Study guide
5.4 Left Ventriculography & Aortography
Registered Cardiovascular Invasive Specialist RCIS
Track your progress as you read
Finished sections get a check mark.
1 Introduction & Exam Overview
1.1 Current RCIS Exam Facts 1.2 Eligibility Pathways & Documentation 1.3 Examination Matrix & Study Plan
2 Pre-Procedural Room Setup & Sterile Technique
2.1 Procedure & Control Room Setup 2.2 Sterile Technique & Universal Precautions 2.3 Personnel Preparation & Cath Lab Roles
3 Pre-Procedural Patient Preparation & Validation
3.1 Preparing the Patient for Invasive Procedures 3.2 Records, Labs, Allergies & Consent 3.3 Time-Out & Universal Protocol
4 Patient Monitoring & Vascular Access
4.1 Intraprocedural Vital Sign Monitoring 4.2 Arterial & Venous Access Sites 4.3 Ultrasound-Guided Access
5 Coronary Anatomy & Left Heart Catheterization
5.1 Coronary Anatomy & Dominance 5.2 Left Heart Catheterization & Coronary Angiography 5.3 Catheter Shapes & Angiographic Projections 5.4 Left Ventriculography & Aortography
6 Right Heart Catheterization & Hemodynamics
6.1 Right Heart Waveforms & Pressures 6.2 Cardiac Output: Fick & Thermodilution 6.3 Valve Area, Bernoulli & Shunt Calculations 6.4 Endomyocardial Biopsy Support
7 ECG & Intracardiac/Intravascular Imaging
7.1 12-Lead ECG in the Cath Lab 7.2 Intracardiac Echocardiography (ICE) 7.3 IVUS & OCT 7.4 Identifying Structures on Cardiac Ultrasound
8 Physiology Assessment & Radiation Safety
8.1 FFR & iFR Studies 8.2 Radiation Safety & ALARA 8.3 Radiographic Equipment Operation 8.4 Image Acquisition & Interpretation
9 Cath Lab Pharmacology
9.1 Anticipating Pharmacologic Needs 9.2 Recognizing Medication Effects & Adverse Reactions 9.3 Anticoagulation & ACT Targets
10 PCI Fundamentals
10.1 Balloon Angioplasty & Stent Platforms 10.2 Coronary Wiring, Lesion Prep & Flow Grades 10.3 Vascular Closure Devices
11 Chapter 11: Complex Coronary Intervention
11.1 CTO, Atherectomy & Alcohol Septal Ablation 11.2 Cardiac Thrombectomy (Mechanical & Aspiration) 11.3 Distal Protection Devices
12 Device Implants & Temporary Support
12.1 Pacemaker, ICD, CRT & Loop Recorders 12.2 Transvenous Temporary Pacemaker 12.3 Pericardiocentesis Support
13 Mechanical Circulatory Support
13.1 IABP & Impella 13.2 ECMO & LVAD Overview for the Cath Lab
14 Peripheral Vascular Interventions
14.1 Peripheral Interventions 14.2 Complex Peripheral Interventions 14.3 Peripheral Thrombectomy & Thrombolysis
15 Structural Heart Interventions
15.1 Nonvalvular Structural Therapies 15.2 Percutaneous Valve Interventions
16 Chapter 16: Emergency Clinical Presentations & Equipment
16.1 STEMI, Cardiogenic Shock & Pulmonary Edema 16.2 Stroke & Neurologic Emergencies in the Lab 16.3 Emergency Equipment Operation
17 Intraprocedural Complications
17.1 Coronary Interventional Complications 17.2 Peripheral & Structural Complications
18 Post-Procedural Activities
18.1 Arterial & Venous Hemostasis 18.2 Access-Site Complications 18.3 Patient Education, Reporting & Transfer of Care
5.4 Left Ventriculography & Aortography
Key Takeaways Explain This Topic AI
Left ventriculography assesses LV ejection fraction, wall motion, and mitral regurgitation severity using a pigtail catheter positioned in the LV body
Standard LV contrast injection uses 30–45 mL of contrast over 10–15 mL/s via power injector or hand injection depending on LV size
Aortography evaluates aortic regurgitation, aneurysm, dissection, and valve morphology with pigtail placement in the ascending aorta or aortic root
Mitral regurgitation severity on ventriculography is graded by contrast reflux into the LA: 1+ faint, 2+ fills LA partially, 3+ fills entire LA, 4+ refluxes into pulmonary veins
Total contrast load must be tracked across coronary, LV, and aortic injections; ventriculography risks include arrhythmia, LV thrombus embolization, and hemodynamic decompensation
Last updated: July 2026
While selective coronary angiography maps the epicardial vessels, left ventriculography and aortography answer questions about pump function, wall motion, and valvular competence. These injections use larger contrast volumes and carry distinct arrhythmia and hemodynamic risks. RCIS technologists must prepare the power injector, confirm catheter position, and monitor the patient closely throughout each shot.
