Study guide
Keratometer/Ophthalmometer: Calibration, Readings & Interpretation
NCLE CLRE Contact Lens Registry Exam Study Guide Exam
Track your progress as you read
Finished sections get a check mark.
1 Introduction & Exam Overview
0.1 About the CLRE Exam 0.2 How to Use This Study Guide
2 Ocular Anatomy: Adnexa
Tear Film: Layers, Glands & Break-Up Time Conjunctiva, Lids, Lashes, Fornix & Canthi Blink Rate, Lid Function & Contact Lens Wear Effects
3 Ocular Anatomy: Anterior & Posterior Segment
Corneal Layers, Curvature, Topography & Metabolism Iris, Pupil, Sclera & Anterior Chamber Crystalline Lens, Retina, Choroid & Posterior Chamber
4 Refractive Errors & Contact Lens Optics
Myopia, Hyperopia, Astigmatism, Presbyopia & Anisometropia Refraction, Prism, Lacrimal Lens, Accommodation & Transposition Vertex Distance, Sagittal Depth & Spherical Equivalent
5 Measurement Instrumentation
Keratometer/Ophthalmometer: Calibration, Readings & Interpretation Corneal Topography & OCT for Contact Lens Fitting Schirmer Tests & Tear Break-Up Time
6 Observation & Verification Instrumentation
Slit Lamp (Biomicroscope) Operation & Illumination Slit Lamp Evaluation With & Without Contact Lenses Radiuscope, Lensometer, Thickness Gauge & ANSI Z80
7 Prefitting Patient Evaluation
6.1 Lifestyle, Occupation, Environment & Motivation 6.2 Medical History, Diseases, Medications & Allergies 6.3 Contraindications & Factors for Successful Wear 6.4 Relating Corneal Measurements, Rx & Lens Parameters
8 Lens Materials, Design & Parameter Selection
Dk, Dk/t, Thermal Conductivity & Holden-Mertz Water Content, Wetting Angle, Modulus, Silicone Hydrogel & Surface Treatments Toric, Multifocal, Aspheric, Hybrid, Scleral, Orthokeratology & Quadrant-Specific Base Curve, Peripheral Curves, Diameter, Thickness, Power & Edge Design
9 Diagnostic Fitting Procedures
8.1 Lens Selection & Application Procedures 8.2 Fluorescein Pattern Evaluation for RGP Lenses 8.3 Soft Lens Movement, Centration & Coverage
10 Fit Evaluation & Over-Refraction
Toric Lens Axis Rotation, LARS Rule & Stabilization Patient Comfort, Visual Acuity & Over-Refraction Modifying Base Curve, Diameter, Sag & Power from Findings
11 Dispensing: Wearing Schedules & Replacement
Daily, Flexible & Extended/Continuous Wear Schedules Daily, Biweekly, Monthly & Conventional Replacement Adaptation Protocols & Patient Compliance Counseling
12 Dispensing: Lens Care Systems
Solution Compatibility with Lens Types Chemical, Heat/UV, Hydrogen Peroxide & Neutralization Surfactant Cleaners, Protein Removers & Weekly Cleaning Solution Sensitivities, Incompatibilities & Adverse Reactions
13 Dispensing: Handling & Hygiene
12.1 Soft Lens Insertion & Removal Techniques 12.2 RGP Insertion, Removal & Emergency Responses 12.3 Hygiene, Soaps, Cosmetics & Storage Containers 12.4 Verifying Lens Parameters Against the Order at Dispensing
14 Follow-Up: Subjective & Objective Findings
13.1 Subjective Complaints: Dryness, Awareness, Blurry Vision, Pain & Photophobia 13.2 Steep vs Flat Lens Fit (Rigid & Soft) & Everted Lens 13.3 Injection, Edema, Discharge, Vision Loss & Lid Problems 13.4 Visual Acuity, Over-Refraction & Gross Evaluation Techniques
15 Follow-Up: Staining Patterns & Complications
14.1 Staining Patterns: 3 & 9 O'Clock, Foreign Body, SPK, Arcuate, Smile & Dimple Veiling 14.2 SEAL, CLARE, VLK, SLK & CLPU 14.3 Microbial Keratitis, Acanthamoeba, Dendritic Ulcer & Infiltrates 14.4 Corneal Molding/Warpage, Dellen & Mucin Balls
16 Follow-Up: Management Protocols
GPC, Blepharitis, Ptosis & Tarsal Plate Evaluation Daily Wear, Extended, Therapeutic, Postsurgical & Pathology Protocols Changing Lens Parameters at Follow-Up & Documenting Records
17 Regulatory & Administrative
FCLCA, FDA, FTC, Contact Lens Rule & Prescriber Verification HIPAA, PHI, OSHA, CDC, ANSI & Universal Precautions Record Retention, Inventory, Ordering & Documentation Methodology
Keratometer/Ophthalmometer: Calibration, Readings & Interpretation
Key Takeaways Explain This Topic AI
A keratometer measures anterior corneal curvature in diopters via tear-film reflection; standard K readings look like 44.00/45.00 @ 090.
Convert diopters to radius with r (mm) = 337.5 / D, so 45.00 D = 7.50 mm.
Average K is ~43.00-44.00 D; the difference between the two principal meridians is the corneal astigmatism.
