Study guide
SI Prefixes and Compound-Unit Conversions
ASQ Certified Calibration Technician CCT
Track your progress as you read
Finished sections get a check mark.
1 SI Definitions and Realizations
The Seven SI Base Units and the 2019 Revision Realizing SI Units in the Laboratory
2 Derived Units, Prefixes, and Constants
Derived Units and Physical Calculations SI Prefixes and Compound-Unit Conversions Physical Constants and Their Applications
3 Measurement Standards and Selection
Types and Roles of Measurement Standards Selecting Standards for an Operational Hierarchy
4 Traceability and Standard Substitution
Traceability Evidence and Impact Review Evaluating and Authorizing Standard Substitution
5 Dimensional and Mechanical Metrology
Linear Displacement and Dimensional Instruments Geometry, Thermal Expansion, and Mass Force, Torque, Pressure, Density, and Viscosity
6 Temperature, Humidity, Volume, and Flow
Temperature Sensors and Thermometer Corrections Humidity, Volume, and Flow
7 Electrical, Optical, and Time Measurements
Voltage, Current, Resistance, and AC Measurement Time, Frequency, Light, Power, and Energy
8 Measurement Methods
Direct, Indirect, and Ratio Methods Transfer, Differential, and UUT Substitution Methods
9 Instrument Characteristics and Measurement Data
Sensitivity, Variability, and Precision Bias, Linearity, Stability, and Hysteresis Data Formats, Readability, and Suitability Resolution, Confidentiality, and Data Integrity
10 Specifications and Error Sources
Interpreting IM&TE Specification Terms Specification Calculations and Operating Qualifiers Instrument, Operator, and Circuit Errors Environmental, Geometric, and Interference Errors
11 Measurement Assurance
Check Standards and Statistical Process Control Proficiency Comparisons and Gage R&R
12 Procedures and Governing Requirements
Procedure Structure and Published Practice Sources Method Development, Deviations, and Handling Regulations, Contracts, and Requirement Precedence
13 Instrument Adjustment and Standardization
Calibration, Zero, and Span Adjustment Linearization, Nulling, and Digital Adjustment
14 Calibration Environment
Temperature and Humidity Control ESD, Dust, Vibration, and Pressure
15 Operations, Roles, and Scheduling
Receiving, As-Found Testing, and Calibration Flow Asset Identification, Tracking, and Release Laboratory Roles and Initial Scheduling Interval Review, Overdue Control, and Workloads
16 Method and Software Validation
Method Verification and Validation Studies Software, Firmware, and Spreadsheet Validation
17 Records and Formal Reporting
Controlled Records and Calibration Certificates Status Labels and Out-of-Tolerance Reporting
18 Uncertainty Concepts and Decision Risk
Error, Uncertainty, and Measurand Definition TAR, TUR, Tolerance Consumption, and Capability Conformity Decisions and False-Accept Risk Guard Bands and Conformity Reporting
19 Uncertainty Models and Budgets
Measurement Models and Uncertainty Contributions Distribution Selection and Budget Assembly
20 Type A and Type B Evaluation
Type A Statistics and Repeatability Outlier Review and Worked Type A Analysis Type B Distribution Models
21 Uncertainty Reporting and Applied Mathematics
Propagation, Covariance, and Effective Degrees of Freedom Coverage Factors and Expanded-Uncertainty Reporting Notation, Unit Conversions, Ratios, and Decibels Rounding, Algebra, Graphs, and Angular Conversions
22 Quality Systems and Tools
Quality-System Components, Planning, and Improvement The Seven Basic Quality Tools
23 Audits and Corrective Action
Audit Types, Planning, and Nonconformance Control Impact Assessment, Root Cause, and Effectiveness
24 Safety, Standards, and Professional Conduct
Electrical, Pressure, Chemical, and Laboratory Hazards PPE, Safety Data Sheets, 6S, and Safe Restoration Quality Standards, GUM, and VIM Accreditation Scope and Professional Ethics
SI Prefixes and Compound-Unit Conversions
Key Takeaways Explain This Topic AI
SI prefixes are case-sensitive and attach directly to the unit symbol.
