4.1 Gas Chromatography and Mass Spectrometry
CIH Certified Industrial Hygienist Exam
Track your progress as you read
Finished sections get a check mark.
1 Chapter 1: The CIH Exam and How to Use This Guide
1.1 CIH Exam Format, Scoring, and Test-Day Logistics 1.2 Eligibility, Application Windows, Fees, and Certification Maintenance 1.3 The BGC Blueprint, Subject Areas, and a Calculation-First Study Plan
2 Chapter 2: Basic Science & Calculations
2.1 Gas Laws and Vapor Calculations 2.2 Unit Conversions: ppm, mg/m³, and Standard Temperature/Pressure 2.3 Fluid Mechanics and Airflow Fundamentals
3 Chapter 3: Air Sampling & Instrumentation
3.1 Active and Passive Air Sampling Strategies 3.2 Particulate and Aerosol Sampling 3.3 Direct-Reading Instruments and Calibration
4 Chapter 4: Analytical Chemistry & Lab Methods
4.1 Gas Chromatography and Mass Spectrometry 4.2 High-Performance Liquid Chromatography and Ion Chromatography 4.3 Spectroscopy and Elemental Analysis 4.4 Gravimetric, Microscopy, and Wet Chemical Methods 4.5 Laboratory Quality Assurance, Chain of Custody, and Desorption Efficiency
5 Chapter 5: Biostatistics & Epidemiology
5.1 Lognormal Distributions and Exposure Data Statistics 5.2 Upper Confidence Limits (UCL95) and Compliance Decision Rules 5.3 Epidemiological Study Designs and Measures of Association 5.4 Sampling Strategy: Sample Size, Power, and Worker Selection
6 Chapter 6: Toxicology & Biological Monitoring
6.1 Dose-Response Relationships and Points of Departure 6.2 Toxicokinetics: Absorption, Distribution, Metabolism, and Elimination 6.3 Aerosol Deposition, Clearance, and Respiratory Tract Dosimetry 6.4 Hepatic, Renal, Neurologic, and Hematologic Target Organ Toxicity 6.5 Pulmonary Toxicology, Carcinogenicity, and Reproductive Hazards 6.6 OEL Frameworks, Classifications, and Excursion Limits 6.7 Chemical Mixtures and Extended Work-Shift OEL Adjustments 6.8 Biological Exposure Indices (BEIs) and Biomonitoring
7 Chapter 7: Biohazards & Indoor Environmental Quality
7.1 Mold, Fungi, and Microbial Contaminants 7.2 Infectious Agents, Bloodborne Pathogens, and Biosafety Levels 7.3 Indoor Environmental Quality (IEQ) Investigations 7.4 Sanitation, Potable Water, and Personal Hygiene Programs
8 Chapter 8: Community Exposure & Environmental Health
8.1 Clean Air Act Framework, NAAQS, and New Source Review 8.2 Hazardous Air Pollutants, CEMS, and EPA Reference Methods 8.3 Atmospheric Stability, Lapse Rates, and Plume Morphology 8.4 Gaussian Plume Modeling, Plume Rise, and Environmental Fate 8.5 HAZWOPER Scope, Mandates, and Training Tiers 8.6 Protection Levels, Site Zoning, Incident Command, and Decontamination 8.7 Water Pollution, Hazardous Waste, and Community Right-to-Know
9 Chapter 9: Health Risk Analysis & Hazard Communication
9.1 GHS Hazard Classification: Health and Physical Hazard Criteria 9.2 GHS Labels, Safety Data Sheets, and Trade Secret Rules 9.3 The AIHA Exposure Assessment Strategy and Similar Exposure Groups 9.4 Qualitative Exposure Profiling and the AIHA Risk Matrix 9.5 Quantitative Health Risk Characterization: HQ, HI, and ELCR 9.6 NIOSH Occupational Exposure Banding 9.7 Risk Communication and Reporting to Affected Parties
10 Chapter 10: Industrial Ventilation Principles & LEV Design
10.1 Elimination, Substitution, Isolation, Containment, and Process Change 10.2 LEV Hood Taxonomy and Capture Velocity Selection 10.3 Hood Airflow Equations, Entry Losses, and Hood Static Pressure 10.4 Particulate Transport Aerodynamics and Duct Sizing 10.5 Duct Friction, Fitting Losses, and System Balancing 10.6 Fan Categories and Pressure Dynamics Across the Fan 10.7 System Curves, the Fan Laws, and Density Corrections
