15.2 Electrotherapy & Therapeutic Ultrasound

BOC Certification Exam for Athletic Trainers ATC

Track your progress as you read

Finished sections get a check mark.

1 Introduction & Exam Overview

1.1 Current BOC ATC Exam Facts 1.2 Item Types, Clinical Reasoning & Study Planning

2 Screening, PPE & Risk Identification

2.1 Pre-Participation Physical Examination 2.2 Injury Surveillance, Epidemiology & Intrinsic/Extrinsic Risk 2.3 Protective Equipment Fitting & Safety Standards

3 Prevention Programs, Facility Safety & Health Literacy

3.1 Evidence-Based Injury Prevention Programs 3.2 Facility, Playing Surface & OSHA Safety 3.3 Health Literacy, Communication & Patient Advocacy

4 Environment & Wellness

4.1 Heat Illness Prevention & WBGT Monitoring 4.2 Lightning, Cold, Altitude & Air Quality 4.3 Nutrition, Hydration & Sleep 4.4 Wellness Promotion & Mental Health

5 History, Exam Framework & Neurological Screen

5.1 Patient History & Interview Techniques 5.2 Physical Examination Framework & Diagnostic Accuracy 5.3 Neurological Screening & Cranial Nerve Assessment

6 Upper Extremity & Cervical/Thoracic Spine Assessment

6.1 Shoulder Complex Assessment 6.2 Elbow, Wrist & Hand Assessment 6.3 Cervical & Thoracic Spine Assessment

7 Lumbar Spine, Hip & Knee Assessment

7.1 Lumbar Spine Assessment 7.2 Hip, Groin & Pelvis Assessment 7.3 Knee Joint Assessment

8 Foot/Ankle, Imaging, General Medical & Diagnosis

8.1 Foot & Ankle Assessment 8.2 Diagnostic Imaging & Clinical Decision Rules 8.3 Cardiovascular, Pulmonary, Abdominal & Dermatologic Assessment 8.4 Clinical Diagnosis, Differential Diagnosis, Plan of Care & Patient Education

9 Emergency Action Plans, Scene Assessment & Triage

9.1 Emergency Action Plans 9.2 Scene Assessment & Safety 9.3 Triage & Primary/Secondary Survey

10 Cardiac, Airway & Medical Collapse

10.1 Sudden Cardiac Arrest, High-Performance CPR & AED 10.2 Airway & Breathing Management 10.3 Medical Collapse: Anaphylaxis, Asthma, Diabetes, Sickle Cell & ECAST

11 Spine, Hemorrhage & Head Trauma

11.1 Cervical Spine Trauma, Equipment Removal & Spine Boarding 11.2 Hemorrhage Control, Shock, Fractures & Dislocations 11.3 Acute Head Trauma & Concussion Sideline Triage

12 Environmental Catastrophe & Mental Health Emergencies

12.1 Exertional Heat Stroke, Hyponatremia & Cold Injury Emergencies 12.2 Mental Health Emergencies, Suicide Risk & Transfer of Care

13 Rehab Design & Therapeutic Exercise

13.1 Tissue Healing Stages & Rehabilitation Plan of Care 13.2 Patient Education, Adherence & Shared Decision-Making 13.3 Therapeutic Exercise: ROM, Flexibility & Strength

14 Neuromuscular Training, Functional Testing & Return-to-Play

14.1 Neuromuscular Control, Plyometrics & Sport-Specific Training 14.2 Functional Testing & Return-to-Play Criteria 14.3 Concussion Return-to-Learn & Return-to-Sport

15 Therapeutic Modalities

15.1 Cryotherapy & Thermotherapy 15.2 Electrotherapy & Therapeutic Ultrasound 15.3 Mechanical Modalities (Traction, Compression, CPM)

16 Manual Therapy, Pharmacology & Medical Management

16.1 Manual Therapy Techniques 16.2 Taping, Bracing & Orthoses 16.3 Pharmacological Principles & Common Medications 16.4 Managing General Medical Conditions During Activity

17 Documentation, Privacy, Policy & Quality

17.1 Clinical Documentation & Medical Records 17.2 HIPAA, FERPA & Privacy 17.3 Policies, Quality Improvement & Budgeting

18 Legal, Ethics & Professional Practice

18.1 State Practice Acts, Scope of Practice & Legal Risk 18.2 BOC Standards, Ethics, EBP & Interprofessional Care

15.2 Electrotherapy & Therapeutic Ultrasound

Key Takeaways Explain This Topic AI

Conventional TENS is high-rate, short-pulse, sensory, gate-control analgesia during treatment. Low-rate TENS is about 1–10 Hz with a motor twitch, endogenous-opioid analgesia, and longer carryover.

