16.3 Artifacts, Markers & Annotation

CAMRT National Certification Exam — Radiological Technology Exam

Track your progress as you read

Finished sections get a check mark.

1 Introduction & Exam Overview

1.1 Current Exam Facts, Format & Scoring 1.2 Competency-Based Study Strategy & How to Use This Guide

2 Professional, Communicator & Collaborator Roles

2.1 Legal Requirements, Privacy & Informed Consent 2.2 Ethics, Integrity, Boundaries & Cultural Safety 2.3 Communication, Documentation & Collaboration

3 Care Provider: Safety & Assessment

3.1 Patient Identity, Transfer & Immobilization 3.2 History Review, Contraindications & Monitoring 3.3 Medical Emergencies & Patient Interventions

4 Care Provider: Infection Control & Education

4.1 Infection Prevention, Control & Sharps 4.2 Patient Education & Support

5 Radiation Safety (RTR.1)

5.1 ALARA, Time, Distance & Shielding 5.2 Radiation Effects, Risks & Dose Quantities 5.3 Personal Dosimetry & Protective Apparel 5.4 Facility Emergencies & Safe Work Environment

6 Radiation Physics & Imaging Systems (RTR.2)

6.1 X-ray Production & Interactions with Matter 6.2 Operating Imaging Systems & Accessory Equipment 6.3 Grids, Collimation, AEC & Exposure Control 6.4 QC Programs, PACS & Reject/Repeat Analysis

7 Clinical Principles (RTR.3)

7.1 Relational Anatomy & Physiology for Imaging 7.2 Pathophysiology Essentials for Radiographers 7.3 Imaging Protocols, Environments & Other Modalities

8 Positioning Fundamentals & Chest Radiography

8.1 Positioning Terminology, Landmarks & Alignment 8.2 Chest & Thorax Radiography 8.3 Ribs, Sternum & Soft-Tissue Neck

9 Upper Extremity Radiography

9.1 Hand, Wrist & Forearm 9.2 Elbow & Humerus 9.3 Shoulder Girdle

10 Lower Extremity & Pelvis Radiography

10.1 Foot, Toes & Ankle 10.2 Leg, Knee & Patella 10.3 Femur & Hip 10.4 Pelvis & SI Joints

11 Spine & Skull Radiography

11.1 Cervical Spine 11.2 Thoracic & Lumbar Spine, Sacrum & Coccyx 11.3 Skull, Facial Bones, Sinuses & Mandible

12 Abdomen & Contrast Studies

12.1 Abdomen & Acute Abdominal Series 12.2 Gastrointestinal Contrast Studies 12.3 Urinary System Contrast Studies

13 Special Procedures & Modalities

13.1 Mobile, Operating Room & Trauma Radiography 13.2 Fluoroscopy & Interventional Overview 13.3 CT, BMD & Mammography Overview

14 Contrast Media & Pharmaceuticals (RTR.5)

14.1 Venipuncture & Pharmaceutical Agents 14.2 Contrast Media: Types, Effects & Routes 14.3 Adverse Reactions & Emergency Response

15 Image Quality Fundamentals (RTR.6)

15.1 Brightness, Density & Receptor Exposure 15.2 Radiographic Contrast 15.3 Spatial Resolution & Distortion 15.4 Exposure Factors & Technical Evaluation

16 Image Critique & Pathology Recognition

16.1 Anatomy Recognition & Normal Variants 16.2 High-Weight Pathologies: Respiratory & Skeletal 16.3 Artifacts, Markers & Annotation 16.4 Repeat Decisions & Post-Processing

16.3 Artifacts, Markers & Annotation

Key Takeaways Explain This Topic AI

RTR.6.5 requires accurate laterality markers and annotations; physical radiopaque markers remain best practice, and digital labels can be wrong, flipped, or copied forward.

Never release an image when side identification is uncertain—wrong-side documentation is a critical safety event.

RTR.6.6 artifacts include motion, grid cutoff, clothing/foreign objects, CR/DR processing defects, double exposure, scatter fog, moiré, and detector hardware faults.

