Study guide

8.1 FFR & iFR Studies

Registered Cardiovascular Invasive Specialist RCIS

Track your progress as you read

Finished sections get a check mark.

1 Introduction & Exam Overview

1.1 Current RCIS Exam Facts 1.2 Eligibility Pathways & Documentation 1.3 Examination Matrix & Study Plan

2 Pre-Procedural Room Setup & Sterile Technique

2.1 Procedure & Control Room Setup 2.2 Sterile Technique & Universal Precautions 2.3 Personnel Preparation & Cath Lab Roles

3 Pre-Procedural Patient Preparation & Validation

3.1 Preparing the Patient for Invasive Procedures 3.2 Records, Labs, Allergies & Consent 3.3 Time-Out & Universal Protocol

4 Patient Monitoring & Vascular Access

4.1 Intraprocedural Vital Sign Monitoring 4.2 Arterial & Venous Access Sites 4.3 Ultrasound-Guided Access

5 Coronary Anatomy & Left Heart Catheterization

5.1 Coronary Anatomy & Dominance 5.2 Left Heart Catheterization & Coronary Angiography 5.3 Catheter Shapes & Angiographic Projections 5.4 Left Ventriculography & Aortography

6 Right Heart Catheterization & Hemodynamics

6.1 Right Heart Waveforms & Pressures 6.2 Cardiac Output: Fick & Thermodilution 6.3 Valve Area, Bernoulli & Shunt Calculations 6.4 Endomyocardial Biopsy Support

7 ECG & Intracardiac/Intravascular Imaging

7.1 12-Lead ECG in the Cath Lab 7.2 Intracardiac Echocardiography (ICE) 7.3 IVUS & OCT 7.4 Identifying Structures on Cardiac Ultrasound

8 Physiology Assessment & Radiation Safety

8.1 FFR & iFR Studies 8.2 Radiation Safety & ALARA 8.3 Radiographic Equipment Operation 8.4 Image Acquisition & Interpretation

9 Cath Lab Pharmacology

9.1 Anticipating Pharmacologic Needs 9.2 Recognizing Medication Effects & Adverse Reactions 9.3 Anticoagulation & ACT Targets

10 PCI Fundamentals

10.1 Balloon Angioplasty & Stent Platforms 10.2 Coronary Wiring, Lesion Prep & Flow Grades 10.3 Vascular Closure Devices

11 Chapter 11: Complex Coronary Intervention

11.1 CTO, Atherectomy & Alcohol Septal Ablation 11.2 Cardiac Thrombectomy (Mechanical & Aspiration) 11.3 Distal Protection Devices

12 Device Implants & Temporary Support

12.1 Pacemaker, ICD, CRT & Loop Recorders 12.2 Transvenous Temporary Pacemaker 12.3 Pericardiocentesis Support

13 Mechanical Circulatory Support

13.1 IABP & Impella 13.2 ECMO & LVAD Overview for the Cath Lab

14 Peripheral Vascular Interventions

14.1 Peripheral Interventions 14.2 Complex Peripheral Interventions 14.3 Peripheral Thrombectomy & Thrombolysis

15 Structural Heart Interventions

15.1 Nonvalvular Structural Therapies 15.2 Percutaneous Valve Interventions

16 Chapter 16: Emergency Clinical Presentations & Equipment

16.1 STEMI, Cardiogenic Shock & Pulmonary Edema 16.2 Stroke & Neurologic Emergencies in the Lab 16.3 Emergency Equipment Operation

17 Intraprocedural Complications

17.1 Coronary Interventional Complications 17.2 Peripheral & Structural Complications

18 Post-Procedural Activities

18.1 Arterial & Venous Hemostasis 18.2 Access-Site Complications 18.3 Patient Education, Reporting & Transfer of Care

8.1 FFR & iFR Studies

Key Takeaways Explain This Topic AI

Fractional flow reserve (FFR) compares distal coronary pressure to aortic pressure during induced hyperemia; an FFR of approximately 0.80 or lower indicates hemodynamically significant stenosis

Instantaneous wave-free ratio (iFR) is a resting, adenosine-free index with a diagnostic cutoff of approximately 0.89

Adenosine (or alternatives such as ATP or regadenoson) is required for standard hyperemic FFR to maximize coronary flow and minimize microvascular resistance

Pressure-wire technique demands meticulous zeroing, equalization, and waveform quality; small errors can shift FFR values enough to change revascularization decisions

Common pitfalls include wire drift, damped aortic pressure, ostial or left main lesions, serial stenoses, microvascular disease, and contrast or air in the pressure line

Last updated: July 2026

8.1 FFR & iFR Studies

Quick Answer: FFR measures pressure across a coronary lesion during pharmacologic hyperemia; ≤ ~0.80 generally indicates ischemia-producing stenosis warranting revascularization. iFR is a resting index measured in the wave-free period with a cutoff of ~0.89 , avoiding adenosine. RCIS candidates must know both thresholds, how adenosine is used, wire technique basics, and the pitfalls that invalidate results.

Physiologic lesion assessment has moved beyond angiography alone. Two pressure-derived indices dominate contemporary cath lab practice: fractional flow reserve (FFR) and instantaneous wave-free ratio (iFR) . Both quantify whether a stenosis meaningfully restricts blood flow, but they differ in timing, pharmacology, and interpretation. As an RCIS, you prepare equipment, monitor the patient during hyperemia, recognize artifact, and support accurate data acquisition.

What FFR Measures

FFR is the ratio of mean distal coronary pressure to mean aortic pressure during maximal hyperemia , when microvascular resistance is minimized and flow is proportional to the pressure gradient across a stenosis.

