Study guide

3.5 Depth Dose Characteristics: PDD, TMR, TPR & TAR Relationships

ARRT Radiation Therapy Exam

Track your progress as you read

Finished sections get a check mark.

1 Chapter 1: Patient Care & Education

1.1 Patient Assessment, History, and Vital Signs 1.2 Psychosocial Support & Coping Mechanisms 1.3 Patient & Family Education on Radiation Therapy 1.4 Radiation Side Effect Management: Skin & Mucosal Reactions 1.5 Radiation Side Effect Management: Systemic & Organ-Specific Side Effects 1.6 Nutritional Support & Management of Dysphagia/Odynophagia 1.7 Medical Emergencies in Radiation Oncology 1.8 Patient Rights, Informed Consent, and Advanced Directives 1.9 Ethical, Legal, and HIPAA Considerations for Radiation Therapists 1.10 Infection Control & Sterile Technique in Radiation Therapy 1.11 Transfer & Positioning Safety for Oncology Patients 1.12 Pharmacology & Contrast Media in Radiation Therapy 1.13 Cancer Epidemiology, Screening & Diagnostic Evaluation

2 Chapter 2: Radiation Physics, Radiobiology & Protection

2.1 Structure of Matter, Electromagnetic & Particulate Radiation 2.2 Radiation Production in Linear Accelerators & X-Ray Tubes 2.3 Photon & Electron Beam Interactions with Matter 2.4 Radiation Measurement Quantities, Units & Dosimeters 2.5 Radiation Protection Principles: ALARA, Time, Distance & Shielding 2.6 NCRP Dose Limits for Occupational & Public Exposure 2.7 Vault Design, Primary & Secondary Protective Barriers 2.8 Radiobiology Principles: Cell Survival Curves & Radiosensitivity (4 R's) 2.9 Linear Energy Transfer (LET) & Relative Biological Effectiveness (RBE) 2.10 Acute & Late Radiation Effects on Normal Tissues 2.11 Radiation Safety Rules, Emergency Procedures & Sealed Source Safety 2.12 MRI Magnetic Field Screening for Radiation Therapy Patients

3 Chapter 5: Treatment Sites, Tumors & Classification

5.1 Brain, Spinal Cord Tumors & Metastatic Patterns 5.2 Head & Neck Sites: Anatomy, Lymphatics & Patterns of Spread 5.3 Breast Cancer: Anatomy, Lymphatic Drainage & Radiation Considerations 5.4 Lung & Thoracic Tumors: Anatomy, Spread & Critical Structures 5.5 Abdomen, Pelvis, GI & GU Sites: Anatomy & Metastatic Patterns 5.6 Reproductive System Tumors: Prostate, Cervix & Related Sites 5.7 Lymphoma, Sarcoma, Skin, Leukemia & Oncologic Emergencies 5.8 Tumor Classification: Histopathology, Grade & Staging Systems

4 Chapter 3: Treatment Planning & Dosimetry

3.1 CT Simulation, Patient Positioning & Target Localization 3.2 Immobilization Devices & Radiopaque Marking Protocols 3.3 ICRU Target Volume Definitions (GTV, CTV, PTV, PRV, OAR) 3.4 Beam Modifiers: Wedges, Compensators, Bolus & MLC Systems 3.5 Depth Dose Characteristics: PDD, TMR, TPR & TAR Relationships 3.6 Monitor Unit (MU) Calculations & Clinical Dosimetry Formulas 3.7 3D Conformal Radiation Therapy (3D-CRT) Beam Arrangements 3.8 IMRT, VMAT & Inverse Planning Optimization Concepts 3.9 SRS, SBRT & Stereotactic Radiosurgery Techniques 3.10 Brachytherapy Principles, HDR, LDR & Isotopes 3.11 Dose Volume Histograms (DVH) & QUANTEC Tissue Tolerances