Squaring or cubing a prefixed length also squares or cubes its conversion factor.
Ronna and ronto represent factors of 10²⁷ and 10⁻²⁷, respectively.
Last updated: October 2026
Modern calibration spans extreme measurement ranges: from measuring optical surface roughness on silicon wafers in picometers ( 10 − 12 m 10^{-12}\text{ m} 1 0 − 12 m ) or femtoseconds in frequency standards ( 10 − 15 s 10^{-15}\text{ s} 1 0 − 15 s ), up to electrical insulation resistances in gigohms ( 10 9 Ω 10^9\ \Omega 1 0 9 Ω ) or terahertz optical frequencies ( 10 12 Hz 10^{12}\text{ Hz} 1 0 12 Hz ). The SI prefix system provides a standardized, decimal-based framework that eliminates ambiguous local terminology (such as "billions" or "trillions") and ensures consistent scientific notation.
The Complete 24 SI Prefixes (Including the 2022 CGPM Additions)
In November 2022, the 27th General Conference on Weights and Measures (CGPM) adopted four new prefixes to accommodate the exponential growth of digital data science, astronomy, and quantum physics: quetta ( 10 30 10^{30} 1 0 30 ), ronna ( 10 27 10^{27} 1 0 27 ), ronto ( 10 − 27 10^{-27} 1 0 − 27 ), and quecto ( 10 − 30 10^{-30} 1 0 − 30 ). This represents the first expansion of the SI prefix system since 1991.
Multiplier Factor Prefix Name Symbol Decimal Value / Representation Common Metrology & Engineering Application
10 30 10^{30} 1 0 30 quetta Q \text{Q} Q 1 , 000 , 000 , 000 , 000 , 000 , 000 , 000 , 000 , 000 , 000 1,000,000,000,000,000,000,000,000,000,000 1 , 000 , 000 , 000 , 000 , 000 , 000 , 000 , 000 , 000 , 000 Astronomical mass calculations, global data
10 27 10^{27} 1 0 27 ronna R \text{R} R 1 , 000 , 000 , 000 , 000 , 000 , 000 , 000 , 000 , 000 1,000,000,000,000,000,000,000,000,000 1 , 000 , 000 , 000 , 000 , 000 , 000 , 000 , 000 , 000 Planetary mass estimates
10 24 10^{24} 1 0 24 yotta Y \text{Y} Y 1 , 000 , 000 , 000 , 000 , 000 , 000 , 000 , 000 1,000,000,000,000,000,000,000,000 1 , 000 , 000 , 000 , 000 , 000 , 000 , 000 , 000 Global computing capacity
10 21 10^{21} 1 0 21 zetta Z \text{Z} Z 1 , 000 , 000 , 000 , 000 , 000 , 000 , 000 1,000,000,000,000,000,000,000 1 , 000 , 000 , 000 , 000 , 000 , 000 , 000 Global digital internet traffic (zettabytes)
10 18 10^{18} 1 0 18 exa E \text{E} E 1 , 000 , 000 , 000 , 000 , 000 , 000 1,000,000,000,000,000,000 1 , 000 , 000 , 000 , 000 , 000 , 000 Exascale supercomputer benchmark calculations
10 15 10^{15} 1 0 15 peta P \text{P} P 1 , 000 , 000 , 000 , 000 , 000 1,000,000,000,000,000 1 , 000 , 000 , 000 , 000 , 000 High-energy laser pulse power (petawatts)
