5.3 Quality Control of Gamma Cameras
ARRT Nuclear Medicine Technology Exam
Track your progress as you read
Finished sections get a check mark.
1 Chapter 1: Patient Assessment, Care & Emergency Procedures
1.1 Patient Assessment, Medical History & Communication 1.2 Venipuncture, Vascular Access & Emergency Response
2 Chapter 2: Ethical, Legal & Professional Standards
2.1 ARRT Code of Ethics & Patient Rights 2.2 Patient Care in Special Populations & Infection Control
3 Chapter 3: Radiation Physics & Biological Effects
3.1 Atomic Structure, Nuclear Decay & Radiation Interactions 3.2 Biological Effects of Radiation & Radiobiology Principles
4 Chapter 4: Radiation Protection, ALARA & NRC Regulations
4.1 Radiation Protection Principles & ALARA Implementation 4.2 NRC Regulations, Occupational Dose Limits & Area Monitoring 4.3 Radioactive Spills, Contamination Control & Medical Events
5 Chapter 5: Gamma Camera Instrumentation & Physics
5.1 Gamma Camera Components & Signal Processing 5.2 Collimation Principles & Types 5.3 Quality Control of Gamma Cameras
6 Chapter 6: Advanced Imaging Modalities: SPECT, PET/CT & QC
6.1 SPECT Instrumentation, Reconstruction & Quality Control 6.2 PET/CT Instrumentation, Coincidence Detection & Crystals 6.3 PET/CT Quality Control, Attenuation Correction & PACS
7 Chapter 7: Radionuclide Production & Generator Systems
7.1 Radionuclide Production Methods 7.2 Mo-99/Tc-99m Generator Operations & Elution 7.3 Generator Quality Control & Breakthrough Testing
8 Chapter 8: Radiopharmaceutical Preparation, QC & Administration
8.1 Radiopharmaceutical Preparation & Chemistry 8.2 Radiochemical Purity & Dose Calibrator Quality Control 8.3 Dose Calculations, Unit Dose Dispensing & Waste Management
9 Chapter 9: Endocrine, Central Nervous System & Gastrointestinal Imaging
9.1 Thyroid & Parathyroid Imaging & Uptake 9.2 Central Nervous System Imaging 9.3 Gastrointestinal & Hepatobiliary Imaging
10 Chapter 10: Cardiovascular Imaging Procedures
10.1 Myocardial Perfusion SPECT Imaging 10.2 Cardiac Stress Protocols & Vascular Anatomy 10.3 Ventricular Function & MUGA Scans 10.4 Myocardial Viability Imaging 10.5 Cardiac Amyloid Imaging (Tc-99m PYP)
11 Chapter 11: Pulmonary, Skeletal & Renal Imaging Procedures
11.1 Pulmonary Ventilation & Perfusion (V/Q) Imaging 11.2 Skeletal System Imaging & 3-Phase Bone Scans 11.3 Renal & Genitourinary Imaging
12 Chapter 12: Tumor, Infection, Hematology & Radiopharmaceutical Therapy
12.1 Oncology PET/CT Imaging with F-18 FDG 12.2 Infection & Tumor Imaging with Special Radiopharmaceuticals 12.3 Therapeutic Radiopharmaceuticals & Radiation Safety Precautions 12.4 Lymphoscintigraphy & Sentinel Lymph Node Mapping
5.3 Quality Control of Gamma Cameras
Key Takeaways Explain This Topic AI
Daily flood field uniformity tests (intrinsic or extrinsic) are essential to verify the camera's ability to produce a uniform image from a uniform source.
Intrinsic uniformity is performed without a collimator using a Tc-99m point source; extrinsic is performed with a collimator using a Co-57 sheet source.
Uniformity tolerances (both integral and differential) are typically maintained below 3-5%.
Spatial resolution and linearity are tested weekly using lead bar phantoms (e.g., four-quadrant or PMM).
Prompt corrective action is required if QC fails, including photomultiplier tube tuning or service calls.
Last updated: July 2026
Quality Control of Gamma Cameras
Quick Answer: Quality Control (QC) ensures the gamma camera operates accurately and reliably. The most critical daily test is flood field uniformity , ensuring the camera sees a uniform source of radiation equally across its field of view (FOV). Weekly tests verify spatial resolution and linearity using bar phantoms.
