Study guide
5.2 Collimation Principles & Types
ARRT Nuclear Medicine Technology Exam
Track your progress as you read
Finished sections get a check mark.
1 Chapter 1: Patient Assessment, Care & Emergency Procedures
1.1 Patient Assessment, Medical History & Communication 1.2 Venipuncture, Vascular Access & Emergency Response
2 Chapter 2: Ethical, Legal & Professional Standards
2.1 ARRT Code of Ethics & Patient Rights 2.2 Patient Care in Special Populations & Infection Control
3 Chapter 3: Radiation Physics & Biological Effects
3.1 Atomic Structure, Nuclear Decay & Radiation Interactions 3.2 Biological Effects of Radiation & Radiobiology Principles
4 Chapter 4: Radiation Protection, ALARA & NRC Regulations
4.1 Radiation Protection Principles & ALARA Implementation 4.2 NRC Regulations, Occupational Dose Limits & Area Monitoring 4.3 Radioactive Spills, Contamination Control & Medical Events
5 Chapter 5: Gamma Camera Instrumentation & Physics
5.1 Gamma Camera Components & Signal Processing 5.2 Collimation Principles & Types 5.3 Quality Control of Gamma Cameras
6 Chapter 6: Advanced Imaging Modalities: SPECT, PET/CT & QC
6.1 SPECT Instrumentation, Reconstruction & Quality Control 6.2 PET/CT Instrumentation, Coincidence Detection & Crystals 6.3 PET/CT Quality Control, Attenuation Correction & PACS
7 Chapter 7: Radionuclide Production & Generator Systems
7.1 Radionuclide Production Methods 7.2 Mo-99/Tc-99m Generator Operations & Elution 7.3 Generator Quality Control & Breakthrough Testing
8 Chapter 8: Radiopharmaceutical Preparation, QC & Administration
8.1 Radiopharmaceutical Preparation & Chemistry 8.2 Radiochemical Purity & Dose Calibrator Quality Control 8.3 Dose Calculations, Unit Dose Dispensing & Waste Management
9 Chapter 9: Endocrine, Central Nervous System & Gastrointestinal Imaging
9.1 Thyroid & Parathyroid Imaging & Uptake 9.2 Central Nervous System Imaging 9.3 Gastrointestinal & Hepatobiliary Imaging
10 Chapter 10: Cardiovascular Imaging Procedures
10.1 Myocardial Perfusion SPECT Imaging 10.2 Cardiac Stress Protocols & Vascular Anatomy 10.3 Ventricular Function & MUGA Scans 10.4 Myocardial Viability Imaging 10.5 Cardiac Amyloid Imaging (Tc-99m PYP)
11 Chapter 11: Pulmonary, Skeletal & Renal Imaging Procedures
11.1 Pulmonary Ventilation & Perfusion (V/Q) Imaging 11.2 Skeletal System Imaging & 3-Phase Bone Scans 11.3 Renal & Genitourinary Imaging
12 Chapter 12: Tumor, Infection, Hematology & Radiopharmaceutical Therapy
12.1 Oncology PET/CT Imaging with F-18 FDG 12.2 Infection & Tumor Imaging with Special Radiopharmaceuticals 12.3 Therapeutic Radiopharmaceuticals & Radiation Safety Precautions 12.4 Lymphoscintigraphy & Sentinel Lymph Node Mapping
5.2 Collimation Principles & Types
Key Takeaways Explain This Topic AI
Collimators are made of lead (or tungsten) and act as the 'lens' of the gamma camera by absorbing gamma rays not traveling in the proper direction.
Parallel hole collimators (LEHR, MEGP, HEGP) are the most common and are chosen based on the isotope's energy.
Pinhole collimators magnify small organs (like the thyroid) and invert the image.
There is always an inverse trade-off between a collimator's spatial resolution and its sensitivity.
Septal penetration occurs when high-energy gamma rays cross through collimator walls, degrading the image.
Last updated: July 2026
5.2 Collimation Principles & Types
Quick Answer: The collimator acts as the directional geometric lens of the gamma camera, restricting incident gamma rays so that only photons traveling perpendicular (or along specific geometric angles) to the crystal face pass through to form an image. Collimation dictates spatial resolution and count-rate sensitivity, governed by strict mathematical trade-offs between hole length ( l l l ), hole diameter ( d d d ), septal thickness ( t t t ), and source distance ( b b b ).
