Corneal Topography & OCT for Contact Lens Fitting

NCLE CLRE Contact Lens Registry Exam Study Guide Exam

Track your progress as you read

Finished sections get a check mark.

1 Introduction & Exam Overview

0.1 About the CLRE Exam 0.2 How to Use This Study Guide

2 Ocular Anatomy: Adnexa

Tear Film: Layers, Glands & Break-Up Time Conjunctiva, Lids, Lashes, Fornix & Canthi Blink Rate, Lid Function & Contact Lens Wear Effects

3 Ocular Anatomy: Anterior & Posterior Segment

Corneal Layers, Curvature, Topography & Metabolism Iris, Pupil, Sclera & Anterior Chamber Crystalline Lens, Retina, Choroid & Posterior Chamber

4 Refractive Errors & Contact Lens Optics

Myopia, Hyperopia, Astigmatism, Presbyopia & Anisometropia Refraction, Prism, Lacrimal Lens, Accommodation & Transposition Vertex Distance, Sagittal Depth & Spherical Equivalent

5 Measurement Instrumentation

Keratometer/Ophthalmometer: Calibration, Readings & Interpretation Corneal Topography & OCT for Contact Lens Fitting Schirmer Tests & Tear Break-Up Time

6 Observation & Verification Instrumentation

Slit Lamp (Biomicroscope) Operation & Illumination Slit Lamp Evaluation With & Without Contact Lenses Radiuscope, Lensometer, Thickness Gauge & ANSI Z80

7 Prefitting Patient Evaluation

6.1 Lifestyle, Occupation, Environment & Motivation 6.2 Medical History, Diseases, Medications & Allergies 6.3 Contraindications & Factors for Successful Wear 6.4 Relating Corneal Measurements, Rx & Lens Parameters

8 Lens Materials, Design & Parameter Selection

Dk, Dk/t, Thermal Conductivity & Holden-Mertz Water Content, Wetting Angle, Modulus, Silicone Hydrogel & Surface Treatments Toric, Multifocal, Aspheric, Hybrid, Scleral, Orthokeratology & Quadrant-Specific Base Curve, Peripheral Curves, Diameter, Thickness, Power & Edge Design

9 Diagnostic Fitting Procedures

8.1 Lens Selection & Application Procedures 8.2 Fluorescein Pattern Evaluation for RGP Lenses 8.3 Soft Lens Movement, Centration & Coverage

10 Fit Evaluation & Over-Refraction

Toric Lens Axis Rotation, LARS Rule & Stabilization Patient Comfort, Visual Acuity & Over-Refraction Modifying Base Curve, Diameter, Sag & Power from Findings

11 Dispensing: Wearing Schedules & Replacement

Daily, Flexible & Extended/Continuous Wear Schedules Daily, Biweekly, Monthly & Conventional Replacement Adaptation Protocols & Patient Compliance Counseling

12 Dispensing: Lens Care Systems

Solution Compatibility with Lens Types Chemical, Heat/UV, Hydrogen Peroxide & Neutralization Surfactant Cleaners, Protein Removers & Weekly Cleaning Solution Sensitivities, Incompatibilities & Adverse Reactions

13 Dispensing: Handling & Hygiene

12.1 Soft Lens Insertion & Removal Techniques 12.2 RGP Insertion, Removal & Emergency Responses 12.3 Hygiene, Soaps, Cosmetics & Storage Containers 12.4 Verifying Lens Parameters Against the Order at Dispensing

14 Follow-Up: Subjective & Objective Findings

13.1 Subjective Complaints: Dryness, Awareness, Blurry Vision, Pain & Photophobia 13.2 Steep vs Flat Lens Fit (Rigid & Soft) & Everted Lens 13.3 Injection, Edema, Discharge, Vision Loss & Lid Problems 13.4 Visual Acuity, Over-Refraction & Gross Evaluation Techniques

15 Follow-Up: Staining Patterns & Complications

14.1 Staining Patterns: 3 & 9 O'Clock, Foreign Body, SPK, Arcuate, Smile & Dimple Veiling 14.2 SEAL, CLARE, VLK, SLK & CLPU 14.3 Microbial Keratitis, Acanthamoeba, Dendritic Ulcer & Infiltrates 14.4 Corneal Molding/Warpage, Dellen & Mucin Balls

16 Follow-Up: Management Protocols

GPC, Blepharitis, Ptosis & Tarsal Plate Evaluation Daily Wear, Extended, Therapeutic, Postsurgical & Pathology Protocols Changing Lens Parameters at Follow-Up & Documenting Records

17 Regulatory & Administrative

FCLCA, FDA, FTC, Contact Lens Rule & Prescriber Verification HIPAA, PHI, OSHA, CDC, ANSI & Universal Precautions Record Retention, Inventory, Ordering & Documentation Methodology

Corneal Topography & OCT for Contact Lens Fitting

Key Takeaways Explain This Topic AI

Corneal topography uses Placido disc reflections to map curvature, power, and elevation across the whole cornea — far beyond the central 3-4 mm a keratometer samples.

Color scales run warm (red/orange = steep) to cool (blue/green = flat); always check the scale step size because smaller steps exaggerate pathology.

Map types include axial/sagittal, tangential, refractive, and elevation; tangential maps localize cones and ortho-K treatment zones.

Asymmetric bow-tie, inferior steepening, and post-LASIK oblate patterns flag ectasia and surgical corneas that need topography-driven designs.

Anterior-segment OCT measures sagittal depth, limbal clearance, and landing zone relationships essential for scleral and ortho-K fitting.

