Study guide

Temperature and Humidity Control

ASQ Certified Calibration Technician CCT

Track your progress as you read

Finished sections get a check mark.

1 SI Definitions and Realizations

The Seven SI Base Units and the 2019 Revision Realizing SI Units in the Laboratory

2 Derived Units, Prefixes, and Constants

Derived Units and Physical Calculations SI Prefixes and Compound-Unit Conversions Physical Constants and Their Applications

3 Measurement Standards and Selection

Types and Roles of Measurement Standards Selecting Standards for an Operational Hierarchy

4 Traceability and Standard Substitution

Traceability Evidence and Impact Review Evaluating and Authorizing Standard Substitution

5 Dimensional and Mechanical Metrology

Linear Displacement and Dimensional Instruments Geometry, Thermal Expansion, and Mass Force, Torque, Pressure, Density, and Viscosity

6 Temperature, Humidity, Volume, and Flow

Temperature Sensors and Thermometer Corrections Humidity, Volume, and Flow

7 Electrical, Optical, and Time Measurements

Voltage, Current, Resistance, and AC Measurement Time, Frequency, Light, Power, and Energy

8 Measurement Methods

Direct, Indirect, and Ratio Methods Transfer, Differential, and UUT Substitution Methods

9 Instrument Characteristics and Measurement Data

Sensitivity, Variability, and Precision Bias, Linearity, Stability, and Hysteresis Data Formats, Readability, and Suitability Resolution, Confidentiality, and Data Integrity

10 Specifications and Error Sources

Interpreting IM&TE Specification Terms Specification Calculations and Operating Qualifiers Instrument, Operator, and Circuit Errors Environmental, Geometric, and Interference Errors

11 Measurement Assurance

Check Standards and Statistical Process Control Proficiency Comparisons and Gage R&R

12 Procedures and Governing Requirements

Procedure Structure and Published Practice Sources Method Development, Deviations, and Handling Regulations, Contracts, and Requirement Precedence

13 Instrument Adjustment and Standardization

Calibration, Zero, and Span Adjustment Linearization, Nulling, and Digital Adjustment

14 Calibration Environment

Temperature and Humidity Control ESD, Dust, Vibration, and Pressure

15 Operations, Roles, and Scheduling

Receiving, As-Found Testing, and Calibration Flow Asset Identification, Tracking, and Release Laboratory Roles and Initial Scheduling Interval Review, Overdue Control, and Workloads

16 Method and Software Validation

Method Verification and Validation Studies Software, Firmware, and Spreadsheet Validation

17 Records and Formal Reporting

Controlled Records and Calibration Certificates Status Labels and Out-of-Tolerance Reporting

18 Uncertainty Concepts and Decision Risk

Error, Uncertainty, and Measurand Definition TAR, TUR, Tolerance Consumption, and Capability Conformity Decisions and False-Accept Risk Guard Bands and Conformity Reporting

19 Uncertainty Models and Budgets

Measurement Models and Uncertainty Contributions Distribution Selection and Budget Assembly

20 Type A and Type B Evaluation

Type A Statistics and Repeatability Outlier Review and Worked Type A Analysis Type B Distribution Models

21 Uncertainty Reporting and Applied Mathematics

Propagation, Covariance, and Effective Degrees of Freedom Coverage Factors and Expanded-Uncertainty Reporting Notation, Unit Conversions, Ratios, and Decibels Rounding, Algebra, Graphs, and Angular Conversions

22 Quality Systems and Tools

Quality-System Components, Planning, and Improvement The Seven Basic Quality Tools

23 Audits and Corrective Action

Audit Types, Planning, and Nonconformance Control Impact Assessment, Root Cause, and Effectiveness

24 Safety, Standards, and Professional Conduct

Electrical, Pressure, Chemical, and Laboratory Hazards PPE, Safety Data Sheets, 6S, and Safe Restoration Quality Standards, GUM, and VIM Accreditation Scope and Professional Ethics

Temperature and Humidity Control

Key Takeaways Explain This Topic AI

ISO 1:2022 specifies 20°C as the standard reference temperature for dimensional properties.

Soak time depends on equilibrium, gradients, equipment, and uncertainty rather than one universal time-per-inch rule.

Humidity limits must address the actual equipment, materials, and measurement requirements.

Last updated: October 2026

The numerical environmental limits and material coefficients below are illustrative examples. The actual method, equipment specifications, uncertainty budget, and observed stability establish required limits. ISO/IEC 17025 does not prescribe one room temperature, RH band, cleanroom class, or soak time for every calibration.

Control, monitor, and record environmental conditions relevant to the measurement and applicable requirements. Select the parameters, limits, sampling, and response from actual influences and the method. ISO/IEC 17025 does not require continuous recording of every conceivable environmental quantity for every calibration.

