Study guide
4.5 Laboratory Quality Assurance, Chain of Custody, and Desorption Efficiency
CIH Certified Industrial Hygienist Exam
Track your progress as you read
Finished sections get a check mark.
1 Chapter 1: The CIH Exam and How to Use This Guide
1.1 CIH Exam Format, Scoring, and Test-Day Logistics 1.2 Eligibility, Application Windows, Fees, and Certification Maintenance 1.3 The BGC Blueprint, Subject Areas, and a Calculation-First Study Plan
2 Chapter 2: Basic Science & Calculations
2.1 Gas Laws and Vapor Calculations 2.2 Unit Conversions: ppm, mg/m³, and Standard Temperature/Pressure 2.3 Fluid Mechanics and Airflow Fundamentals
3 Chapter 3: Air Sampling & Instrumentation
3.1 Active and Passive Air Sampling Strategies 3.2 Particulate and Aerosol Sampling 3.3 Direct-Reading Instruments and Calibration
4 Chapter 4: Analytical Chemistry & Lab Methods
4.1 Gas Chromatography and Mass Spectrometry 4.2 High-Performance Liquid Chromatography and Ion Chromatography 4.3 Spectroscopy and Elemental Analysis 4.4 Gravimetric, Microscopy, and Wet Chemical Methods 4.5 Laboratory Quality Assurance, Chain of Custody, and Desorption Efficiency
5 Chapter 5: Biostatistics & Epidemiology
5.1 Lognormal Distributions and Exposure Data Statistics 5.2 Upper Confidence Limits (UCL95) and Compliance Decision Rules 5.3 Epidemiological Study Designs and Measures of Association 5.4 Sampling Strategy: Sample Size, Power, and Worker Selection
6 Chapter 6: Toxicology & Biological Monitoring
6.1 Dose-Response Relationships and Points of Departure 6.2 Toxicokinetics: Absorption, Distribution, Metabolism, and Elimination 6.3 Aerosol Deposition, Clearance, and Respiratory Tract Dosimetry 6.4 Hepatic, Renal, Neurologic, and Hematologic Target Organ Toxicity 6.5 Pulmonary Toxicology, Carcinogenicity, and Reproductive Hazards 6.6 OEL Frameworks, Classifications, and Excursion Limits 6.7 Chemical Mixtures and Extended Work-Shift OEL Adjustments 6.8 Biological Exposure Indices (BEIs) and Biomonitoring
7 Chapter 7: Biohazards & Indoor Environmental Quality
7.1 Mold, Fungi, and Microbial Contaminants 7.2 Infectious Agents, Bloodborne Pathogens, and Biosafety Levels 7.3 Indoor Environmental Quality (IEQ) Investigations 7.4 Sanitation, Potable Water, and Personal Hygiene Programs
8 Chapter 8: Community Exposure & Environmental Health
8.1 Clean Air Act Framework, NAAQS, and New Source Review 8.2 Hazardous Air Pollutants, CEMS, and EPA Reference Methods 8.3 Atmospheric Stability, Lapse Rates, and Plume Morphology 8.4 Gaussian Plume Modeling, Plume Rise, and Environmental Fate 8.5 HAZWOPER Scope, Mandates, and Training Tiers 8.6 Protection Levels, Site Zoning, Incident Command, and Decontamination 8.7 Water Pollution, Hazardous Waste, and Community Right-to-Know
9 Chapter 9: Health Risk Analysis & Hazard Communication
9.1 GHS Hazard Classification: Health and Physical Hazard Criteria 9.2 GHS Labels, Safety Data Sheets, and Trade Secret Rules 9.3 The AIHA Exposure Assessment Strategy and Similar Exposure Groups 9.4 Qualitative Exposure Profiling and the AIHA Risk Matrix 9.5 Quantitative Health Risk Characterization: HQ, HI, and ELCR 9.6 NIOSH Occupational Exposure Banding 9.7 Risk Communication and Reporting to Affected Parties
10 Chapter 10: Industrial Ventilation Principles & LEV Design
10.1 Elimination, Substitution, Isolation, Containment, and Process Change 10.2 LEV Hood Taxonomy and Capture Velocity Selection 10.3 Hood Airflow Equations, Entry Losses, and Hood Static Pressure 10.4 Particulate Transport Aerodynamics and Duct Sizing 10.5 Duct Friction, Fitting Losses, and System Balancing 10.6 Fan Categories and Pressure Dynamics Across the Fan 10.7 System Curves, the Fan Laws, and Density Corrections
11 Chapter 11: General Ventilation, Air Cleaning & Diagnostics
11.1 Dilution Ventilation Principles and Steady-State Modeling 11.2 Transient Purging and Air Distribution Geometry 11.3 Particulate Air Cleaning: Cyclones, Baghouses, Scrubbers, and ESPs 11.4 Gaseous Emission Control and Air Cleaner Selection 11.5 Ventilation System Troubleshooting and Testing
12 Chapter 12: Noise Measurement & Hearing Conservation
12.1 Acoustic Physics and Decibel Mathematics 12.2 Sound Propagation, Octave Bands, and Frequency Weighting 12.3 Noise Dose Metrics and OSHA versus ACGIH Criteria 12.4 Exchange Rates, Extended Shifts, and Dosimetry Field Protocols 12.5 Engineering Noise Controls and Hearing Protector Attenuation 12.6 Hearing Conservation Programs and Standard Threshold Shifts 12.7 Hand-Arm and Whole-Body Vibration
13 Chapter 13: Thermal Stress Assessment & Control
