2.1 Procedure & Control Room Setup

Registered Cardiovascular Invasive Specialist RCIS

Track your progress as you read

Finished sections get a check mark.

1 Introduction & Exam Overview

1.1 Current RCIS Exam Facts 1.2 Eligibility Pathways & Documentation 1.3 Examination Matrix & Study Plan

2 Pre-Procedural Room Setup & Sterile Technique

2.1 Procedure & Control Room Setup 2.2 Sterile Technique & Universal Precautions 2.3 Personnel Preparation & Cath Lab Roles

3 Pre-Procedural Patient Preparation & Validation

3.1 Preparing the Patient for Invasive Procedures 3.2 Records, Labs, Allergies & Consent 3.3 Time-Out & Universal Protocol

4 Patient Monitoring & Vascular Access

4.1 Intraprocedural Vital Sign Monitoring 4.2 Arterial & Venous Access Sites 4.3 Ultrasound-Guided Access

5 Coronary Anatomy & Left Heart Catheterization

5.1 Coronary Anatomy & Dominance 5.2 Left Heart Catheterization & Coronary Angiography 5.3 Catheter Shapes & Angiographic Projections 5.4 Left Ventriculography & Aortography

6 Right Heart Catheterization & Hemodynamics

6.1 Right Heart Waveforms & Pressures 6.2 Cardiac Output: Fick & Thermodilution 6.3 Valve Area, Bernoulli & Shunt Calculations 6.4 Endomyocardial Biopsy Support

7 ECG & Intracardiac/Intravascular Imaging

7.1 12-Lead ECG in the Cath Lab 7.2 Intracardiac Echocardiography (ICE) 7.3 IVUS & OCT 7.4 Identifying Structures on Cardiac Ultrasound

8 Physiology Assessment & Radiation Safety

8.1 FFR & iFR Studies 8.2 Radiation Safety & ALARA 8.3 Radiographic Equipment Operation 8.4 Image Acquisition & Interpretation

9 Cath Lab Pharmacology

9.1 Anticipating Pharmacologic Needs 9.2 Recognizing Medication Effects & Adverse Reactions 9.3 Anticoagulation & ACT Targets

10 PCI Fundamentals

10.1 Balloon Angioplasty & Stent Platforms 10.2 Coronary Wiring, Lesion Prep & Flow Grades 10.3 Vascular Closure Devices

11 Chapter 11: Complex Coronary Intervention

11.1 CTO, Atherectomy & Alcohol Septal Ablation 11.2 Cardiac Thrombectomy (Mechanical & Aspiration) 11.3 Distal Protection Devices

12 Device Implants & Temporary Support

12.1 Pacemaker, ICD, CRT & Loop Recorders 12.2 Transvenous Temporary Pacemaker 12.3 Pericardiocentesis Support

13 Mechanical Circulatory Support

13.1 IABP & Impella 13.2 ECMO & LVAD Overview for the Cath Lab

14 Peripheral Vascular Interventions

14.1 Peripheral Interventions 14.2 Complex Peripheral Interventions 14.3 Peripheral Thrombectomy & Thrombolysis

15 Structural Heart Interventions

15.1 Nonvalvular Structural Therapies 15.2 Percutaneous Valve Interventions

16 Chapter 16: Emergency Clinical Presentations & Equipment

16.1 STEMI, Cardiogenic Shock & Pulmonary Edema 16.2 Stroke & Neurologic Emergencies in the Lab 16.3 Emergency Equipment Operation

17 Intraprocedural Complications

17.1 Coronary Interventional Complications 17.2 Peripheral & Structural Complications

18 Post-Procedural Activities

18.1 Arterial & Venous Hemostasis 18.2 Access-Site Complications 18.3 Patient Education, Reporting & Transfer of Care

2.1 Procedure & Control Room Setup

Key Takeaways Explain This Topic AI

CCI Domain A Task 1 requires verifying procedure-room and control-room readiness before every case — equipment function, radiation safety, and emergency supplies are non-negotiable

