Study guide
4.7 Pacemaker Rhythmias, Spike Analysis & Pacemaker Malfunctions
CCI Certified Cardiographic Technician Exam
Track your progress as you read
Finished sections get a check mark.
1 Chapter 1: Cardiac Anatomy, Electrophysiology & Pre-Procedural Patient Care
1.1 Cardiac Anatomy & Blood Flow Pathway 1.2 Cardiac Electrophysiology & Conduction System 1.3 Pre-Procedural Patient Care, Identification & Infection Control 1.4 Cardiovascular Pharmacology & Cardiac Medications
2 Chapter 2: 12-Lead ECG Technique, Lead Placement & Technical Quality
2.1 Bipolar & Augmented Limb Leads (Einthoven's Triangle) 2.2 Precordial Lead Placement (V1-V6) & Special Lead Configurations (V7-V9, Right-Sided, Dextrocardia, Pediatric, 15-Lead) 2.3 ECG Paper Calibration, Grid Standards & Speed Settings 2.4 ECG Artifact Recognition, Signal Interference & Technical Troubleshooting
3 Chapter 3: Analyzing the 12-Lead ECG (Axis, Hypertrophy & Ischemia)
3.1 Cardiac Axis Determination & Axis Deviations 3.2 Chamber Hypertrophy (Atrial & Ventricular) 3.3 Ischemic Changes, Injury & Infarction Patterns
4 Chapter 4: ECG Waveform Analysis & Rhythm Recognition
4.1 Normal Sinus Rhythm & Sinus Node Dysrhythmias 4.2 Atrial Dysrhythmias (PACs, Atrial Fibrillation, Atrial Flutter, SVT) 4.3 Junctional Rhythmias & AV Nodal Reentry 4.4 Ventricular Dysrhythmias (PVCs, Ventricular Tachycardia, Ventricular Fibrillation, Torsades de Pointes) 4.5 Atrioventricular (AV) Blocks (1st Degree, 2nd Degree Type I/II, 3rd Degree Complete) 4.6 Intraventricular Conduction Delay, Bundle Branch Blocks & WPW Pattern 4.7 Pacemaker Rhythmias, Spike Analysis & Pacemaker Malfunctions
5 Chapter 5: Ambulatory ECG Monitoring (Holter, Event & Telemetry)
5.1 Holter Monitoring (24-48 Hour Continuous Recording) & Lead Placement 5.2 Cardiac Event Monitors, Mobile Cardiac Telemetry (MCT), Patch Recorders & Implantable Loop Recorders 5.3 Patient Diary Correlation, Symptom Tracking & Data Analysis
6 Chapter 6: Exercise & Pharmacologic Cardiac Stress Testing
6.1 Treadmill Stress Test Protocols 6.2 Target Heart Rate Calculation, Hemodynamic Monitoring & Test Termination Criteria 6.3 Pharmacologic Stress Testing Agents 6.4 Complications & Emergency Protocols During Stress Testing
7 Chapter 7: Equipment Quality Control, Calibration & Compliance Standards
7.1 Equipment Calibration, Voltage Standards & Frequency Filters 7.2 Electrical Safety, Patient Grounding & Equipment Maintenance 7.3 Quality Assurance, Documentation & HIPAA/Regulatory Compliance
4.7 Pacemaker Rhythmias, Spike Analysis & Pacemaker Malfunctions
Key Takeaways Explain This Topic AI
The NBG Pacemaker Code uses positions I-III to indicate the chamber paced, chamber sensed, and the response to sensing (I=Inhibits, T=Triggers, D=Dual).
An atrial pacing spike should be immediately followed by a P wave, indicating atrial capture.
A ventricular pacing spike should be immediately followed by a wide QRS complex, indicating ventricular capture.
Failure to pace is recognized by the absence of expected pacing spikes when the intrinsic heart rate falls below the programmed rate.
Failure to capture occurs when a pacing spike is present but is not followed by the appropriate cardiac depolarization (P wave or QRS).
Last updated: July 2026
Pacemaker Rhythmias and Analysis
Artificial pacemakers are devices implanted to deliver electrical impulses to the heart when the intrinsic conduction system is too slow or fails (e.g., symptomatic bradycardia, Mobitz II, third-degree heart block). Evaluating pacemaker function on an ECG is a core competency, as identifying malfunctions rapidly can prevent severe patient decompensation.
