13.4 Pacing & Sensing Malfunctions, Electromagnetic Interference & Temporary Pacing

Registered Cardiac Electrophysiology Specialist RCES

Track your progress as you read

Finished sections get a check mark.

1 Chapter 1: Pre-Procedural Assessment & Patient Preparation

1.1 Procedure Room Preparation, Equipment Verification & Aseptic Readiness 1.2 Pre-Procedural Clinical Evaluation, Labs, Consents & The Universal Time-Out 1.3 Patient Identification, Transport, HIPAA & Procedural Documentation 1.4 Patient Monitoring Setup, Defibrillator Pads & Pre-Procedural CIED Interrogation 1.5 Vascular Access Planning, Skin Preparation & Sterile Draping

2 Chapter 2: Cardiac Anatomy, Conduction & Electrophysiologic Foundations

2.1 Cardiac Anatomy for the EP Lab: Chambers, Access Routes & Coronary Supply 2.2 The Cardiac Conduction System, Action Potentials & Ion Channel Physiology 2.3 Cardiac Vector Analysis, 12-Lead ECG Fundamentals & Axis Determination

3 Chapter 3: EP Lab Instrumentation, Connectology & Lab Safety

3.1 EP Recording System Architecture, Signal Amplification & Frequency Filters 3.2 Cardiac Stimulator Operation, Junction Box Routing & Noise Troubleshooting 3.3 Electrical Safety, Electrical Units, Electrocautery & Fire Safety 3.4 Radiation Safety Physics, ALARA Principles, Shielding & Dose Limits 3.5 Procedural Supplies, Catheter Inventory, Surgical Instrumentation & Wound Closure

4 Chapter 4: Intra-Procedural Imaging, Sedation & Physiologic Monitoring

4.1 Fluoroscopic Projections (RAO, LAO, AP) & X-ray Tube Optimization 4.2 Intracardiac Echocardiography (ICE) Principles, Views & Catheter Navigation 4.3 Non-Radiographic Imaging: Cardiac CT/MRI, Image Fusion & Ultrasound-Guided Access 4.4 Conscious Sedation Protocols, Airway Management & Hemodynamic Monitoring 4.5 Vital Signs, Normal Laboratory Values & Intra-Procedural Lab Monitoring

5 Chapter 5: EP Pharmacology, Anticoagulation & Emergency Drugs

5.1 Vaughan Williams Antiarrhythmic Classification & EP Electrophysiology 5.2 Diagnostic & Provocative Drugs: Adenosine, Isoproterenol, Epinephrine, Atropine 5.3 Sedation & Analgesia Pharmacology, Reversal Agents & Recovery Scoring 5.4 Heparin Protocol, ACT Monitoring & Emergency Reversal with Protamine 5.5 Oral Anticoagulants, Antiplatelets & Peri-Procedural Management 5.6 ACLS Algorithms & Emergency Medications in the EP Lab

6 Chapter 6: Catheter Navigation, Normal Electrograms & Baseline Intervals

6.1 Standard Diagnostic Catheter Positions & Fluoroscopic/Anatomical Landmarks 6.2 Intracardiac Electrogram Interpretation & Normal Baseline Intervals (PA, AH, HV) 6.3 His Bundle Electrogram Recording, Split His Potentials & Infranodal Conduction

7 Chapter 7: Programmed Electrical Stimulation & Refractory Periods

7.1 Pacing Stimulation Fundamentals, Capture Thresholds & Drive Trains 7.2 Refractory Period Determination (ERP, RRP, FRP) & Sinus Node Function Testing (SNRT, cSNRT) 7.3 AV Nodal Decremental Conduction & Dual AV Nodal Physiology

8 Chapter 8: Diagnostic Evaluation of Supraventricular & Ventricular Tachycardias

8.1 Narrow-Complex Tachycardias: Differential Diagnosis of AVNRT, AVRT & AT 8.2 Diagnostic Pacing Maneuvers: Entrainment, PPI-TCL, V-A-V vs V-A-A-V & His-Refractory PVCs 8.3 Wide-Complex Tachycardias & Ventricular Arrhythmia Mapping Criteria

9 Chapter 9: Bradyarrhythmias, AV Block & Atrial Arrhythmia Recognition

9.1 Bradyarrhythmias, AV Block Classification & Conduction System Disease 9.2 Atrial Fibrillation & Atrial Flutter: Classification, ECG Recognition & Stroke Risk

10 Chapter 10: 3D Electroanatomic Mapping & Ablation Biophysics

10.1 3D Electroanatomic Mapping Systems (Carto, EnSite, Rhythmia) Setup & Troubleshooting 10.2 Activation, Propagation & Bipolar Voltage Substrate Mapping 10.3 Ablation Biophysics: Radiofrequency, Cryoablation & Pulsed Field Ablation (PFA)

