Study guide

Humidity, Volume, and Flow

ASQ Certified Calibration Technician CCT

Track your progress as you read

Finished sections get a check mark.

1 SI Definitions and Realizations

The Seven SI Base Units and the 2019 Revision Realizing SI Units in the Laboratory

2 Derived Units, Prefixes, and Constants

Derived Units and Physical Calculations SI Prefixes and Compound-Unit Conversions Physical Constants and Their Applications

3 Measurement Standards and Selection

Types and Roles of Measurement Standards Selecting Standards for an Operational Hierarchy

4 Traceability and Standard Substitution

Traceability Evidence and Impact Review Evaluating and Authorizing Standard Substitution

5 Dimensional and Mechanical Metrology

Linear Displacement and Dimensional Instruments Geometry, Thermal Expansion, and Mass Force, Torque, Pressure, Density, and Viscosity

6 Temperature, Humidity, Volume, and Flow

Temperature Sensors and Thermometer Corrections Humidity, Volume, and Flow

7 Electrical, Optical, and Time Measurements

Voltage, Current, Resistance, and AC Measurement Time, Frequency, Light, Power, and Energy

8 Measurement Methods

Direct, Indirect, and Ratio Methods Transfer, Differential, and UUT Substitution Methods

9 Instrument Characteristics and Measurement Data

Sensitivity, Variability, and Precision Bias, Linearity, Stability, and Hysteresis Data Formats, Readability, and Suitability Resolution, Confidentiality, and Data Integrity

10 Specifications and Error Sources

Interpreting IM&TE Specification Terms Specification Calculations and Operating Qualifiers Instrument, Operator, and Circuit Errors Environmental, Geometric, and Interference Errors

11 Measurement Assurance

Check Standards and Statistical Process Control Proficiency Comparisons and Gage R&R

12 Procedures and Governing Requirements

Procedure Structure and Published Practice Sources Method Development, Deviations, and Handling Regulations, Contracts, and Requirement Precedence

13 Instrument Adjustment and Standardization

Calibration, Zero, and Span Adjustment Linearization, Nulling, and Digital Adjustment

14 Calibration Environment

Temperature and Humidity Control ESD, Dust, Vibration, and Pressure

15 Operations, Roles, and Scheduling

Receiving, As-Found Testing, and Calibration Flow Asset Identification, Tracking, and Release Laboratory Roles and Initial Scheduling Interval Review, Overdue Control, and Workloads

16 Method and Software Validation

Method Verification and Validation Studies Software, Firmware, and Spreadsheet Validation

17 Records and Formal Reporting

Controlled Records and Calibration Certificates Status Labels and Out-of-Tolerance Reporting

18 Uncertainty Concepts and Decision Risk

Error, Uncertainty, and Measurand Definition TAR, TUR, Tolerance Consumption, and Capability Conformity Decisions and False-Accept Risk Guard Bands and Conformity Reporting

19 Uncertainty Models and Budgets

Measurement Models and Uncertainty Contributions Distribution Selection and Budget Assembly

20 Type A and Type B Evaluation

Type A Statistics and Repeatability Outlier Review and Worked Type A Analysis Type B Distribution Models

21 Uncertainty Reporting and Applied Mathematics

Propagation, Covariance, and Effective Degrees of Freedom Coverage Factors and Expanded-Uncertainty Reporting Notation, Unit Conversions, Ratios, and Decibels Rounding, Algebra, Graphs, and Angular Conversions

22 Quality Systems and Tools

Quality-System Components, Planning, and Improvement The Seven Basic Quality Tools

23 Audits and Corrective Action

Audit Types, Planning, and Nonconformance Control Impact Assessment, Root Cause, and Effectiveness

24 Safety, Standards, and Professional Conduct

Electrical, Pressure, Chemical, and Laboratory Hazards PPE, Safety Data Sheets, 6S, and Safe Restoration Quality Standards, GUM, and VIM Accreditation Scope and Professional Ethics

Humidity, Volume, and Flow

Key Takeaways Explain This Topic AI

Relative humidity depends on both moisture content and temperature.

Gravimetric volume determination uses liquid density and relevant buoyancy and apparatus corrections.

Flow meters have different installation and fluid limitations; Coriolis measurement is not immune to every disturbance.

