Study guide
3.5 Procedural Supplies, Catheter Inventory, Surgical Instrumentation & Wound Closure
Registered Cardiac Electrophysiology Specialist RCES
Track your progress as you read
Finished sections get a check mark.
1 Chapter 1: Pre-Procedural Assessment & Patient Preparation
1.1 Procedure Room Preparation, Equipment Verification & Aseptic Readiness 1.2 Pre-Procedural Clinical Evaluation, Labs, Consents & The Universal Time-Out 1.3 Patient Identification, Transport, HIPAA & Procedural Documentation 1.4 Patient Monitoring Setup, Defibrillator Pads & Pre-Procedural CIED Interrogation 1.5 Vascular Access Planning, Skin Preparation & Sterile Draping
2 Chapter 2: Cardiac Anatomy, Conduction & Electrophysiologic Foundations
2.1 Cardiac Anatomy for the EP Lab: Chambers, Access Routes & Coronary Supply 2.2 The Cardiac Conduction System, Action Potentials & Ion Channel Physiology 2.3 Cardiac Vector Analysis, 12-Lead ECG Fundamentals & Axis Determination
3 Chapter 3: EP Lab Instrumentation, Connectology & Lab Safety
3.1 EP Recording System Architecture, Signal Amplification & Frequency Filters 3.2 Cardiac Stimulator Operation, Junction Box Routing & Noise Troubleshooting 3.3 Electrical Safety, Electrical Units, Electrocautery & Fire Safety 3.4 Radiation Safety Physics, ALARA Principles, Shielding & Dose Limits 3.5 Procedural Supplies, Catheter Inventory, Surgical Instrumentation & Wound Closure
4 Chapter 4: Intra-Procedural Imaging, Sedation & Physiologic Monitoring
4.1 Fluoroscopic Projections (RAO, LAO, AP) & X-ray Tube Optimization 4.2 Intracardiac Echocardiography (ICE) Principles, Views & Catheter Navigation 4.3 Non-Radiographic Imaging: Cardiac CT/MRI, Image Fusion & Ultrasound-Guided Access 4.4 Conscious Sedation Protocols, Airway Management & Hemodynamic Monitoring 4.5 Vital Signs, Normal Laboratory Values & Intra-Procedural Lab Monitoring
5 Chapter 5: EP Pharmacology, Anticoagulation & Emergency Drugs
5.1 Vaughan Williams Antiarrhythmic Classification & EP Electrophysiology 5.2 Diagnostic & Provocative Drugs: Adenosine, Isoproterenol, Epinephrine, Atropine 5.3 Sedation & Analgesia Pharmacology, Reversal Agents & Recovery Scoring 5.4 Heparin Protocol, ACT Monitoring & Emergency Reversal with Protamine 5.5 Oral Anticoagulants, Antiplatelets & Peri-Procedural Management 5.6 ACLS Algorithms & Emergency Medications in the EP Lab
6 Chapter 6: Catheter Navigation, Normal Electrograms & Baseline Intervals
6.1 Standard Diagnostic Catheter Positions & Fluoroscopic/Anatomical Landmarks 6.2 Intracardiac Electrogram Interpretation & Normal Baseline Intervals (PA, AH, HV) 6.3 His Bundle Electrogram Recording, Split His Potentials & Infranodal Conduction
7 Chapter 7: Programmed Electrical Stimulation & Refractory Periods
7.1 Pacing Stimulation Fundamentals, Capture Thresholds & Drive Trains 7.2 Refractory Period Determination (ERP, RRP, FRP) & Sinus Node Function Testing (SNRT, cSNRT) 7.3 AV Nodal Decremental Conduction & Dual AV Nodal Physiology
8 Chapter 8: Diagnostic Evaluation of Supraventricular & Ventricular Tachycardias
8.1 Narrow-Complex Tachycardias: Differential Diagnosis of AVNRT, AVRT & AT 8.2 Diagnostic Pacing Maneuvers: Entrainment, PPI-TCL, V-A-V vs V-A-A-V & His-Refractory PVCs 8.3 Wide-Complex Tachycardias & Ventricular Arrhythmia Mapping Criteria
