Study guide
Type B Distribution Models
ASQ Certified Calibration Technician CCT
Track your progress as you read
Finished sections get a check mark.
1 SI Definitions and Realizations
The Seven SI Base Units and the 2019 Revision Realizing SI Units in the Laboratory
2 Derived Units, Prefixes, and Constants
Derived Units and Physical Calculations SI Prefixes and Compound-Unit Conversions Physical Constants and Their Applications
3 Measurement Standards and Selection
Types and Roles of Measurement Standards Selecting Standards for an Operational Hierarchy
4 Traceability and Standard Substitution
Traceability Evidence and Impact Review Evaluating and Authorizing Standard Substitution
5 Dimensional and Mechanical Metrology
Linear Displacement and Dimensional Instruments Geometry, Thermal Expansion, and Mass Force, Torque, Pressure, Density, and Viscosity
6 Temperature, Humidity, Volume, and Flow
Temperature Sensors and Thermometer Corrections Humidity, Volume, and Flow
7 Electrical, Optical, and Time Measurements
Voltage, Current, Resistance, and AC Measurement Time, Frequency, Light, Power, and Energy
8 Measurement Methods
Direct, Indirect, and Ratio Methods Transfer, Differential, and UUT Substitution Methods
9 Instrument Characteristics and Measurement Data
Sensitivity, Variability, and Precision Bias, Linearity, Stability, and Hysteresis Data Formats, Readability, and Suitability Resolution, Confidentiality, and Data Integrity
10 Specifications and Error Sources
Interpreting IM&TE Specification Terms Specification Calculations and Operating Qualifiers Instrument, Operator, and Circuit Errors Environmental, Geometric, and Interference Errors
11 Measurement Assurance
Check Standards and Statistical Process Control Proficiency Comparisons and Gage R&R
12 Procedures and Governing Requirements
Procedure Structure and Published Practice Sources Method Development, Deviations, and Handling Regulations, Contracts, and Requirement Precedence
13 Instrument Adjustment and Standardization
Calibration, Zero, and Span Adjustment Linearization, Nulling, and Digital Adjustment
14 Calibration Environment
Temperature and Humidity Control ESD, Dust, Vibration, and Pressure
15 Operations, Roles, and Scheduling
Receiving, As-Found Testing, and Calibration Flow Asset Identification, Tracking, and Release Laboratory Roles and Initial Scheduling Interval Review, Overdue Control, and Workloads
16 Method and Software Validation
Method Verification and Validation Studies Software, Firmware, and Spreadsheet Validation
17 Records and Formal Reporting
Controlled Records and Calibration Certificates Status Labels and Out-of-Tolerance Reporting
18 Uncertainty Concepts and Decision Risk
Error, Uncertainty, and Measurand Definition TAR, TUR, Tolerance Consumption, and Capability Conformity Decisions and False-Accept Risk Guard Bands and Conformity Reporting
19 Uncertainty Models and Budgets
Measurement Models and Uncertainty Contributions Distribution Selection and Budget Assembly
20 Type A and Type B Evaluation
Type A Statistics and Repeatability Outlier Review and Worked Type A Analysis Type B Distribution Models
21 Uncertainty Reporting and Applied Mathematics
Propagation, Covariance, and Effective Degrees of Freedom Coverage Factors and Expanded-Uncertainty Reporting Notation, Unit Conversions, Ratios, and Decibels Rounding, Algebra, Graphs, and Angular Conversions
22 Quality Systems and Tools
Quality-System Components, Planning, and Improvement The Seven Basic Quality Tools
23 Audits and Corrective Action
Audit Types, Planning, and Nonconformance Control Impact Assessment, Root Cause, and Effectiveness
24 Safety, Standards, and Professional Conduct
Electrical, Pressure, Chemical, and Laboratory Hazards PPE, Safety Data Sheets, 6S, and Safe Restoration Quality Standards, GUM, and VIM Accreditation Scope and Professional Ethics
Type B Distribution Models
Key Takeaways Explain This Topic AI
Symmetric rectangular, triangular, and arcsine bounds give standard uncertainties a/√3, a/√6, and a/√2.
