Study guide
TAR, TUR, Tolerance Consumption, and Capability
ASQ Certified Calibration Technician CCT
Track your progress as you read
Finished sections get a check mark.
1 SI Definitions and Realizations
The Seven SI Base Units and the 2019 Revision Realizing SI Units in the Laboratory
2 Derived Units, Prefixes, and Constants
Derived Units and Physical Calculations SI Prefixes and Compound-Unit Conversions Physical Constants and Their Applications
3 Measurement Standards and Selection
Types and Roles of Measurement Standards Selecting Standards for an Operational Hierarchy
4 Traceability and Standard Substitution
Traceability Evidence and Impact Review Evaluating and Authorizing Standard Substitution
5 Dimensional and Mechanical Metrology
Linear Displacement and Dimensional Instruments Geometry, Thermal Expansion, and Mass Force, Torque, Pressure, Density, and Viscosity
6 Temperature, Humidity, Volume, and Flow
Temperature Sensors and Thermometer Corrections Humidity, Volume, and Flow
7 Electrical, Optical, and Time Measurements
Voltage, Current, Resistance, and AC Measurement Time, Frequency, Light, Power, and Energy
8 Measurement Methods
Direct, Indirect, and Ratio Methods Transfer, Differential, and UUT Substitution Methods
9 Instrument Characteristics and Measurement Data
Sensitivity, Variability, and Precision Bias, Linearity, Stability, and Hysteresis Data Formats, Readability, and Suitability Resolution, Confidentiality, and Data Integrity
10 Specifications and Error Sources
Interpreting IM&TE Specification Terms Specification Calculations and Operating Qualifiers Instrument, Operator, and Circuit Errors Environmental, Geometric, and Interference Errors
11 Measurement Assurance
Check Standards and Statistical Process Control Proficiency Comparisons and Gage R&R
12 Procedures and Governing Requirements
Procedure Structure and Published Practice Sources Method Development, Deviations, and Handling Regulations, Contracts, and Requirement Precedence
13 Instrument Adjustment and Standardization
Calibration, Zero, and Span Adjustment Linearization, Nulling, and Digital Adjustment
14 Calibration Environment
Temperature and Humidity Control ESD, Dust, Vibration, and Pressure
15 Operations, Roles, and Scheduling
Receiving, As-Found Testing, and Calibration Flow Asset Identification, Tracking, and Release Laboratory Roles and Initial Scheduling Interval Review, Overdue Control, and Workloads
16 Method and Software Validation
Method Verification and Validation Studies Software, Firmware, and Spreadsheet Validation
17 Records and Formal Reporting
Controlled Records and Calibration Certificates Status Labels and Out-of-Tolerance Reporting
18 Uncertainty Concepts and Decision Risk
Error, Uncertainty, and Measurand Definition TAR, TUR, Tolerance Consumption, and Capability Conformity Decisions and False-Accept Risk Guard Bands and Conformity Reporting
19 Uncertainty Models and Budgets
Measurement Models and Uncertainty Contributions Distribution Selection and Budget Assembly
20 Type A and Type B Evaluation
Type A Statistics and Repeatability Outlier Review and Worked Type A Analysis Type B Distribution Models
21 Uncertainty Reporting and Applied Mathematics
Propagation, Covariance, and Effective Degrees of Freedom Coverage Factors and Expanded-Uncertainty Reporting Notation, Unit Conversions, Ratios, and Decibels Rounding, Algebra, Graphs, and Angular Conversions
22 Quality Systems and Tools
Quality-System Components, Planning, and Improvement The Seven Basic Quality Tools
23 Audits and Corrective Action
Audit Types, Planning, and Nonconformance Control Impact Assessment, Root Cause, and Effectiveness
24 Safety, Standards, and Professional Conduct
Electrical, Pressure, Chemical, and Laboratory Hazards PPE, Safety Data Sheets, 6S, and Safe Restoration Quality Standards, GUM, and VIM Accreditation Scope and Professional Ethics
TAR, TUR, Tolerance Consumption, and Capability
Key Takeaways Explain This Topic AI
TAR uses an accuracy bound, whereas TUR uses evaluated process uncertainty under a stated convention.
A symmetric ±0.100 bar tolerance and 0.025 bar expanded uncertainty give TUR 4:1 using half-widths.
