Study guide

15.2 Laboratory Biosafety, Chemical Hygiene & Waste

ASCP Molecular Biology Technologist MB

Track your progress as you read

Finished sections get a check mark.

1 Introduction & Exam Strategy

1.1 ASCP MB Exam Structure, Eligibility Routes & Policies 1.2 Computer Adaptive Testing (CAT) & Scoring Mechanics 1.3 Blueprint Navigation & Preparation Strategy

2 Nucleic Acid Chemistry & Molecular Structure

2.1 DNA Structure, Nitrogenous Bases & Thermodynamic Stability 2.2 RNA Types, Secondary Structure & Chemical Lability 2.3 Chromatin Organization, Histone Modifications & DNA Methylation

3 Central Dogma, Genome Organization & Cytogenetics

3.1 DNA Replication Machinery & Enzymology 3.2 Transcription, Translation & Genetic Code 3.3 Mutations, Polymorphisms & Chromosomal Aberrations

4 Biochemical Reagents & Enzymes

4.1 DNA Polymerases, Reverse Transcriptases & Proofreading 4.2 Restriction Endonucleases, Ligases & Nucleases 4.3 Reaction Buffers, Cations & Master Mix Additives

5 Nucleic Acid Isolation & Purification

5.1 Organic & Inorganic Extraction Chemistries 5.2 Solid-Phase Silica Columns & Magnetic Bead Isolation 5.3 Specimen-Specific Lysis & Challenging Matrices

6 Nucleic Acid Quantification & Electrophoresis

6.1 UV Spectrophotometry & Fluorometry 6.2 Agarose & Polyacrylamide Gel Electrophoresis 6.3 Capillary Electrophoresis & Microfluidic Bioanalyzers 6.4 Mass Spectrometry (MALDI-TOF) in Molecular Diagnostics

7 PCR Principles & Primer Design Optimization

7.1 Conventional PCR Thermodynamics & Cycling Kinetics 7.2 Primer Design Criteria, Melting Temp & Secondary Structures 7.3 PCR Inhibition, Facilitators & Troubleshooting

8 Advanced Amplification Technologies

8.1 Real-Time qPCR Probe Chemistries & Fluorescent Dyes 8.2 qPCR Quantification (Standard Curve, ΔΔCt) & Dissociation Melt Curve Analysis 8.3 Digital PCR (dPCR), Droplet Partitioning & Poisson Absolute Quantification 8.4 Isothermal & Target Multiplex Amplification

9 DNA Sequencing Technologies

9.1 Sanger Dideoxy Sequencing & Electropherogram Interpretation 9.2 NGS Library Preparation, Fragmentation & Target Enrichment 9.3 Massively Parallel NGS Platforms & Chemistries 9.4 NGS Quality Metrics, Coverage Depth & Bioinformatics

10 Hybridization, Blotting & Cytogenetic Assays

10.1 Southern, Northern & Dot Blot Chemistries 10.2 Fluorescence In Situ Hybridization (FISH) & CISH 10.3 Microarrays & Array Comparative Genomic Hybridization (aCGH)

11 Molecular Infectious Disease Diagnostics

11.1 Quantitative Viral Load Monitoring (HIV-1, HCV, HBV, CMV) 11.2 Qualitative Multiplex Syndromic Pathogen Panels 11.3 Antimicrobial Resistance Genotyping 11.4 Molecular Epidemiology & Pathogen Typing

12 Molecular Oncology & Hematopathology

12.1 Hematologic Malignancies & Translocation Assays 12.2 Solid Tumor Somatic Driver Mutations 12.3 Microsatellite Instability & DNA Mismatch Repair 12.4 Liquid Biopsy, ctDNA & Minimal Residual Disease

13 Medical Genetics, Pharmacogenomics & Identity Testing

13.1 Single-Gene Mendelian Disorders & CFTR Testing 13.2 Trinucleotide Repeat Expansion Disorders 13.3 Clinical Pharmacogenomics (PGx) Core Alleles 13.4 Human Identity Testing & Post-Transplant Chimerism

14 Quality Assurance, Controls & Method Validation

14.1 Quality Assurance Controls & Reference Materials 14.2 Clinical Method Validation & Performance Characteristics 14.3 Quality Management, Proficiency Testing & Competency

15 Laboratory Operations, Safety & Clinical Variant Reporting

15.1 Physical Facility Design & Unidirectional Workflow 15.2 Laboratory Biosafety, Chemical Hygiene & Waste 15.3 Clinical Calculations, ACMG Variant Classification & HGVS

15.2 Laboratory Biosafety, Chemical Hygiene & Waste

Key Takeaways Explain This Topic AI

Biosafety Levels (BSL-1 to BSL-4) define physical containment based on agent infectivity, transmission mode, and severity, with clinical diagnostic molecular labs predominantly operating under BSL-2 standards.