Daily fluoroscopy QC includes beam alignment, collimation, kVp/mAs consistency, and dose-display accuracy; out-of-tolerance results require service before clinical use

Dual dosimetry (collar outside lead, waist inside lead) must be worn and logged; ALARA is applied through Time, Distance, and Shielding before the first image

Hemodynamic transducers are zeroed to atmospheric pressure and leveled at the phlebostatic axis (4th intercostal space, mid-axillary line) before pressure monitoring

Emergency trays — crash cart, defibrillator, pericardiocentesis kit, IABP/Impella supplies — are checked, charged, and positioned within arm's reach before patient entry

Last updated: July 2026

Procedure & Control Room Setup

Quick Answer: Before any patient enters the cath lab, the RCIS verifies that the procedure room, control room, fluoroscopy system, hemodynamic monitors, contrast injectors, and emergency equipment are functional, calibrated, and positioned for safe case flow. Radiation QC/QA checks and ALARA readiness are part of the same pre-case checklist — not optional add-ons.

Pre-procedural room and equipment setup maps directly to CCI Domain A, Task 1 : prepare the procedure room and control room, confirm radiation equipment readiness, and perform quality control/quality assurance checks. Although the pre-procedural domain is only 8% of the RCIS exam matrix, setup errors propagate into every subsequent task — a mis-leveled transducer skews hemodynamics, an uncalibrated dose display violates ALARA documentation, and a missing crash-cart pad costs seconds in cardiac arrest. The RCIS owns the operational readiness of the environment even when a biomedical engineer maintains the equipment long-term.

Cath Lab Zones: Procedure Room vs. Control Room

Modern cardiac catheterization suites are divided into distinct functional zones. Understanding what belongs where prevents contamination, radiation exposure, and workflow bottlenecks.

Zone Primary Function Key Equipment RCIS Responsibilities

Procedure room Sterile field, patient care, fluoroscopy Angiography table, C-arm/ceiling mount, hemodynamic console, contrast injector, sterile supplies Verify table locks, radiation shields, monitor cables, sterile back-table layout

Control room Image acquisition, recording, dose monitoring Fluoroscopy console, hemodynamic recorder, PACS workstation, intercom Confirm image display, recording mode, patient ID linkage, dose tracking active

Scrub area Hand hygiene and gowning Scrub sinks, gown/glove dispensers, trash receptacles Ensure soap, towels, and PPE stocked before first scrub

Equipment storage Non-sterile inventory Catheter carts, guidewire racks, closure devices Restock case-specific trays; verify lot numbers and expiration dates

The procedure room must be cleaned and decontaminated between cases per institutional infection-control policy. High-touch surfaces — table rails, control panels, injector keypads — are wiped with an EPA-registered disinfectant. The control room is a non-sterile zone; personnel who circulate between zones must observe hand hygiene when returning to the sterile field.

Fluoroscopy & Imaging System Readiness

The fluoroscopy/angiography system is the central tool of invasive cardiology. Pre-case checks go beyond "does it turn on?" — they confirm image quality, safety interlocks, and dose reporting.

Pre-case fluoroscopy checklist:

Power-on self-test (POST) — System completes initialization without error codes; tube and detector temperatures within operating range.

Image quality — Test pattern or phantom image shows uniform brightness, no dead pixels, adequate resolution for vessel detail.

Collimation — Manual and automatic collimators open/close symmetrically; last-image-hold (LIH) and stored fluoroscopy loops function.

Pedal function — Fluoroscopy and cine pedals respond; foot-switch lockout prevents unintended exposure.

Table and C-arm movement — Table tilt, cradle, and gantry rotation move smoothly; locks engage; no cable pinch points.

Dose display — Cumulative air kerma (CAK) and kerma-area product (KAP/PKA) reset to zero for the new case; dose-rate display active.

Recording — Cine and DSA modes capture to PACS; patient demographics match the scheduled case.