Gage capability indices compare variation and bias with chosen study criteria; they do not supply universal approval thresholds.
Last updated: October 2026
Test Accuracy Ratio (TAR) vs. Test Uncertainty Ratio (TUR)
Calibration laboratories must quantitatively evaluate whether their calibration process has sufficient capability to verify the tolerance of a Unit Under Test (UUT). Historically, this was evaluated using the Test Accuracy Ratio (TAR) . In modern accredited calibration (ISO/IEC 17025 and ANSI/NCSL Z540.3), TAR has been largely superseded by the Test Uncertainty Ratio (TUR) .
Test Accuracy Ratio (TAR)
Originating in military standards such as MIL-STD-45662A, TAR is defined as the ratio of the tolerance limit of the UUT to the tolerance limit of the calibration standard: TAR = Tolerance UUT Tolerance STD = Span UUT Span STD \text{TAR} = \frac{\text{Tolerance}_{\text{UUT}}}{\text{Tolerance}_{\text{STD}}} = \frac{\text{Span}_{\text{UUT}}}{\text{Span}_{\text{STD}}} TAR = Tolerance STD Tolerance UUT = Span STD Span UUT
For two-sided symmetric tolerances ( ± T UUT \pm T_{\text{UUT}} ± T UUT and ± T STD \pm T_{\text{STD}} ± T STD ): TAR = T UUT T STD \text{TAR} = \frac{T_{\text{UUT}}}{T_{\text{STD}}} TAR = T STD T UUT
The Fatal Limitation of TAR:
TAR compares stated accuracy or error bounds using a declared convention. It does not assume the reference has zero error; its limitation is that a reference bound alone omits other process uncertainty contributions. Evaluate the setup, corrections, reference uncertainty, and UUT effects instead of substituting a catalog ratio for that analysis.
It does not account for the standard's calibration drift since its last calibration.
It completely ignores ambient environmental fluctuations (such as temperature gradients in the calibration room).
It omits the resolution limit of the UUT and the comparator.
It omits operator technique, fixturing deformation, and lead resistance.
Consequently, a TAR of 4:1 does not guarantee that the calibration process has an uncertainty 4 times smaller than the UUT tolerance.
Test Uncertainty Ratio (TUR)
To address the deficiencies of TAR, national and international standards—including ANSI/NCSL Z540.3-2006 and ILAC-G8:09/2019 —describe uncertainty-aware capability and decision approaches; TUR is a useful ratio, not a universal ISO mandate. TUR is defined as the ratio of the tolerance span of the UUT attribute being measured to twice the expanded measurement uncertainty ( k = 2 k = 2 k = 2 ) of the calibration process: TUR = Tolerance Span UUT 2 ⋅ U = USL − LSL 2 ⋅ U \text{TUR} = \frac{\text{Tolerance Span}_{\text{UUT}}}{2 \cdot U} = \frac{\text{USL} - \text{LSL}}{2 \cdot U} TUR = 2 ⋅ U Tolerance Span UUT = 2 ⋅ U USL − LSL
where:
USL \text{USL} USL is the Upper Specification Limit of the UUT.
LSL \text{LSL} LSL is the Lower Specification Limit of the UUT.
Tolerance Span UUT = USL − LSL \text{Tolerance Span}_{\text{UUT}} = \text{USL} - \text{LSL} Tolerance Span UUT = USL − LSL (for a symmetric tolerance of ± T \pm T ± T , the span is 2 T 2T 2 T ).
U U U is the expanded measurement uncertainty of the calibration process evaluated at a 95 % 95\% 95% confidence level (conventionally coverage factor k = 2 k = 2 k = 2 ).
For a symmetric two-sided tolerance ( ± T \pm T ± T ): TUR = 2 ⋅ T 2 ⋅ U = T U \text{TUR} = \frac{2 \cdot T}{2 \cdot U} = \frac{T}{U} TUR = 2 ⋅ U 2 ⋅ T = U T
For the following one-sided illustration, explicitly define a nominal-to-limit distance and use that distance in the ratio. Other conventions can differ; state numerator width and uncertainty convention. A ratio alone is not a conformity decision or risk probability. TUR = ∣ Specification Limit − Nominal ∣ U \text{TUR} = \frac{|\text{Specification Limit} - \text{Nominal}|}{U} TUR = U ∣ Specification Limit − Nominal ∣
Crucially, the expanded uncertainty U U U in the denominator is not just the uncertainty of the standard; it is the total combined expanded uncertainty of the calibration process , encompassing the reference standard, environmental influences, UUT resolution, and short-term repeatability.