Because gamma rays cannot be refracted or focused by optical lenses, planar and SPECT gamma cameras depend entirely on absorptive collimation . Made of lead ( Z = 82 Z=82 Z = 82 ) or tungsten ( Z = 74 Z=74 Z = 74 ), collimators absorb off-axis photons while allowing parallel or geometrically directed photons to pass through thousands of precisely arrayed channels (bores).
Mathematical Fundamentals of Collimator Performance
Collimator performance is characterized by two competing parameters: Collimator Spatial Resolution ( R c R_c R c ) and Geometric Sensitivity ( g g g ) .
1. Collimator Spatial Resolution ( R c R_c R c )
For a parallel-hole collimator, geometric spatial resolution ( R c R_c R c , defined as Full Width at Half Maximum [FWHM] of a line source response) is given by: R c = d ⋅ ( l + b ) l = d + d ⋅ ( b l ) R_c = \frac{d \cdot (l + b)}{l} = d + d \cdot \left(\frac{b}{l}\right) R c = l d ⋅ ( l + b ) = d + d ⋅ ( l b )
Where:
- d d d = Hole diameter (bore size)
- l l l = Effective hole length (bore height)
- b b b = Distance from the front face of the collimator to the target radioactive object
Total System Resolution ( R sys R_{\text{sys}} R sys ): Overall spatial resolution combines intrinsic crystal resolution ( R intr R_{\text{intr}} R intr ) and collimator resolution ( R c R_c R c ) in quadrature: R sys = R intr 2 + R c 2 R_{\text{sys}} = \sqrt{R_{\text{intr}}^2 + R_c^2} R sys = R intr 2 + R c 2
Clinical Rule: Because R c R_c R c degrades linearly as distance b b b increases, the collimator face must be positioned as close to the patient's body as physically possible . Doubling the distance from 5 cm 5\ \text{cm} 5 cm to 10 cm 10\ \text{cm} 10 cm significantly increases R c R_c R c (poorer resolution), blurring fine anatomical structures.
2. Geometric Sensitivity ( g g g )
Geometric sensitivity represents the fraction of emitted gamma photons that pass through the collimator holes without striking septa. For parallel-hole collimators with round or hexagonal holes: g = K 2 ⋅ ( d 2 l ⋅ ( d + t ) ) 2 g = K^2 \cdot \left( \frac{d^2}{l \cdot (d + t)} \right)^2 g = K 2 ⋅ ( l ⋅ ( d + t ) d 2 ) 2
Where:
- K K K = Geometric constant based on hole array packing (hexagonal vs. square)
- t t t = Septal thickness
- The Distance Independence Paradox
A fundamental property of parallel-hole collimators is that for a uniform, extended source larger than the field of view, geometric sensitivity ( g g g ) is independent of distance ( b b b ) . As distance increases, the count flux from any single point decreases according to the Inverse Square Law ( 1 / b 2 1/b^2 1/ b 2 ), but the area of the source viewed by each hole expands proportionally ( b 2 b^2 b 2 ). The two effects cancel exactly. However, spatial resolution still degrades with distance.
3. The Resolution vs. Sensitivity Trade-Off
To increase spatial resolution: Make holes longer ( l ↑ l \uparrow l ↑ ) or narrower ( d ↓ d \downarrow d ↓ ). This restricts the angular acceptance cone. However, sensitivity drops dramatically ( g ∝ d 4 / l 2 g \propto d^4 / l^2 g ∝ d 4 / l 2 ).
To increase sensitivity: Make holes shorter ( l ↓ l \downarrow l ↓ ) or wider ( d ↑ d \uparrow d ↑ ). Count rates increase, but off-axis photons enter, increasing R c R_c R c and blurring the image.
Septal Penetration Physics
Septal penetration occurs when high-energy photons pass through lead septa between holes rather than being absorbed. Penetrating photons strike the crystal at incorrect spatial coordinates, causing low-frequency background fog, loss of image contrast, and "starburst" artifacts.