11 Chapter 11: General Ventilation, Air Cleaning & Diagnostics
11.1 Dilution Ventilation Principles and Steady-State Modeling 11.2 Transient Purging and Air Distribution Geometry 11.3 Particulate Air Cleaning: Cyclones, Baghouses, Scrubbers, and ESPs 11.4 Gaseous Emission Control and Air Cleaner Selection 11.5 Ventilation System Troubleshooting and Testing
12 Chapter 12: Noise Measurement & Hearing Conservation
12.1 Acoustic Physics and Decibel Mathematics 12.2 Sound Propagation, Octave Bands, and Frequency Weighting 12.3 Noise Dose Metrics and OSHA versus ACGIH Criteria 12.4 Exchange Rates, Extended Shifts, and Dosimetry Field Protocols 12.5 Engineering Noise Controls and Hearing Protector Attenuation 12.6 Hearing Conservation Programs and Standard Threshold Shifts 12.7 Hand-Arm and Whole-Body Vibration
13 Chapter 13: Thermal Stress Assessment & Control
13.1 Wet Bulb Globe Temperature (WBGT) and Heat Stress Indices 13.2 Heat-Related Disorders, Acclimatization, and Work/Rest Regimens 13.3 Cold Stress Evaluation and Wind Chill
14 Chapter 14: Ionizing Radiation Protection
14.1 Ionizing Radiation Physics and Radioactive Decay Kinetics 14.2 Radiation Units, Radiobiology, and Health Effects 14.3 ALARA, Stay-Time, and the Inverse Square Law 14.4 Shielding Physics, Material Selection, and Gamma Constants 14.5 Radiation Detection Instruments and Spectrometry 14.6 Personnel Dosimetry, Dose Limits, and Contamination Surveys
15 Chapter 15: Non-Ionizing Radiation & Laser Safety
15.1 The Non-Ionizing Spectrum and Static/ELF Field Exposures 15.2 Radiofrequency and Microwave Exposure: SAR, Standards, and Surveys 15.3 Ultraviolet, Visible, and Infrared Radiation Hazards 15.4 Laser Safety Classifications and Optical Density Calculations 15.5 Illumination, Lighting Design, and Visual Performance
16 Chapter 16: Non-Engineering Controls & Personal Protective Equipment
16.1 Respiratory Protection Programs and Assigned Protection Factors 16.2 Maximum Use Concentration, Oxygen Deficiency, and Breathing Air 16.3 Particulate Filters, Sorbent Cartridges, and Change Schedules 16.4 Respirator Fit Testing Protocols 16.5 Chemical Protective Clothing and Dermal Exposure Control
17 Chapter 17: Ergonomics & Workstation Design
17.1 Soft-Tissue Biomechanics and WMSD Pathophysiology 17.2 Ergonomic Risk Factors and Their Interaction Effects 17.3 The Revised NIOSH Lifting Equation 17.4 RULA, REBA, the Strain Index, and the ACGIH HAL TLV 17.5 Anthropometry and Workstation Geometry 17.6 VDT Workstations, Hand Tool Design, and Administrative Ergonomics
18 Chapter 18: Work Environments & Industrial Processes
18.1 Welding Fume Emissions, Metal Fume Fever, and Cr(VI)/Manganese Toxicology 18.2 Foundry Operations and Hot Work Fire Prevention 18.3 Vapor Degreasing, Pickling, Electroplating, and Cyanide Baths 18.4 Spray Coatings, Isocyanates, and Semiconductor Manufacturing 18.5 Confined Spaces and High-Hazard Industrial Operations 18.6 Asbestos, Lead, and Silica: Regulated Substance Programs 18.7 Abrasive Blasting and Surface Preparation
19 Chapter 19: IH Program Management & Professional Ethics
19.1 The BGC Code of Ethics: Canons and Enforcement 19.2 Professional Liability, Standard of Care, and Expert Testimony 19.3 Occupational Health & Safety Management Systems (ISO 45001, ANSI Z10) 19.4 Exposure Records Access and Medical Surveillance 19.5 OSHA Injury and Illness Recordkeeping and Enforcement
4.1 Gas Chromatography and Mass Spectrometry
Key Takeaways Explain This Topic AI
Gas Chromatography (GC) separates volatile organic compounds based on differential partitioning between a gaseous mobile phase (He, N2, or H2) and a stationary phase liquid film inside a capillary column.
