Study guide
Measurement Models and Uncertainty Contributions
ASQ Certified Calibration Technician CCT
Track your progress as you read
Finished sections get a check mark.
1 SI Definitions and Realizations
The Seven SI Base Units and the 2019 Revision Realizing SI Units in the Laboratory
2 Derived Units, Prefixes, and Constants
Derived Units and Physical Calculations SI Prefixes and Compound-Unit Conversions Physical Constants and Their Applications
3 Measurement Standards and Selection
Types and Roles of Measurement Standards Selecting Standards for an Operational Hierarchy
4 Traceability and Standard Substitution
Traceability Evidence and Impact Review Evaluating and Authorizing Standard Substitution
5 Dimensional and Mechanical Metrology
Linear Displacement and Dimensional Instruments Geometry, Thermal Expansion, and Mass Force, Torque, Pressure, Density, and Viscosity
6 Temperature, Humidity, Volume, and Flow
Temperature Sensors and Thermometer Corrections Humidity, Volume, and Flow
7 Electrical, Optical, and Time Measurements
Voltage, Current, Resistance, and AC Measurement Time, Frequency, Light, Power, and Energy
8 Measurement Methods
Direct, Indirect, and Ratio Methods Transfer, Differential, and UUT Substitution Methods
9 Instrument Characteristics and Measurement Data
Sensitivity, Variability, and Precision Bias, Linearity, Stability, and Hysteresis Data Formats, Readability, and Suitability Resolution, Confidentiality, and Data Integrity
10 Specifications and Error Sources
Interpreting IM&TE Specification Terms Specification Calculations and Operating Qualifiers Instrument, Operator, and Circuit Errors Environmental, Geometric, and Interference Errors
11 Measurement Assurance
Check Standards and Statistical Process Control Proficiency Comparisons and Gage R&R
12 Procedures and Governing Requirements
Procedure Structure and Published Practice Sources Method Development, Deviations, and Handling Regulations, Contracts, and Requirement Precedence
13 Instrument Adjustment and Standardization
Calibration, Zero, and Span Adjustment Linearization, Nulling, and Digital Adjustment
14 Calibration Environment
Temperature and Humidity Control ESD, Dust, Vibration, and Pressure
15 Operations, Roles, and Scheduling
Receiving, As-Found Testing, and Calibration Flow Asset Identification, Tracking, and Release Laboratory Roles and Initial Scheduling Interval Review, Overdue Control, and Workloads
16 Method and Software Validation
Method Verification and Validation Studies Software, Firmware, and Spreadsheet Validation
17 Records and Formal Reporting
Controlled Records and Calibration Certificates Status Labels and Out-of-Tolerance Reporting
18 Uncertainty Concepts and Decision Risk
Error, Uncertainty, and Measurand Definition TAR, TUR, Tolerance Consumption, and Capability Conformity Decisions and False-Accept Risk Guard Bands and Conformity Reporting
19 Uncertainty Models and Budgets
Measurement Models and Uncertainty Contributions Distribution Selection and Budget Assembly
20 Type A and Type B Evaluation
Type A Statistics and Repeatability Outlier Review and Worked Type A Analysis Type B Distribution Models
21 Uncertainty Reporting and Applied Mathematics
Propagation, Covariance, and Effective Degrees of Freedom Coverage Factors and Expanded-Uncertainty Reporting Notation, Unit Conversions, Ratios, and Decibels Rounding, Algebra, Graphs, and Angular Conversions
22 Quality Systems and Tools
Quality-System Components, Planning, and Improvement The Seven Basic Quality Tools
23 Audits and Corrective Action
Audit Types, Planning, and Nonconformance Control Impact Assessment, Root Cause, and Effectiveness
24 Safety, Standards, and Professional Conduct
Electrical, Pressure, Chemical, and Laboratory Hazards PPE, Safety Data Sheets, 6S, and Safe Restoration Quality Standards, GUM, and VIM Accreditation Scope and Professional Ethics
Measurement Models and Uncertainty Contributions
Key Takeaways Explain This Topic AI
Type A and Type B describe evaluation methods rather than random and systematic physical sources.
Sensitivity coefficients convert input uncertainties into the output quantity’s units.