Class II Biological Safety Cabinets (BSCs) provide personnel, product, and environmental protection via vertical laminar HEPA filtration; laminar flow clean benches protect ONLY the product and must NEVER be used with human pathogens.

Class II Type B2 (100% total exhaust, 0% recirculation) BSCs or chemical fume hoods are mandatory when biological procedures generate volatile toxic chemical vapors or utilize radionuclides.

Hazardous molecular reagents possess specific toxicities: phenol causes painless chemical burns and systemic toxicity; unpolymerized acrylamide is a cumulative neurotoxin; and ethidium bromide is a potent intercalating mutagen.

Guanidine thiocyanate or guanidinium hydrochloride in lysis buffers must NEVER be mixed with sodium hypochlorite bleach, as the reaction generates lethal toxic hydrogen cyanide (HCN) gas.

Last updated: August 2026

15.2 Laboratory Biosafety, Chemical Hygiene & Waste

Quick Summary: Molecular biology laboratories routinely manipulate high-risk infectious pathogens (e.g., HIV, Hepatitis B/C, Mycobacterium tuberculosis , SARS-CoV-2) alongside hazardous chemical reagents (e.g., phenol, chloroform, ethidium bromide, formamide, acrylamide, guanidinium salts). Strict compliance with Biosafety Levels (BSL-1 through BSL-4) , OSHA Chemical Hygiene Plans (29 CFR 1910.1450) , and EPA Resource Conservation and Recovery Act (RCRA) standards is mandatory. Technologists must distinguish between Class II Biological Safety Cabinets (BSCs) —which protect personnel, product, and environment—and laminar flow clean benches , which protect only the product and blow hazardous bioaerosols directly into the operator's breathing zone. A paramount safety rule is that guanidine-containing lysis buffers must NEVER be mixed with sodium hypochlorite (bleach) due to the immediate liberation of deadly hydrogen cyanide ( H C N HCN H CN ) gas .

1. Biosafety Levels (BSL-1 to BSL-4) & Containment Principles

The Centers for Disease Control and Prevention (CDC) and National Institutes of Health (NIH) define four Biosafety Levels ( Biosafety in Microbiological and Biomedical Laboratories , BMBL 6th Edition) based on pathogenic risk: +----------------------------------------------------------------------------------------------------+

  • | CDC/NIH BIOSAFETY LEVELS (BSL-1 TO BSL-4) MATRIX |
  • +-------+-------------------+-------------------+--------------------+-------------------------------+
  • | Level | Risk Group / Pathogen| Transmission Route| Primary Containment| Facility Engineering & |
  • | | Characteristics | & Disease Severity| & Safety Equipment | Secondary Barriers |
  • +-------+-------------------+-------------------+--------------------+-------------------------------+
  • |**BSL-1| Well-characterized| Non-pathogenic to | Open benchtop; PPE | Sink for handwashing; standard|
  • | | agents (non-patho-| immunocompetent | (lab coat, gloves, | doors; easily cleanable |
  • | | genic *E. coli* | adults; minimal | eye protection as | benchtops; no special |
  • | | K12, *B. subtilis*)| hazard | needed) | containment engineering |
  • +-------+-------------------+-------------------+--------------------+-------------------------------+
  • |**BSL-2| Moderate-risk human| Percutaneous | **Class II BSC** | Autoclave available; biohazard|
  • | | pathogens (HBV, | injury, ingestion,| for aerosol- | signage; self-closing doors; |
  • | | HCV, HIV, *S. | mucous membrane | generating tasks; | eyewash station; certified BSC|
  • | | aureus*, HSV) | contact | PPE (gloves, coat) | annual inspection |
  • +-------+-------------------+-------------------+--------------------+-------------------------------+
  • |**BSL-3| Indigenous/exotic | **Aerosol / | **Class II or III | Controlled access; double-door|
  • | | agents (*M. | inhalation**; | BSC**; respiratory | entry airlock; negative air; |
  • | | tuberculosis*, | serious or fatal | protection (N95 / | 100% HEPA-filtered exhaust; |
  • | | *B. anthracis*) | systemic disease | PAPR); full PPE | non-recirculating airflow |
  • +-------+-------------------+-------------------+--------------------+-------------------------------+
  • |**BSL-4| Dangerous/exotic | **Aerosol-borne / | **Class III BSC** | Isolated facility; dedicated |
  • | | agents (Ebola, | unknown**; lethal;| or positive- | supply/exhaust HEPA; air lock |
  • | | Marburg, Lassa, | **no vaccine or | pressure suit with | decon shower; sealed conduits |
  • | | Smallpox virus) | specific therapy**| dedicated air line | and pass-through autoclave |
  • +-------+-------------------+-------------------+--------------------+-------------------------------+
  • Universal Precautions & Bloodborne Pathogens Standard

