Study guide

4.3 Non-Radiographic Imaging: Cardiac CT/MRI, Image Fusion & Ultrasound-Guided Access

Registered Cardiac Electrophysiology Specialist RCES

Track your progress as you read

Finished sections get a check mark.

1 Chapter 1: Pre-Procedural Assessment & Patient Preparation

1.1 Procedure Room Preparation, Equipment Verification & Aseptic Readiness 1.2 Pre-Procedural Clinical Evaluation, Labs, Consents & The Universal Time-Out 1.3 Patient Identification, Transport, HIPAA & Procedural Documentation 1.4 Patient Monitoring Setup, Defibrillator Pads & Pre-Procedural CIED Interrogation 1.5 Vascular Access Planning, Skin Preparation & Sterile Draping

2 Chapter 2: Cardiac Anatomy, Conduction & Electrophysiologic Foundations

2.1 Cardiac Anatomy for the EP Lab: Chambers, Access Routes & Coronary Supply 2.2 The Cardiac Conduction System, Action Potentials & Ion Channel Physiology 2.3 Cardiac Vector Analysis, 12-Lead ECG Fundamentals & Axis Determination

3 Chapter 3: EP Lab Instrumentation, Connectology & Lab Safety

3.1 EP Recording System Architecture, Signal Amplification & Frequency Filters 3.2 Cardiac Stimulator Operation, Junction Box Routing & Noise Troubleshooting 3.3 Electrical Safety, Electrical Units, Electrocautery & Fire Safety 3.4 Radiation Safety Physics, ALARA Principles, Shielding & Dose Limits 3.5 Procedural Supplies, Catheter Inventory, Surgical Instrumentation & Wound Closure

4 Chapter 4: Intra-Procedural Imaging, Sedation & Physiologic Monitoring

4.1 Fluoroscopic Projections (RAO, LAO, AP) & X-ray Tube Optimization 4.2 Intracardiac Echocardiography (ICE) Principles, Views & Catheter Navigation 4.3 Non-Radiographic Imaging: Cardiac CT/MRI, Image Fusion & Ultrasound-Guided Access 4.4 Conscious Sedation Protocols, Airway Management & Hemodynamic Monitoring 4.5 Vital Signs, Normal Laboratory Values & Intra-Procedural Lab Monitoring

5 Chapter 5: EP Pharmacology, Anticoagulation & Emergency Drugs

5.1 Vaughan Williams Antiarrhythmic Classification & EP Electrophysiology 5.2 Diagnostic & Provocative Drugs: Adenosine, Isoproterenol, Epinephrine, Atropine 5.3 Sedation & Analgesia Pharmacology, Reversal Agents & Recovery Scoring 5.4 Heparin Protocol, ACT Monitoring & Emergency Reversal with Protamine 5.5 Oral Anticoagulants, Antiplatelets & Peri-Procedural Management 5.6 ACLS Algorithms & Emergency Medications in the EP Lab

6 Chapter 6: Catheter Navigation, Normal Electrograms & Baseline Intervals

6.1 Standard Diagnostic Catheter Positions & Fluoroscopic/Anatomical Landmarks 6.2 Intracardiac Electrogram Interpretation & Normal Baseline Intervals (PA, AH, HV) 6.3 His Bundle Electrogram Recording, Split His Potentials & Infranodal Conduction

7 Chapter 7: Programmed Electrical Stimulation & Refractory Periods

7.1 Pacing Stimulation Fundamentals, Capture Thresholds & Drive Trains 7.2 Refractory Period Determination (ERP, RRP, FRP) & Sinus Node Function Testing (SNRT, cSNRT) 7.3 AV Nodal Decremental Conduction & Dual AV Nodal Physiology

8 Chapter 8: Diagnostic Evaluation of Supraventricular & Ventricular Tachycardias

8.1 Narrow-Complex Tachycardias: Differential Diagnosis of AVNRT, AVRT & AT 8.2 Diagnostic Pacing Maneuvers: Entrainment, PPI-TCL, V-A-V vs V-A-A-V & His-Refractory PVCs 8.3 Wide-Complex Tachycardias & Ventricular Arrhythmia Mapping Criteria

9 Chapter 9: Bradyarrhythmias, AV Block & Atrial Arrhythmia Recognition

9.1 Bradyarrhythmias, AV Block Classification & Conduction System Disease 9.2 Atrial Fibrillation & Atrial Flutter: Classification, ECG Recognition & Stroke Risk

10 Chapter 10: 3D Electroanatomic Mapping & Ablation Biophysics

10.1 3D Electroanatomic Mapping Systems (Carto, EnSite, Rhythmia) Setup & Troubleshooting 10.2 Activation, Propagation & Bipolar Voltage Substrate Mapping 10.3 Ablation Biophysics: Radiofrequency, Cryoablation & Pulsed Field Ablation (PFA)

11 Chapter 11: Transseptal Access, Catheter Ablation Targets & Endpoints

11.1 Transseptal Puncture: Anatomy, Technique, Imaging Guidance & Complications 11.2 Catheter Ablation of AVNRT Slow Pathway & Accessory Pathways (WPW) 11.3 Cavotricuspid Isthmus (CTI) Ablation for Typical Atrial Flutter 11.4 Pulmonary Vein Isolation (PVI) for AFib & Ventricular Tachycardia Ablation