Dexamethasone sodium phosphate is negatively charged—place it under the cathode. NMES/Russian current is for re-education and strengthening, with a rest-heavy on:off ratio to limit fatigue.

Ultrasound: 1 MHz for deeper targets (~2–5 cm), 3 MHz for superficial (~1–2 cm). Continuous (100%) duty cycle is thermal; pulsed (often 20%) is nonthermal. Know BNR, ERA, coupling, and a moving soundhead.

Do not electrically stimulate a demand pacemaker, the carotid sinus, the pregnant abdomen/low back, DVT, or malignancy. Ultrasound also avoids eyes, heart, exposed CNS after laminectomy, and thermal dose over growth plates.

Trap: stationary 1 MHz at high intensity over a bony prominence, or TENS on a patient with a demand pacemaker. Phonophoresis and LLLT evidence for many musculoskeletal conditions is limited or mixed.

Last updated: August 2026

Quick Answer: Conventional TENS (high rate, sensory) uses gate control . Low-rate TENS (about 1–10 Hz , motor twitch) uses endogenous opioids . Dexamethasone is negative —put it on the cathode . 1 MHz ultrasound is deeper ( 2–5 cm ); 3 MHz is superficial ( 1–2 cm ). Trap: TENS on a demand pacemaker, or a stationary high-intensity 1 MHz head on bone.

PA8 task 0404 includes electrical stimulating currents and acoustic energy. The exam tests parameters you can defend and contraindications you never skip . It does not require you to claim that ultrasound heals every tendon.

TENS: two mechanisms, two recipes

Transcutaneous electrical nerve stimulation (TENS) is for pain , not for a 10-repetition-maximum quad set.

Gate control (Melzack and Wall, as commonly taught): large-diameter A-beta (touch/pressure) input closes the substantia gelatinosa 'gate' and reduces nociceptive traffic to the brain. That is conventional / high-rate / sensory TENS :

  • Parameter Conventional (high-rate) TENS Low-rate (acupuncture-like) TENS
  • Frequency About 80–150 pps (many units 50–150) About 1–10 pps (often 2–4)
  • Pulse duration Short, about 50–80 μs Longer, about 150–300 μs
  • Amplitude Strong sensory tingling; no muscle twitch Visible motor twitch (somewhat noxious)
  • Typical time 20–30 minutes, can be longer/as needed About 20–45 minutes, not all day
  • Proposed mechanism Segmental gate (A-beta) Endogenous opioids (endorphins/enkephalins)
  • Carryover Mostly during stimulation Often hours after
  • Avoid if — Muscle contraction is unsafe (unstable fracture, acute tear you are protecting)

Burst TENS packages low-rate-like bursts into a more comfortable waveform; think opioid-style effects with less sting. Electrode placement: around the painful site, along the dermatome, or at related trigger/acupuncture points. Bipolar is common; interferential current (IFC) is the 'bigger, deeper TENS' cousin.

IFC: two medium-frequency alternating currents (classically ~4000 Hz and ~4100 Hz) interfere in the tissue. The beat frequency is the difference (1–150 Hz) and is what you dose like a TENS frequency. Medium frequency meets less skin impedance , so athletes often tolerate it over a larger area (low back, thigh). Quadripolar setup produces a cloverleaf field; vector scan can 'stir' the field. Use IFC for pain (and, in some protocols, edema) with the same systemic contraindications as TENS.

NMES and Russian current: make muscle fire

Neuromuscular electrical stimulation (NMES) is for re-education and strengthening when the athlete cannot volitionally recruit (arthrogenic inhibition after ACL reconstruction or TKA is the classic AT example). You need a tetanic contraction, not a sensory tingle.

Commonly taught NMES parameters:

  • Frequency about 35–80 Hz (50 Hz is a workhorse) for tetany
  • Pulse duration about 200–400 μs for large muscles

On:off about 1:5 for strength (for example 10 s on / 50 s off ) so the muscle can recover; shorter rest if the goal is endurance or a pump, accepting more fatigue

  • Ramp 1–2 s so the contraction is not a slap
  • Amplitude to a strong, tolerated contraction in the target muscle

Russian current (Kots, as taught): 2500 Hz carrier burst-modulated at 50 bursts/s , classically 10/50/10 (10 s contraction, 50 s rest, 10 minutes). Use it when the unit offers it for quad re-ed . It is still electricity across the chest and neck—pacemaker and carotid rules still apply.

Place electrodes over the motor point and along the muscle belly, not over the carotid sinus, not across the heart, and not on broken skin. Combine NMES with voluntary effort when the athlete can—the goal is a brain-muscle loop, not a passive twitch show.

Iontophoresis: polarity is the exam item

Iontophoresis uses direct current (DC) to repel a charged drug through the skin. Like charges repel. The electrode that holds the drug is the active electrode.