Flame Ionization Detectors (FID) offer universal response and wide linear dynamic range (10^7) for hydrocarbons, while Electron Capture Detectors (ECD) provide extreme sensitivity (picogram level) specifically for halogenated compounds.
Mass Spectrometry operates standard electron ionization (EI) at 70 eV to yield reproducible fragmentation spectra; Selected Ion Monitoring (SIM) improves sensitivity 10- to 100-fold over Total Ion Chromatogram (TIC) scanning.
Carbon disulfide (CS2) is the standard solvent for charcoal tube desorption due to high desorption efficiency (>90%) and near-zero response on FID, whereas Thermal Desorption (TD) achieves ppb/ppt sensitivity without solvent dilution.
Front and back sorbent tube sections must be analyzed separately; analyte mass on the back section exceeding 10% of the front section indicates breakthrough and invalidates sample quantification unless adjusted per standard protocols.
Last updated: August 2026
3.1 Gas Chromatography and Mass Spectrometry
Gas chromatography (GC) coupled with specialized detectors represents the primary analytical backbone of industrial hygiene for identifying and quantifying volatile and semi-volatile organic compounds (VOCs and SVOCs) in workplace air. An industrial hygienist must understand both the physical principles governing chromatographic separation and the operating characteristics of different detector systems to properly select sampling media, interpret laboratory analytical reports, and evaluate data quality.
1. Principles of Gas Chromatographic Separation
Gas chromatography separates complex chemical mixtures based on the differential partitioning of vaporized analytes between a flowing gaseous mobile phase (carrier gas) and an immobilized stationary phase coated onto the inner walls of a capillary column or packed within a tube. [Carrier Gas Supply] ──> [Injector / Splitter] ──> [Capillary Column in Oven] ──> [Detector (FID/ECD/MS)] ──> [Data System]
- (He, N2, H2) (Flash Vaporization) (Controlled Temp Ramp) (Signal Transduction) (Chromatogram)
- Carrier Gases (Mobile Phase)
The carrier gas must be chemically inert toward both the sample analytes and the stationary phase. Common carrier gases include helium (He) , nitrogen (N₂) , and hydrogen (H₂) . Their operational performance is governed by the Van Deemter Equation , which relates the Height Equivalent to a Theoretical Plate (H or HETP) to the linear carrier gas velocity (u): H = A + B u + C ⋅ u H = A + \frac{B}{u} + C \cdot u H = A + u B + C ⋅ u
Where:
- A = Eddy diffusion (multipath effect; negligible in open-tubular capillary columns)
- B = Longitudinal molecular diffusion of analyte in the mobile phase
- C = Resistance to mass transfer in both the stationary and mobile phases
- u = Linear velocity of the carrier gas (cm/s)
Carrier Gas Optimum Linear Velocity (uopt) Advantages Limitations / Operational Concerns
Helium (He) 30 - 40 cm/s Excellent compromise of high chromatographic efficiency and speed; non-flammable; safe standard for GC-MS. Finite global supply; high operational cost; requires 99.999% purity with moisture/oxygen traps.