Master Reference Table: Laboratory Environmental Parameters

The following table outlines the primary environmental influences in calibration facilities, their operational boundaries, and their direct metrological impacts:

Environmental Parameter Target Operating Band Dominant Metrological Impact Applicable Standards

Temperature (Dimensional) 20 ∘ C ± 0.5 ∘ C 20^\circ\text{C} \pm 0.5^\circ\text{C} 2 0 ∘ C ± 0. 5 ∘ C (Primary); 20 ∘ C ± 1.0 ∘ C 20^\circ\text{C} \pm 1.0^\circ\text{C} 2 0 ∘ C ± 1. 0 ∘ C (Secondary) Dimensional thermal expansion/contraction of standards and workpieces; thermal gradients across granite surface plates. ISO 1:2022

Temperature (General Lab) 23 ∘ C ± 2.0 ∘ C 23^\circ\text{C} \pm 2.0^\circ\text{C} 2 3 ∘ C ± 2. 0 ∘ C ( 73.4 ∘ F ± 3.6 ∘ F 73.4^\circ\text{F} \pm 3.6^\circ\text{F} 73. 4 ∘ F ± 3. 6 ∘ F ) Electronic component drift (Zener references, precision shunts); thermal EMFs at dissimilar metal junctions. General Metrology Practices

Rate of Temp Change < 0.5 ∘ C / hour < 0.5^\circ\text{C} / \text{hour} < 0. 5 ∘ C / hour Non-uniform internal thermal gradients in large gage blocks, optical flats, and mass standards. NCSL RP-7 / ISO 17025

Relative humidity Method-specific limits; 30–50% RH is an illustrative laboratory band, not a universal standard requirement. Low humidity can increase charging; moisture can affect corrosion and leakage. Condensation occurs when a surface is below dew point. Apply the relevant equipment and environmental procedure.

Electrostatic Discharge Static dissipative surfaces: 10 6 to 10 9 Ω 10^6\text{ to }10^9\ \Omega 1 0 6 to 1 0 9 Ω Latent or catastrophic gate-oxide destruction of CMOS, FET, and microelectronic circuits. ANSI/ESD S20.20 ; IEC 61340-5-1

Airborne Particulates ISO Class 7 or Class 8 (ISO Class 5 for optical/mass) Particulate entrapment on gage block faces prevents optical wringing; bearing wear on analytical balances. ISO 14644-1:2015

Room Differential Pressure + 10 to + 25 Pa +10\text{ to }+25\text{ Pa} + 10 to + 25 Pa relative to corridors Prevents dust, pollen, and turbulent non-conditioned air ingress when laboratory doors open. Cleanroom Metrology

Mechanical vibration Use the relevant frequency-dependent limits and measurement method. Balance noise, optical fringe motion, or roundness errors. VC-A through VC-E are distinct criteria, not all a 1–3 µm/s band. Validate the actual isolation and measurement environment.

Barometric Pressure Ambient ( 80 to 105 kPa 80\text{ to }105\text{ kPa} 80 to 105 kPa ), continuously logged Air buoyancy corrections for deadweight testers and mass standards; refractive index of air ( n air n_{\text{air}} n air ​ ) in laser interferometry. Edlén Formula / OIML R 111

Temperature Control, ISO 1, and Thermal Stabilization

Temperature is the single most pervasive source of measurement uncertainty across physical, dimensional, and mechanical metrology.

ISO 1 Standard Reference Temperature

ISO 1:2022 ( Geometrical product specifications — Standard reference temperature for the specification of geometrical and dimensional properties ) defines: T ref = 20 ∘ C ( 68 ∘ F = 293.15 K ) T_{\text{ref}} = 20^\circ\text{C} \quad (68^\circ\text{F} = 293.15\text{ K}) T ref ​ = 2 0 ∘ C ( 6 8 ∘ F = 293.15 K )

ISO 1 specifies 20°C as the standard reference temperature for geometrical and dimensional properties. Measurements can be made at other temperatures when the method accounts for expansion and uncertainty. A drawing can state a different specified reference condition; there is no blanket rule that every dimension away from 20°C is legally invalid.

Linear Thermal Expansion

When an object is heated or cooled, its physical length changes according to the linear expansion law: Δ L = L 0 ⋅ α ⋅ Δ T = L 0 ⋅ α ⋅ ( T − 20 ∘ C ) \Delta L = L_0 \cdot \alpha \cdot \Delta T = L_0 \cdot \alpha \cdot (T - 20^\circ\text{C}) Δ L = L 0 ​ ⋅ α ⋅ Δ T = L 0 ​ ⋅ α ⋅ ( T − 2 0 ∘ C )

Where:

L 0 L_0 L 0 ​ is the nominal length at 20 ∘ C 20^\circ\text{C} 2 0 ∘ C (meters or millimeters)

