pulses. The number of echoes collected per TR is the Echo Train Length (ETL) or Turbo Factor . The formula is: T a = TR × N y × NEX ETL T_a = \frac{\text{TR} \times N_y \times \text{NEX}}{\text{ETL}} T a = ETL TR × N y × NEX
Worked Example: A T2-weighted FSE sequence is programmed with a TR of 4000 ms, a phase matrix of 256, a NEX of 2, and an ETL of 16. T a = 4.0 s × 256 × 2 16 = 2048 16 = 128 seconds = 2 minutes and 8 seconds T_a = \frac{4.0\text{ s} \times 256 \times 2}{16} = \frac{2048}{16} = 128\text{ seconds} = 2\text{ minutes and 8 seconds} T a = 16 4.0 s × 256 × 2 = 16 2048 = 128 seconds = 2 minutes and 8 seconds
3. 3D Volume Acquisition
3D sequences excite an entire volume of tissue rather than individual slices. Spatial localization along the slice axis requires an additional phase encoding step, called slice encoding (or partitions). The formula is: T a = TR × N y × N z × NEX T_a = \text{TR} \times N_y \times N_z \times \text{NEX} T a = TR × N y × N z × NEX
Where N z N_z N z is the number of slice-encoding partitions.
Worked Example: A 3D gradient echo sequence has a TR of 20 ms, a phase matrix of 256, 64 slice partitions, and a NEX of 1. T a = 0.020 s × 256 × 64 × 1 = 327.68 seconds = 5 minutes and 28 seconds T_a = 0.020\text{ s} \times 256 \times 64 \times 1 = 327.68\text{ seconds} = 5\text{ minutes and 28 seconds} T a = 0.020 s × 256 × 64 × 1 = 327.68 seconds = 5 minutes and 28 seconds
Spatial Resolution Factors
Spatial resolution is defined as the ability to distinguish two separate points in tissue as distinct structures. It is determined by the size of the voxels: smaller voxels result in higher spatial resolution.
Voxel Dimensions
Voxel size is determined by three parameters:
Pixel Width (Frequency axis): FOV frequency ÷ Frequency Matrix \text{FOV}_{\text{frequency}} \div \text{Frequency Matrix} FOV frequency ÷ Frequency Matrix
Pixel Length (Phase axis): FOV phase ÷ Phase Matrix \text{FOV}_{\text{phase}} \div \text{Phase Matrix} FOV phase ÷ Phase Matrix
Voxel Depth (Slice axis): Slice thickness
Parameter Adjustments for High Resolution:
Reduce FOV: Decreasing the FOV reduces the pixel size, increasing spatial resolution, but decreases SNR.
Increase Matrix Size: Increasing matrix size (e.g., from 192 to 256 phase steps) decreases pixel size, increasing spatial resolution.
Decrease Slice Thickness: Thinner slices reduce voxel depth, increasing spatial resolution along the z-axis and reducing partial volume averaging.
Phase Matrix vs. Frequency Matrix Tradeoffs
Understanding how the matrix dimensions affect scan time is critical:
Frequency Encoding Matrix: Increasing the frequency matrix size (e.g., from 256 to 512) increases spatial resolution along the readout axis but has no effect on scan time . Frequency encoding occurs during readout, which takes milliseconds.
Phase Encoding Matrix: Increasing the phase matrix size (e.g., from 256 to 512) increases spatial resolution along the phase axis but increases scan time . Each phase encoding line requires a separate TR interval.
Rectangular FOV / Asymmetric FOV
To save scan time when imaging anatomy that is longer in one direction (e.g., spine or extremities), a rectangular FOV (rFOV) is used.
By reducing the phase FOV (e.g., to 50% of the frequency FOV), the number of phase encoding steps ( N y N_y N y ) can be cut in half while maintaining the same pixel size (spatial resolution).
This reduces the scan time by 50% without altering spatial resolution or causing aliasing, provided the anatomy does not exceed the borders of the reduced phase FOV.
Advanced Acceleration Techniques
When clinical needs require short scan times (e.g., for pediatric, dyspneic, or claustrophobic patients), technologists use advanced acceleration options:
1. Fractional NEX (Partial Fourier)
The Physics: k-space has conjugate symmetry, meaning the top half is a mirror image of the bottom half.
The Method: The scanner collects only a fraction of the phase lines (e.g., 60% or 0.75 NEX) and mathematically calculates the remaining lines.
Tradeoff: Scan time is reduced proportionally. However, because less total signal is collected, the SNR decreases by fraction collected \sqrt{\text{fraction collected}} fraction collected .
2. Parallel Imaging (SENSE / GRAPPA)
The Physics: Uses spatial sensitivity profiles of multi-channel phased-array receiver coils to reconstruct the spatial origin of signals.
The Method: The scanner skips lines of phase encoding during acquisition, reducing the number of phase lines by the acceleration factor ( R R R ).
Tradeoffs:
Scan time is reduced by a factor of R R R .
SNR is reduced due to the reduction in data collection, calculated as: SNR Parallel = SNR base g R \text{SNR}_{\text{Parallel}} = \frac{\text{SNR}_{\text{base}}}{g \sqrt{R}} SNR Parallel = g R SNR base where g g g is the geometry factor of the receiver coil array.
If the acceleration factor ( R R R ) is set too high, unfolding artifacts (aliasing) can occur.
Clinical Optimization Strategies
Technologists must select parameters based on the patient's clinical situation:
Motion-Prone Patients (Pediatric, Claustrophobic): Prioritize scan speed. Increase parallel imaging acceleration ( R R R ), increase ETL on FSE sequences, use rectangular FOV, and reduce NEX.
High-Resolution Needs (IACs, Pituitary): Prioritize spatial resolution. Use a small FOV, high matrix, and thin slices. Compensate for the low SNR by increasing the NEX (accepting longer scan times), using dedicated local coils, or scanning at a higher field strength ( 3.0 T 3.0\text{ T} 3.0 T ).
Resolution and Scan Time Summary Table
- Parameter Change Spatial Resolution Scan Time Signal-to-Noise Ratio (SNR)
- Increase FOV Decrease No Effect Increase
- Increase Phase Matrix Increase Increase Decrease
- Increase Frequency Matrix Increase No Effect Decrease
- Increase Slice Thickness Decrease No Effect Increase
- Increase NEX No Effect Increase Increase ( NEX \sqrt{\text{NEX}} NEX )
- Increase ETL Decrease (Blurring) Decrease Decrease
- Enable Parallel Imaging No Effect Decrease Decrease