Evaluate relevant standard, UUT, method, environment, operator, and material contributions without double counting.
Last updated: October 2026
The international standard for evaluating and expressing measurement uncertainty is JCGM 100:2008 , commonly known as the GUM (Guide to the Expression of Uncertainty in Measurement) . Under the GUM, evaluating uncertainty is not a subjective guessing game; it is a structured, mathematically rigorous engineering process. An uncertainty budget is the formal document that itemizes all known input components affecting a measurement, quantifies their standard uncertainties, translates them through sensitivity coefficients, and synthesizes them into a single, defensible combined expanded uncertainty.
The Measurement Model Equation: Y = f ( X 1 , X 2 , … , X n ) Y = f(X_1, X_2, \dots, X_n) Y = f ( X 1 , X 2 , … , X n )
In metrology, a measurement result is rarely obtained from a single, isolated instrument reading. Instead, the measurand (output quantity Y Y Y ) is determined from several other input quantities ( X 1 , X 2 , … , X n X_1, X_2, \dots, X_n X 1 , X 2 , … , X n ) through a functional relationship: Y = f ( X 1 , X 2 , … , X n ) Y = f(X_1, X_2, \dots, X_n) Y = f ( X 1 , X 2 , … , X n )
Input Estimates and the Output Estimate:
Each input quantity X i X_i X i has an estimated value denoted by x i x_i x i .
The output estimate of the measurand y y y is obtained by evaluating the functional relationship using the input estimates: y = f ( x 1 , x 2 , … , x n ) y = f(x_1, x_2, \dots, x_n) y = f ( x 1 , x 2 , … , x n )
Direct vs. Indirect Measurement Models:
Direct Measurement Model (Additive Comparison): When an instrument reading is directly compared to a reference standard, the functional relationship is typically additive: Y = X STD + δ X res + δ X rep + δ X temp + δ X drift Y = X_{\text{STD}} + \delta X_{\text{res}} + \delta X_{\text{rep}} + \delta X_{\text{temp}} + \delta X_{\text{drift}} Y = X STD + δ X res + δ X rep + δ X temp + δ X drift
Each residual influence term has an estimated value, often zero when the model justifies that estimate, and associated uncertainty. Apply significant estimated nonzero corrections in the model rather than hiding them inside zero-centered terms. Unknown residual effects and uncertainties in corrections remain after compensation.
2. Indirect Measurement Model (Non-linear Physics): When the measurand is calculated from disparate physical dimensions (e.g., electrical power P = V ⋅ I P = V \cdot I P = V ⋅ I , pressure P = F / A P = F / A P = F / A , or torque τ = F ⋅ L \tau = F \cdot L τ = F ⋅ L ), the functional relationship is multiplicative or non-linear.
Sensitivity Coefficients ( c i c_i c i ):
To determine how an uncertainty in an input quantity X i X_i X i affects the output measurand Y Y Y , metrologists calculate the partial derivative of the model function with respect to X i X_i X i , evaluated at the input estimates: c i = ∂ f ∂ X i ∣ X 1 = x 1 , … , X n = x n c_i = \frac{\partial f}{\partial X_i} \Bigg|_{X_1 = x_1, \dots, X_n = x_n} c i = ∂ X i ∂ f X 1 = x 1 , … , X n = x n
The sensitivity coefficient c i c_i c i acts as a scale factor or conversion multiplier. It converts the units of the input uncertainty u ( x i ) u(x_i) u ( x i ) into the units of the output measurand:
For an additive model Y = X 1 − X 2 Y = X_1 - X_2 Y = X 1 − X 2 : c 1 = ∂ Y ∂ X 1 = + 1 c_1 = \frac{\partial Y}{\partial X_1} = +1 c 1 = ∂ X 1 ∂ Y = + 1 , c 2 = ∂ Y ∂ X 2 = − 1 c_2 = \frac{\partial Y}{\partial X_2} = -1 c 2 = ∂ X 2 ∂ Y = − 1 .
For a multiplicative model P = V ⋅ I P = V \cdot I P = V ⋅ I : c V = ∂ P ∂ V = I c_V = \frac{\partial P}{\partial V} = I c V = ∂ V ∂ P = I , and c I = ∂ P ∂ I = V c_I = \frac{\partial P}{\partial I} = V c I = ∂ I ∂ P = V .