For the ARRT exam, understanding the distinction between intrinsic and extrinsic testing, the specific sources used, and the quantitative limits for QC failure is mandatory.
Flood Field Uniformity
Uniformity evaluates the camera's ability to produce an image of uniform density when exposed to a uniform source of radiation. Over time, PMTs can drift in voltage, or the crystal can degrade, leading to "hot" (hyperactive) or "cold" (hypoactive) spots on the image.
Uniformity testing is performed daily before any patient imaging occurs. It can be performed in two ways: intrinsically or extrinsically.
Intrinsic Uniformity
Collimator: OFF.
Source: A small Tc-99m point source (typically ~10-20 MBq or a few hundred microcuries).
Setup: The source must be placed at a distance of at least 5 times the largest dimension of the detector (usually the diagonal) from the uncollimated crystal. This ensures the gamma rays arrive relatively parallel and uniformly across the FOV.
Purpose: Tests the crystal and the electronics (PMTs, PHA) without the influence of the collimator.
Extrinsic Uniformity
Collimator: ON (usually LEHR).
Source: A Co-57 sheet source (a flat, radioactive source that covers the entire FOV).
Setup: The sheet source is placed directly on the collimator face.
Purpose: Tests the entire system, including the integrity of the collimator (checking for defects, dents, or manufacturing flaws).
Note: Co-57 is used because its principal gamma energy (122 keV) is very close to Tc-99m (140 keV), and its long half-life (271 days) means the source only needs to be replaced every 1-2 years.
Quantifying Uniformity
Modern cameras use computers to analyze the flood image and calculate uniformity numerically. The calculation compares the pixel with the highest counts (Max) to the pixel with the lowest counts (Min).
There are two main metrics evaluated:
Integral Uniformity: Evaluates the variation over the entire field of view. It finds the maximum and minimum pixels anywhere in the useful FOV.
Differential Uniformity: Evaluates the rate of change over small, local areas (typically comparing any 5 consecutive pixels). This is often more visually sensitive to rapid changes in density.
Tolerance: For most clinical systems, both integral and differential uniformity should be less than 3% to 5% . If the value exceeds 5%, the camera should not be used for patient imaging until serviced or recalibrated (e.g., PMT tuning).
Spatial Resolution and Linearity
While uniformity is tested daily, spatial resolution (ability to separate distinct objects) and linearity (ability to draw a straight line straight) are typically tested weekly .
These parameters are evaluated using a lead bar phantom . The phantom consists of lead bars embedded in plastic. A uniform radiation source is placed behind the phantom.
Types of Bar Phantoms
Four-Quadrant Bar Phantom: The most common. It has four sections, each with a different bar width and spacing. You can only evaluate resolution accurately in one quadrant at a time, so the phantom often needs to be rotated 90 degrees and imaged multiple times.
Parallel-Line Equal-Space (PLES): All bars are the same width and spacing across the entire phantom. Two images at 90 degrees evaluate the whole FOV.
Orthogonal Hole Pattern (OHP): An array of holes in a lead sheet.
Evaluating the Image
Resolution: Is determined by the smallest set of bars the camera can clearly distinguish (resolve). If you can see the 2 mm bars today, but next week they are blurred, resolution has degraded.
Linearity: Is evaluated by looking at the lines of the bars across the FOV. They should appear perfectly straight. If the lines appear wavy, curved, or "barrel" shaped, there is a failure in the positioning logic, requiring immediate corrective action.
Corrective Action Protocols
If a daily uniformity flood fails (e.g., a visible "bullseye" artifact, cold tubes, or quantitative values >5%):
Check the obvious: Is the wrong energy window set? Is the source cracked or placed incorrectly?
Perform a "tune" or PMT calibration if the software allows.
If the failure persists, the camera must be taken out of service, and an engineer must be called. Document the failure and actions in the QC log.
- QC Test Frequency Collimator State Source Used Key Metric
- Intrinsic Uniformity Daily OFF Tc-99m Point Source Integral/Diff < 5%
- Extrinsic Uniformity Daily ON Co-57 Sheet Source Integral/Diff < 5%
- Resolution & Linearity Weekly ON or OFF Co-57 or Tc-99m + Bar Phantom Visual inspection of bars
- Center of Rotation (COR) Monthly/Quarterly ON Tc-99m Point Source Crucial for SPECT imaging