With-the-rule astigmatism has the steep meridian near 090; against-the-rule has the steep meridian near 180.
Keratometry samples only the central 3-4 mm, assumes regular astigmatism, and is calibrated with a steel ball of known curvature.
Last updated: July 2026
Keratometry: Calibration, Readings & Interpretation
Quick Answer: A keratometer (ophthalmometer) measures the anterior corneal curvature in diopters by reflecting mires off the tear film. Standard K readings look like 44.00/45.00 @ 090 . Convert diopters to radius (mm) with r = 337.5 / D , so 45.00 D = 7.50 mm. Average K is ~43-44 D, and the difference between the two meridians is the corneal astigmatism. Keratometry only samples the central 3-4 mm, so it assumes regular astigmatism and misses peripheral pathology.
Principle
The keratometer treats the cornea as a convex mirror . Mires (target objects) are projected onto the precorneal tear film and the reflected image size is measured. Because image size depends on corneal curvature, the instrument reports curvature in diopters (D) of refractive power and/or millimeters (mm) of radius of curvature. The corneal refractive index used by most keratometers is 1.3375 , which is why 337.5 appears in conversion formulas.
Instrument Types
Type Features
Bausch & Lomb (one-position) Mires doubled via prism; both meridians read without rotating; most common in US clinics
Javal-Schiotz (two-position) Doubled images via Wollaston prism; instrument rotated 90° between meridians; mires are fixed shape
- Automatic keratometers Auto-alignment, digital readout; common in auto-refractors
- Calibration
Calibration is verified with a precision steel ball of known curvature (commonly 42.50 D). Place the ball on the chin-rest cradle, focus the mires, and confirm the reading matches the engraved value. Tolerance is typically ±0.25 D. Mire clarity also confirms instrument focus; distorted mires on the steel ball suggest optical misalignment or a dirty objective.
Measurement Procedure
Seat the patient comfortably; adjust chair and instrument height.
Instruct the patient to fixate on the reflected image of their own eye (or the mire target).
Focus the eyepiece first by turning the eyepiece ring until the mires are sharpest — this corrects for the examiner's refraction.
Align the mires: superimpose and center the plus/minus signs or circles.
Read the horizontal meridian, then rotate 90° (one-position instruments read both simultaneously).
Record as flat K / steep K @ axis of steep meridian .
Reading Format & Interpretation
A K reading is recorded as: 44.00 / 45.00 @ 090 .
- 44.00 D = flatter meridian
- 45.00 D = steeper meridian
- 090 = axis of the steeper meridian in degrees
With-the-rule (WTR) astigmatism : steeper vertical meridian (~090), e.g., 44.00/45.00 @ 090. Common in younger patients.
Against-the-rule (ATR) astigmatism : steeper horizontal meridian (~180), e.g., 44.00/45.00 @ 180. Common with age.
Oblique : steep axis near 045 or 135.
Average K is ~43.00-44.00 D. Values >46.00 D should raise suspicion for keratoconus or ectasia, especially when asymmetric between eyes. The amount of corneal astigmatism equals steep K − flat K (here 1.00 D).
Diopter-to-Millimeter Conversion
The relationship: r (mm) = 337.5 / D .
- K (D) r (mm)
- 40.50 8.33
- 42.50 7.94
- 43.00 7.85
- 44.00 7.67
- 45.00 7.50
- 46.00 7.34
- 48.00 7.03
Worked example: a K of 45.00 D → r = 337.5 / 45.00 = 7.50 mm . Conversely, a 7.80 mm radius → D = 337.5 / 7.80 = 43.27 D .
Contact Lens Application
For initial base curve (BC) selection:
Soft lenses : typically 0.8-1.2 mm flatter than flat K (or near flat K). A common rule is on-K or 0.2-0.4 mm flatter than flat K for corneal soft lenses; modern sag-based designs use sagittal depth instead of K alone.
RGP corneal lenses : fitted ~0.1 mm flatter than flat K to produce an apical clearance or three-point touch pattern; use trial lenses and over-refract.
Scleral lenses : vault the cornea entirely; K is only a starting reference, with the scleral landing zone shaped by OCT and topography.
Worked Fitting Example
A patient presents with K readings 43.00 / 44.50 @ 090 .
Astigmatism : 44.50 − 43.00 = 1.50 D WTR .
Steep meridian radius : r = 337.5 / 44.50 = 7.58 mm .
Flat meridian radius : r = 337.5 / 43.00 = 7.85 mm .
Soft lens starting BC : 0.8-1.2 mm flatter than flat K → 8.6-9.1 mm (select an 8.6 or 8.9 mm BC trial).
RGP starting BC : ~0.1 mm flatter than flat K → ~7.75-7.85 mm trial lens, then evaluate fluorescein pattern.
If the mires appear distorted or broken, suspect an irregular cornea and proceed to topography before any lens selection.
Limitations
Measures only the central 3-4 mm corneal zone — peripheral pathology is missed.
Assumes regular astigmatism ; irregular corneas (keratoconus, post-surgical) produce distorted mires that may be unmeasurable.
Tear-film quality affects mire clarity; dry eye distorts readings.
One-position instruments assume the two principal meridians are exactly 90° apart — not valid in irregular astigmatism.
Keratometry cannot detect early keratoconus; an apparently normal K pair with asymmetric bow-tie on topography may still represent forme fruste keratoconus.