For the same half-width, the arcsine model gives greater standard uncertainty than the rectangular or triangular model.
Convert a reported expanded uncertainty to standard uncertainty using its stated coverage factor.
Last updated: October 2026
Many contributions, such as certificate uncertainty or specification bounds, are evaluated using Type B methods . The classification follows the evaluation method. Environmental variation, drift, or repeatability can instead be evaluated statistically when adequate observation data are available; a physical source is not inherently Type A or Type B.
According to GUM Clause 4.3, a Type B evaluation relies on the metrologist's scientific judgment using all available information, including:
- Previous calibration data and historical control charts
- Manufacturer technical specifications
- Calibrated values and expanded uncertainties from accredited calibration certificates
- Uncertainties assigned to reference data taken from physical handbooks
- Digital display resolution and quantization limits
To combine a Type B uncertainty component with Type A components in a combined uncertainty budget, the non-statistical tolerance or limit ± a \pm a ± a must be converted into an equivalent standard uncertainty u ( x i ) u(x_i) u ( x i ) , which represents one standard deviation ( σ \sigma σ ). This conversion requires selecting an appropriate mathematical model: the probability distribution .
Master Summary of Probability Distributions in Metrology
Distribution Shape Description Half-Width Symbol Divisor to Obtain u ( x i ) u(x_i) u ( x i ) Formula for Standard Uncertainty u ( x i ) u(x_i) u ( x i ) Typical Metrological Applications
Rectangular (Uniform) Flat plateau between − a -a − a and + a +a + a ; zero outside a a a 3 ≈ 1.73205 \sqrt{3} \approx 1.73205 3 ≈ 1.73205 u = a 3 ≈ 0.577 a u = \frac{a}{\sqrt{3}} \approx 0.577 a u = 3 a ≈ 0.577 a Digital resolution ( ± 0.5 LSD \pm 0.5\text{ LSD} ± 0.5 LSD ), environmental chamber stability ( ± Δ T \pm \Delta T ± Δ T ), manufacturer specs without distribution or confidence level
Triangular Symmetrical triangle peaking at center, tapering to zero at ± a \pm a ± a a a a 6 ≈ 2.44949 \sqrt{6} \approx 2.44949 6 ≈ 2.44949 u = a 6 ≈ 0.408 a u = \frac{a}{\sqrt{6}} \approx 0.408 a u = 6 a ≈ 0.408 a Two independent equal uniform contributions, or other evidence supporting a triangular model
U-Shaped (Arc-Sine) Bimodal bathtub curve peaking at ± a \pm a ± a , minimum at center a a a 2 ≈ 1.41421 \sqrt{2} \approx 1.41421 2 ≈ 1.41421 u = a 2 ≈ 0.707 a u = \frac{a}{\sqrt{2}} \approx 0.707 a u = 2 a ≈ 0.707 a RF and microwave transmission line mismatch, sinusoidal power fluctuations, uniform random phase of a sinusoidal influence
Normal (Gaussian, 95%) Bell curve with stated 95 % 95\% 95% confidence interval a a a 1.960 ≈ 2.000 1.960 \approx 2.000 1.960 ≈ 2.000 u = a 1.96 ≈ a 2 u = \frac{a}{1.96} \approx \frac{a}{2} u = 1.96 a ≈ 2 a Manufacturer specifications quoting 95 % 95\% 95% confidence, expanded uncertainty from accredited certificates with k = 2 k=2 k = 2
Normal (Gaussian, 99.73%) Bell curve with stated 3 σ 3\sigma 3 σ ( 99.73 % 99.73\% 99.73% ) limits a a a 3.000 3.000 3.000 u = a 3 ≈ 0.333 a u = \frac{a}{3} \approx 0.333 a u = 3 a ≈ 0.333 a Conservative manufacturer tolerances stated as 3 σ 3\sigma 3 σ or 99.73 % 99.73\% 99.73% limits
- Detailed Mathematical Derivations and Applications
- The Rectangular (Uniform) Distribution
The rectangular distribution is the default model under the GUM when a quantity is known to lie within upper and lower bounds [ − a , + a ] [-a, +a] [ − a , + a ] , but there is no specific knowledge about the probability of values within those bounds.