11 Chapter 11: Transseptal Access, Catheter Ablation Targets & Endpoints

11.1 Transseptal Puncture: Anatomy, Technique, Imaging Guidance & Complications 11.2 Catheter Ablation of AVNRT Slow Pathway & Accessory Pathways (WPW) 11.3 Cavotricuspid Isthmus (CTI) Ablation for Typical Atrial Flutter 11.4 Pulmonary Vein Isolation (PVI) for AFib & Ventricular Tachycardia Ablation

12 Chapter 12: Cardiac Implantable Electronic Devices: Implantation & Testing

12.1 Pacemaker & Transvenous ICD Implantation Techniques & Safety 12.2 Intra-Procedural Electrical Testing: Sensing, Pacing Thresholds, Impedance & DFT 12.3 Cardiac Resynchronization Therapy (CRT) & Conduction System Pacing (HBP, LBBAP)

13 Chapter 13: CIED Interrogation, Programming & Troubleshooting

13.1 NBG Pacemaker Codes, Pacing Modes & Timing Cycles 13.2 CIED Interrogation Workflow, Programmer Operation & Peri-Procedural Reprogramming 13.3 Interpreting & Conveying Interrogated Data: Diagnostics, Counters & Stored Episodes 13.4 Pacing & Sensing Malfunctions, Electromagnetic Interference & Temporary Pacing

14 Chapter 14: Specialized Devices, Extraction & Acute Procedural Emergencies

14.1 Subcutaneous ICD (S-ICD), Leadless Pacemakers & Left Atrial Appendage Closure 14.2 Transvenous Lead Extraction Tools, Techniques & Safety Precautions 14.3 Acute Procedural Emergencies: Perforation, Tamponade, Pericardiocentesis & Air Embolism

15 Chapter 15: Post-Procedural Care, Hemostasis & Complication Management

15.1 Femoral Hemostasis Techniques: Manual Compression, Figure-of-8 & Closure Devices 15.2 Post-Procedural Complications: Vascular Injury, AV Block, Esophageal Injury & Neurologic Deficits 15.3 Post-Procedural CIED Assessment, Device Follow-Up & Wound Care 15.4 Post-Procedure Patient Recovery, Discharge Education & Clinical Handover

13.4 Pacing & Sensing Malfunctions, Electromagnetic Interference & Temporary Pacing

Key Takeaways Explain This Topic AI

Failure to capture shows a pacing spike with no depolarization; failure to output shows no spike at all where one was expected.

Undersensing produces inappropriate pacing spikes that fall too early, while oversensing produces inappropriate pauses with no spike at all.

A higher programmed sensitivity number means a less sensitive device, so raising the millivolt value corrects oversensing and lowering it corrects undersensing.

Hyperkalemia, antiarrhythmic drugs, myocardial infarction at the lead tip, and acidosis all raise the capture threshold and can cause acute loss of capture.

Temporary transvenous pacing output is set at two to three times threshold, with sensitivity set to reliably sense the intrinsic R wave in demand mode.

Last updated: September 2026

13.4 Pacing & Sensing Malfunctions, Electromagnetic Interference & Temporary Pacing

CCI lists "Evaluation of temporary and permanent pacing malfunctions" , "CIED troubleshooting" , and "Temporary pacemaker operation" as three separate knowledge areas. Troubleshooting is a small, closed set of problems, and the fastest route to the answer is a fixed framework.

1. The Four Malfunctions

Malfunction ECG appearance Meaning

Failure to capture Spike present, no depolarization follows The stimulus reached the tissue but did not depolarize it

Failure to output No spike where one was expected The device did not deliver a stimulus

Undersensing Inappropriate spikes appearing too soon after an intrinsic beat The device did not see an intrinsic event

Oversensing Inappropriate pauses with no spike The device saw something that was not an intrinsic cardiac event and inhibited

The two most reliable discriminators: failure to capture has a spike, failure to output does not ; and undersensing adds spikes, oversensing removes them .

Failure to capture — causes

  • Category Examples
  • Lead Dislodgement, micro-dislodgement, perforation, poor contact

Threshold rise Hyperkalemia , acidosis, antiarrhythmic drugs (class IC especially) , myocardial infarction at the lead tip, exit block from fibrosis

  • Device Output programmed too low, battery depletion, header connection problem
  • Circuit Conductor fracture with intermittent contact, loose set screw

Hyperkalemia is the classic acute answer because it raises the resting membrane potential, inactivates sodium channels, and both raises threshold and widens the QRS at the same time.

Failure to output — causes

Complete conductor fracture (open circuit, very high impedance), a loose or disconnected set screw, battery at end of life, oversensing that inhibits output (which is really a sensing problem presenting as no output), or the device being programmed to a rate below the intrinsic rate so no output was ever expected — the last of which is not a malfunction at all.

Undersensing — causes

Sensitivity set too low (numerically too high a millivolt threshold), lead dislodgement, insulation or conductor failure, a low-amplitude intrinsic signal (as in the substrate change of infarction), a signal falling within a refractory or blanking period , or a change in the intrinsic rhythm's morphology such as new bundle branch block or ventricular ectopy.