5 Chapter 4: Treatment Delivery, Image Guidance & QA

4.1 Pre-Treatment Chart Check & Plan Verification Protocols 4.2 Daily Patient Positioning, Setup Tolerances & Alignment Verification 4.3 Image-Guided Radiation Therapy (IGRT): Portal Imaging & CBCT Verification 4.4 Linear Accelerator Safety Systems, Interlocks & Gantry Controls 4.5 Special Techniques: TBI, TSI, SGRT & Gated Radiotherapy 4.6 Daily Linac Quality Assurance per AAPM TG-142 Specifications 4.7 Monthly & Annual Linac QA Requirements & Dosimetric Calibration 4.8 Patient-Specific IMRT QA & In-Vivo Dosimetry Verification 4.9 Machine Malfunctions, Override Protocols & Emergency Off Procedures 4.10 Record and Verify (R&V) Systems, EHR Integration & Treatment Documentation

3.5 Depth Dose Characteristics: PDD, TMR, TPR & TAR Relationships

Key Takeaways Explain This Topic AI

Percentage Depth Dose (PDD) increases with increasing beam energy, increases with increasing field size, and increases with increasing SSD.

The depth of maximum dose (dmax) deepens with higher beam energy: Co-60 at 0.5 cm, 6 MV at 1.5 cm, 10 MV at 2.5 cm, and 18 MV at 3.3 cm.

Mayneord F-Factor calculates PDD changes when converting between different Source-Skin Distances (SSD), overestimating PDD increases for large fields at deep depths.

Tissue Maximum Ratio (TMR) is the ratio of dose at a depth to dose at dmax at a constant distance from the source, making it entirely independent of SSD.

Tissue Air Ratio (TAR) is independent of SSD and is used primarily for low-energy photons, cobalt-60, and rotational SAD calculations.

Last updated: July 2026

Central Axis Depth Dose Physics

As a high-energy photon beam traverses biological tissue, energy is transferred to electrons via photoelectric absorption, Compton scattering, and pair production. These energetic recoil electrons travel through tissue, depositing absorbed dose along their paths. The dose profile along the central axis of the beam exhibits a characteristic curve comprising a build-up region , a point of maximum dose ( d max ⁡ d_{\max} d m a x ​ ) , and an exponential attenuation tail at depth. 100% | + dmax

  • | / \
  • 80% | / \ Exponential Attenuation Tail

| / \...

60% | / \...

| / Build-up \...

40% | / Region \...

  • +----------------------------------------> Depth (cm)
  • 0 dmax (1.5 cm) 10 cm 20 cm
  • Depth of Maximum Dose ( d max ⁡ d_{\max} d m a x ​ ) and Electronic Equilibrium

d max ⁡ d_{\max} d m a x ​ represents the depth at which kerma (Kinetic Energy Released per Unit Mass) equals absorbed dose, establishing longitudinal charged particle equilibrium (CPE) . In the build-up region between the skin surface and d max ⁡ d_{\max} d m a x ​ , kerma is maximum at the surface and decreases exponentially, while absorbed dose rises rapidly as secondary electrons are generated and travel forward into deeper tissue layers. Beyond d max ⁡ d_{\max} d m a x ​ , absorbed dose decreases exponentially due to photon attenuation.

Photon Energy / Beam Quality Depth of Maximum Dose ( d max ⁡ d_{\max} d m a x ​ ) Surface Dose (% of d max ⁡ d_{\max} d m a x ​ ) P D D ( 10 cm , 10 × 10 cm 2 , 100 cm SSD ) PDD(10\text{ cm}, 10\times10\text{ cm}^2, 100\text{ cm SSD}) P DD ( 10 cm , 10 × 10 cm 2 , 100 cm SSD )

  • Orthovoltage (250 kVp) 0.0 cm (Skin surface) 100% 35.0%
  • Cobalt-60 ( 60 Co ^{60}\text{Co} 60 Co ) 0.5 cm 50.0% 56.4%
  • 4 MV Photons 1.0 cm 40.0% 61.0%
  • 6 MV Photons 1.5 cm 25.0% – 30.0% 67.0%
  • 10 MV Photons 2.5 cm 20.0% 73.0%
  • 15 MV Photons 3.0 cm 15.0% 77.0%
  • 18 MV Photons 3.3 cm – 3.5 cm 10.0% – 12.0% 80.0%
  • Percentage Depth Dose (PDD) Characteristics
  • Mathematical Definition