13.1 Wet Bulb Globe Temperature (WBGT) and Heat Stress Indices 13.2 Heat-Related Disorders, Acclimatization, and Work/Rest Regimens 13.3 Cold Stress Evaluation and Wind Chill
14 Chapter 14: Ionizing Radiation Protection
14.1 Ionizing Radiation Physics and Radioactive Decay Kinetics 14.2 Radiation Units, Radiobiology, and Health Effects 14.3 ALARA, Stay-Time, and the Inverse Square Law 14.4 Shielding Physics, Material Selection, and Gamma Constants 14.5 Radiation Detection Instruments and Spectrometry 14.6 Personnel Dosimetry, Dose Limits, and Contamination Surveys
15 Chapter 15: Non-Ionizing Radiation & Laser Safety
15.1 The Non-Ionizing Spectrum and Static/ELF Field Exposures 15.2 Radiofrequency and Microwave Exposure: SAR, Standards, and Surveys 15.3 Ultraviolet, Visible, and Infrared Radiation Hazards 15.4 Laser Safety Classifications and Optical Density Calculations 15.5 Illumination, Lighting Design, and Visual Performance
16 Chapter 16: Non-Engineering Controls & Personal Protective Equipment
16.1 Respiratory Protection Programs and Assigned Protection Factors 16.2 Maximum Use Concentration, Oxygen Deficiency, and Breathing Air 16.3 Particulate Filters, Sorbent Cartridges, and Change Schedules 16.4 Respirator Fit Testing Protocols 16.5 Chemical Protective Clothing and Dermal Exposure Control
17 Chapter 17: Ergonomics & Workstation Design
17.1 Soft-Tissue Biomechanics and WMSD Pathophysiology 17.2 Ergonomic Risk Factors and Their Interaction Effects 17.3 The Revised NIOSH Lifting Equation 17.4 RULA, REBA, the Strain Index, and the ACGIH HAL TLV 17.5 Anthropometry and Workstation Geometry 17.6 VDT Workstations, Hand Tool Design, and Administrative Ergonomics
18 Chapter 18: Work Environments & Industrial Processes
18.1 Welding Fume Emissions, Metal Fume Fever, and Cr(VI)/Manganese Toxicology 18.2 Foundry Operations and Hot Work Fire Prevention 18.3 Vapor Degreasing, Pickling, Electroplating, and Cyanide Baths 18.4 Spray Coatings, Isocyanates, and Semiconductor Manufacturing 18.5 Confined Spaces and High-Hazard Industrial Operations 18.6 Asbestos, Lead, and Silica: Regulated Substance Programs 18.7 Abrasive Blasting and Surface Preparation
19 Chapter 19: IH Program Management & Professional Ethics
19.1 The BGC Code of Ethics: Canons and Enforcement 19.2 Professional Liability, Standard of Care, and Expert Testimony 19.3 Occupational Health & Safety Management Systems (ISO 45001, ANSI Z10) 19.4 Exposure Records Access and Medical Surveillance 19.5 OSHA Injury and Illness Recordkeeping and Enforcement
4.5 Laboratory Quality Assurance, Chain of Custody, and Desorption Efficiency
Key Takeaways Explain This Topic AI
Desorption efficiency is the fraction of a known spiked mass recovered from the sampling medium, and reported results must be corrected by dividing the analysed mass by the DE.
A desorption efficiency below about 75% signals a poor medium/solvent match and should trigger method reconsideration rather than a large correction factor.
Field blanks travel with the samples and are handled identically but draw no air; media blanks stay sealed; trip blanks address transport contamination.
Chain of custody documents unbroken possession from collection to analysis and is what makes a result legally defensible in enforcement or litigation.
Laboratory competence is demonstrated by AIHA-LAP accreditation and by proficiency testing such as the Industrial Hygiene Proficiency Analytical Testing program.
Last updated: August 2026
Laboratory Quality Assurance, Chain of Custody, and Desorption Efficiency
An exposure result is a legal and medical document. It may determine whether an employer is cited, whether a worker is removed from a job, or whether a plaintiff prevails twenty years later. The quality system described in this section is what separates a defensible number from a number that a competent opposing expert will destroy.
1. Desorption Efficiency
Sorbent sampling has an inherent loss: not all of the analyte adsorbed onto charcoal or silica gel comes back off in the desorbing solvent. Desorption efficiency (DE) quantifies that loss.
Determination
A known mass of analyte is spiked directly onto the same lot of sampling medium, allowed to equilibrate (typically overnight), then desorbed and analysed exactly as a field sample would be: DE = mass recovered from the spiked medium mass spiked \text{DE} = \frac{\text{mass recovered from the spiked medium}}{\text{mass spiked}} DE = mass spiked mass recovered from the spiked medium
DE is determined at several loading levels bracketing the expected field mass, because it is not constant across the working range — it typically falls at very low loadings where irreversible adsorption is proportionally more significant.