Mathematical Derivation of Variance
The probability density function (PDF) is constant across the interval: p ( x ) = { 1 2 a , − a ≤ x ≤ + a 0 , ∣ x ∣ > a p(x) = \begin{cases} \frac{1}{2a}, & -a \le x \le +a \\ 0, & |x| > a \end{cases} p ( x ) = { 2 a 1 , 0 , − a ≤ x ≤ + a ∣ x ∣ > a
Because the distribution is symmetric, the mean is zero ( E [ X ] = 0 E[X] = 0 E [ X ] = 0 ). The variance is the expected value of x 2 x^2 x 2 : V ( x ) = σ 2 = ∫ − a + a x 2 p ( x ) d x = ∫ − a + a x 2 ( 1 2 a ) d x = 1 2 a [ x 3 3 ] − a + a V(x) = \sigma^2 = \int_{-a}^{+a} x^2 p(x) dx = \int_{-a}^{+a} x^2 \left(\frac{1}{2a}\right) dx = \frac{1}{2a} \left[ \frac{x^3}{3} \right]_{-a}^{+a} V ( x ) = σ 2 = ∫ − a + a x 2 p ( x ) d x = ∫ − a + a x 2 ( 2 a 1 ) d x = 2 a 1 [ 3 x 3 ] − a + a V ( x ) = 1 2 a ( a 3 3 − ( − a ) 3 3 ) = 1 2 a ( 2 a 3 3 ) = a 2 3 V(x) = \frac{1}{2a} \left( \frac{a^3}{3} - \frac{(-a)^3}{3} \right) = \frac{1}{2a} \left( \frac{2a^3}{3} \right) = \frac{a^2}{3} V ( x ) = 2 a 1 ( 3 a 3 − 3 ( − a ) 3 ) = 2 a 1 ( 3 2 a 3 ) = 3 a 2
Taking the square root yields the standard uncertainty: u ( x ) = a 2 3 = a 3 ≈ a 1.73205 ≈ 0.5774 a u(x) = \sqrt{\frac{a^2}{3}} = \frac{a}{\sqrt{3}} \approx \frac{a}{1.73205} \approx 0.5774 a u ( x ) = 3 a 2 = 3 a ≈ 1.73205 a ≈ 0.5774 a
Primary Calibration Applications
Digital Display Resolution (Quantization): A digital instrument rounds its internal estimate to the nearest display increment (least significant digit, LSD, or resolution d d d ). The maximum rounding error is half of the resolution: a = d / 2 = 0.5 LSD a = d/2 = 0.5\text{ LSD} a = d /2 = 0.5 LSD . u resolution = d / 2 3 = d 2 3 = d 12 ≈ 0.2887 d u_{\text{resolution}} = \frac{d / 2}{\sqrt{3}} = \frac{d}{2\sqrt{3}} = \frac{d}{\sqrt{12}} \approx 0.2887 d u resolution = 3 d /2 = 2 3 d = 12 d ≈ 0.2887 d
Example: A digital indicator displays readings to 0.001 mm 0.001\text{ mm} 0.001 mm ( 1 μ m 1\ \mu\text{m} 1 μ m ). The semi-range is a = 0.0005 mm a = 0.0005\text{ mm} a = 0.0005 mm . The standard uncertainty due to resolution is: u res = 0.0005 mm 3 = 0.000289 mm = 0.289 μ m u_{\text{res}} = \frac{0.0005\text{ mm}}{\sqrt{3}} = 0.000289\text{ mm} = 0.289\ \mu\text{m} u res = 3 0.0005 mm = 0.000289 mm = 0.289 μ m
Manufacturer Specifications Without Stated Confidence: When an equipment manual quotes an accuracy of " ± 0.05 % \pm 0.05\% ± 0.05% of reading" with no mention of a normal distribution or confidence level, a rectangular model can be justified when reliable bounds are known and no better distribution information is available with half-width a = 0.05 % × Reading a = 0.05\% \times \text{Reading} a = 0.05% × Reading .