Formula (conceptual): FFR = Pd / Pa (during hyperemia)

  • FFR Value Typical Interpretation
  • ~0.80 or lower Hemodynamically significant; lesion likely causes ischemia
  • > ~0.80 Not hemodynamically significant by FFR criteria
  • ~0.75–0.80 Gray zone in some protocols; institution-specific

Landmark trials (DEFER, FAME) established that deferring PCI for FFR > 0.80 is safe, while revascularizing FFR ≤ 0.80 improves outcomes. The ~0.80 cutoff is the exam anchor number.

Hyperemia and Adenosine

Standard FFR requires pharmacologic hyperemia so microvascular resistance falls and the pressure gradient reflects the stenosis alone. Without hyperemia, resting flow may not expose the full functional severity of moderate lesions.

Adenosine is the most common agent in the United States:

Intracoronary (IC): Rapid bolus through the guiding catheter (e.g., 40–100 mcg depending on vessel and protocol); quick onset, short duration; useful when systemic effects must be minimized

Intravenous (IV): Continuous infusion (commonly 140 mcg/kg/min through a dedicated line); sustained hyperemia for pull-back assessments and multiple vessels

Alternatives include ATP (similar mechanism, IV infusion) and regadenoson (selective A2A agonist, IV bolus). Each causes transient side effects—flushing, dyspnea, chest discomfort, AV block, bronchospasm in reactive airways—that the RCIS must anticipate and communicate to the physician.

RCIS responsibilities during hyperemia:

  • Confirm dedicated IV access and infusion pump settings when IV adenosine is used
  • Monitor HR, rhythm, BP, and SpO₂; adenosine can provoke bradycardia, AV block, or hypotension
  • Ensure the pressure wire signal remains stable before and during hyperemia
  • Document hemodynamic tracings and any arrhythmia or patient symptoms
  • iFR: Resting Assessment Without Adenosine

iFR (instantaneous wave-free ratio) is measured during the wave-free period of diastole, when microvascular resistance is naturally lower and distal pressure more reliably reflects flow through the stenosis. Because hyperemia is not required, iFR reduces cost, time, and patient discomfort.

  • Index Hyperemia Required? Approximate Cutoff
  • FFR Yes (adenosine/ATP/regadenoson) ≤ ~0.80 significant
  • iFR No (resting) ≤ ~0.89 significant

An iFR ≤ ~0.89 correlates with ischemia in validation studies (VERIFY, DEFINE-FLAIR). Values above ~0.89 generally support deferral. Hybrid strategies (iFR-first, FFR if borderline) appear in many labs; know both numbers for the exam.

Pressure-Wire Technique Essentials

Accurate FFR/iFR depends on signal quality. The RCIS supports setup and troubleshooting:

Zeroing: Zero the transducer at atmospheric pressure with the wire in a neutral reference (often the catheter hub or saline flush port per manufacturer protocol)

Equalization: Match wire and guide pressures in the aorta before crossing the lesion—waveforms should overlay

Crossing: Advance the pressure wire distal to the lesion; confirm d waveforms are not damped and not in a side branch

Recording: Acquire stable tracings during hyperemia (FFR) or in the resting wave-free window (iFR)

Pull-back: Slowly withdraw the wire during sustained hyperemia to identify pressure drops across tandem lesions

Waveform inspection is as important as the numeric result. Equalization mismatch, damped aortic pressure, or ventricularized distal tracings invalidate the index.

Pitfalls and Sources of Error

  • Pitfall Effect RCIS Response
  • Wire drift / miscalibration False positive or negative FFR Re-zero, re-equalize, replace wire if needed
  • Damped Pa (guide against wall) Artificially low FFR Reposition guide in aortic root; flush and aspirate
  • Ostial / left main disease Difficult equalization; unreliable Pd Recognize limitation; notify physician
  • Serial stenoses Single FFR reflects combined lesion burden Pull-back during hyperemia to localize gradients
  • Microvascular disease Low FFR without epicardial stenosis Clinical context; may need additional testing
  • Air or contrast in pressure line Artifact spikes Flush system; eliminate bubbles
  • Incomplete hyperemia Falsely elevated FFR Confirm adenosine dose, infusion rate, and duration
  • Ectopy / arrhythmia during measurement Unstable mean pressures Wait for stable rhythm; repeat acquisition

FFR vs. iFR selection pitfalls: Applying the 0.80 FFR threshold to iFR values (or vice versa) is a common exam trap. iFR is not adenosine-free FFR—they are different indices with different cutoffs.

Clinical Integration in the Cath Lab

Physiologic assessment guides PCI versus medical therapy. When FFR ≤ ~0.80 or iFR ≤ ~0.89, stenting typically improves symptoms and outcomes; when values are above threshold, deferral is often appropriate even if angiography looks moderate.

The RCIS role extends beyond button presses: maintain infusion lines, monitor for bronchospasm or high-grade AV block during adenosine, archive hemodynamic records, and verbalize when waveforms look damped. Catching artifact before the physician commits to revascularization is a high-value safety contribution.

  • Exam Focus Checklist
  • FFR cutoff ~0.80 during hyperemia
  • iFR cutoff ~0.89 at rest, no adenosine
  • Adenosine routes (IC bolus vs IV infusion) and side effects
  • Equalization, zeroing, pull-back technique
  • Pitfalls: drift, damped pressures, serial lesions, microvascular disease

Master these concepts and you can anticipate physician needs, protect data integrity, and answer RCIS items on functional coronary assessment with confidence.