Percentage Depth Dose (PDD) is defined as the ratio of absorbed dose at depth ( d d d ) to absorbed dose at reference depth ( d max ⁡ d_{\max} d m a x ​ ) along the central axis of the beam, expressed as a percentage. PDD is measured at a fixed Source-to-Skin Distance (SSD) (typically 100 cm SSD) in a water phantom. PDD ( d , r , f ) = ( D d D d max ⁡ ) × 100 % \text{PDD}(d, r, f) = \left( \frac{D_d}{D_{d_{\max}}} \right) \times 100\% PDD ( d , r , f ) = ( D d m a x ​ ​ D d ​ ​ ) × 100%

Where d d d is depth, r r r is field size at the skin surface, and f f f is SSD.

Physical Parameters Influencing PDD

Beam Energy (Directly Proportional): Higher energy photons possess greater penetrating power (lower linear attenuation coefficient μ \mu μ ), shifting d max ⁡ d_{\max} d m a x ​ deeper into tissue and increasing PDD at depth.

Depth (Inversely Proportional): Beyond d max ⁡ d_{\max} d m a x ​ , photon attenuation and beam divergence cause PDD to decay exponentially with increasing depth.

Field Size (Directly Proportional): As field size increases, a larger volume of surrounding phantom tissue is irradiated, generating increased phantom scatter radiation reaching the central axis, which increases PDD .

Source-Skin Distance (Directly Proportional): Increasing SSD reduces beam divergence according to the Inverse Square Law over the depth interval between d max ⁡ d_{\max} d m a x ​ and depth d d d , which increases PDD .

Mayneord F-Factor for Extended Distance Calculations

When a patient is treated at an extended SSD (e.g., treating a total body irradiation or whole-leg sarcoma field at 120 cm SSD instead of 100 cm SSD), PDD increases. The Mayneord F-Factor is derived from the Inverse Square Law to calculate the new PDD ( P D D 2 PDD_2 P D D 2 ​ ) at extended distance f 2 f_2 f 2 ​ from the known PDD ( P D D 1 PDD_1 P D D 1 ​ ) at distance f 1 f_1 f 1 ​ : F = ( f 2 + d max ⁡ f 1 + d max ⁡ ) 2 × ( f 1 + d f 2 + d ) 2 F = \left( \frac{f_2 + d_{\max}}{f_1 + d_{\max}} \right)^2 \times \left( \frac{f_1 + d}{f_2 + d} \right)^2 F = ( f 1 ​ + d m a x ​ f 2 ​ + d m a x ​ ​ ) 2 × ( f 2 ​ + d f 1 ​ + d ​ ) 2 P D D 2 = P D D 1 × F PDD_2 = PDD_1 \times F P D D 2 ​ = P D D 1 ​ × F

Clinical Worked Example: A 6 MV photon beam ( d max ⁡ = 1.5 cm d_{\max} = 1.5\text{ cm} d m a x ​ = 1.5 cm ) delivers a PDD of 67.0% at depth d = 10 cm d = 10\text{ cm} d = 10 cm for f 1 = 100 cm f_1 = 100\text{ cm} f 1 ​ = 100 cm SSD. Calculate the new PDD at an extended distance f 2 = 120 cm f_2 = 120\text{ cm} f 2 ​ = 120 cm SSD: F = ( 120 + 1.5 100 + 1.5 ) 2 × ( 100 + 10 120 + 10 ) 2 = ( 121.5 101.5 ) 2 × ( 110 130 ) 2 = ( 1.197 ) 2 × ( 0.846 ) 2 = 1.433 × 0.716 = 1.026 F = \left( \frac{120 + 1.5}{100 + 1.5} \right)^2 \times \left( \frac{100 + 10}{120 + 10} \right)^2 = \left( \frac{121.5}{101.5} \right)^2 \times \left( \frac{110}{130} \right)^2 = (1.197)^2 \times (0.846)^2 = 1.433 \times 0.716 = 1.026 F = ( 100 + 1.5 120 + 1.5 ​ ) 2 × ( 120 + 10 100 + 10 ​ ) 2 = ( 101.5 121.5 ​ ) 2 × ( 130 110 ​ ) 2 = ( 1.197 ) 2 × ( 0.846 ) 2 = 1.433 × 0.716 = 1.026 P D D 2 = 67.0 % × 1.026 = 68.7 % PDD_2 = 67.0\% \times 1.026 = 68.7\% P D D 2 ​ = 67.0% × 1.026 = 68.7%