Environmental Temperature Stability: A laboratory thermostat maintains room temperature within ± 1.5 ∘ C \pm 1.5^\circ\text{C} ± 1. 5 ∘ C of setpoint. If no continuous logging histogram is available, assume a rectangular distribution with a = 1.5 ∘ C a = 1.5^\circ\text{C} a = 1. 5 ∘ C : u = 1.5 / 3 = 0.866 ∘ C u = 1.5 / \sqrt{3} = 0.866^\circ\text{C} u = 1.5/ 3 = 0.86 6 ∘ C .
The Triangular Distribution
The triangular distribution applies when values are known to cluster near the center of the interval [ − a , + a ] [-a, +a] [ − a , + a ] and diminish linearly toward the extremes.
Mathematical Derivation of Variance
By convolution, the sum (sum or difference) of two independent rectangular distributions with identical half-widths a / 2 a/2 a /2 forms an exact symmetric triangle with base [ − a , + a ] [-a, +a] [ − a , + a ] .
The probability density function is: p ( x ) = { 1 a ( 1 − ∣ x ∣ a ) , − a ≤ x ≤ + a 0 , ∣ x ∣ > a p(x) = \begin{cases} \frac{1}{a} \left(1 - \frac{|x|}{a}\right), & -a \le x \le +a \\ 0, & |x| > a \end{cases} p ( x ) = { a 1 ( 1 − a ∣ x ∣ ) , 0 , − a ≤ x ≤ + a ∣ x ∣ > a
Integrating x 2 p ( x ) x^2 p(x) x 2 p ( x ) across the interval yields the variance: V ( x ) = σ 2 = ∫ − a + a x 2 1 a ( 1 − ∣ x ∣ a ) d x = 2 ∫ 0 a ( x 2 a − x 3 a 2 ) d x = 2 [ a 3 3 a − a 4 4 a 2 ] = 2 ( a 2 3 − a 2 4 ) = a 2 6 V(x) = \sigma^2 = \int_{-a}^{+a} x^2 \frac{1}{a}\left(1 - \frac{|x|}{a}\right) dx = 2 \int_0^a \left(\frac{x^2}{a} - \frac{x^3}{a^2}\right) dx = 2 \left[ \frac{a^3}{3a} - \frac{a^4}{4a^2} \right] = 2 \left( \frac{a^2}{3} - \frac{a^2}{4} \right) = \frac{a^2}{6} V ( x ) = σ 2 = ∫ − a + a x 2 a 1 ( 1 − a ∣ x ∣ ) d x = 2 ∫ 0 a ( a x 2 − a 2 x 3 ) d x = 2 [ 3 a a 3 − 4 a 2 a 4 ] = 2 ( 3 a 2 − 4 a 2 ) = 6 a 2
Taking the square root yields the standard uncertainty: u ( x ) = a 2 6 = a 6 ≈ a 2.44949 ≈ 0.4082 a u(x) = \sqrt{\frac{a^2}{6}} = \frac{a}{\sqrt{6}} \approx \frac{a}{2.44949} \approx 0.4082 a u ( x ) = 6 a 2 = 6 a ≈ 2.44949 a ≈ 0.4082 a
A supported triangular model: Two independent, equal uniform components sum to an exact triangle. A backlash mechanism or controlled bath does not automatically have that shape; inspect the physics and data before selecting the divisor.
The U-Shaped (Arc-Sine) Distribution
The U-shaped distribution applies when a physical parameter varies cyclically or sinusoidally over time, dwelling primarily near the turnaround peaks and passing rapidly through the center.