Hydrogen (H₂) 40 - 50 cm/s Flattest Van Deemter curve at high linear velocities; enables the fastest run times without loss of chromatographic resolution. Flammability and explosive risk (lower explosive limit 4.0%); requires hydrogen generators or in-oven leak sensors.
Nitrogen (N₂) 10 - 15 cm/s Lowest minimum plate height (Hmin) at low velocities; highly economical and readily available. Extremely steep efficiency loss at velocities above 15 cm/s; results in excessively long chromatographic run times.
Stationary Phase Chemistry
Modern industrial hygiene laboratories utilize Wall-Coated Open Tubular (WCOT) fused-silica capillary columns. Standard dimensions range from 15 m to 60 m in length, 0.18 mm to 0.53 mm internal diameter (ID), with a stationary phase film thickness (df) between 0.10 µm and 5.0 µm.
The chemical rule of thumb "like dissolves like" dictates phase selection:
Non-Polar Phases (100% Dimethylpolysiloxane, e.g., DB-1, HP-1): Analytes elute predominantly in order of increasing boiling point . Excellent for aliphatic hydrocarbons, gasoline range organics, and non-polar solvents.
Intermediate / Semi-Polar Phases (5% Phenyl / 95% Dimethylpolysiloxane, e.g., DB-5, HP-5MS): The standard general-purpose column for environmental and industrial hygiene VOC/SVOC screening and GC-MS analysis.
Polar Phases (Polyethylene Glycol / Carbowax, e.g., DB-WAX): Strongly retain polar molecules via hydrogen bonding and dipole-dipole interactions. Ideal for free alcohols, glycols, aldehydes, and aromatic isomer separations (e.g., separating meta- and para-xylene).
Key Chromatographic Parameters
Retention Time (tR): The total elapsed time from sample injection to the apex of the analyte detector peak. Dead time (t0 or tM) is the transit time required for an unretained compound (e.g., methane or air) to traverse the column.
Retention Factor (k'): A dimensionless measure of column retention independent of column length and flow rate: k ′ = t R − t 0 t 0 k' = \frac{t_R - t_0}{t_0} k ′ = t 0 t R − t 0 Ideal analytical values for k' range between 2 and 10.
Selectivity / Separation Factor (α): The relative retention of two adjacent peaks: α = k 2 ′ k 1 ′ = t R 2 − t 0 t R 1 − t 0 ( where α > 1.0 ) \alpha = \frac{k'_2}{k'_1} = \frac{t_{R2} - t_0}{t_{R1} - t_0} \quad (\text{where } \alpha > 1.0) α = k 1 ′ k 2 ′ = t R 1 − t 0 t R 2 − t 0 ( where α > 1.0 )
Column Efficiency (N) & Plate Height (H): N = 16 ( t R W ) 2 = 5.545 ( t R W 1 / 2 ) 2 H = L N N = 16 \left(\frac{t_R}{W}\right)^2 = 5.545 \left(\frac{t_R}{W_{1/2}}\right)^2 \qquad H = \frac{L}{N} N = 16 ( W t R ) 2 = 5.545 ( W 1/2 t R ) 2 H = N L Where W is peak width at baseline, W(1/2) is peak width at half-maximum height, and L is column length.
Chromatographic Resolution (Rs): R s = 2 ( t R 2 − t R 1 ) W 1 + W 2 = 1.177 ( t R 2 − t R 1 ) W 1 / 2 , 1 + W 1 / 2 , 2 R_s = \frac{2 (t_{R2} - t_{R1})}{W_1 + W_2} = \frac{1.177 (t_{R2} - t_{R1})}{W_{1/2,1} + W_{1/2,2}} R s = W 1 + W 2 2 ( t R 2 − t R 1 ) = W 1/2 , 1 + W 1/2 , 2 1.177 ( t R 2 − t R 1 ) A resolution of Rs ≥ 1.5 represents complete baseline separation between two adjacent peaks.