- Feature / Metric Test Accuracy Ratio (TAR) Test Uncertainty Ratio (TUR)
- Governing Standards MIL-STD-45662A (historical), NAVAIR 17-35QAC-01 ANSI/NCSL Z540.3, ISO/IEC 17025, ILAC-G8
Numerator UUT tolerance band ( USL − LSL \text{USL} - \text{LSL} USL − LSL ) UUT tolerance band ( USL − LSL \text{USL} - \text{LSL} USL − LSL )
Denominator Calibration standard manufacturer accuracy tolerance ( 2 ⋅ T STD 2 \cdot T_{\text{STD}} 2 ⋅ T STD ) Twice the calibration process expanded uncertainty ( 2 ⋅ U 2 \cdot U 2 ⋅ U at k = 2 k = 2 k = 2 )
Includes Process Effects? No (ignores environment, resolution, fixturing) Yes (full GUM uncertainty budget included)
Statistical Meaning Simple ratio of catalog specifications Capability ratio; risk requires additional assumptions
- Conformity Defensibility Weak under ISO/IEC 17025 audits Internationally accepted and legally defensible
- Capability targets and their limits
A 4:1 TUR corresponds to U = T / 4 U=T/4 U = T /4 for a symmetric tolerance; a 10:1 TUR corresponds to U = T / 10 U=T/10 U = T /10 . These numbers can be program targets, but neither proves a universal false-accept probability. Tolerance is a permissible limit, not a probability distribution or variance. Squaring a ratio of uncertainty to tolerance does not give a universal fraction of “tolerance variance.”
The decision rule and risk model must account for result location and uncertainty. A result exactly at a specification boundary can have substantial specific risk even at high TUR. Low TUR can be managed through an agreed guard band, better measurement capability, adjusted scope, or reporting results without a conformity statement. Determine what the contract permits; ISO/IEC 17025 does not automatically ban every process below 4:1.
Percent of Tolerance Consumed
During routine calibration, technicians calculate the percent of tolerance consumed to assess how close an indicated reading is to the allowable specification limits: % of Tolerance = ∣ Measured Error ∣ Allowable Tolerance × 100 % = ∣ y − x ref ∣ T allowable × 100 % \% \text{ of Tolerance} = \frac{|\text{Measured Error}|}{\text{Allowable Tolerance}} \times 100\% = \frac{|y - x_{\text{ref}}|}{T_{\text{allowable}}} \times 100\% % of Tolerance = Allowable Tolerance ∣ Measured Error ∣ × 100% = T allowable ∣ y − x ref ∣ × 100%
The ratio indicates the distance from nominal relative to the chosen tolerance, not future stability or legal permission to use the instrument. For E = + 0.007 mm E=+0.007\text{ mm} E = + 0.007 mm and a symmetric tolerance of ± 0.010 mm \pm0.010\text{ mm} ± 0.010 mm , 70% is consumed. If U = 0.004 mm U=0.004\text{ mm} U = 0.004 mm and the agreed acceptance rule requires ∣ E ∣ + U ≤ T |E|+U\le T ∣ E ∣ + U ≤ T , the result is outside the acceptance zone despite its error being inside tolerance.
A laboratory can choose alert levels to prompt interval review or adjustment, but 70% is not a universal action threshold. Trend, duty cycle, environment, and historical data inform the next interval. An observed error beyond the stated tolerance triggers the applicable nonconforming-work evaluation and decision rule.
Gage Capability Indices ( C g C_g C g and C g k C_{gk} C g k )
While process capability indices ( C p , C p k C_p, C_{pk} C p , C p k ) evaluate manufacturing processes against part tolerances, gage capability indices ( C g , C g k C_g, C_{gk} C g , C g k ) evaluate the suitability of the measurement equipment itself. A Type 1 Gage Study isolates the measurement system from part-to-part variation by repeatedly measuring a single calibrated reference standard.