- Left Ventriculography
- Purpose
Left ventriculography provides:
- LV ejection fraction (LVEF) estimation
- Regional wall motion assessment (hypokinesis, akinesis, dyskinesis, aneurysm)
- Mitral regurgitation (MR) severity by contrast reflux into the left atrium
- Intracardiac mass or thrombus detection (contrast filling defects)
- VSD identification in post-MI mechanical complications
Ventriculography is performed when echocardiography is inconclusive, when MR severity must be quantified invasively, or when wall motion must be correlated with coronary findings in the same session.
Catheter and Positioning
The pigtail catheter is advanced across the aortic valve into the mid-LV cavity —typically away from the apex (to avoid ectopy) and away from the mitral apparatus (to avoid artifact). Position is confirmed in both LAO and RAO projections before injection.
Alternative: multipurpose (MP) catheter for ventriculography in some labs, though pigtail remains standard.
Injection Technique
- Parameter Typical Range Notes
- Contrast volume 30–45 mL Reduce in small LV or renal impairment
- Injection rate 10–15 mL/s Power injector preferred for consistent opacification
- Projection RAO 30° common Shows LV cavity and MR reflux profile
- Breath hold End-expiration or end-inspiration per lab Reduces motion artifact
Adequate opacification requires dense contrast in the LV cavity during systole. Suboptimal positioning (apical placement, mitral chordae contact) produces incomplete fills and inaccurate EF estimation.
Wall Motion Scoring
Wall motion is graded by segment (17-segment model correlates with coronary territories):
- Grade Description
- Normal Symmetric thickening and contraction
- Hypokinesis Reduced contraction
- Akinesis Absent contraction
- Dyskinesis Outward paradoxical motion (aneurysm)
Correlate anterior/apical akinesis with LAD disease, lateral with LCx, and inferior with RCA/PDA territory.
Mitral Regurgitation Grading on Ventriculography
- Grade Angiographic Finding
- 1+ (mild) Faint LA opacification; clears with each beat
- 2+ (moderate) LA opacification that does not fill entire chamber
- 3+ (moderate-severe) Complete LA opacification equal to LV density
- 4+ (severe) LA opacification with reflux into pulmonary veins
Ventriculographic MR grading is semi-quantitative; echocardiography and hemodynamics (large V wave on wedge tracing) provide complementary data.
- Aortography
- Purpose
Aortography evaluates:
- Aortic regurgitation (AR) severity
- Aortic root and ascending aorta dimensions
- Aortic aneurysm and dissection
- Aortic valve morphology (bicuspid, stenosis concomitant assessment)
- Coronary ostia relationship to sinuses (before TAVR or root surgery)
- Catheter and Positioning
The pigtail catheter is positioned in the ascending aorta or aortic root , just above the valve. For root shots, the tip sits below the sinotubular junction; for ascending aortography, the tip is higher to include the tubular aorta.
Injection Technique
- Parameter Typical Range Notes
- Contrast volume 40–60 mL Larger volumes for aneurysm run-off
- Injection rate 15–20 mL/s Rapid fill needed to see AR during diastole
- Projection LAO 45–60° or RAO Opens aortic root; valve profile visible
- Aortic Regurgitation Grading on Aortography
- Grade Angiographic Finding
- 1+ (mild) Faint LV opacification; clears each cycle
- 2+ (moderate) LV opacification that clears slowly
- 3+ (moderate-severe) Dense LV opacification equal to aortic root
- 4+ (severe) Complete LV opacification with reflux lasting throughout diastole
Combined with hemodynamics, a wide pulse pressure and low diastolic pressure support severe AR.
Contrast Management
Total contrast exposure accumulates across the case:
- Procedure Component Approximate Volume
- LCA angiography 6–12 mL per injection × multiple views
- RCA angiography 4–8 mL per injection × multiple views
- LV ventriculography 30–45 mL
- Aortography 40–60 mL
Track cumulative contrast volume , especially when eGFR is reduced. Standard prophylaxis includes IV hydration per protocol, minimizing unnecessary injections, and coordinating with the physician on whether ventriculography can be deferred to echocardiography.
Complications and RCIS Monitoring
Complication Mechanism RCIS Response
Ventricular arrhythmia (PVCs, VT) Contrast irritates endocardium; apical catheter contact Notify physician; prepare antiarrhythmics per protocol
Hypotension Contrast-induced vasodilation, MR/AR hemodynamic burden Monitor BP; IV fluids; pressors if ordered
Thrombus embolization Contrast dislodges apical thrombus Avoid aggressive apical positioning; maintain awareness in anterior MI patients
- Contrast reaction Anaphylactoid response Stop injection; airway and epinephrine per ACLS
- Complete heart block Rare during deep catheter manipulation in AV node territory Temporary pacing readiness
After ventriculography, watch for delayed hypotension and tachycardia from contrast load. Document LVEF estimate, wall motion abnormalities, MR/AR grade, and total contrast volume in the procedure record.
When to Skip or Limit Ventriculography
Ventriculography may be omitted or minimized when:
- Recent high-quality echocardiography provides EF and MR data
- Renal function is severely impaired
- LV thrombus is suspected (risk of embolization)
- Patient is hemodynamically unstable
The RCIS role includes having this equipment ready when ordered while supporting contrast-sparing decisions when the physician defers the shot.