Flat-panel detectors have largely replaced image intensifiers, but RCIS candidates should know both: image intensifiers use input phosphor and electron optics; flat panels use direct or indirect digital conversion with higher dynamic range and lower patient dose at equivalent signal.

Radiation QC/QA Program

Regulatory and accreditation bodies (The Joint Commission, state radiation control agencies, FDA under 21 CFR 1020.30) require a documented quality control (QC) program for fluoroscopic equipment. Quality assurance (QA) is the broader management process ensuring QC results are reviewed, trends tracked, and corrective actions taken.

  • QC Test Frequency Acceptance Criteria (Typical) Action if Failed
  • Beam alignment & central ray Daily or per manufacturer ±2% of expected field size Service before clinical use
  • kVp accuracy Monthly ±5% of set value Recalibrate or service
  • mAs linearity Monthly ±10% across range Service
  • Dose-rate display accuracy Annual (physicist) ±20% per IEC 60601 Adjust or replace display module
  • High-contrast resolution Semi-annual Meets manufacturer spec Detector service
  • Low-contrast detectability Semi-annual Meets manufacturer spec Detector service
  • Lead apron integrity (fluoroscopy) Annual No cracks >2 mm on fluoroscopy Replace apron

The RCIS participates in daily constancy checks — verifying image quality, collimation, and dose-display reset — and reports anomalies to the radiation safety officer (RSO) or medical physicist. ALARA (As Low As Reasonably Achievable) is applied from the first moment of setup: ceiling-suspended lead shields, table-mounted skirts, and portable barriers are positioned before the patient arrives. Personnel wear 0.5 mm Pb-equivalent aprons minimum, thyroid collars, and leaded glasses when available.

Dosimetry: Best practice uses dual badges — one at the collar (outside the lead apron) to measure dose to the eyes, thyroid, and unshielded head, and one at the waist (under the apron) to assess whole-body effective dose. Ring dosimeters may be required for high-volume operators. Badges are exchanged monthly (or per program policy) and must never be shared or left in the procedure room between cases.

Hemodynamic, Injector & Support Systems

Pressure monitoring accuracy depends on transducer setup performed before the first waveform appears on screen.

Transducer zeroing and leveling protocol:

Close the stopcock to the patient and open to air (atmosphere).

Press Zero on the monitor — the waveform should flatline at 0 mmHg.

Level the transducer stopcock at the phlebostatic axis : 4th intercostal space, mid-axillary line (same height as the right atrium in supine patients).

Open to the patient and confirm a physiologic waveform (RA ~2–6 mmHg, PA ~15–30/4–12 mmHg).

The power contrast injector is checked for syringe size, contrast type (low-osmolar or iso-osmolar non-ionic preferred), air-bubble detection, pressure-limit settings, and IV connection integrity. Hemodynamic consoles, 12-lead ECG, pulse oximetry, and non-invasive blood pressure cuffs are verified against the patient monitor module.

Emergency Equipment Verification

Every cath lab case carries risk of ventricular fibrillation, tamponade, allergic reaction, and vascular catastrophe. Emergency equipment is verified before the patient enters — not retrieved after an event begins.

Emergency Resource Pre-Case Check

Crash cart / code cart Defibrillator charged; pads connected; epinephrine, amiodarone, atropine, calcium, sodium bicarbonate present and unexpired

  • Defibrillator Self-test passed; biphasic energy available (200 J typical first shock for VF/pulseless VT)
  • Pericardiocentesis tray Needle, guide wire, dilator, pigtail catheter, drainage bag ready
  • Temporary pacing kit Pacing generator tested; transvenous leads available

IABP / mechanical support Console self-test; helium tank pressure adequate; correct size balloon on cart for planned case

Airway supplies Ambu bag, suction, supplemental O₂ flow confirmed

The RCIS documents completion of the pre-case checklist per institutional policy. Any equipment deficiency is resolved or the case is delayed until the deficiency is corrected — patient safety overrides schedule pressure.