Mathematical Derivation of Variance
Consider a variable undergoing simple harmonic oscillation with amplitude a a a : x ( t ) = a sin ( ω t ) x(t) = a \sin(\omega t) x ( t ) = a sin ( ω t )
The probability of observing a value x x x in an infinitesimal window d x dx d x is inversely proportional to the speed of the wave: ∣ d x / d t ∣ = a ω cos ( ω t ) = ω a 2 − x 2 |dx/dt| = a\omega \cos(\omega t) = \omega \sqrt{a^2 - x^2} ∣ d x / d t ∣ = aω cos ( ω t ) = ω a 2 − x 2 . Normalizing over the domain ( − a , + a ) (-a, +a) ( − a , + a ) gives the arc-sine probability density function: p ( x ) = 1 π a 2 − x 2 , − a < x < + a p(x) = \frac{1}{\pi \sqrt{a^2 - x^2}}, \quad -a < x < +a p ( x ) = π a 2 − x 2 1 , − a < x < + a
The variance is obtained by integrating: V ( x ) = σ 2 = ∫ − a + a x 2 π a 2 − x 2 d x V(x) = \sigma^2 = \int_{-a}^{+a} \frac{x^2}{\pi \sqrt{a^2 - x^2}} dx V ( x ) = σ 2 = ∫ − a + a π a 2 − x 2 x 2 d x
Substituting x = a sin θ x = a \sin\theta x = a sin θ (where d x = a cos θ d θ dx = a \cos\theta d\theta d x = a cos θ d θ and a 2 − x 2 = a cos θ \sqrt{a^2 - x^2} = a \cos\theta a 2 − x 2 = a cos θ ): V ( x ) = 1 π ∫ − π / 2 + π / 2 ( a 2 sin 2 θ ) d θ = a 2 π ∫ − π / 2 + π / 2 ( 1 − cos 2 θ 2 ) d θ = a 2 π [ θ 2 ] − π / 2 + π / 2 = a 2 2 V(x) = \frac{1}{\pi} \int_{-\pi/2}^{+\pi/2} (a^2 \sin^2\theta) d\theta = \frac{a^2}{\pi} \int_{-\pi/2}^{+\pi/2} \left(\frac{1 - \cos 2\theta}{2}\right) d\theta = \frac{a^2}{\pi} \left[ \frac{\theta}{2} \right]_{-\pi/2}^{+\pi/2} = \frac{a^2}{2} V ( x ) = π 1 ∫ − π /2 + π /2 ( a 2 sin 2 θ ) d θ = π a 2 ∫ − π /2 + π /2 ( 2 1 − cos 2 θ ) d θ = π a 2 [ 2 θ ] − π /2 + π /2 = 2 a 2
Taking the square root yields the standard uncertainty: u ( x ) = a 2 2 = a 2 ≈ a 1.41421 ≈ 0.7071 a u(x) = \sqrt{\frac{a^2}{2}} = \frac{a}{\sqrt{2}} \approx \frac{a}{1.41421} \approx 0.7071 a u ( x ) = 2 a 2 = 2 a ≈ 1.41421 a ≈ 0.7071 a
Environmental cycling: A sinusoidal temperature variation sampled at a uniformly random phase has an arcsine distribution. On/off thermostat control alone does not prove this model; an exponential heating/cooling cycle can have another distribution.
Normal Distribution and Calibration Certificate Conversions
When a specification or calibration certificate quotes an uncertainty with a stated confidence interval, use the stated distribution and coverage information; a confidence statement alone does not prove a normal model.
Converting Expanded Uncertainty from a Calibration Certificate
Use the certificate’s stated expanded uncertainty, coverage factor, and coverage information. For a justified normal model with adequate degrees of freedom, k = 2 k=2 k = 2 corresponds to about 95.45% central coverage; it is not the interpretation of every certificate or distribution.