The uncertainty contribution of the i i i -th component to the output is: u i ( y ) = ∣ c i ∣ ⋅ u ( x i ) u_i(y) = |c_i| \cdot u(x_i) u i ( y ) = ∣ c i ∣ ⋅ u ( x i )
Classification of Uncertainty: Type A vs. Type B Evaluations
A cornerstone principle of the GUM is the classification of evaluation methods into Type A and Type B .
Important
Type A and Type B refer to the method of evaluation , not the type of error. Both Type A and Type B components represent physical variations modeled by probability distributions, and both are quantified by standard uncertainties (standard deviations). Do not equate Type A with "random error" and Type B with "systematic error."
Type A Evaluation (Statistical Analysis)
A Type A evaluation of standard uncertainty is obtained by the statistical analysis of a series of repeated independent observations under repeatable or reproducible conditions:
Sample Mean ( q ˉ \bar{q} q ˉ ): The best estimate of the quantity from n n n repeated readings: q ˉ = 1 n ∑ k = 1 n q k \bar{q} = \frac{1}{n} \sum_{k=1}^n q_k q ˉ = n 1 k = 1 ∑ n q k
Sample Standard Deviation ( s ( q k ) s(q_k) s ( q k ) ): Characterizes the repeatability dispersion of individual readings: s ( q k ) = 1 n − 1 ∑ k = 1 n ( q k − q ˉ ) 2 s(q_k) = \sqrt{\frac{1}{n - 1} \sum_{k=1}^n (q_k - \bar{q})^2} s ( q k ) = n − 1 1 k = 1 ∑ n ( q k − q ˉ ) 2
Standard Uncertainty of the Mean ( u ( q ˉ ) u(\bar{q}) u ( q ˉ ) ): If the reported calibration result is the average of the n n n readings, the standard uncertainty is the standard deviation of the mean (standard error): u ( q ˉ ) = s ( q k ) n u(\bar{q}) = \frac{s(q_k)}{\sqrt{n}} u ( q ˉ ) = n s ( q k )
(Note: If a single reading is used in routine service, the standard uncertainty is s ( q k ) s(q_k) s ( q k ) , not divided by n \sqrt{n} n ).
Degrees of Freedom ( ν \nu ν ): For a Type A evaluation of n n n observations, the degrees of freedom are: ν = n − 1 \nu = n - 1 ν = n − 1
Type B Evaluation (Non-Statistical Analysis)
A Type B evaluation of standard uncertainty is obtained by methods other than the statistical analysis of current observations . Type B evaluations draw upon scientific judgment, engineering principles, and available documentation:
Calibration certificates and reports of higher-echelon reference standards.
Manufacturer technical specifications, instrument manuals, and tolerance limits.
Reference data from certified standard reference materials (SRMs) or physical handbooks.
Historical drift data from laboratory control charts.
Knowledge of the physical behavior of materials, thermal coefficients, and fixturing mechanics.
The Six Primary Sources of Calibration Uncertainty
Consider relevant contributions from the following six categories. They are a useful checklist, not an exhaustive or universally mandatory six-part budget; include effects that matter for the actual measurement model:
Reference Standard Uncertainty and Drift
Calibration Uncertainty: The reference standard's calibration certificate provides an expanded uncertainty U STD U_{\text{STD}} U STD at a stated coverage factor k k k (typically k = 2 k = 2 k = 2 ). The standard uncertainty is u ( x STD ) = U STD / k u(x_{\text{STD}}) = U_{\text{STD}} / k u ( x STD ) = U STD / k .
Reference drift: Evaluate stability evidence, elapsed time, and the applicable model. A reliable symmetric drift bound with no better information can justify a rectangular component u d r i f t = a / 3 u_{drift}=a/\sqrt{3} u d r i f t = a / 3 ; observed drift may support a different distribution or an explicit correction. Include uncertainty in predictions and do not assume every drift process is uniform.