10 12 10^{12} 1 0 12 tera T \text{T} T 1 , 000 , 000 , 000 , 000 1,000,000,000,000 1 , 000 , 000 , 000 , 000 Optical spectroscopy, terahertz radiation ( THz \text{THz} THz )
10 9 10^9 1 0 9 giga G \text{G} G 1 , 000 , 000 , 000 1,000,000,000 1 , 000 , 000 , 000 Microwave frequency ( GHz \text{GHz} GHz ), insulation resistance ( G Ω \text{G}\Omega G Ω )
10 6 10^6 1 0 6 mega M \text{M} M 1 , 000 , 000 1,000,000 1 , 000 , 000 Hydraulic pressure ( MPa \text{MPa} MPa ), RF broadcast ( MHz \text{MHz} MHz )
10 3 10^3 1 0 3 kilo k \text{k} k 1 , 000 1,000 1 , 000 Mass ( kg \text{kg} kg ), voltage ( kV \text{kV} kV ), pressure ( kPa \text{kPa} kPa )
- 10 2 10^2 1 0 2 hecto h \text{h} h 100 100 100 Atmospheric barometric pressure ( hPa \text{hPa} hPa )
- 10 1 10^1 1 0 1 deka da \text{da} da 10 10 10 Industrial force testing (decanewton, daN \text{daN} daN )
- 10 − 1 10^{-1} 1 0 − 1 deci d \text{d} d 0.1 0.1 0.1 Acoustics and power ratios (decibels, dB \text{dB} dB )
10 − 2 10^{-2} 1 0 − 2 centi c \text{c} c 0.01 0.01 0.01 Dimensional metrology ( cm \text{cm} cm ), volume ( cL \text{cL} cL )
10 − 3 10^{-3} 1 0 − 3 milli m \text{m} m 0.001 0.001 0.001 Current loops ( mA \text{mA} mA ), length ( mm \text{mm} mm ), voltage ( mV \text{mV} mV )
10 − 6 10^{-6} 1 0 − 6 micro μ \mu μ 0.000 001 0.000\ 001 0.000 001 Microvolts ( μ V \mu\text{V} μ V ), micrometers ( μ m \mu\text{m} μ m ), capacitance ( μ F \mu\text{F} μ F )
10 − 9 10^{-9} 1 0 − 9 nano n \text{n} n 0.000 000 001 0.000\ 000\ 001 0.000 000 001 Laser wavelength ( nm \text{nm} nm ), capacitance ( nF \text{nF} nF ), time ( ns \text{ns} ns )
10 − 12 10^{-12} 1 0 − 12 pico p \text{p} p 0.000 000 000 001 0.000\ 000\ 000\ 001 0.000 000 000 001 Accelerometer charge sensitivity ( pC/g \text{pC/g} pC/g ), capacitance ( pF \text{pF} pF )
10 − 15 10^{-15} 1 0 − 15 femto f \text{f} f 0.000 000 000 000 001 0.000\ 000\ 000\ 000\ 001 0.000 000 000 000 001 Femtosecond optical frequency comb pulses ( fs \text{fs} fs )
10 − 18 10^{-18} 1 0 − 18 atto a \text{a} a 0.000 000 000 000 000 001 0.000\ 000\ 000\ 000\ 000\ 001 0.000 000 000 000 000 001 Ultra-low current measurement (attoamperes, aA \text{aA} aA )
10 − 21 10^{-21} 1 0 − 21 zepto z \text{z} z 0.000 000 000 000 000 000 001 0.000\ 000\ 000\ 000\ 000\ 000\ 001 0.000 000 000 000 000 000 001 Subatomic particle charge physics
10 − 24 10^{-24} 1 0 − 24 yocto y \text{y} y 0.000 000 000 000 000 000 000 001 0.000\ 000\ 000\ 000\ 000\ 000\ 000\ 001 0.000 000 000 000 000 000 000 001 Mass of subatomic nucleons
10 − 27 10^{-27} 1 0 − 27 ronto r \text{r} r 0.000 000 000 000 000 000 000 000 001 0.000\ 000\ 000\ 000\ 000\ 000\ 000\ 000\ 001 0.000 000 000 000 000 000 000 000 001 Mass of an electron ( ≈ 0.91 rg \approx 0.91\text{ rg} ≈ 0.91 rg )