Note on Mayneord F-Factor Limitations: The Mayneord F-Factor works well for small field sizes at moderate depths, but overestimates the PDD increase for large field sizes and deep depths because it ignores the increased phantom scatter generated at extended distances.

  • Iso-Distance Dosimetric Parameters: TMR, TPR, and TAR
  • Tissue Maximum Ratio (TMR)

In Source-to-Axis Distance (SAD) setups (isocentric technique), the machine isocenter remains at a fixed distance (100 cm) while patient depth varies. Using PDD is cumbersome because SSD changes with gantry rotation. Tissue Maximum Ratio (TMR) is defined as the ratio of absorbed dose at depth d d d to absorbed dose at d max ⁡ d_{\max} d m a x ​ , measured at a constant distance from the radiation source . TMR ( d , r d ) = D d D d max ⁡ \text{TMR}(d, r_d) = \frac{D_d}{D_{d_{\max}}} TMR ( d , r d ​ ) = D d m a x ​ ​ D d ​ ​

Fundamental Physics Principle: TMR is completely independent of Source-to-Skin Distance (SSD) . TMR depends exclusively on beam energy, depth d d d , and field size projected at depth r d r_d r d ​ .

Tissue Phantom Ratio (TPR) and Beam Quality Specifier ( T P R 20 , 10 TPR_{20,10} TP R 20 , 10 ​ )

Tissue Phantom Ratio (TPR) is the ratio of dose at depth d d d to dose at a fixed reference depth d ref d_{\text{ref}} d ref ​ (typically 5 cm or 10 cm) at a constant source-to-detector distance. TPR ( d , d ref , r d ) = D d D d ref \text{TPR}(d, d_{\text{ref}}, r_d) = \frac{D_d}{D_{d_{\text{ref}}}} TPR ( d , d ref ​ , r d ​ ) = D d ref ​ ​ D d ​ ​

The ratio of TPR at 20 cm depth to TPR at 10 cm depth in water ( T P R 20 , 10 TPR_{20,10} TP R 20 , 10 ​ ) serves as the international standard beam quality index under IAEA TRS-398 and AAPM TG-51 protocols to specify photon beam energy independently of electron contamination.

Tissue Air Ratio (TAR)

Tissue Air Ratio (TAR) is the ratio of absorbed dose at depth d d d in phantom to absorbed dose at the same point in free space (air). TAR is independent of SSD and is used primarily for low-energy photons, Cobalt-60, and radionuclide calculations. At high photon energies (> 10 MV), measuring dose in free space becomes physically unreliable due to large electron contamination cap requirements.

  • Parameter Reference Setup Distance Constraint SSD Dependence Clinical Application
  • PDD Fixed SSD (100 cm) Surface at 100 cm Dependent on SSD Non-isocentric SSD setups, single electron fields
  • TMR Fixed SAD (100 cm) Detector at 100 cm Independent of SSD Isocentric 3D-CRT, IMRT, VMAT MU calculations

TPR Fixed SAD ( d ref d_{\text{ref}} d ref ​ ) Detector at 100 cm Independent of SSD High-energy photon reference dosimetry ( T P R 20 , 10 TPR_{20,10} TP R 20 , 10 ​ )

TAR Fixed SAD (Free space) Detector at 100 cm Independent of SSD Orthovoltage, 60 Co ^{60}\text{Co} 60 Co , rotational arc dosimetry