Isothermal vs. Temperature-Programmed GC
Isothermal Analysis: Column oven is maintained at a constant temperature. Suitable only for mixtures containing compounds with very narrow boiling point ranges. For wide-boiling mixtures, early-eluting peaks suffer from poor resolution while late-eluting peaks exhibit extreme band broadening and excessively long run times (the general elution problem ).
Temperature Programming: Oven temperature is linearly increased during the chromatographic run (e.g., hold at 40°C for 3 min, ramp at 10°C/min to 260°C, hold for 5 min). This sharpens late peaks, decreases detection limits, and compresses total analysis time.
2. Gas Chromatography Detectors in Industrial Hygiene
Industrial hygiene methods match specific detectors to target analytes based on chemical composition, sensitivity requirements, and potential matrix interferences.
Detector Type Principle of Operation Selectivity & Target Analytes Linear Range Destructive?
Flame Ionization Detector (FID) Pyrolysis of organic carbon in an H2 / Air diffusion flame forming CHO⁺ ions and electrons; measured as a picoamp current. Universal for organic compounds containing carbon-carbon or carbon-hydrogen bonds. Insensitive to CO, CO2, CS2, H2O, N2, O2, SO2. 10⁷ (widest dynamic range) Yes
Electron Capture Detector (ECD) Radioactive ⁶³Ni beta source ionizes make-up gas creating a standing current; electronegative molecules capture thermal electrons, reducing current. Highly selective for electronegative species: halogenated solvents (TCE, PCE, CCl4, CHCl3), PCBs, organochlorine pesticides, and SF6 tracer gas. 10³ - 10⁴ (narrow; sensitive to contamination) No
Photoionization Detector (PID) Ultraviolet lamp (9.8, 10.6, or 11.7 eV) ionizes molecules having an Ionization Potential (IP) lower than the photon energy (R + hν → R⁺ + e⁻). Selective for compounds with IP ≤ lamp energy: aromatics (benzene, toluene, xylene), olefins, chlorinated alkenes, amines. Alkanes (< C4) are not detected by 10.6 eV. 10⁵ - 10⁶ No
Mass Spectrometer (MS) High-vacuum electron ionization fragmentation followed by mass-to-charge separation in a quadrupole or time-of-flight analyzer. Universal identification and targeted quantification for complex multi-analyte industrial solvent mixtures, unknowns, and regulatory compliance. 10⁵ - 10⁶ Yes
3. Mass Spectrometry (GC-MS) Fundamentals
Gas chromatography coupled to mass spectrometry (GC-MS) combines the physical separation power of capillary GC with the absolute structural identification and quantification capabilities of mass spectrometry.
Electron Ionization (EI)
Standard Ionization Energy: The universal standard for GC-MS analytical libraries is 70 eV electron impact ionization.
Mechanism: High-energy electrons emitted from a heated rhenium or tungsten filament bombard vaporized analyte molecules (M) exiting the GC transfer line under high vacuum (10⁻⁵ to 10⁻⁶ Torr): M + e − ( 70 eV ) ⟶ M + ∙ + 2 e − M + e^- (70 \text{ eV}) \longrightarrow M^{+\bullet} + 2e^- M + e − ( 70 eV ) ⟶ M +∙ + 2 e −
Fragmentation: Because 70 eV substantially exceeds the covalent bond energies of organic molecules (3 - 10 eV), extensive, highly reproducible bond cleavage occurs, generating a characteristic "fingerprint" distribution of fragment ions.
Library Matching: Spectra generated at 70 eV are directly comparable against standardized reference databases (NIST, EPA/NIH, Wiley).
Mass Analyzers & Operational Acquisition Modes
Quadrupole mass filters separate ions based on their mass-to-charge ratio (m/z) using oscillating radiofrequency (RF) and direct-current (DC) electric fields. [GC Effluent] ──> [Ion Source (EI @ 70 eV)] ──> [Quadrupole Rods (RF/DC Filter)] ──> [Electron Multiplier Detector]
M + e- -> M+• + 2e- Filters specific m/z ions Generates signal pulse
Full Scan / Total Ion Chromatogram (TIC):
Scans a continuous mass range (e.g., m/z 35 to 450) continuously throughout the run.