10 − 30 10^{-30} 1 0 − 30 quecto q \text{q} q 0.000 000 000 000 000 000 000 000 000 001 0.000\ 000\ 000\ 000\ 000\ 000\ 000\ 000\ 000\ 001 0.000 000 000 000 000 000 000 000 000 001 Quantum particle mass limits
BIPM SI Grammar and Prefix Arithmetic Rules
Adhering to correct SI syntax is not merely an editorial preference; on the ASQ CCT exam and in accredited calibration certificates, syntax errors can create severe ambiguities. The BIPM SI Brochure establishes strict conventions:
Attachment and Spacing
Prefix symbols are printed in upright (roman) type and attached directly to the unit symbol without an intervening space or punctuation. A single space is always placed between the numerical value and the unit symbol:
Correct: 25.4 mm 25.4\text{ mm} 25.4 mm , 10.0 k Ω 10.0\text{ k}\Omega 10.0 k Ω , 5.0 μ V 5.0\ \mu\text{V} 5.0 μ V
Incorrect: 25.4 m m 25.4\text{m m} 25.4 m m , 10.0 k Ω 10.0\text{k}\Omega 10.0 k Ω , 5.0 μ V 5.0\mu\text{V} 5.0 μ V (missing space between number and unit)
Strict Case Sensitivity
Case indicates entirely different orders of magnitude:
Lowercase m \text{m} m = milli ( 10 − 3 10^{-3} 1 0 − 3 ), while uppercase M \text{M} M = mega ( 10 6 10^6 1 0 6 ). Writing 10 mV 10\text{ mV} 10 mV represents ten millivolts; writing 10 MV 10\text{ MV} 10 MV represents ten megavolts—a catastrophic error of 10 9 10^9 1 0 9 (one billion times)!
Lowercase p \text{p} p = pico ( 10 − 12 10^{-12} 1 0 − 12 ), while uppercase P \text{P} P = peta ( 10 15 10^{15} 1 0 15 ).
Lowercase k \text{k} k = kilo ( 10 3 10^3 1 0 3 ). Writing uppercase K \text{K} K is reserved exclusively for the kelvin.
Prohibition of Compound (Double) Prefixes
Only one prefix symbol may be attached to a unit. Compound prefixes formed by juxtaposing two prefixes are strictly forbidden:
Do NOT write: 10 μ μ F 10\ \mu\mu\text{F} 10 μμ F (micro-microfarads) → \rightarrow → Correct: 10 pF 10\text{ pF} 10 pF (picofarads)
Do NOT write: 5 m μ m 5\text{ m}\mu\text{m} 5 m μ m (milli-micrometers) → \rightarrow → Correct: 5 nm 5\text{ nm} 5 nm (nanometers)
- Do NOT write: 2 kMHz 2\text{ k}\text{M}\text{Hz} 2 k M Hz → \rightarrow → Correct: 2 GHz 2\text{ GHz} 2 GHz
- The Kilogram Rule (Forming Multiples of Mass)