To limit septal penetration ( P P P ) to less than 5 % 5\% 5% , the minimum required septal thickness ( t t t ) is calculated as: t ≥ 2 ⋅ d ⋅ μ l − 2 ⋅ μ t \ge \frac{2 \cdot d \cdot \mu}{l - 2 \cdot \mu} t ≥ l − 2 ⋅ μ 2 ⋅ d ⋅ μ
Where μ \mu μ is the linear attenuation coefficient of the collimator material (lead/tungsten) at the specific gamma energy. Higher energy isotopes (e.g., I-131 at 364 keV 364\ \text{keV} 364 keV ) require substantially thicker septa than Tc-99m ( 140 keV 140\ \text{keV} 140 keV ).
Collimator Class Energy Range Typical Isotope Septal Thickness ( t t t ) Hole Length ( l l l ) Clinical Uses
LEHR (Low Energy High Res) < 150 keV <150\ \text{keV} < 150 keV Tc-99m, Tl-201 ∼ 0.16 − 0.20 mm \sim 0.16-0.20\ \text{mm} ∼ 0.16 − 0.20 mm 25 − 35 mm 25-35\ \text{mm} 25 − 35 mm Bone scans, cardiac SPECT, renal imaging
LEAP / LEGP (General Purpose) < 150 keV <150\ \text{keV} < 150 keV Tc-99m ∼ 0.20 − 0.25 mm \sim 0.20-0.25\ \text{mm} ∼ 0.20 − 0.25 mm 20 − 25 mm 20-25\ \text{mm} 20 − 25 mm High count rate dynamic studies (first-pass cardiac)
MEGP (Medium Energy) 150 − 300 keV 150-300\ \text{keV} 150 − 300 keV In-111, Ga-67, I-123 ∼ 1.0 − 1.5 mm \sim 1.0-1.5\ \text{mm} ∼ 1.0 − 1.5 mm 30 − 40 mm 30-40\ \text{mm} 30 − 40 mm Indium-111 WBC, Gallium-67 tumor/infection
HEGP (High Energy) 300 − 400 keV 300-400\ \text{keV} 300 − 400 keV I-131 ∼ 2.5 − 3.5 mm \sim 2.5-3.5\ \text{mm} ∼ 2.5 − 3.5 mm 40 − 60 mm 40-60\ \text{mm} 40 − 60 mm Post-therapy Iodine-131 thyroid cancer whole body
- Specialized Collimator Geometries
- 1. Pinhole Collimator
Features a single conical lead horn terminating in a small interchangeable aperture ( 1 mm , 3 mm , 4 mm 1\ \text{mm}, 3\ \text{mm}, 4\ \text{mm} 1 mm , 3 mm , 4 mm tungsten insert).
Geometric Magnification: M = f / b M = f / b M = f / b , where f f f is focal length (aperture to crystal) and b b b is source distance. Objects closer than f f f are magnified ( M > 1 M > 1 M > 1 ).
Inversion: Images are inverted along both horizontal and vertical axes (upside down and right-to-left flipped).
Use: Unmatched spatial resolution for tiny organs (thyroid, parathyroid, pediatric hip joint).
Limitation: Sensitivity decreases with the square of distance from the aperture.
2. Converging Collimator
Holes are angled inward toward a focal point in front of the camera face.
Magnification: Image is magnified ( M > 1 M > 1 M > 1 ) without image inversion.
Sensitivity: Increases as the target organ approaches the focal point.
Use: Imaging small structures (e.g., pediatric cardiac/renal) on standard large-field-of-view detectors.
3. Diverging Collimator
Holes are angled outward from a focal point behind the detector.
Minification: Image is minified ( M < 1 M < 1 M < 1 ), fitting large anatomical structures (e.g., adult lung perfusion) onto smaller gamma camera detectors.
4. Fan-Beam & Cone-Beam Collimators
Designed for SPECT brain and cardiac imaging. Fan-beam collimators converge in the transverse plane (magnifying the brain across the head) while remaining parallel in the axial plane.
- Collimator Type Hole Geometry Image Size vs. Object Primary Application
- LEHR Parallel, thin septae 1:1 (Actual size) Routine Tc-99m imaging (Bone, Myocardial)
- MEGP Parallel, medium septae 1:1 (Actual size) In-111, Ga-67 studies
- HEGP Parallel, thick septae 1:1 (Actual size) I-131 therapy/imaging
- Pinhole Single aperture, cone Magnified (and Inverted) Thyroid, parathyroid, small joints
- Converging Angled inward Magnified Brain, pediatric imaging
- Diverging Angled outward Minified Large organs on small detectors