OSHA Bloodborne Pathogens Standard (29 CFR 1910.1030) : Mandates that all human blood, bone marrow, tissues, and body fluids be treated as potentially infectious for Human Immunodeficiency Virus (HIV), Hepatitis B Virus (HBV), Hepatitis C Virus (HCV), and other bloodborne pathogens.

Engineering and Work Practice Controls : Puncture-resistant sharps containers located at point of use, plastic blood collection tubes instead of glass, needle-recapping prohibition (or one-handed scoop technique only if unavoidable), and mandatory HBV vaccination offered free to all at-risk personnel.

2. Containment Equipment: BSCs, Clean Benches & Chemical Fume Hoods

Choosing the correct engineering containment device is vital for operator survival and sample integrity. CONTAINMENT EQUIPMENT AIRFLOW COMPARISON

  • [ Class II Type A2 BSC ] [ Laminar Flow Clean Bench ]
  • (Protects You & Product!) (DANGER: Protects Product ONLY!)
  • Room Air Inflow (Grille) HEPA Filter at Back
  • | |
  • v v
  • Vertical HEPA Downflow (Air Curtain) Horizontal Clean Airflow
  • | |
  • v v
  • Work Surface (Split Airflow) Blows ACROSS Work Surface
  • / \ |
  • v v v
  • 70% Recirculated 30% Exhausted *** BLOWS PATHOGENS & AEROSOLS ***

(via Supply HEPA) (via Exhaust HEPA) *** DIRECTLY INTO OPERATOR FACE! *** +----------------------------------------------------------------------------------------------------+

  • | CONTAINMENT AND HOOD MECHANISM COMPARISON |
  • +-------------------+-------------------+-------------------+-------------------+--------------------+
  • | Device Type | Personnel | Product (Sample) | Exhaust Air Flow | Safe for Toxic |
  • | | Protection | Protection | Destination | Chemicals / Bio? |
  • +-------------------+-------------------+-------------------+-------------------+--------------------+
  • | **Class I BSC** | **YES** | **NO** (Room air | HEPA-filtered to | Biologicals only; |
  • | | (Inward face | sweeps across | room or facility | NO volatile toxic |
  • | | velocity 75 fpm) | dirty sample) | exhaust | chemicals |
  • +-------------------+-------------------+-------------------+-------------------+--------------------+
  • | **Class II BSC | **YES** | **YES** | 70% recirculated, | Minute volatile |
  • | (Type A1 / A2)** | (Inward air | (Vertical laminar | 30% HEPA exhaust | chemicals ONLY if |
  • | | barrier curtain) | HEPA downflow) | to room / thimble | thimble-ducted |
  • +-------------------+-------------------+-------------------+-------------------+--------------------+
  • | **Class II BSC | **YES** | **YES** | **100% Exhausted**| **YES**: Volatile |
  • | (Type B2 - Total | (Inward velocity | (Vertical laminar | to outdoors via | toxic chemicals & |
  • | Exhaust)** | 100 fpm) | HEPA downflow) | dedicated duct | radionuclides |
  • +-------------------+-------------------+-------------------+-------------------+--------------------+
  • | **Class III BSC | **YES (Maximum)** | **YES (Maximum)** | 100% double-HEPA | BSL-4 pathogens & |
  • | (Glove Box)** | Gas-tight seal, | Fully enclosed, | or incinerated | high-toxicity bio- |
  • | | heavy gloves | sterile chamber | exhaust | hazard agents |
  • +-------------------+-------------------+-------------------+-------------------+--------------------+
  • | **Laminar Flow | **NO (ZERO!)** | **YES** | Blows directly | **NEVER USE FOR |
  • | Clean Bench** | *Exposes worker | (Horizontal / | into room and | PATHOGENS, BLOOD, |
  • | | to aerosols!* | vertical HEPA) | operator face | OR CHEMICALS!* |
  • +-------------------+-------------------+-------------------+-------------------+--------------------+
  • | **Chemical Fume | **YES** | **NO** | 100% exhausted to | **YES**: Toxic and |
  • | Hood** | (Captures toxic | (Turbulent un- | outdoors (NO | flammable chemical |
  • | | chemical vapors) | filtered airflow) | HEPA filter) | vapors ONLY |
  • +-------------------+-------------------+-------------------+-------------------+--------------------+
  • Critical Operational Rules for Biosafety Cabinets

Purge Time : Run blower for 4 to 5 minutes prior to beginning work and after completion to purge air particulates.