Last updated: October 2026

Humidity and Moisture Metrology

Humidity defines water vapor content in a gas mixture. Relative humidity (%RH) is the ratio of water vapor partial pressure e e e to the saturation vapor pressure e w ( T ) e_w(T) e w ​ ( T ) at that gas temperature: % RH = e e w ( T ) × 100 % \%\text{RH} = \frac{e}{e_w(T)} \times 100\% % RH = e w ​ ( T ) e ​ × 100%

Humidity Instrumentation

Chilled Mirror Hygrometers (Calibrated Optical Reference): A gas sample flows across a polished rhodium or platinum mirror cooled by a thermoelectric Peltier cooler. A light source reflects off the mirror into an optical photodetector. When the mirror cools to the gas condensation temperature, dew or frost forms, scattering light. An analog or digital servo loop controls mirror temperature to maintain a constant condensate layer thickness. The mirror temperature, measured by a calibrated embedded miniature PRT, gives the true dew point ( T d T_d T d ​ ) or frost point ( T f T_f T f ​ ) .

Below freezing, establish whether the condensate is liquid water or ice. Dew point and frost point correspond to different saturation vapor-pressure relations; the temperature difference is not a fixed 0.5°C across a range. Determine phase and use the appropriate relation and uncertainty.

Capacitive Thin-Film Polymer Hygrometers: Fast-response field instruments. A thin hygroscopic polymer film changes dielectric constant as water vapor is absorbed. Susceptible to hysteresis between wetting and drying cycles, and temperature-dependent drift.

Aspiration Psychrometers: Utilize paired thermometers—one dry bulb measuring ambient temperature, and one wet bulb wrapped in a wetted cotton wick. Forced air aspiration ( ≥ 3 to 5 m/s \ge 3\text{ to }5\text{ m/s} ≥ 3 to 5 m/s ) produces evaporative cooling. Relative humidity is derived from the dry-wet depression, barometric pressure, and psychrometric tables.

Saturated Salt Solutions (ASTM E104): Sealed enclosures containing saturated salt slurries in equilibrium with air generate fixed, reproducible relative humidity values at 25 ∘ C 25^\circ\text{C} 2 5 ∘ C :

  • Lithium Chloride ( LiCl \text{LiCl} LiCl ): 11.3 % ± 0.3 % 11.3\% \pm 0.3\% 11.3% ± 0.3%
  • Magnesium Chloride ( MgCl 2 \text{MgCl}_2 MgCl 2 ​ ): 32.8 % ± 0.2 % 32.8\% \pm 0.2\% 32.8% ± 0.2%
  • Sodium Chloride ( NaCl \text{NaCl} NaCl ): 75.3 % ± 0.1 % 75.3\% \pm 0.1\% 75.3% ± 0.1%
  • Potassium Chloride ( KCl \text{KCl} KCl ): 84.3 % ± 0.3 % 84.3\% \pm 0.3\% 84.3% ± 0.3%

Potassium sulfate is an example of a high-humidity salt reference near 97.3% RH at 25°C. Use the evaluated temperature-dependent value and uncertainty; salt composition, equilibration, contamination, and temperature gradients affect realized humidity.

Rule: Slurries must maintain excess solid undissolved crystals, and enclosure temperature must remain strictly isothermal (a 1 ∘ C 1^\circ\text{C} 1 ∘ C temperature difference between liquid and air induces up to 6 % RH 6\%\text{RH} 6% RH error).

Fluid Flow Measurement

Fluid flow characterizes the motion of liquids and gases in closed conduits. Calibration technicians distinguish between Volumetric Flow ( Q Q Q ) and Mass Flow ( m ˙ \dot{m} m ˙ ) : m ˙ = ρ ⋅ Q \dot{m} = \rho \cdot Q m ˙ = ρ ⋅ Q

Closed-Conduit Flowmeter Technologies

Flowmeter Principle Direct Measured Parameter Operating Formula / Mechanism Metrological Considerations

Orifice Plate (DP) Differential Pressure ( Δ P \Delta P Δ P ) Q = C d A 2 1 − β 4 2 Δ P ρ Q = C_d \frac{A_2}{\sqrt{1-\beta^4}} \sqrt{\frac{2\Delta P}{\rho}} Q = C d ​ 1 − β 4 ​ A 2 ​ ​ ρ 2Δ P ​ ​ Square-root extraction; edge rounding error; high permanent head loss

Venturi Tube (DP) Differential Pressure ( Δ P \Delta P Δ P ) Convergent cone & diffuser Low head loss; high pressure recovery; resistant to slurry wear

Turbine Flowmeter Rotor Blade Frequency ( f f f ) Q = f / K Q = f / K Q = f / K ( K K K = pulses per unit volume) Viscosity & Reynolds sensitive; straight lengths depend on the model and upstream disturbances