Because the base unit of mass ( kg \text{kg} kg ) already contains the prefix "kilo", names and symbols for decimal multiples and submultiples of mass are formed by attaching prefixes to the gram ( g \text{g} g ), never to the kilogram: 10 − 6 kg = 10 − 3 g = 1 mg ( NOT 1 μ kg ) 10^{-6}\text{ kg} = 10^{-3}\text{ g} = 1\text{ mg} \quad (\text{NOT } 1\ \mu\text{kg}) 1 0 − 6 kg = 1 0 − 3 g = 1 mg ( NOT 1 μ kg ) 10 − 9 kg = 10 − 6 g = 1 μ g ( NOT 1 nkg ) 10^{-9}\text{ kg} = 10^{-6}\text{ g} = 1\ \mu\text{g} \quad (\text{NOT } 1\text{ nkg}) 1 0 − 9 kg = 1 0 − 6 g = 1 μ g ( NOT 1 nkg ) 10 3 kg = 1 Mg = 1 metric ton (t) ( NOT 1 kkg ) 10^3\text{ kg} = 1\text{ Mg} = 1\text{ metric ton (t)} \quad (\text{NOT } 1\text{ kkg}) 1 0 3 kg = 1 Mg = 1 metric ton (t) ( NOT 1 kkg )
Exponentiation of Prefixed Units
When a prefixed unit is modified by an exponent, the exponent applies to the entire compound unit (both the multiplying prefix and the base unit): 1 cm 3 = ( 1 cm ) 3 = ( 10 − 2 m ) 3 = 10 − 6 m 3 ( NOT 10 − 2 m 3 ) 1\text{ cm}^3 = (1\text{ cm})^3 = (10^{-2}\text{ m})^3 = 10^{-6}\text{ m}^3 \quad (\text{NOT } 10^{-2}\text{ m}^3) 1 cm 3 = ( 1 cm ) 3 = ( 1 0 − 2 m ) 3 = 1 0 − 6 m 3 ( NOT 1 0 − 2 m 3 ) 1 mm 2 = ( 1 mm ) 2 = ( 10 − 3 m ) 2 = 10 − 6 m 2 ( NOT 10 − 3 m 2 ) 1\text{ mm}^2 = (1\text{ mm})^2 = (10^{-3}\text{ m})^2 = 10^{-6}\text{ m}^2 \quad (\text{NOT } 10^{-3}\text{ m}^2) 1 mm 2 = ( 1 mm ) 2 = ( 1 0 − 3 m ) 2 = 1 0 − 6 m 2 ( NOT 1 0 − 3 m 2 ) 1 km 2 = ( 1 km ) 2 = ( 10 3 m ) 2 = 10 6 m 2 ( NOT 10 3 m 2 ) 1\text{ km}^2 = (1\text{ km})^2 = (10^3\text{ m})^2 = 10^6\text{ m}^2 \quad (\text{NOT } 10^3\text{ m}^2) 1 km 2 = ( 1 km ) 2 = ( 1 0 3 m ) 2 = 1 0 6 m 2 ( NOT 1 0 3 m 2 )
Dimensional Analysis: The Factor-Label Method
In calibration laboratories, technicians must frequently convert complex compound measurement units (such as volumetric flow, density, dynamic viscosity, or torque). The factor-label method (also known as unit factor analysis or the chain-ratio method) treats units as algebraic quantities that can be multiplied, divided, and canceled.
Core Rules of the Factor-Label Method
Write down the initial quantity with its given units as a fraction over 1.
Identify the target unit.
Form conversion factors from known mathematical equalities such that each factor equals unity ( 1 1 1 ).
Arrange each fraction so that the unit to be cancelled appears on the opposite side of the fraction bar (numerator vs. denominator).
Multiply all numerators, multiply all denominators, cancel identical units algebraically, and verify that the remaining unit matches the desired target.
Worked Calibration Example 1: Gas Volumetric Flow Rate Conversion
A gas flow calibrator measures a volumetric leak rate of 0.0750 cm 3 / s 0.0750\text{ cm}^3/\text{s} 0.0750 cm 3 / s . A customer calibration procedure requires this value to be documented in cubic meters per hour ( m 3 / h \text{m}^3/\text{h} m 3 / h ).