Produces complete fragmentation spectra for every chromatographic peak.
Application: Screening for unknown contaminants, identifying unexpected solvent adulterants, and untargeted indoor environmental quality (IEQ) investigations.
Selected Ion Monitoring (SIM):
The quadrupole is programmed to monitor only 2 to 4 predetermined characteristic ions (a quantifier ion and one or more qualifier ions ) at specific retention time windows.
Dwell time on target ions is dramatically increased while ignoring matrix background ions.
Performance: Achieves 10- to 100-fold higher sensitivity (signal-to-noise ratio) compared to full scan mode.
Application: Ultra-trace targeted compliance monitoring (e.g., sub-ppb quantification of benzene, 1,3-butadiene, vinyl chloride, or ethylene oxide).
4. Sample Preparation: Solvent Extraction vs. Thermal Desorption
Airborne organic vapors captured on solid sorbent sampling tubes must be released into the gas phase for GC analysis.
Solvent Desorption (NIOSH Method 1500/1501 Series)
Solid Sorbent: Activated coconut shell charcoal (typically 100 mg front section, 50 mg back section separated by polyurethane foam).
Desorption Solvent: Carbon disulfide (CS2) is the classic solvent of choice for non-polar organics because:
It provides exceptionally high desorption efficiency (DE > 90%) for aliphatic and aromatic hydrocarbons.
It has a low boiling point (46.3°C).
Crucially, CS2 produces virtually zero response in an FID , preventing solvent peak masking of early-eluting target analytes.
Polar Modifiers: For polar analytes collected on silica gel or specialized sorbents, solvent mixtures such as 1% - 5% methanol or N,N-dimethylformamide (DMF) in methylene chloride or CS2 are required to overcome polar adsorption forces.
Desorption Efficiency (DE): The fraction of analyte recovered from the sorbent matrix: D E = Mass Recovered from Spiked Media Mass Spiked onto Media DE = \frac{\text{Mass Recovered from Spiked Media}}{\text{Mass Spiked onto Media}} D E = Mass Spiked onto Media Mass Recovered from Spiked Media M corrected = M measured − M blank D E M_{\text{corrected}} = \frac{M_{\text{measured}} - M_{\text{blank}}}{DE} M corrected = D E M measured − M blank DE curves must be determined across multiple loading levels; standard NIOSH guidelines require DE ≥ 0.75 (75%).
Thermal Desorption (TD) (EPA Methods TO-17, NIOSH 2549)
Solid Sorbent: Specially packed multi-bed tubes (e.g., Tenax TA, Carbopack, Carboxen).
Mechanism: Sorbent tubes are heated to 250°C - 350°C while being swept with inert carrier gas. Desorbed analytes are focused onto an electrically cooled Peltier cryo-trap (-30°C to -100°C), which is then flash-heated to inject a narrow plug onto the GC column.
Comparison:
Solvent Desorption: Allows multiple replicate injections from the 1.0 mL extract; sample can be re-analyzed; but dilutes the sample (reducing absolute method sensitivity) and generates hazardous CS2 waste.
Thermal Desorption: 100% of the collected analyte is transferred to the analytical column, yielding sub-ppb detection limits (1,000x more sensitive); zero solvent toxicity; but historically single-shot (destructive) unless automated split/re-collection tubes are used.
Breakthrough Evaluation in Sorbent Sampling
Standard solid sorbent tubes consist of a primary front section (A) and a secondary backup section (B) . Breakthrough (%) = Mass on Backup Section (B) Mass on Front Section (A) + Mass on Backup Section (B) × 100 % \text{Breakthrough (\%)} = \frac{\text{Mass on Backup Section (B)}}{\text{Mass on Front Section (A)} + \text{Mass on Backup Section (B)}} \times 100\% Breakthrough (%) = Mass on Front Section (A) + Mass on Backup Section (B) Mass on Backup Section (B) × 100%
Breakthrough Rule: If MassB > 0.10 × MassA (i.e., > 10% on the backup section), the sorbent capacity was exceeded during sampling, indicating probable loss of analyte into the sampling pump. The true workplace concentration was underestimated, and the sample is legally invalid for compliance defense.