Repeatability Index ( C g C_g C g )
The C g C_g C g index evaluates the inherent precision (repeatability dispersion) of the measuring instrument relative to the total product tolerance band: C g = 0.20 ⋅ T 6 ⋅ s g C_g = \frac{0.20 \cdot T}{6 \cdot s_g} C g = 6 ⋅ s g 0.20 ⋅ T
where:
T = USL − LSL T = \text{USL} - \text{LSL} T = USL − LSL is the total tolerance band of the attribute to be measured.
The 20% tolerance allocation in this illustrative capability convention is not a universal industry or ISO requirement. Declare the adopted formula, units, tolerance width, study conditions, and acceptance criterion before interpreting an index.
s g s_g s g is the sample standard deviation of the number of repeated measurements performed on the reference standard under repeatable conditions.
6 ⋅ s g 6 \cdot s_g 6 ⋅ s g represents the full 6 σ 6\sigma 6 σ spread of the gage's repeatability distribution ( 99.73 % 99.73\% 99.73% coverage).
Repeatability and Bias Index ( C g k C_{gk} C g k )
The C g k C_{gk} C g k index evaluates both the precision (repeatability) and the trueness (systematic bias) of the measuring instrument: C g k = 0.10 ⋅ T − ∣ x ˉ g − x ref ∣ 3 ⋅ s g C_{gk} = \frac{0.10 \cdot T - |\bar{x}_g - x_{\text{ref}}|}{3 \cdot s_g} C g k = 3 ⋅ s g 0.10 ⋅ T − ∣ x ˉ g − x ref ∣
where:
0.10 ⋅ T 0.10 \cdot T 0.10 ⋅ T represents half of the allowable gage variation ( 10 % 10\% 10% of product tolerance allocated to each side).
x ˉ g \bar{x}_g x ˉ g is the arithmetic mean of the repeated gage measurements: x ˉ g = 1 n ∑ i = 1 n x g , i \bar{x}_g = \frac{1}{n}\sum_{i=1}^n x_{g,i} x ˉ g = n 1 ∑ i = 1 n x g , i .
x ref x_{\text{ref}} x ref is the accepted calibrated value of the reference standard.
∣ x ˉ g − x ref ∣ |\bar{x}_g - x_{\text{ref}}| ∣ x ˉ g − x ref ∣ represents the absolute systematic bias of the gage.
3 ⋅ s g 3 \cdot s_g 3 ⋅ s g represents the half-width ( 3 σ 3\sigma 3 σ ) of the gage repeatability spread.
Acceptance Criteria
If the applicable study plan adopts a 20% tolerance allocation and 1.33 thresholds, interpret the results as follows; these are study-policy choices, not universal requirements:
C g ≥ 1.33 C_g \ge 1.33 C g ≥ 1.33 and C g k ≥ 1.33 C_{gk} \ge 1.33 C g k ≥ 1.33 : The gage is capable and approved for production inspection.
C g < 1.33 C_g < 1.33 C g < 1.33 : The gage has excessive random variation (inadequate repeatability or resolution).
C g ≥ 1.33 C_g \ge 1.33 C g ≥ 1.33 but C g k < 1.33 C_{gk} < 1.33 C g k < 1.33 : The gage has excellent repeatability but possesses a significant uncorrected systematic bias (offset), requiring zero adjustment, recalibration, or mechanical re-alignment.
Common Calibration Traps & CCT Exam Pitfalls
Warning
Trap 1: Confusing TAR with TUR in Specification Review Never substitute catalog accuracy for process uncertainty. If an auditor asks for a TUR calculation and you compute the ratio using the standard's catalog accuracy tolerance, you have calculated TAR. A true TUR calculation must incorporate the standard's expanded uncertainty, environmental gradients, UUT resolution, and operator repeatability.
Caution
Trap 2: Attempting to "Correct" Measurement Uncertainty A frequent question on the ASQ CCT exam asks how a technician can correct for measurement uncertainty. The answer: you cannot. Systematic error can be corrected by applying an offset; random error can be reduced by averaging; but measurement uncertainty is a measure of incomplete knowledge that can only be evaluated and reported.
Tip
Trap 3: Forgetting UUT Resolution in TUR Calculations Evaluate relevant UUT resolution in the process model. A rectangular rounding contribution u r e s = d / 12 u_{res}=d/\sqrt{12} u res = d / 12 can be justified for a step d d d , but is not mandatory for every observation process. Avoid double counting resolution effects already represented in repeatability and use the uncertainty of the reported result.