To enter this component into your working calibration uncertainty budget as a standard uncertainty u ( x i ) u(x_i) u ( x i ) , simply divide by the stated coverage factor: u ( x i ) = U k u(x_i) = \frac{U}{k} u ( x i ) = k U
Hypothetical exercise: A reference resistor certificate gives the following value, uncertainty, and coverage factor. These are invented teaching inputs rather than quotations from an actual NIST report: R std = 10 , 000.042 Ω , U = 0.024 Ω ( k = 2.00 , approx. 95 % confidence ) R_{\text{std}} = 10,000.042\ \Omega, \quad U = 0.024\ \Omega \quad (k = 2.00, \text{ approx. } 95\% \text{ confidence}) R std = 10 , 000.042 Ω , U = 0.024 Ω ( k = 2.00 , approx. 95% confidence )
The standard uncertainty to enter into the laboratory's uncertainty budget is: u ( R std ) = 0.024 Ω 2.00 = 0.012 Ω u(R_{\text{std}}) = \frac{0.024\ \Omega}{2.00} = 0.012\ \Omega u ( R std ) = 2.00 0.024 Ω = 0.012 Ω
Converting Manufacturer Specifications with Stated Confidence Levels
If the stated interval is central coverage under a justified normal model, use the corresponding normal quantile. Confidence alone does not establish normality, infinite degrees of freedom, or an exact divisor. If a certificate supplies its coverage factor, use that factor and its stated interpretation. The following quantiles are normal-model examples:
90.00 % 90.00\% 90.00% Confidence: Divisor k = 1.645 ⟹ u = a / 1.645 k = 1.645 \implies u = a / 1.645 k = 1.645 ⟹ u = a /1.645
95.00 % 95.00\% 95.00% Confidence: Divisor k = 1.960 ⟹ u = a / 1.960 k = 1.960 \implies u = a / 1.960 k = 1.960 ⟹ u = a /1.960
95.45 % 95.45\% 95.45% Confidence ( 2 σ 2\sigma 2 σ ): Divisor k = 2.000 ⟹ u = a / 2.000 k = 2.000 \implies u = a / 2.000 k = 2.000 ⟹ u = a /2.000
99.00 % 99.00\% 99.00% Confidence: Divisor k = 2.576 ⟹ u = a / 2.576 k = 2.576 \implies u = a / 2.576 k = 2.576 ⟹ u = a /2.576
99.73 % 99.73\% 99.73% Confidence ( 3 σ 3\sigma 3 σ ): Divisor k = 3.000 ⟹ u = a / 3.000 k = 3.000 \implies u = a / 3.000 k = 3.000 ⟹ u = a /3.000
Comparative Worked Calculation: Impact of Divisor Selection
To illustrate how the choice of probability distribution affects calibration budgets, consider an instrument tolerance limit of ± a = ± 12.0 μ m \pm a = \pm 12.0\ \mu\text{m} ± a = ± 12.0 μ m . Let us calculate the resulting standard uncertainty under all four models:
Important
Critical Metrological Insight: For the exact same physical tolerance band ( ± 12.0 μ m \pm 12.0\ \mu\text{m} ± 12.0 μ m ):
The U-shaped distribution yields the largest standard uncertainty ( 8.49 μ m 8.49\ \mu\text{m} 8.49 μ m ) because the data concentrates near the extreme limits.
The triangular distribution yields a smaller standard uncertainty ( 4.90 μ m 4.90\ \mu\text{m} 4.90 μ m ) than the rectangular distribution because data concentrates near zero.
Mistakenly applying a triangular divisor ( 6 \sqrt{6} 6 ) to a rectangular process underestimates the uncertainty by nearly 30 % 30\% 30% ( 4.90 4.90 4.90 vs 6.93 μ m 6.93\ \mu\text{m} 6.93 μ m ), potentially causing false-accept audit nonconformances!
Technical reference checked October 10, 2026: JCGM uncertainty and vocabulary publications .