9 Chapter 9: Bradyarrhythmias, AV Block & Atrial Arrhythmia Recognition
9.1 Bradyarrhythmias, AV Block Classification & Conduction System Disease 9.2 Atrial Fibrillation & Atrial Flutter: Classification, ECG Recognition & Stroke Risk
10 Chapter 10: 3D Electroanatomic Mapping & Ablation Biophysics
10.1 3D Electroanatomic Mapping Systems (Carto, EnSite, Rhythmia) Setup & Troubleshooting 10.2 Activation, Propagation & Bipolar Voltage Substrate Mapping 10.3 Ablation Biophysics: Radiofrequency, Cryoablation & Pulsed Field Ablation (PFA)
11 Chapter 11: Transseptal Access, Catheter Ablation Targets & Endpoints
11.1 Transseptal Puncture: Anatomy, Technique, Imaging Guidance & Complications 11.2 Catheter Ablation of AVNRT Slow Pathway & Accessory Pathways (WPW) 11.3 Cavotricuspid Isthmus (CTI) Ablation for Typical Atrial Flutter 11.4 Pulmonary Vein Isolation (PVI) for AFib & Ventricular Tachycardia Ablation
12 Chapter 12: Cardiac Implantable Electronic Devices: Implantation & Testing
12.1 Pacemaker & Transvenous ICD Implantation Techniques & Safety 12.2 Intra-Procedural Electrical Testing: Sensing, Pacing Thresholds, Impedance & DFT 12.3 Cardiac Resynchronization Therapy (CRT) & Conduction System Pacing (HBP, LBBAP)
13 Chapter 13: CIED Interrogation, Programming & Troubleshooting
13.1 NBG Pacemaker Codes, Pacing Modes & Timing Cycles 13.2 CIED Interrogation Workflow, Programmer Operation & Peri-Procedural Reprogramming 13.3 Interpreting & Conveying Interrogated Data: Diagnostics, Counters & Stored Episodes 13.4 Pacing & Sensing Malfunctions, Electromagnetic Interference & Temporary Pacing
14 Chapter 14: Specialized Devices, Extraction & Acute Procedural Emergencies
14.1 Subcutaneous ICD (S-ICD), Leadless Pacemakers & Left Atrial Appendage Closure 14.2 Transvenous Lead Extraction Tools, Techniques & Safety Precautions 14.3 Acute Procedural Emergencies: Perforation, Tamponade, Pericardiocentesis & Air Embolism
15 Chapter 15: Post-Procedural Care, Hemostasis & Complication Management
15.1 Femoral Hemostasis Techniques: Manual Compression, Figure-of-8 & Closure Devices 15.2 Post-Procedural Complications: Vascular Injury, AV Block, Esophageal Injury & Neurologic Deficits 15.3 Post-Procedural CIED Assessment, Device Follow-Up & Wound Care 15.4 Post-Procedure Patient Recovery, Discharge Education & Clinical Handover
3.5 Procedural Supplies, Catheter Inventory, Surgical Instrumentation & Wound Closure
Key Takeaways Explain This Topic AI
French size divided by three equals the outer diameter in millimeters, so an 8.5 Fr transseptal sheath is roughly 2.8 mm across.
Sheath sizing describes inner diameter, so a catheter of a given French size passes through a sheath labelled with the same French size.
Absorbable sutures such as polyglactin and poliglecaprone close deep pocket layers; non-absorbable sutures such as polypropylene and silk anchor leads and skin.
Suture sizing runs 2-0, 3-0, 4-0 with each additional zero indicating a thinner strand; 2-0 silk anchors leads and 4-0 absorbable monofilament closes subcuticular skin.
Wound healing proceeds through hemostasis, inflammation (days 1-4), proliferation (days 4-21), and remodeling (up to a year), and pocket hematoma is the strongest predictor of CIED infection.