Grille Integrity : Never block front intake grilles or rear exhaust grilles with arms, pipettes, or absorbent towels. Blocking grilles breaks the laminar air barrier curtain, allowing room contaminants in and escaping aerosols out.

Surface Disinfection : Wipe interior walls, glass sash, and stainless steel drain pan with 10% bleach followed by 70% ethanol before and after each session.

Certification : BSCs must be certified upon installation, annually, and whenever moved, verifying HEPA filter integrity, downflow velocity, and face inflow velocity (minimum 75 − 100 feet per minute 75 - 100\text{ feet per minute} 75 − 100 feet per minute ).

3. Chemical Hygiene & Hazardous Molecular Biology Reagents

The OSHA Laboratory Standard (29 CFR 1910.1450) requires every clinical molecular laboratory to maintain a written Chemical Hygiene Plan (CHP) , appoint a qualified Chemical Hygiene Officer (CHO) , and provide Safety Data Sheets ( SDS , containing 16 standardized GHS sections) for all hazardous chemicals. +----------------------------------------------------------------------------------------------------+

  • | HAZARDOUS MOLECULAR REAGENTS & CLINICAL TOXICOLOGY |
  • +-------------------+-------------------+-------------------+----------------------------------------+
  • | Chemical Reagent | Molecular Function| Toxicology & Risk | Safe Handling, PPE & Disposal Protocol |
  • +-------------------+-------------------+-------------------+----------------------------------------+
  • | **Phenol / | Phase separation | **Severe Chemical | Handled strictly in **chemical fume |
  • | Chloroform / | during organic | Burns & Systemic | hood**; double nitrile gloves; if skin |
  • | Isoamyl Alcohol** | RNA/DNA extraction| Neuro/Cardiotoxin**| exposed, rinse with PEG 300/400 or water|
  • | (25:24:1) | | Chloroform: hepatic| Collect in **Halogenated Organic Waste**|
  • +-------------------+-------------------+-------------------+----------------------------------------+
  • | **Ethidium | Intercalates dsDNA| **Potent Mutagen | Handle with nitrile gloves; decontaminate|
  • | Bromide (EtBr)** | for fluorescence | & Teratogen** | with activated charcoal (GreenBag) or |
  • | | under UV light | Binds nuclear DNA | chemical destruction; hazardous waste |
  • +-------------------+-------------------+-------------------+----------------------------------------+
  • | **Guanidine | Chaotropic protein| **DEADLY TOXIC GAS| NEVER MIX WITH BLEACH (forms toxic HCN |
  • | Thiocyanate / | denaturant in RNA | HAZARD**: Reacts | gas!); flush spills with copious water |
  • | Guanidinium HCl** | lysis buffers | with bleach -> HCN| & detergent; dispose as non-halogenated|
  • +-------------------+-------------------+-------------------+----------------------------------------+
  • | **Formamide** | Lowers DNA melting| **Reproductive & | Fume hood handling; nitrile gloves; |
  • | | temp ($T_m$) in | Teratogenic Toxin**| strictly prohibited for pregnant staff |
  • | | FISH / sequencing | Fetal damage | without specialized containment |
  • +-------------------+-------------------+-------------------+----------------------------------------+
  • | **Unpolymerized | Monomer matrix for| **Cumulative Potent| Wear nitrile gloves, lab coat, face |
  • | Acrylamide** | high-resolution | Neurotoxin** | shield; prepare in fume hood; polymerized|
  • | (Powder / Soln) | PAGE gels | Absorbs via skin | polyacrylamide gels are non-toxic |
  • +-------------------+-------------------+-------------------+----------------------------------------+
  • Critical Chemical Incompatibilities: Guanidinium and Bleach