Coriolis Mass Meter Tube Vibration Phase Shift ( Δ t \Delta t Δ t ) m ˙ = C ⋅ Δ t \dot{m} = C \cdot \Delta t m ˙ = C ⋅ Δ t True mass flow ; density from resonant f 0 f_0 f 0 ​ ; follow model-specific installation requirements

Variable Area (Rotameter) Float Equilibrium Height Drag + Buoyancy = Float Weight Calibrated for specific gas/liquid density; read the manufacturer-specified float reference

Coriolis Mass Flow Advantages

Coriolis meters represent a major advancement in fluid metrology. When fluid travels through vibrating tubes (excited at natural resonance by an electromagnetic drive coil), the fluid mass experiences Coriolis accelerations directed opposite to the vibration at the inlet and in the direction of vibration at the outlet. The resulting twist creates a phase/time shift ( Δ t \Delta t Δ t ) between upstream and downstream optical or magnetic pickoff sensors. The phase shift is directly proportional to true mass flow, less dependent on a developed velocity profile than many volumetric meters. Temperature, installation stress, entrained gas, two-phase flow, zero stability, and fluid conditions can still affect results; follow the meter’s documented limits.

Fluid Volume Calibration and the Gravimetric Method

Liquid volume standards include volumetric flasks, transfer pipettes, burettes, and graduated cylinders.

Class A and Class B: Consult the applicable glassware standard, vessel type, capacity, material, reference temperature, and to-contain/to-deliver designation. Some standards give Class A half the Class B tolerance, but that relationship and one glass composition are not universal for every volumetric item. Use actual certificate corrections and uncertainty when needed.

TC ("To Contain" / In): Calibrated to contain the specified volume at 20 ∘ C 20^\circ\text{C} 2 0 ∘ C (e.g., volumetric flasks).

TD ("To Deliver" / Ex): Calibrated to deliver the specified volume at 20 ∘ C 20^\circ\text{C} 2 0 ∘ C , accounting for the wetting film adhering to the internal walls (e.g., pipettes, burettes).

Gravimetric Calibration of Piston Pipettes (ISO 8655)

Under ISO 8655 , micro-pipettes are calibrated gravimetrically using an analytical balance, high-purity degassed distilled water, and an evaporation trap.

Use the gravimetric conversion appropriate to the apparatus. Density and air-buoyancy terms convert weighed water mass to volume; a thermal-expansion factor applies only when the validated method requires normalization of the apparatus to a reference temperature. Do not apply a glassware expansion factor indiscriminately to piston pipettes. The following illustrates such a reference-temperature model: V 20 = ( m L − m E ) × Z V_{20} = (m_L - m_E) \times Z V 20 ​ = ( m L ​ − m E ​ ) × Z

Where:

m L − m E m_L - m_E m L ​ − m E ​ = net apparent mass of dispensed water (loaded mass minus tare mass)

Z Z Z = conversion factor combining water density at temperature t t t , air buoyancy, and container thermal expansion: Z = 1 ρ w ( t ) − ρ a ( 1 − ρ a ρ b ) [ 1 − γ ( t − 20 ∘ C ) ] Z = \frac{1}{\rho_w(t) - \rho_a} \left( 1 - \frac{\rho_a}{\rho_b} \right) \left[ 1 - \gamma (t - 20^\circ\text{C}) \right] Z = ρ w ​ ( t ) − ρ a ​ 1 ​ ( 1 − ρ b ​ ρ a ​ ​ ) [ 1 − γ ( t − 2 0 ∘ C ) ]

Where:

  • ρ w ( t ) \rho_w(t) ρ w ​ ( t ) = density of pure water at calibration temperature t t t
  • ρ a \rho_a ρ a ​ = air density based on barometric pressure, temperature, and relative humidity
  • ρ b \rho_b ρ b ​ = density of balance calibration weights ( 8000 kg/m 3 8000\text{ kg/m}^3 8000 kg/m 3 )
  • γ \gamma γ = cubical thermal expansion coefficient of the pipette material
  • Pipetting technique

Use the method and manufacturer guidance for the pipette volume, liquid, tips, and aspiration/dispense mode. Prewetting, vertical aspiration, immersion depth, dwell, blowout, and contact with the receiving vessel can affect delivered volume. Do not prescribe one angle or immersion depth for every instrument. Record water temperature, evaporation controls, weighing conditions, and the selected gravimetric conversion.