Last updated: September 2026
3.5 Procedural Supplies, Catheter Inventory, Surgical Instrumentation & Wound Closure
Task B8 — Align supplies and equipment with procedure looks trivial until the transseptal needle is opened and no matching dilator is on the field. The RCES specialist is the person who knows what a case will need before the operator asks, and the exam tests that inventory knowledge directly, along with surgical instrumentation , suture techniques and materials , wound closure , wound assessment , and wound healing from the published knowledge list.
1. French Sizing and Catheter Dimensions
The French (Fr) scale measures circumference-derived outer diameter : Outer diameter (mm) = French size 3 French size = OD (mm) × 3 \text{Outer diameter (mm)} = \frac{\text{French size}}{3} \qquad \text{French size} = \text{OD (mm)} \times 3 Outer diameter (mm) = 3 French size French size = OD (mm) × 3
- French Outer diameter
- 5 Fr 1.67 mm
- 6 Fr 2.0 mm
- 7 Fr 2.33 mm
- 8 Fr 2.67 mm
- 8.5 Fr 2.83 mm
- 11 Fr 3.67 mm
- 12 Fr 4.0 mm
The critical convention: a sheath's French size describes its inner diameter , while a catheter's French size describes its outer diameter . A 7 Fr catheter therefore passes through a 7 Fr sheath. The sheath's own outer diameter is roughly 1.5 to 2 Fr larger, which is what the vessel actually accommodates — a fact that matters when planning hemostasis for a 12 Fr ICE sheath.
Guidewires are sized by inch diameter (0.032", 0.035", 0.038") and by length in centimeters; the standard access wire is 0.035" × 145 cm, while extraction and exchange work uses 260 cm wires.
2. The EP Lab Inventory
Category Typical items
Access Micropuncture kit (21 G needle, 0.018" wire, 4-5 Fr transitional sheath), 18 G access needle, 0.035" J-wire, 6-8 Fr short sheaths, sterile ultrasound probe cover
Diagnostic catheters Quadripolar (RA, RV, His), decapolar or duodecapolar CS/Halo, steerable and fixed-curve, 5-7 Fr with 2-10 mm interelectrode spacing
Mapping Multipolar high-density catheters (basket, grid, spline), circular mapping catheter for PVI, patient reference patches
Ablation Non-irrigated, open-irrigated, and contact-force RF catheters; cryoablation console and balloon; PFA catheter and generator; irrigation pump and tubing
Transseptal Brockenbrough-type needle, transseptal dilator and sheath (typically 8-8.5 Fr), pressure tubing, contrast
Imaging ICE catheter (8-10 Fr) and sheath; contrast; pressure transducer
CIED implant Introducer sets with peel-away sheaths, leads, generator, PSA cables, torque wrench, header plugs, tunneling tool
Emergency Pericardiocentesis kit, chest tube tray, temporary pacing wire and external generator, protamine, code cart, endotracheal intubation kit
The specialist's operational discipline is to open the minimum sterile inventory and to keep the case-specific extras — a second transseptal set, a larger sheath, the pericardiocentesis kit — unopened but immediately at hand. Opened, unused disposables are pure cost; unavailable disposables during a tamponade are a patient safety event.
Sterile field maintenance rules that the exam tests:
Only the top of a draped table is sterile; anything below table level is contaminated and is never brought back up.
A gowned person is sterile from chest to sterile-field level, and from the hands to just above the elbow ; the back is never sterile.
Sterile packages are opened away from the body, and the last flap is pulled toward the opener.
Wetness wicks bacteria — strike-through of a drape or gown contaminates it.
Movement is face-to-face or back-to-back; sterile personnel do not turn their backs to the field or pass between two sterile fields.
The circulator pours solution into a basin held at the field's edge without leaning over it.