Step 1: Establish the unit equalities: 1 cm = 10 − 2 m ⟹ ( 1 cm ) 3 = ( 10 − 2 m ) 3 ⟹ 1 cm 3 = 10 − 6 m 3 1\text{ cm} = 10^{-2}\text{ m} \implies (1\text{ cm})^3 = (10^{-2}\text{ m})^3 \implies 1\text{ cm}^3 = 10^{-6}\text{ m}^3 1 cm = 1 0 − 2 m ⟹ ( 1 cm ) 3 = ( 1 0 − 2 m ) 3 ⟹ 1 cm 3 = 1 0 − 6 m 3 1 h = 3 , 600 s 1\text{ h} = 3,600\text{ s} 1 h = 3 , 600 s
Step 2: Construct the chain-ratio calculation: Flow Rate = ( 0.0750 cm 3 1 s ) × ( 10 − 6 m 3 1 cm 3 ) × ( 3 , 600 s 1 h ) \text{Flow Rate} = \left(\frac{0.0750\text{ cm}^3}{1\text{ s}}\right) \times \left(\frac{10^{-6}\text{ m}^3}{1\text{ cm}^3}\right) \times \left(\frac{3,600\text{ s}}{1\text{ h}}\right) Flow Rate = ( 1 s 0.0750 cm 3 ) × ( 1 cm 3 1 0 − 6 m 3 ) × ( 1 h 3 , 600 s )
Step 3: Cancel units algebraically ( cm 3 \text{cm}^3 cm 3 and s \text{s} s ): Flow Rate = 0.0750 × 10 − 6 × 3 , 600 m 3 / h \text{Flow Rate} = 0.0750 \times 10^{-6} \times 3,600\text{ m}^3/\text{h} Flow Rate = 0.0750 × 1 0 − 6 × 3 , 600 m 3 / h Flow Rate = 270.0 × 10 − 6 m 3 / h = 2.70 × 10 − 4 m 3 / h \text{Flow Rate} = 270.0 \times 10^{-6}\text{ m}^3/\text{h} = 2.70 \times 10^{-4}\text{ m}^3/\text{h} Flow Rate = 270.0 × 1 0 − 6 m 3 / h = 2.70 × 1 0 − 4 m 3 / h
Worked Calibration Example 2: Density of Hydraulic Fluid
A hydrometer calibration requires converting a measured hydraulic fluid density of 0.875 g/cm 3 0.875\text{ g/cm}^3 0.875 g/cm 3 into SI base units ( kg/m 3 \text{kg/m}^3 kg/m 3 ).
Step 1: Set up conversion factors for mass ( 1 kg = 10 3 g 1\text{ kg} = 10^3\text{ g} 1 kg = 1 0 3 g ) and volume ( 1 cm 3 = 10 − 6 m 3 1\text{ cm}^3 = 10^{-6}\text{ m}^3 1 cm 3 = 1 0 − 6 m 3 ): ρ = ( 0.875 g 1 cm 3 ) × ( 1 kg 10 3 g ) × ( 1 cm 3 10 − 6 m 3 ) \rho = \left(\frac{0.875\text{ g}}{1\text{ cm}^3}\right) \times \left(\frac{1\text{ kg}}{10^3\text{ g}}\right) \times \left(\frac{1\text{ cm}^3}{10^{-6}\text{ m}^3}\right) ρ = ( 1 cm 3 0.875 g ) × ( 1 0 3 g 1 kg ) × ( 1 0 − 6 m 3 1 cm 3 )
Step 2: Cancel units and solve: ρ = 0.875 × 10 − 3 kg 10 − 6 m 3 = 0.875 × 10 3 kg/m 3 = 875 kg/m 3 \rho = 0.875 \times \frac{10^{-3}\text{ kg}}{10^{-6}\text{ m}^3} = 0.875 \times 10^3\text{ kg/m}^3 = 875\text{ kg/m}^3 ρ = 0.875 × 1 0 − 6 m 3 1 0 − 3 kg = 0.875 × 1 0 3 kg/m 3 = 875 kg/m 3
Worked Calibration Example 3: Torque Wrench Calibration
A mechanical torque wrench is calibrated on a bench analyzer that reads in newton meters ( N ⋅ m \text{N}\cdot\text{m} N ⋅ m ). The customer specification sheet lists a torque requirement of 150.0 lbf ⋅ in 150.0\text{ lbf}\cdot\text{in} 150.0 lbf ⋅ in (pound-force inches). Convert this value to newton meters.