12 Chapter 12: Cardiac Implantable Electronic Devices: Implantation & Testing

12.1 Pacemaker & Transvenous ICD Implantation Techniques & Safety 12.2 Intra-Procedural Electrical Testing: Sensing, Pacing Thresholds, Impedance & DFT 12.3 Cardiac Resynchronization Therapy (CRT) & Conduction System Pacing (HBP, LBBAP)

13 Chapter 13: CIED Interrogation, Programming & Troubleshooting

13.1 NBG Pacemaker Codes, Pacing Modes & Timing Cycles 13.2 CIED Interrogation Workflow, Programmer Operation & Peri-Procedural Reprogramming 13.3 Interpreting & Conveying Interrogated Data: Diagnostics, Counters & Stored Episodes 13.4 Pacing & Sensing Malfunctions, Electromagnetic Interference & Temporary Pacing

14 Chapter 14: Specialized Devices, Extraction & Acute Procedural Emergencies

14.1 Subcutaneous ICD (S-ICD), Leadless Pacemakers & Left Atrial Appendage Closure 14.2 Transvenous Lead Extraction Tools, Techniques & Safety Precautions 14.3 Acute Procedural Emergencies: Perforation, Tamponade, Pericardiocentesis & Air Embolism

15 Chapter 15: Post-Procedural Care, Hemostasis & Complication Management

15.1 Femoral Hemostasis Techniques: Manual Compression, Figure-of-8 & Closure Devices 15.2 Post-Procedural Complications: Vascular Injury, AV Block, Esophageal Injury & Neurologic Deficits 15.3 Post-Procedural CIED Assessment, Device Follow-Up & Wound Care 15.4 Post-Procedure Patient Recovery, Discharge Education & Clinical Handover

4.3 Non-Radiographic Imaging: Cardiac CT/MRI, Image Fusion & Ultrasound-Guided Access

Key Takeaways Explain This Topic AI

Task B15 requires acquiring and interpreting both radiographic and non-radiographic images, naming MRI, ultrasound, and the 3D mapping system explicitly.

Cardiac CT is acquired at 60-80 percent of the R-R interval for a rate-controlled patient and defines pulmonary vein ostial anatomy, common ostia, and left atrial appendage morphology.

Late gadolinium enhancement on cardiac MRI identifies ventricular scar and border zone that predicts the ventricular tachycardia substrate before the catheter enters the chamber.

Ultrasound-guided venous access reduces access-site complications and confirms vein compressibility, patency, and the artery-vein relationship before the needle moves.

An MRI-conditional CIED system requires the generator and every lead to be labelled conditional, with the device placed in an MRI mode before scanning and restored afterward.

Last updated: September 2026

4.3 Non-Radiographic Imaging: Cardiac CT/MRI, Image Fusion & Ultrasound-Guided Access

CCI task B15 — Acquire/interpret radiographic and non-radiographic (e.g., MRI, ultrasound, 3D mapping system) image is explicit that the RCES specialist works with more than fluoroscopy. Modern EP practice front-loads anatomy into cross-sectional imaging and then imports it into the mapping system, so the catheter time in the chamber is spent confirming rather than discovering.

1. Cardiac Computed Tomography

Acquisition principles. A contrast-enhanced, ECG-gated cardiac CT reconstructs the heart at a chosen point in the cardiac cycle.

Parameter Typical practice Rationale

Gating phase 60-80% of the R-R interval (mid-to-late diastole) Least cardiac motion in a rate-controlled patient

  • Heart rate target Often < 65 bpm, beta blockade if needed Reduces motion blur and permits prospective gating
  • Contrast Iodinated, timed by bolus tracking to the left atrium Opacifies the chamber of interest
  • Reconstruction 0.5-0.75 mm slices, then 3D segmentation Enables surface-model export

In atrial fibrillation the R-R interval varies, so a prospectively gated scan may misfire; retrospective gating with dose modulation or a wide-detector single-beat acquisition is used instead, at a higher dose cost.

What CT answers before an ablation:

Pulmonary vein anatomy — number, ostial dimensions, and variants. A left common ostium occurs in roughly 20-30% of patients and changes both catheter selection and cryoballoon strategy; a right middle vein is a common accessory.

Left atrial appendage morphology for closure planning — chicken-wing, windsock, cactus, cauliflower — plus ostial dimensions and landing-zone depth.

Thrombus exclusion in the appendage (though transesophageal echocardiography remains the reference standard).

Esophageal course relative to the posterior left atrial wall.

Left atrial volume and wall thickness , which predict ablation outcome.

Coronary and phrenic relationships to planned lesion sets.

2. Cardiac Magnetic Resonance

MRI's contribution is tissue characterization , which CT cannot provide.