Correct the root cause (geometry, collimation, gowning, detector hygiene, exposure time) rather than only windowing over the problem.

If an artifact obscures the region of interest or laterality is compromised, reject/correct under ALARA and escalate equipment failures through QC/service pathways.

Last updated: July 2026

16.3 Artifacts, Markers & Annotation

Quick Answer: Correct laterality markers and accurate annotation are patient-safety requirements (RTR.6.5). Artifacts (RTR.6.6)—motion, grid cutoff, foreign objects/clothing, CR/DR processing problems, double exposure, scatter fog—must be recognized, explained, and corrected . Digital tools make annotation errors easier, not rarer. If an artifact obscures the clinical region or laterality is uncertain, do not release until fixed or properly escalated.

Image analysis is incomplete if you only judge mAs and rotation. Legal identification and artifactual densities decide whether an image is usable and whether the wrong side of a patient could be treated.

  • Markers & Annotation Accuracy (RTR.6.5)
  • Physical lead markers remain the gold standard habit

Best practice in Canadian departments still emphasizes radiopaque side markers placed in the primary beam (outside essential anatomy when possible) at exposure:

  • Rule Why it matters
  • Correct side (R / L) Wrong-side surgery and treatment risk
  • Correct patient contact side Marker on the side being imaged, not the opposite cassette edge by habit alone
  • Not over critical anatomy Marker over fracture line or lung apex can force a repeat
  • Visible after collimation Tight collimation that clips the marker creates incomplete legal image

Supplemental markers as protocol Erect/supine, weight-bearing, inspiration/expiration, minutes post-contrast, mobile

Digital annotation risks

Electronic annotation is useful for add-on labels (portable, upright, post-void) but creates unique failure modes:

  • Post-exposure side labels applied to the wrong side after the fact
  • Copy-forward annotations from a prior study on the wrong patient
  • Image flip/rotate after annotation so the “R” no longer matches anatomy
  • Missing annotation on stitched or reformatted secondaries
  • Over-reliance on default laterality from the worklist without visual confirmation

Exam principle: A digital “R” is not automatically trustworthy if the physical marker is absent and anatomy is ambiguous (e.g., AP supine abdomen, unilateral extremity without landmarks). When laterality is uncertain, investigate before release .

  • Annotation content that supports care
  • Projection (PA/AP/lateral/oblique) when not obvious
  • Patient position (erect, semi-erect, supine, decubitus with side down)
  • Behavioral/technique notes required by protocol (inspiration/expiration)
  • Time-sensitive series labels (pre/post void, post-reduction)

Avoid cluttering the image with redundant text that covers pathology.

Artifacts (RTR.6.6): Recognition Table & Corrective Response

Artifact Appearance / cues Common causes Corrective response

Motion Blur, ghosting of edges, repeated contours Long exposure, patient pain/tremor, kids, respiration, equipment vibration Shorten exposure time (↑ mA), immobilize, coach breath-hold, supports; pediatric distraction; check tube/stand locks

Grid cutoff (lateral decentering) Overall or unilateral density loss; underexposed look across field CR angled across lead strips; grid upside-down; off-level; wrong focal range SID Re-center, match SID to grid focal distance, verify grid orientation, use appropriate ratio

Grid cutoff (upside-down focused grid) Severe density loss laterally both sides or characteristic pattern Focused grid inverted Flip grid to correct orientation; repeat

Grid lines Fine parallel lines Stationary grid visible; wrong grid use; moving grid failure Use moving Bucky when indicated; check mechanism; accept only if lines do not hide ROI and protocol allows

Clothing / buttons / zippers Opaque shapes, fabrics, sparkles Incomplete gowning Remove radiopaque objects from FOV; re-gown; explain to patient

Jewelry / hair pins / ECG leads High-density discrete objects Not removed/relocated Remove or reposition leads outside ROI when safe

Foreign objects on IR/detector Dust, cracks, scratches, glue Dirty plate, damaged detector Clean per manufacturer; take QC flat field; remove from service if hardware defect

Double exposure Superimposed anatomies, dark/overexposed combined image Cassette reused (CR); wrong detector assignment; procedural error Reject; QC process; re-image once with clear detector workflow