5. Worked Example: Quantitative GC-FID Air Exposure Calculation
Worked Problem: An industrial hygienist samples for airborne toluene (Molecular Weight = 92.14 g/mol) in a paint spray booth. An air sample is collected using a standard coconut shell charcoal tube (100/50 mg) at a calibrated flow rate of 0.200 L/min for 4.0 hours (240 minutes). Laboratory analysis via GC-FID with CS2 desorption yields the following data:
- Front Section Mass (MA): 1.420 mg
- Backup Section Mass (MB): 0.035 mg
- Field Blank Mass (Mblank): 0.005 mg
- Desorption Efficiency (DE) at this loading: 0.94 (94%)
- Sampling Conditions: 25°C, 1 atm (760 mmHg)
Tasks:
Determine if breakthrough occurred.
Calculate the total corrected mass of toluene collected (Mcorr in mg).
Calculate the sampled air volume in liters and cubic meters.
Determine the 4-hour airborne toluene concentration in mg/m³ and ppm.
Step-by-Step Solution:
Step 1: Check for Sorbent Breakthrough M B M A = 0.035 mg 1.420 mg = 0.0246 = 2.46 % \frac{M_B}{M_A} = \frac{0.035 \text{ mg}}{1.420 \text{ mg}} = 0.0246 = 2.46\% M A M B = 1.420 mg 0.035 mg = 0.0246 = 2.46% Because 2.46% < 10%, no significant breakthrough occurred. The backup section successfully captured migration, and the total mass can be reliably quantified.
Step 2: Calculate Corrected Analyte Mass (Mcorr) Total measured mass: Mtotal = MA + MB = 1.420 mg + 0.035 mg = 1.455 mg
Blank-corrected mass: Mnet = 1.455 mg - 0.005 mg = 1.450 mg
Applying Desorption Efficiency (DE): M c o r r = M n e t D E = 1.450 mg 0.94 = 1.5426 mg M_{corr} = \frac{M_{net}}{DE} = \frac{1.450 \text{ mg}}{0.94} = 1.5426 \text{ mg} M corr = D E M n e t = 0.94 1.450 mg = 1.5426 mg
Step 3: Calculate Air Sampling Volume (V) V = Flow Rate × Time = 0.200 L/min × 240 min = 48.0 L = 0.0480 m 3 V = \text{Flow Rate} \times \text{Time} = 0.200 \text{ L/min} \times 240 \text{ min} = 48.0 \text{ L} = 0.0480 \text{ m}^3 V = Flow Rate × Time = 0.200 L/min × 240 min = 48.0 L = 0.0480 m 3
Step 4: Calculate Airborne Concentration (C) In milligrams per cubic meter: C = M c o r r V = 1.5426 mg 0.0480 m 3 = 32.14 mg/m 3 C = \frac{M_{corr}}{V} = \frac{1.5426 \text{ mg}}{0.0480 \text{ m}^3} = 32.14 \text{ mg/m}^3 C = V M corr = 0.0480 m 3 1.5426 mg = 32.14 mg/m 3 Converting to parts per million (ppm) at 25°C and 1 atm: ppm = C ( mg/m 3 ) × 24.45 MW = 32.14 × 24.45 92.14 = 785.82 92.14 = 8.53 ppm \text{ppm} = \frac{C (\text{mg/m}^3) \times 24.45}{\text{MW}} = \frac{32.14 \times 24.45}{92.14} = \frac{785.82}{92.14} = 8.53 \text{ ppm} ppm = MW C ( mg/m 3 ) × 24.45 = 92.14 32.14 × 24.45 = 92.14 785.82 = 8.53 ppm
Conclusion: The worker's 4-hour exposure is 32.14 mg/m³ (8.53 ppm).