Dexamethasone sodium phosphate is negatively charged at physiologic pH. Place it under the cathode (−). Delivery from the anode is the historical error (Glass mixed competing ions and used the anode); modern teaching and in-vitro data (Kalia and colleagues) favor the cathode . Lidocaine is positive (anode). Acetic acid is negative (calcium deposits, as taught).

Dosage is mA × minutes . Commonly taught target: about 40–80 mA·min (example: 4 mA × 20 min = 80). Current is usually 1–4 mA , limited by electrode size ( current density ) and sting. DC causes electrochemical burns if you over-dose, leave a dry electrode, or treat broken skin. Do not mix dexamethasone and lidocaine in the same electrode 'to cancel charges'—you add competing ions and wreck delivery.

Therapeutic ultrasound: frequency, duty cycle, BNR, ERA

Ultrasound is high-frequency sound (in AT units, 1 MHz or 3 MHz ). The crystal vibrates; the beam is absorbed more in protein-rich tissue (tendon, ligament, muscle) than in fat, and bone and nerve absorb still more —which is why a stationary head over the malleolus is a periosteal-pain machine.

  • Parameter Commonly taught meaning Clinical choice
  • 1 MHz Deeper: about 2–5 cm Hip, deep rotator cuff, lumbar paravertebrals
  • 3 MHz Superficial: about 1–2.5 cm ; heats about 3× faster (Draper) Patellar tendon, epicondyle, MCL
  • Duty cycle 100% (continuous) Thermal Stiffness, before stretch; not for acute hot tissue

Duty cycle ~20% (pulsed) Nonthermal (cavitation, acoustic streaming); little net heat Acute/irritable tissue if you use US at all

Intensity Often 0.5–2.0 W/cm² SATP Thermal doses commonly ~1.0–1.5 W/cm²; patient report of deep ache = too hot

ERA Effective radiating area (crystal that actually emits; usually smaller than the faceplate) Treat about 2× ERA , not the whole thigh with a 5-cm² head

BNR Beam nonuniformity ratio = spatial peak / average Typical units 2:1–6:1 ; higher BNR = hotter spots. FDA-labeled; a perfect 1:1 beam does not exist.

Draper/Lehmann-style heating-rate teaching (textbook-conventional, not a BOC-stamped law): at 1 MHz , about 0.2°C/min per 1.0 W/cm² ; at 3 MHz , about 0.6°C/min per 1.0 W/cm² . Thermal goals often taught: +1°C metabolism, +2–3°C blood flow/spasm, +4°C collagen extensibility.

Coupling is mandatory. Air reflects almost all of the beam. Use gel, or underwater treatment about 1 cm from irregular bony contours (wipe bubbles). Move the soundhead in overlapping circles at about 4 cm/s . A stationary head, especially 1 MHz, high intensity, over bone , creates a standing-wave hot spot .

Precautions / do-not-treat fields: eyes , heart , pregnant uterus , malignancy , DVT , pacemaker/electronics in the field, exposed CNS after laminectomy , acute fracture (thermal US; low-intensity pulsed US for fractures is a different, device-specific prescription), and thermal dose over open growth plates (precaution—do not cook an epiphysis). Cemented or plastic implants can overheat with thermal US—know what is in the limb.

Evidence honesty: therapeutic ultrasound parameters are still taught because units are everywhere and the physics is testable. For many musculoskeletal conditions, clinical-outcome evidence is mixed or modest (Cochrane and subsequent reviews). Do not claim US 'heals' a chronic rotator cuff. Phonophoresis (US plus hydrocortisone in the gel) has limited evidence that meaningful drug reaches the target—do not treat it as a steroid injection. Low-level laser therapy (LLLT) evidence for common AT diagnoses is mixed ; if you use it, protect the eyes and do not claim FDA-untrue cures. Shortwave diathermy heats large volumes (27.12 MHz is the classic ISM frequency); metal, pregnancy, and pacemakers are hard stops. Many AT clinics no longer own a diathermy unit; know the contraindications anyway.

Shared electrical and US contraindications, and the traps

Do not electrically stimulate:

  • Demand pacemaker / implanted defibrillator (and generally implanted stimulators in the field)
  • Carotid sinus (dysrhythmia, hypotension)
  • Pregnancy over the abdomen or low back
  • DVT / thrombophlebitis
  • Malignancy in the field

Active hemorrhage, over the eyes, through the chest in unstable cardiac disease, and over broken skin for DC (iontophoresis)

Exam traps: (1) TENS on a patient with a demand pacemaker —sensory intensity does not make it safe. (2) Stationary 1 MHz at high intensity over a bony prominence. (3) Putting dexamethasone on the anode. (4) Treating phonophoresis as proven transdermal steroid delivery.

Loading diagram... Electrotherapy and ultrasound parameter gates