The NBG Pacemaker Code
Pacemaker functions are described using a standardized 3- to 5-letter code developed by the North American Society of Pacing and Electrophysiology (NASPE) and the British Pacing and Electrophysiology Group (BPEG) (NBG code). The first three positions are the most critical for ECG interpretation:
Comparison Table: NBG 5-Letter Pacemaker Code
Position I: Chamber Paced Position II: Chamber Sensed Position III: Response to Sensing Position IV: Rate Modulation Position V: Multisite Pacing
O = None O = None O = None O = None O = None
A = Atrium A = Atrium I = Inhibited (Pacing stops if beat sensed) R = Rate Modulation (Adjusts HR to activity) A = Atrium
- V = Ventricle V = Ventricle T = Triggered (Sensing triggers a paced beat) V = Ventricle
- D = Dual (A + V) D = Dual (A + V) D = Dual (T + I) D = Dual (A + V)
Example: VVI - Paces the Ventricle, Senses the Ventricle, and Inhibits pacing if an intrinsic ventricular beat is sensed. The classic demand pacemaker.
Example: DDD - Paces both Atrium and Ventricle, Senses both, and can Trigger or Inhibit based on intrinsic activity. Closely mimics physiologic pacing.
Normal Pacemaker Function on ECG
Pacemaker activity is visible on the ECG as thin, vertical, straight lines called pacer spikes . The location of the spike relative to the cardiac cycle indicates which chamber is being paced.
Atrial Pacing: A pacing spike appears just before the P wave. The resulting P wave may look slightly different from a native sinus P wave.
Ventricular Pacing: A pacing spike appears just before the QRS complex. Because the impulse originates in the ventricular muscle (usually the right ventricular apex) rather than the normal conduction system, the resulting QRS complex is wide (≥ 0.12s) and often resembles a Left Bundle Branch Block pattern.
Dual-Chamber Pacing (A-V Sequential): Two spikes are seen in one cardiac cycle. The first spike is followed by a P wave (atrial capture), and the second spike, occurring after a programmed PR interval, is followed by a wide QRS complex (ventricular capture).
Pacemaker Malfunctions
Identifying pacemaker failure is critical. There are three primary types of malfunction identifiable on an ECG. Each carries specific clinical implications and requires distinct troubleshooting approaches.
Comparison Table: Pacemaker Malfunctions
Malfunction ECG Finding Common Cause Clinical Impact Intervention
Failure to Pace (Fire) Complete absence of pacing spikes when HR drops below programmed set rate. Dead battery, broken lead wire, over-sensing. Bradycardia, syncope, or asystole. Apply magnet, check connections, pacing.
Failure to Capture Pacing spike is visible but not followed by P wave or QRS complex. Lead dislodgment, low voltage output, electrolyte imbalance, ischemia. Decreased cardiac output. Pacemaker fires but heart does not respond. Increase output voltage, correct electrolytes.
Failure to Sense (Under-sensing) Pacing spikes occur inappropriately, falling randomly on intrinsic beats (P waves, QRS, or T waves). Lead dislodgment, sensitivity threshold set too low, fibrous tissue on lead. Competes with intrinsic rhythm. R-on-T phenomenon risk , triggering VF. Increase sensitivity setting.
1. Failure to Pace (Failure to Fire)
ECG Finding: Complete absence of pacing spikes where they should be expected based on the programmed rate.
Result: The heart rate drops below the pacemaker's set rate, leading to bradycardia or asystole if no intrinsic escape rhythm emerges.
Causes: Dead battery, broken lead wire, over-sensing (the pacemaker inappropriately senses noise and inhibits pacing).
2. Failure to Capture
ECG Finding: A pacing spike is clearly visible, but it is not followed by the appropriate waveform (no P wave after an atrial spike, or no QRS complex after a ventricular spike).
Result: The pacemaker fires, but the electrical energy is insufficient to depolarize the myocardium.
Causes: Lead dislodgment, low voltage output, increased myocardial stimulation threshold (e.g., due to ischemia or electrolyte imbalance).
3. Failure to Sense (Under-sensing)
ECG Finding: Pacing spikes occur inappropriately, firing regardless of the patient's intrinsic electrical activity. You will see pacing spikes falling on top of native P waves, QRS complexes, or dangerously, on T waves (R-on-T phenomenon).
Result: The pacemaker fails to "see" the patient's own heartbeat and competes with the intrinsic rhythm. This is particularly dangerous if a ventricular pacing spike lands on a T wave.
Causes: Lead dislodgment, sensitivity set too low, fibrous tissue around the lead tip.