Correction
Every reported field mass must be corrected: Corrected mass = mass found by analysis DE \text{Corrected mass} = \frac{\text{mass found by analysis}}{\text{DE}} Corrected mass = DE mass found by analysis
Failing to apply the correction biases every result low , which is the direction that matters: it under-reports exposure.
Worked example. A charcoal tube is analysed and 0.180 mg of toluene is found on the front section. The laboratory reports a desorption efficiency of 0.90 for toluene on that lot. The corrected mass is 0.180 / 0.90 = 0.200 mg . If the sample volume was 12 L, the concentration is 0.200 mg / 0.012 m³ = 16.7 mg/m³ — an 11% difference from the uncorrected value.
Acceptance
A DE around 0.90 to 1.00 is typical for a well-matched analyte/sorbent/solvent combination. NIOSH methods generally expect DE at or above about 0.75 . Below that, the correction factor is doing too much work: small variations in DE propagate into large errors in the reported result, and the correct response is to reconsider the medium or the desorbing solvent, not to apply a factor of 2.
2. Blanks: Three Kinds, Three Purposes
Blank type How it is handled What it detects
Media (lot) blank Unopened medium from the same lot, sent directly to the lab Contamination present in the medium as manufactured
Field blank Opened at the sampling site, handled and transported identically to samples, no air drawn Contamination from handling, site atmosphere, storage, and transport
Trip blank Travels with the sample set but is never opened Contamination during shipping alone (used mainly for volatile organics)
Field blanks are the workhorse. The usual rule is at least 10% of the sample set, with a minimum of two and typically no more than ten per sample set. The mean field blank mass is subtracted from each sample mass before the DE correction is applied. A field blank carrying substantial analyte invalidates the set — you cannot subtract your way out of a contaminated sampling process.
Spiked (control) samples and duplicates complete the picture: spikes verify accuracy against a known quantity, and side-by-side duplicates verify precision.
3. Chain of Custody
Chain of custody is the documented, unbroken record of who possessed each sample from the moment of collection until analysis and disposal. A sample is "in custody" when it is in your physical possession, in your view, secured by you, or in a designated secure area.
A defensible chain of custody record contains:
- Unique sample identifier and collection date, time, and location
- Sampler's name and signature
- Analyte requested and analytical method specified
- Sample matrix, medium type, and lot number
- Volume sampled, or flow rate and duration
- Preservation applied and required holding time
- Every transfer , with relinquishing and receiving signatures, date, and time
Condition of the shipping container and seals on receipt at the laboratory +-------------------------------------------------------------+
- | COLLECTION -> SEALED -> SHIPPED -> RECEIVED -> ANALYSED |
- | sign seal sign sign sign |
- | |
- | A gap anywhere in this line is a gap in the defensibility |
- | of every number that comes out of the laboratory. |
- +-------------------------------------------------------------+
Custody seals on the shipping container let the receiving laboratory certify that the samples were not accessible in transit. A broken seal is documented as a discrepancy, not quietly ignored.
4. Preservation and Holding Times
Sample integrity degrades with time and temperature. Common requirements:
- Refrigeration for thermally labile analytes and for most bioaerosol and derivatised samples
- Protection from light for photodegradable analytes
- Prompt analysis of hexavalent chromium and derivatised isocyanate samples
Sealed, capped, and separately bagged sorbent tubes to prevent cross-contamination from high-concentration samples in the same shipment
Exceeding a method holding time is a data qualification , not a discretionary matter: the result is flagged, and its usability for a compliance or medical decision must be assessed explicitly.
5. Laboratory Competence
Two independent lines of evidence establish that a laboratory can do the work:
Accreditation. The AIHA Laboratory Accreditation Programs (AIHA-LAP) accredit laboratories against ISO/IEC 17025 for defined fields of testing — industrial hygiene, environmental lead, environmental microbiology. Accreditation is scope-specific: a laboratory accredited for lead is not thereby accredited for asbestos.
Proficiency testing. Blind samples of known concentration are analysed at intervals and scored against reference values. The Industrial Hygiene Proficiency Analytical Testing (IHPAT) program and its analogues for asbestos and lead provide the ongoing performance record.
Some regulatory standards require accreditation and proficiency participation explicitly, and for enforcement or litigation purposes the absence of either is a straightforward line of attack on the data.
6. Limits of Detection and Quantitation
Two figures of merit bound what the number means:
Limit of detection (LOD): the smallest mass distinguishable from the blank, conventionally three times the standard deviation of the blank response.
Limit of quantitation (LOQ): the smallest mass that can be reported with acceptable precision, conventionally ten times that standard deviation.
A result reported as "less than the LOD" is not zero . For statistical treatment of censored data, a common convention is to substitute LOD divided by the square root of 2, though maximum-likelihood and regression-on-order-statistics methods are preferred when a substantial fraction of results are non-detects. Reporting non-detects as zero biases the mean low and understates exposure.