Last updated: July 2026

Corneal Topography & OCT for Contact Lens Fitting

Quick Answer: Corneal topography uses Placido disc reflections to map curvature, power, and elevation across the entire corneal surface — far more than the central 3-4 mm a keratometer samples. Color scales run warm (steep) to cool (flat) , and patterns such as bow-tie, asymmetric bow-tie, or inferior steepening flag astigmatism and keratoconus. Anterior-segment OCT adds cross-sectional imaging that lets fitters measure sagittal depth , limbal clearance , and landing zone relationships — essential for scleral and ortho-K fitting.

Topography: Principle

A Placido disc projects a series of concentric light rings onto the tear film. The system analyzes the spacing and shape of the reflected rings: closely spaced rings indicate a steep cornea; widely spaced rings indicate a flat cornea. Modern systems combine Placido imaging with Scheimpflug cameras or slit-scanning to add true elevation data.

Color Scale

Standardized color scales use:

  • Warm colors (red/orange) = steep (high power, short radius)
  • Cool colors (blue/green) = flat (low power, long radius)
  • A green central zone is typical of normal curvature (~43-44 D)

Always check the scale step size (1.0 D vs 0.5 D) — smaller steps exaggerate pathology. Absolute scales aid comparison across visits; relative scales highlight local variation.

Map Types

  • Map What it shows Best use
  • Axial / Sagittal Curvature referenced to the visual axis; smoother, more global Overall pattern, screening

Tangential Local curvature at each point; more sensitive to focal changes Cone localization, ortho-K zone mapping

  • Refractive Corneal refractive power accounting for Stiles-Crawford Visual impact of curvature
  • Elevation Height above/below a reference sphere Scleral lens vault, RGP 3-point touch
  • Pachymetry Corneal thickness Keratoconus staging
  • Patterns & Pathology

Round / oval pattern : regular astigmatism, symmetric bow-tie.

Symmetric bow-tie : with-the-rule or against-the-rule regular astigmatism.

Asymmetric bow-tie : skewed axis, one lobe larger — early keratoconus suspect.

Inferior steepening ("claw" or dropping bow-tie) : classic keratoconus pattern.

Superior steepening : contact lens-induced warpage or superior ectasia.

Post-LASIK pattern : central flat zone (blue) surrounded by annular steepening (red donut); oblate pattern.

Post-RK pattern : radial flat spokes with steep incision bridges.

Post-PK pattern : irregular, often high astigmatism; topography mandatory before fitting.

Clinical Uses in Contact Lens Practice

Keratoconus screening — topography detects sub-clinical ectasia before slit-lamp signs. Indices such as Sim-K , I-S value (>1.4 D suspicious), and KISA% aid diagnosis.

RGP fitting — align the lens with the flattest meridian; tangential maps show where the lens will bear.

Scleral lens fitting — elevation maps guide vault; OCT confirms clearance.

Ortho-K fitting and monitoring — tangential maps show the treatment zone (central flat) and reverse curve; confirm centration and assess for decentration.

Post-PK and post-refractive fitting — irregular corneas require topography-driven designs.

Anterior Segment OCT

OCT uses low-coherence interferometry to produce cross-sectional images of the cornea, tear film, and conjunctiva/sclera. For contact lens fitting, AS-OCT provides:

Sagittal depth (vault) of scleral lenses — central clearance target ~200-300 µm after settling.

Landing zone relationship — confirms the lens lands on conjunctiva, not on the limbus.

Limbal clearance — prevents limbal staining and vascular constriction.

Tear film thickness — helps evaluate dry eye and lens dehydration.

Post-lens tear reservoir — verifies fluid reservoir under scleral lenses.

Pachymetry — corneal thickness for edema monitoring in extended wear.

Topography vs Keratometry

  • Feature Keratometry Topography
  • Coverage Central 3-4 mm Whole cornea
  • Astigmatism Assumes regular Detects irregular
  • Output Two K values Color map + indices
  • Keratoconus Often unmeasurable Detects sub-clinical
  • Use Soft lens BC starting point RGP/scleral/ortho-K/medical
  • Ortho-K and Post-Surgical Considerations

For ortho-K , the tangential map is the workhorse: it shows the central flat treatment zone, the reverse curve (steep ring), and the alignment zone. A well-centered treatment zone 5-6 mm wide with a symmetric reverse curve indicates a good fit. Decentration, smiley-face patterns, or central islands signal lens design or positioning problems that need re-design before continued wear.

For post-PK (penetrating keratoplasty) corneas, topography is mandatory before any fitting attempt because the graft often has high, irregular astigmatism that keratometry cannot capture. Elevation maps guide RGP and scleral designs that vault the graft-host junction. Fitting should wait until sutures are out and curvature has stabilized — typically 3-6 months post-suture removal.

For post-LASIK / PRK patients, the oblate cornea requires a reverse-geometry RGP or a scleral lens. Topography reveals the central flat treatment zone and transition zone; the lens must vault the central flat area and land on the midperipheral steep ring. Misreading a post-LASIK map as a normal oblate pattern leads to central bearing and staining.

Limitations of Topography

Tear film disruption degrades the reflected Placido image.

Placido-only systems miss true elevation data (need Scheimpflug or slit-scanning).

Color scales can mislead if not standardized.

Reference sphere selection changes elevation maps; always compare to a normal baseline.

Topography measures the anterior surface only; posterior ectasia (e.g., posterior keratoconus) requires Scheimpflug or OCT.