Unit Under Test (UUT) Resolution and Repeatability
Digital Display Resolution ( δ x \delta x δ x ): A digital display can only resolve values to within its least significant digit δ x \delta x δ x . Under a nearest-rounding quantization model, the rounding error is assumed equally likely within the interval [ − δ x / 2 , + δ x / 2 ] [-\delta x/2, +\delta x/2] [ − δ x /2 , + δ x /2 ] . The semi-range (half-width) is a = δ x / 2 a = \delta x / 2 a = δ x /2 . Modeled as a rectangular distribution: u res = δ x / 2 3 = δ x 2 3 ≈ 0.289 ⋅ δ x u_{\text{res}} = \frac{\delta x / 2}{\sqrt{3}} = \frac{\delta x}{2\sqrt{3}} \approx 0.289 \cdot \delta x u res = 3 δ x /2 = 2 3 δ x ≈ 0.289 ⋅ δ x
Analog Scale Resolution: For an analog pointer and dial scale, the half-width is typically taken as half of the smallest scale graduation (or 1 / 5 1/5 1/5 to 1 / 10 1/10 1/10 of a division if optical interpolation is verified), divided by 3 \sqrt{3} 3 or 6 \sqrt{6} 6 .
UUT Repeatability: Short-term repeatability observed during the calibration run, evaluated as Type A.
Environmental Influence Quantities
Environmental parameters alter both the standard and the UUT:
Temperature Fluctuation ( Δ T \Delta T Δ T ): Deviations from the 20.0 ∘ C 20.0^\circ\text{C} 20. 0 ∘ C reference baseline cause differential thermal expansion in dimensional metrology.
Barometric Pressure ( P amb P_{\text{amb}} P amb ): Atmospheric pressure changes affect air density ( ρ air \rho_{\text{air}} ρ air ), altering the buoyant force in precision mass calibration and deadweight pressure balances.
Relative Humidity ( R H RH R H ): Variations influence surface moisture adsorption on mass standards, dielectric constants in electrical capacitance standards, and dimensional stability of polymers.
Measurement Method and Fixturing Artifacts
Electrical Calibrations: Lead resistance in 2-wire vs. 4-wire connections, thermal electromotive forces (thermal EMFs / Seebeck effect) generated at bimetallic junctions, AC common-mode leakage, and electromagnetic interference (EMI).
Dimensional Calibrations: Cosine error caused by angular misalignment between the measuring axis and the artifact axis ( Δ L = L ( 1 − cos θ ) ≈ L θ 2 / 2 \Delta L = L(1 - \cos\theta) \approx L\theta^2 / 2 Δ L = L ( 1 − cos θ ) ≈ L θ 2 /2 ), Abbe offset errors ( E = d ⋅ tan θ E = d \cdot \tan\theta E = d ⋅ tan θ ), and elastic Hertzian contact deformation under measuring probe forces.
Pressure Calibrations: Head height fluid column offsets ( Δ P = ρ g Δ h \Delta P = \rho g \Delta h Δ P = ρ g Δ h ) and fluid meniscus surface tension.
Operator Technique and Interaction
Human parallax error when reading analog scale pointers from an oblique angle.
Clamping torque variability on mechanical micrometers lacking calibrated friction thimbles.
Hand-heat transfer warming precision steel gauge blocks during prolonged manual manipulation.
Material Thermo-Physical Properties
In dimensional comparisons between a standard and a UUT made of different materials, the differential thermal expansion equation is: Δ L = L ⋅ [ ( α UUT − α STD ) ⋅ Δ T + α UUT ⋅ δ T diff ] \Delta L = L \cdot [(\alpha_{\text{UUT}} - \alpha_{\text{STD}}) \cdot \Delta T + \alpha_{\text{UUT}} \cdot \delta T_{\text{diff}}] Δ L = L ⋅ [( α UUT − α STD ) ⋅ Δ T + α UUT ⋅ δ T diff ]
where α \alpha α is the coefficient of thermal expansion (CTE), Δ T \Delta T Δ T is the deviation of the standard temperature from 20 ∘ C 20^\circ\text{C} 2 0 ∘ C , and δ T diff \delta T_{\text{diff}} δ T diff is the temperature difference between the two artifacts.
Even if temperatures are measured, uncertainty in the literature values of CTE ( ± 10 % \pm 10\% ± 10% to ± 15 % \pm 15\% ± 15% on α \alpha α ) introduces a non-negligible Type B uncertainty component.