Most commercial silica-column and magnetic-bead viral and cellular lysis buffers (e.g., Qiagen QIAamp, Roche MagNA Pure) utilize concentrated guanidine thiocyanate ( G u S C N GuSCN G u SCN ) or guanidine hydrochloride ( G u H C l GuHCl G u H Cl ) to denature RNases and cellular structures. Guanidine Thiocyanate [ C 2 H 6 N 4 S ] + Sodium Hypochlorite [ N a O C l ] ⟶ Hydrogen Cyanide Gas [ H C N ↑ ] + Toxic Chlorine/Sulfur Byproducts \text{Guanidine Thiocyanate } [C_2H_6N_4S] + \text{Sodium Hypochlorite } [NaOCl] \longrightarrow \text{Hydrogen Cyanide Gas } [HCN \uparrow] + \text{Toxic Chlorine/Sulfur Byproducts} Guanidine Thiocyanate [ C 2 ​ H 6 ​ N 4 ​ S ] + Sodium Hypochlorite [ N a OCl ] ⟶ Hydrogen Cyanide Gas [ H CN ↑ ] + Toxic Chlorine/Sulfur Byproducts

EMERGENCY HAZARD WARNING: When sodium hypochlorite (bleach) contacts guanidine thiocyanate, an immediate chemical reaction liberates hydrogen cyanide ( H C N HCN H CN ) gas , a potent cellular asphyxiant that inhibits cytochrome c oxidase in mitochondria, causing rapid loss of consciousness and death.

Spill Protocol: If a guanidine-containing lysis buffer spills, DO NOT APPLY BLEACH . Clean the spill using absorbent pads, mild detergent, and copious sterile water. Autoclaving guanidine solutions is also strictly prohibited because heat decomposes thiocyanate into toxic gases.

Specific Chemical First-Aid Protocols

Phenol Exposure : Phenol acts as a local anesthetic, causing deep, painless, white coagulative necrotic burns while rapidly absorbing into systemic circulation to cause cardiac arrhythmias and renal failure.

Immediate Action : Flush the affected skin immediately with Polyethylene Glycol (PEG 300 or PEG 400) or copious volumes of running water for at least 15 minutes . Seek immediate emergency medical care.

Ethidium Bromide Spills & Alternatives :

Filter liquid EtBr solutions through activated charcoal cartridges (e.g., Extran or GreenBag systems) before aqueous discharge, or incinerate as hazardous chemical waste.

Modern molecular laboratories substitute EtBr with safer, less mutagenic dyes such as GelGreen , GelRed , or SYBR Safe , which do not readily permeate intact eukaryotic cell membranes.

4. Hazardous Waste Management & Segregation

Under EPA Resource Conservation and Recovery Act ( RCRA ) regulations, laboratories must segregate waste streams at the point of generation to prevent catastrophic reactions and environmental contamination. LABORATORY WASTE STREAM SEGREGATION

  • [ Waste Generated at Molecular Bench ]
  • |
  • +---------------+---------------+---------------+
  • | | |
  • v v v
  • [ Biohazardous Waste ] [ Chemical Waste ] [ Sharps Waste ]
  • Red biohazard bags - Halogenated vs. - Rigid puncture-proof
  • Blood tubes, tips, swabs Non-Halogenated plastic container
  • Autoclave: 121°C, 15 psi - Labeled with GHS - Needles, scalpel
  • Spore test: *G. stearo-* - EPA manifest / blades, glass slides

*thermophilus* validated hazardous disposal - Discard when 2/3 full!

Biological vs. Chemical Waste Protocols

Regulated Medical Waste (Biohazard) :

Disposed of in heavy-duty, leak-proof red biohazard bags bearing the international biohazard symbol.

Decontaminated on-site via steam autoclaving at 121 ∘ C 121^\circ\text{C} 12 1 ∘ C ( 250 ∘ F 250^\circ\text{F} 25 0 ∘ F ) at 15 psi 15\text{ psi} 15 psi for 30 to 60 minutes .

Biological Autoclave Validation : Verified monthly using heat-resistant endospores of Geobacillus stearothermophilus (incubated at 56 − 60 ∘ C 56 - 60^\circ\text{C} 56 − 6 0 ∘ C ). Chemical indicator tape (Bowie-Dick) confirms heat exposure but does not guarantee sterility.

Hazardous Chemical Waste Streams :

Halogenated Organic Solvents : Chloroform, dichloromethane, chlorobenzene. Must be stored in dedicated solvent safety cans labeled "Halogenated Organics."

Non-Halogenated Organic Solvents : Ethanol, methanol, isopropanol, xylene, acetone. Must never be mixed with halogenated waste.