3. Surgical Instrumentation for CIED Implantation
- Instrument Purpose
- #10 / #15 scalpel blade Skin incision (#15 for finer work)
- Metzenbaum scissors Blunt/sharp dissection of tissue planes
- Mayo scissors Heavier tissue and suture cutting
- Adson forceps (with/without teeth) Tissue handling
- Weitlaner / Gelpi self-retaining retractor Holds the pocket open
- Senn or Army-Navy retractor Hand-held retraction
- Kelly / mosquito hemostat Clamping vessels
- Needle driver (Mayo-Hegar) Suture placement
- Vein pick / lifter Cephalic vein cutdown
- Torque wrench Tightening header set screws to the manufacturer's calibrated torque
- Tunneling tool Routing an S-ICD or extra-pocket lead
Counts — sponges, sharps, and instruments — are performed before the procedure, before wound closure, and at skin closure, and a discrepancy triggers a search and a radiograph before the patient leaves the room.
4. Sutures and Wound Closure
Suture size runs from largest to smallest as the number of zeros increases: 2-0 is thicker than 3-0, which is thicker than 4-0 .
- Suture Type Absorbable Typical EP use
- Silk 2-0 Braided, natural No Lead anchoring sleeve tie-down; sheath securement
- Polypropylene (Prolene) Monofilament synthetic No Skin closure needing removal; vascular repair
- Polyglactin 910 (Vicryl) Braided synthetic Yes (~60-90 days) Deep pocket and fascial layers
- Poliglecaprone (Monocryl) Monofilament synthetic Yes (~90-120 days) Subcuticular skin closure
- Polydioxanone (PDS) Monofilament synthetic Yes (~180-210 days) Fascia requiring prolonged strength
- Nylon Monofilament synthetic No Interrupted skin closure
Monofilament versus braided is the exam's favorite distinction: braided sutures handle and knot better but have interstices that harbor bacteria, so monofilament is preferred where infection risk matters — precisely the CIED pocket. Silk remains the standard for lead anchoring sleeve ties because of its handling, and the tie is placed around the suture sleeve, never directly on the lead body , which would crush the insulation and conductor.
Closure techniques: interrupted (each stitch independent, tolerant of one failure), continuous/running (fast, even tension distribution), subcuticular (buried, best cosmetic result, no removal needed), purse-string, and the figure-of-eight (Z-stitch) used for immediate venous hemostasis around large sheaths in a fully anticoagulated patient. Alternatives include skin adhesive, adhesive strips, and staples.
Removal timing for non-absorbable skin sutures: face 3-5 days, trunk and upper extremity 7-10 days, and the pectoral CIED pocket typically 7-10 days .
5. Wound Healing and Assessment
- Phase Timing Events
- Hemostasis Immediate Vasoconstriction, platelet plug, fibrin clot
Inflammatory Days 1-4 Neutrophils then macrophages debride; erythema, warmth, and mild edema are expected , not infection
Proliferative Days 4-21 Fibroblasts lay collagen III, angiogenesis, granulation, epithelialization
Remodeling / maturation Week 3 to 1 year Collagen III replaced by collagen I; the wound reaches only about 80% of original tensile strength
Healing categories: primary intention (clean, approximated edges — the normal pocket closure), secondary intention (left open to granulate), and tertiary/delayed primary intention (left open, then closed).
Post-procedure wound assessment documents: the incision line and approximation, drainage character and volume, surrounding erythema and its extent, induration, fluctuance, tenderness beyond expectation, and any dehiscence or exposed hardware.
Distinguish the two most-tested findings:
- Expected pocket appearance Concerning pocket appearance
- Soft, mildly ecchymotic swelling that is stable in size Tense, expanding, fluctuant hematoma
- Thin serosanguinous drainage in the first 24-48 h Purulent drainage, or any drainage after 48 h
- Localized erythema at the incision line, fading Spreading erythema, warmth, fever, exposed device
A pocket hematoma is the single strongest predictor of subsequent CIED infection , and the correct management is pressure, correction of anticoagulation where appropriate, and surgical evacuation if it is tense or expanding — never needle aspiration , which introduces organisms into a closed prosthetic space. Erosion or exposure of any part of the system mandates complete extraction of the generator and all leads ; the device cannot be salvaged in place.
Healing is delayed by diabetes and hyperglycemia, corticosteroids and immunosuppression, malnutrition and low albumin, smoking, obesity, renal failure, radiation to the field, and continued anticoagulation with hematoma formation.