A specific and testable point: an intrinsic beat that falls inside a programmed refractory period is not undersensing. It is normal, designed behavior, and correcting it means adjusting the refractory period, not the sensitivity.

Oversensing — causes

  • Source Typical signature
  • T-wave oversensing Counted rate is exactly double the true rate
  • Myopotential oversensing Pectoral muscle activity or deep breathing; more common with unipolar sensing
  • Far-field R-wave sensing on the atrial channel Ventricular events counted as atrial events
  • Lead fracture / insulation failure noise Non-physiologic, very short intervals; impedance abnormality
  • Crosstalk Ventricular channel senses the atrial output
  • Electromagnetic interference Correlates with an external source
  • 2. The Sensitivity Convention

This is the single most frequently inverted concept in device troubleshooting.

Sensitivity is programmed as a voltage threshold in millivolts, and the device only counts signals larger than that threshold.

  • Programmed value Device behavior Corrects
  • Lower number (e.g. 0.5 mV) More sensitive — detects smaller signals Undersensing
  • Higher number (e.g. 5.0 mV) Less sensitive — ignores smaller signals Oversensing

So the counterintuitive rule is: to make a device more sensitive, program a smaller number. A patient with T-wave oversensing needs the sensitivity value raised (made numerically larger, so the smaller T wave is ignored); a patient undersensing a small R wave needs it lowered .

Sensing configuration matters too. Bipolar sensing uses the closely spaced tip and ring, producing a small antenna that rejects far-field signals and myopotentials. Unipolar sensing uses the tip against the generator can, producing a large antenna that is far more vulnerable to myopotential and electromagnetic oversensing. Converting a unipolar system to bipolar is a standard fix for myopotential oversensing.

3. Electromagnetic Interference

Source Risk Practical guidance

Monopolar electrocautery High — oversensing, inhibition, inappropriate shocks, power-on reset Bipolar cautery, short bursts, asynchronous programming, therapy suspension

MRI High if the system is not conditional Conditional systems only, MRI mode, monitored scan

External defibrillation High Pads ≥ 8 cm from the generator, anteroposterior orientation, interrogate afterward

Radiation therapy Cumulative dose damage and single-event upsets Dose limits; relocate the generator if it is in the field

Radiofrequency ablation Moderate Program asynchronously if dependent, suspend tachytherapy, position the dispersive pad away from the device

  • TENS units, electrolysis, electroconvulsive therapy Moderate Case-specific evaluation
  • Airport / retail security gates Low Walk through at a normal pace; do not linger
  • Cell phones Low Keep ≥ 15 cm (6 inches) from the generator; use the opposite ear

Consumer devices with strong magnets Low but real Keep certain phones, phone cases, wireless chargers, and smart-watch bands ≥ 15 cm from the generator

Household appliances, microwaves, induction hobs Negligible to low Normal use is safe; keep induction cooktops at arm's length

  • Arc welding, large industrial motors, high-power radio transmitters High Occupational assessment required
  • 4. Temporary Pacing
  • Temporary pacing appears in the knowledge list explicitly and in task D8 as "emergency pacing."
  • Modalities

Modality Speed Notes

Transcutaneous Fastest Pads on the chest; painful , requires sedation; verify mechanical capture with a pulse, not just electrical capture on the monitor

  • Transvenous Minutes Balloon-tipped or standard wire to the RV apex; reliable and tolerated
  • Epicardial Post-cardiac-surgery Wires already in place

Through an existing EP catheter Immediate in the lab Connect the RV diagnostic catheter to the stimulator or a temporary generator

Operating a temporary transvenous generator

Three controls, each with a rule:

Rate — set above the intrinsic rate to achieve consistent pacing, typically 60-80 bpm, or higher for overdrive suppression of a pause-dependent arrhythmia.

Output (mA) — determine the threshold by decrementing until capture is lost, then set the output at 2 to 3 times threshold . A typical acute ventricular threshold is under 1 mA, so a typical setting is 2-5 mA.

Sensitivity (mV) — in demand (VVI) mode, set low enough (a smaller number) to sense the intrinsic R wave reliably. Turning the dial fully to asynchronous disables sensing and paces at a fixed rate regardless of intrinsic activity — appropriate only for specific situations such as cautery interference, and hazardous otherwise because of R-on-T risk.

Safety practice: insulate all exposed wire terminals (a glove over the connector block is the classic method) so that stray current cannot reach the endocardium — a microshock hazard at as little as 10-100 µA. Secure the wire, obtain a chest radiograph to confirm position, check thresholds at least daily because they rise over the first days, and monitor continuously.

Complications: RV perforation (thresholds rise, a pericardial rub or effusion appears, and pacing may capture the diaphragm), lead dislodgement with loss of capture, infection at the access site, arrhythmia during placement, and thrombosis. Thresholds that rise steadily over hours to days are expected; an abrupt rise with new diaphragmatic pacing suggests perforation.