Corrosive Acids & Bases : Concentrated hydrochloric acid, acetic acid, sodium hydroxide. Stored in dedicated corrosive cabinets in secondary containment trays.

Sharps Disposal :

Scalpel blades, hypodermic needles, glass Pasteur pipettes, and broken capillary tubes must be immediately discarded into rigid, puncture-resistant, leak-proof sharps containers .

Containers must be sealed and replaced when they reach the fill line ( 2 / 3 2/3 2/3 to 3 / 4 3/4 3/4 full ); never overfill or force sharps into a container.

5. Hazard Communication, NFPA 704 Diamond & Laboratory Safety Equipment

The National Fire Protection Association ( NFPA 704 ) Standard System identifies chemical hazards during emergency responses. THE NFPA 704 HAZARD DIAMOND

  • [ RED: Flammability ]
  • 0 = Will not burn
  • 1 = Flash point > 200°F
  • 2 = Flash point 100-200°F
  • 3 = Flash point 73-100°F
  • 4 = Flash point < 73°F
  • /\
  • / \
  • [ BLUE: Health ] / \ [ YELLOW: Instability ]
  • 0 = Normal material / \ / \ 0 = Normally stable
  • 1 = Slightly hazardous / \/ \ 1 = Unstable if heated
  • 2 = Hazardous / 2 /\ 3 \ 2 = Violent chemical change
  • 3 = Extreme danger \ / \ / 3 = Shock / heat may detonate
  • 4 = Deadly \ / 0 \ / 4 = May detonate
  • \/ W \/
  • [ WHITE: Special ]
  • W = Reacts violently with water
  • OX = Strong oxidizer
  • SA = Simple asphyxiant gas
  • BIO / COR / ACID / ALK
  • Emergency Eyewash & Safety Shower Requirements (ANSI Z358.1)

Accessibility : Located within a 10-second unobstructed walking path ( ≤ 55 feet \le 55\text{ feet} ≤ 55 feet ) of any hazardous chemical work zone on the same floor.

Flushing Parameters : Must deliver continuous, hands-free tepid fluid ( 16 − 38 ∘ C 16 - 38^\circ\text{C} 16 − 3 8 ∘ C / 60 − 100 ∘ F 60 - 100^\circ\text{F} 60 − 10 0 ∘ F ) for a minimum duration of 15 minutes at a flow rate of ≥ 0.4 gpm \ge 0.4\text{ gpm} ≥ 0.4 gpm (eyewashes) or ≥ 20 gpm \ge 20\text{ gpm} ≥ 20 gpm (safety showers).

Maintenance : Eyewashes must be activated and flushed weekly to clear microbial biofilm and sediment; safety showers tested annually.

Fire Extinguisher Classification +----------------------------------------------------------------------------------------------------+

  • | FIRE EXTINGUISHER CLASSIFICATION & APPLICATION |
  • +-------+-------------------+-------------------+----------------------------------------------------+
  • | Class | Fuel Type / Source| Extinguishing | Laboratory Examples & Mechanism |
  • | | | Agent | |
  • +-------+-------------------+-------------------+----------------------------------------------------+
  • | **Class A**| Ordinary combustibles| Water, Dry chemical| Paper, cardboard, wood, cotton lab coats; |
  • | | (leave ash) | | cools fuel below ignition temperature |
  • +-------+-------------------+-------------------+----------------------------------------------------+
  • | **Class B**| Flammable liquids | $ ext{CO}_2$, Dry| Ethanol, isopropanol, xylene, phenol, ether; |
  • | | and gases | chemical, foam | smothers oxygen and suppresses chemical radicals |
  • +-------+-------------------+-------------------+----------------------------------------------------+
  • | **Class C**| Energized electrical| $ ext{CO}_2$, Dry| Thermocyclers, sequencers, power supplies, PCs; |
  • | | equipment | chemical | non-conductive agent prevents operator electrocution|
  • +-------+-------------------+-------------------+----------------------------------------------------+
  • | **Class D**| Combustible metals| Dry powder (NaCl, | Sodium, magnesium, potassium, lithium; |
  • | | | copper powder) | smothers metal without explosive hydrogen release |
  • +-------+-------------------+-------------------+----------------------------------------------------+
  • | **Class K**| Commercial cooking| Wet chemical | Kitchen fats and oils (rare in molecular labs) |
  • | | oils and fats | (potassium acetate)| saponifies fats into non-flammable soap layer |
  • +-------+-------------------+-------------------+----------------------------------------------------+

Loading diagram... Laboratory Engineering Controls and Safety Selection Tree