Scatter fog Reduced contrast, gray veil Poor collimation, no grid when needed, high kVp large field, open back of cassette (film era concepts still tested), patient scatter Collimate, grid when indicated, optimize kVp, control field size

CR plate reader artifacts Lines, skip, plate erasure incompleteness Reader rollers, incomplete erase, light leak Reprocess/erase plate; service if persistent; do not keep sending bad plates

DR processing / flat-field issues Coring, dropout, structured noise, dead pixels clusters Calibration overdue, dropped detector, software LUT errors Recalibrate; service; alternate detector; do not “fix” hardware failure with edge enhance

Moire / aliasing Wavy interference pattern Grid frequency vs sampling; electronic interference Adjust grid/processing; change orientation if protocol allows; service evaluation

Saturation / clipping Pure white/black loss of data in regions Extreme overexposure or processing Cannot recover lost data; repeat with corrected technique

  • Quantum mottle Grainy noise Underexposure Repeat with increased exposure as needed—not just windowing
  • Pajama folds / skin folds Linear densities Soft tissue Reposition, smooth skin, note portable limitations

Cast / splint / backboard Uniform or structured opacity Trauma immobilization Increase exposure appropriately; annotate; do not remove unstable immobilization without clinical order

Motion Artifact — Special Emphasis

Motion is still one of the most common avoidable repeats.

Types:

  • Voluntary — breathing, movement (coaching, supports, clear instructions)
  • Involuntary — tremor, spasm, cardiac (minimize time, stabilize)
  • Equipment — unlocked tube arm, moving table, grid mechanism

Pediatric tip: immobilization devices and caregiver assistance per policy beat multiple failed “please hold still” exposures. ALARA prefers one good immobilized image over three blurred ones.

Grid Cutoff Patterns Worth Memorizing

  • Error Typical density pattern
  • Off-level grid Overall decrease in exposure across image
  • Lateral decentering More density loss toward one side
  • Upside-down focused grid Severe bilateral lateral underexposure
  • Wrong SID outside focal range Density loss toward periphery

Do not compensate solely by raising mAs without fixing geometry—you increase dose and may still fail.

Foreign Objects & Clothing: Workflow Discipline

Build a pre-exposure checklist into muscle memory:

  • Gown / remove artifacts from region
  • Empty pockets
  • Move oxygen tubing, NG slack, monitor cables when safe
  • Explain why metal must move (“it can hide a fracture”)
  • Document non-removable objects (jewelry that cannot come off, permanent medical devices)

Piercings and religious items: use culturally safe communication; if item cannot be removed, annotate and collimate/angle when possible without losing required anatomy.

Double Exposure & Workflow Mix-ups

Modern DR reduces classic double exposure but it still appears in stems and in CR-heavy sites:

  • Two body parts on one image
  • “Ghost” of previous anatomy
  • Unexpected darkness

Response: reject image, secure correct detector/plate, re-identify patient, single controlled re-exposure, report system issues if persistent.

Scatter Fog vs Pathology Veil

A gray, low-contrast image may be:

  • Scatter fog (collimation/grid problem)
  • Correct technique on a large soft-tissue patient
  • True pathologic veil (edema, effusion layering)

Fix what you control first (collimation, grid, kVp, field size), reassess, then consider pathology recognition (16.2).

  • Corrective Response Hierarchy
  • Identify artifact type (do not guess randomly)

Decide impact: Does it obscure ROI or create wrong-side risk?

  • Correct cause (not only symptoms)
  • Repeat or add image under ALARA if needed (16.4)
  • Escalate equipment problems to service/QC
  • Document when non-removable artifacts remain
  • CAMRT-Style Stems

Expect items such as:

“Density loss on both lateral margins after using a focused grid—most likely?” → upside-down grid

“Electronic L marker placed after exposure on a hand without physical marker—risk?” → wrong laterality annotation

  • “Wavy pattern over lumbar spine with stationary grid—what is it?” → moiré/grid-line interference
  • “Blurred trabeculae with correct EI—cause?” → motion, not mAs

Master tables above until recognition is automatic under exam timing.