α \alpha α is the Coefficient of Thermal Expansion (CTE) in 1 / ∘ C 1/^\circ\text{C} 1 / ∘ C or ppm / ∘ C \text{ppm}/^\circ\text{C} ppm / ∘ C ( × 10 − 6 / ∘ C \times 10^{-6}/^\circ\text{C} × 1 0 − 6 / ∘ C )

Δ T = T − 20 ∘ C \Delta T = T - 20^\circ\text{C} Δ T = T − 2 0 ∘ C is the temperature deviation from the standard reference temperature

Differential Thermal Expansion Equation

In precision dimensional calibrations, both the Unit Under Test (workpiece) and the reference standard expand. The differential thermal expansion error ( δ L \delta L δ L ) is given by: δ L = L 0 [ α work ( T work − 20 ∘ C ) − α std ( T std − 20 ∘ C ) ] \delta L = L_0 \left[ \alpha_{\text{work}} (T_{\text{work}} - 20^\circ\text{C}) - \alpha_{\text{std}} (T_{\text{std}} - 20^\circ\text{C}) \right] δ L = L 0 ​ [ α work ​ ( T work ​ − 2 0 ∘ C ) − α std ​ ( T std ​ − 2 0 ∘ C ) ]

Material Nominal CTE ( α \alpha α ) in ppm / ∘ C \text{ppm}/^\circ\text{C} ppm / ∘ C ( 10 − 6 / ∘ C 10^{-6}/^\circ\text{C} 1 0 − 6 / ∘ C )

  • Tungsten Carbide 4.5 ± 0.5 4.5 \pm 0.5 4.5 ± 0.5
  • Gage Block Steel 11.5 ± 1.0 11.5 \pm 1.0 11.5 ± 1.0
  • Cast Iron / Tool Steel 11.0 ± 1.0 11.0 \pm 1.0 11.0 ± 1.0
  • Granite 6.0 ± 1.5 6.0 \pm 1.5 6.0 ± 1.5
  • Stainless Steel (304/316) 16.0 ± 1.0 16.0 \pm 1.0 16.0 ± 1.0
  • Brass / Bronze 18.5 ± 1.5 18.5 \pm 1.5 18.5 ± 1.5
  • Aluminum Alloys 23.0 ± 1.5 23.0 \pm 1.5 23.0 ± 1.5

Low-expansion materials Fused silica and Invar have different, temperature-dependent expansion coefficients; use the material-specific evaluated value rather than assigning both one coefficient.

Thermal Soak and Stabilization Dynamics

Bringing an instrument into a temperature-controlled laboratory is not instantaneous. Heat transfers via conduction, convection, and radiation. The time required for an object's core temperature to equilibrate with the ambient room temperature is known as the thermal soak time .

Select soak and warm-up requirements from the method and equipment information, then verify sufficient stability. These are planning considerations rather than universal minimum times:

Tip

Handling Tip: Hand heat can introduce thermal gradients in metal standards. Use suitable insulating handling aids and allow equilibrium to recover. The change depends on contact, material, geometry, and time; a fixed one-micrometer change after ten seconds is not universal.

Relative Humidity (RH) Control and Monitoring

Select relative-humidity limits for the equipment, method, uncertainty, materials, and ESD controls. A laboratory may choose a 30–50% RH operating band, but ISO/IEC 17025 does not prescribe that universal window. Assess the following mechanisms:

Humidity and Surface-Temperature Risks

Corrosion and Oxidation: Atmospheric moisture condenses on polished steel surfaces (gage blocks, micrometer anvils, mass standards), initiating pitting and surface oxidation.

Surface Leakage Currents: Moisture films form microscopic conductive pathways across insulating surfaces, causing severe leakage errors when calibrating high-resistance standards ( > 10 M Ω > 10\text{ M}\Omega > 10 M Ω to 10 T Ω 10\text{ T}\Omega 10 T Ω ).

Hygroscopic Swelling: Materials such as paper, nylon, phenolic composites, and natural fibers absorb atmospheric water, altering their physical dimensions and mass.

Low Humidity Risks ( < 30 % RH < 30\%\text{ RH} < 30% RH )

Electrostatic Charge Accumulation: Dry air acts as an exceptional electrical insulator. Frictional contact (triboelectric charging) generates electrostatic potentials exceeding 15 , 000 V 15,000\text{ V} 15 , 000 V , leading to catastrophic ESD destruction of semiconductor components.

Humidity Measurement Instrumentation

Chilled-mirror hygrometer: A calibrated mirror thermometer measures the temperature at which the observed condensate forms. Establish water or ice phase, cleanliness, gradients, pressure, and realized uncertainty. Optical condensation detection alone does not make the instrument a primary standard or supply an exact true dew point.

Capacitive Thin-Film Polymer Hygrometer: Fast, reliable secondary working sensor measuring dielectric constant changes of a polymer film ( ± 1.5 % to ± 2.0 % RH \pm 1.5\%\text{ to }\pm 2.0\%\text{ RH} ± 1.5% to ± 2.0% RH ).