Step 1: Identify conversion constants: 1 lbf = 4.448222 N 1\text{ lbf} = 4.448222\text{ N} 1 lbf = 4.448222 N 1 in = 0.0254 m 1\text{ in} = 0.0254\text{ m} 1 in = 0.0254 m
Step 2: Set up the chain equation: τ = 150.0 lbf ⋅ in × ( 4.448222 N 1 lbf ) × ( 0.0254 m 1 in ) \tau = 150.0\text{ lbf}\cdot\text{in} \times \left(\frac{4.448222\text{ N}}{1\text{ lbf}}\right) \times \left(\frac{0.0254\text{ m}}{1\text{ in}}\right) τ = 150.0 lbf ⋅ in × ( 1 lbf 4.448222 N ) × ( 1 in 0.0254 m ) τ = 150.0 × 4.448222 × 0.0254 N ⋅ m = 16.948 N ⋅ m \tau = 150.0 \times 4.448222 \times 0.0254\text{ N}\cdot\text{m} = 16.948\text{ N}\cdot\text{m} τ = 150.0 × 4.448222 × 0.0254 N ⋅ m = 16.948 N ⋅ m
Parts Per Million (PPM) and Dimensionless Ratios in Calibration
Instrument specifications and calibration tolerances are frequently expressed as fractional multipliers: parts per million (ppm) or parts per billion (ppb) . 1 ppm = 1 × 10 − 6 = 0.0001 % = 1 μ V/V = 1 μ Ω / Ω = 1 mg/kg 1\text{ ppm} = 1 \times 10^{-6} = 0.0001\% = 1\ \mu\text{V/V} = 1\ \mu\Omega/\Omega = 1\text{ mg/kg} 1 ppm = 1 × 1 0 − 6 = 0.0001% = 1 μ V/V = 1 μ Ω/Ω = 1 mg/kg 1 ppb = 1 × 10 − 9 = 0.0000001 % = 1 nV/V 1\text{ ppb} = 1 \times 10^{-9} = 0.0000001\% = 1\text{ nV/V} 1 ppb = 1 × 1 0 − 9 = 0.0000001% = 1 nV/V
Tip
BIPM Policy on PPM/PPB: Because the word "billion" historically represented 10 9 10^9 1 0 9 in American English and 10 12 10^{12} 1 0 12 in British/European French, the BIPM strongly discourages the terms ppm and ppb in official calibration reports. Technicians should report tolerances using unambiguous SI ratios: μ V/V \mu\text{V/V} μ V/V , mg/kg \text{mg/kg} mg/kg , or engineering scientific notation ( 10 − 6 10^{-6} 1 0 − 6 ).
Worked example: hypothetical voltmeter
Assume the following 1-year specification on the 10 V DC range. These are exercise inputs, not a claimed specification for a named commercial model: Tolerance = ± ( 3.0 ppm of Reading + 0.5 ppm of Range ) \text{Tolerance} = \pm (3.0\text{ ppm of Reading} + 0.5\text{ ppm of Range}) Tolerance = ± ( 3.0 ppm of Reading + 0.5 ppm of Range )
If the technician reads a calibrated standard cell of 1.018000 V 1.018000\text{ V} 1.018000 V on this 10.000000 V 10.000000\text{ V} 10.000000 V range, calculate the total permissible tolerance in microvolts ( μ V \mu\text{V} μ V ):
Reading Component: Reading error = 1.018000 V × ( 3.0 × 10 − 6 ) = 3.054 × 10 − 6 V = 3.054 μ V \text{Reading error} = 1.018000\text{ V} \times (3.0 \times 10^{-6}) = 3.054 \times 10^{-6}\text{ V} = 3.054\ \mu\text{V} Reading error = 1.018000 V × ( 3.0 × 1 0 − 6 ) = 3.054 × 1 0 − 6 V = 3.054 μ V
Range Component: Range error = 10.000000 V × ( 0.5 × 10 − 6 ) = 5.000 × 10 − 6 V = 5.000 μ V \text{Range error} = 10.000000\text{ V} \times (0.5 \times 10^{-6}) = 5.000 \times 10^{-6}\text{ V} = 5.000\ \mu\text{V} Range error = 10.000000 V × ( 0.5 × 1 0 − 6 ) = 5.000 × 1 0 − 6 V = 5.000 μ V
Total Permissible Tolerance: Permissible Tolerance = ± ( 3.054 μ V + 5.000 μ V ) = ± 8.054 μ V \text{Permissible Tolerance} = \pm (3.054\ \mu\text{V} + 5.000\ \mu\text{V}) = \pm 8.054\ \mu\text{V} Permissible Tolerance = ± ( 3.054 μ V + 5.000 μ V ) = ± 8.054 μ V
At small readings relative to range, the range term can dominate the stated specification bound. Select an appropriate valid range after checking resolution, uncertainty, loading, bandwidth, and ratings; the lowest selectable range is not automatically best for every method.