Sequence What it shows EP application

Late gadolinium enhancement (LGE) Fibrosis and scar (gadolinium accumulates in expanded extracellular space and washes out slowly) Ventricular tachycardia substrate: dense scar core, border zone, and conducting channels

Cine steady-state free precession Wall motion, volumes, ejection fraction Cardiomyopathy assessment, device candidacy

  • T1/T2 mapping and T2-weighted Edema, inflammation, infiltration Myocarditis, cardiac sarcoidosis, amyloid
  • MR angiography Vascular anatomy Pulmonary veins without iodinated contrast

The scar distribution pattern is itself diagnostic and directs the mapping approach:

Pattern Typical cause Mapping implication

Subendocardial or transmural in a coronary territory Ischemic infarction Endocardial substrate mapping usually sufficient

Mid-wall septal stripe Non-ischemic dilated cardiomyopathy Deep septal substrate; may need bipolar or septal approaches

  • Subepicardial lateral Myocarditis, sarcoid, Chagas Anticipate an epicardial approach
  • Patchy multifocal with edema Active sarcoidosis Treat inflammation; substrate may evolve

A pre-procedure LGE map lets the team predict where late potentials and channels will be found, and the segmented scar can be imported into the electroanatomic map as a shell overlay.

3. Image Integration and Fusion

Segmentation converts a CT or MR volume into a 3D surface model of the chamber of interest. Registration aligns that model to the patient's real-time position within the mapping system's coordinate space.

Registration method How it works Weakness

Landmark (fiducial) registration Match discrete anatomic points — vein ostia, appendage, CS os — acquired with the catheter to the same points on the model Depends on the accuracy of each point

Surface registration Least-squares fit of a cloud of acquired points to the model surface Requires broad, even sampling

ICE-based Contours traced from intracardiac echo build or align the shell in real time Operator-dependent tracing

Sources of registration error are heavily tested because they are the practical failure mode: the CT was acquired days earlier in a different rhythm (sinus versus AF), at a different volume status , at a different respiratory phase , and in a different body position . A shell registered in sinus rhythm will misalign once the patient is in atrial fibrillation with a dilated atrium. Registration is verified — not assumed — by touching a known landmark and confirming the catheter icon lands on it, and it is re-verified after any patient movement, cardioversion, or large fluid administration.

The reference patch or reference catheter must not move; if it does, the entire acquired geometry shifts, producing duplicate or displaced anatomy.

4. Ultrasound-Guided Vascular Access

Ultrasound guidance has become the default for femoral access in the EP lab because it converts a landmark-and-palpation procedure into a visualized one.

Distinguishing artery from vein in the short axis:

  • Feature Vein Artery
  • Compressibility Collapses with light probe pressure Remains round, resists compression
  • Pulsatility Non-pulsatile (may transmit) Pulsatile
  • Wall Thin Thick, echogenic
  • Doppler Low-velocity, phasic with respiration High-velocity, pulsatile
  • Response to Valsalva Distends No change
  • Position at the groin Medial to the artery Lateral

Technique. Sterile probe cover and sterile gel; identify the common femoral vein above the bifurcation and below the inguinal ligament ; use a short-axis (transverse) view for orientation and a long-axis view to track needle depth; advance under continuous visualization until the tip tents and then enters the anterior wall; confirm the wire is intraluminal in long axis before dilating.

Value: reduced inadvertent arterial puncture, fewer attempts, fewer hematomas, and reliable success where anatomy is difficult — obesity, prior scarring, chronic occlusion, or a small vein. Ultrasound also detects an anatomic variant such as an anteriorly overlying artery, chronic femoral vein thrombosis, or an unexpectedly high bifurcation before the needle is used, which no landmark technique can do.

5. MRI in Patients With Cardiac Implantable Electronic Devices

Historically an absolute contraindication, MRI is now routinely performed under controlled protocols. The distinction the exam tests is between MRI-conditional and legacy systems.

MRI-conditional requires that every component — generator and every lead — is labelled MRI-conditional by the manufacturer , and that the specific scanner field strength, gradient, and SAR conditions in the labelling are met. A conditional generator with one legacy lead is not a conditional system. An abandoned or capped lead makes a system non-conditional because it cannot be programmed and can heat at its free end.

Hazard Mechanism Mitigation

Lead-tip heating RF energy couples into the lead, concentrating at the electrode Conditional lead design; SAR limits; scan-region restrictions

Oversensing / inhibition Gradient and RF noise interpreted as intrinsic activity Program asynchronous pacing in a dependent patient

Inappropriate ICD therapy Noise detected as ventricular arrhythmia Suspend tachytherapy for the scan

Force and torque Static field acting on ferromagnetic components Modern components are largely non-ferromagnetic

Reed switch behavior Static field may close a magnet-sensitive switch Conditional devices use Hall sensors and a defined MRI mode

Workflow: interrogate and document baseline parameters; confirm the patient is not pacing-dependent, or program to an asynchronous mode if they are; enable MRI mode; monitor continuously with ECG and pulse oximetry during the scan with trained personnel and a resuscitation plan; then exit MRI mode and re-interrogate , restoring the original programming and verifying thresholds, sensing, and impedances. Leaving a device in MRI mode after the scan is a documented and preventable adverse event.