7.5 Blood Pressure Measurement

CNA Certified Nursing Assistant

Track your progress as you read

Finished sections get a check mark.

1 Introduction to the CNA Exam

1.1 About the CNA Certification Exam 1.2 Exam Content Areas 1.3 Clinical Skills Evaluation 1.4 How to Use This Course

2 Role and Responsibilities

2.1 CNA Scope of Practice 2.2 The Healthcare Team 2.3 Delegation and Supervision 2.4 Documentation and Reporting 2.5 Time Management and Organization 2.6 Professionalism and Work Ethics

3 Communication and Interpersonal Skills

3.1 Verbal Communication 3.2 Nonverbal Communication 3.3 Active Listening 3.4 Communication with Special Needs Patients 3.5 Cultural Sensitivity and Diversity 3.6 Handling Difficult Communication Situations

4 Infection Control

4.1 Chain of Infection 4.2 Hand Hygiene 4.3 Personal Protective Equipment (PPE) 4.4 Standard Precautions 4.5 Transmission-Based Precautions 4.6 Cleaning, Disinfection, and Sterilization

5 Safety and Emergency Procedures

5.1 Fall Prevention 5.2 Fire Safety 5.3 Body Mechanics and Safe Lifting 5.4 Restraints and Alternatives 5.5 Emergency Response Procedures 5.6 Disaster Preparedness

6 Personal Care Skills

6.1 Bathing and Showering 6.2 Oral Hygiene and Denture Care 6.3 Hair Care, Grooming, and Shaving 6.4 Dressing and Undressing 6.5 Perineal Care 6.6 Toileting and Incontinence Care

7 Basic Nursing Skills

7.1 Vital Signs Overview 7.2 Temperature Measurement 7.3 Pulse Measurement 7.4 Respirations Measurement 7.5 Blood Pressure Measurement 7.6 Positioning, Transfers, and Mobility

8 Mental Health and Cognitive Care

8.1 Understanding Dementia 8.2 Communication with Dementia Residents 8.3 Dementia Care Strategies 8.4 Depression and Anxiety in the Elderly 8.5 Abuse Recognition and Reporting

9 Legal and Ethical Considerations

9.1 Residents' Rights 9.2 HIPAA and Confidentiality 9.3 Advance Directives and End-of-Life Care 9.4 CNA Scope of Practice 9.5 Ethical Principles in Healthcare 9.6 Documentation and Reporting

7.5 Blood Pressure Measurement

Key Takeaways Explain This Topic AI

Blood pressure is systolic (heart contracts) over diastolic (heart rests), normal 90-120 / 60-80 mmHg

A cuff that is too small reads falsely HIGH; too large reads falsely LOW

Support the arm at heart level and rest the patient 5 minutes; arm below heart reads high, above heart reads low

The first Korotkoff sound is systolic; the disappearance of sound is diastolic; deflate 2-4 mmHg per second

Report systolic above 160 or below 90 and diastolic above 100 or below 60; never use an arm with an IV, dialysis access, or on the mastectomy side

Last updated: June 2026

Understanding the Two Numbers

Blood pressure (BP) is the force of blood pushing on artery walls, written as systolic over diastolic in millimeters of mercury (mmHg).

  • Component Definition Normal Range
  • Systolic Pressure as the heart contracts (top number) 90-120 mmHg
  • Diastolic Pressure as the heart rests between beats (bottom) 60-80 mmHg
  • Pulse pressure Systolic minus diastolic 30-50 mmHg
  • A reading of 120/80 is spoken "one-twenty over eighty."
  • Adult Classifications (ACC/AHA)

The diagnostic threshold for hypertension is 130/80 mmHg , unchanged in the 2025 ACC/AHA guideline, which keeps the same categories below.

  • Category Systolic Diastolic
  • Normal Below 120 and below 80
  • Elevated 120-129 and below 80
  • Hypertension Stage 1 130-139 or 80-89
  • Hypertension Stage 2 140+ or 90+
  • Hypertensive crisis Above 180 and/or above 120
  • Hypotension Below 90 or below 60
  • Equipment and Cuff Size

The sphygmomanometer has a cuff (with an inner bladder), a gauge, and a bulb-and-valve. The stethoscope's diaphragm sits over the brachial artery.

  • Cuff Size Arm Circumference
  • Small adult 22-26 cm
  • Adult 27-34 cm
  • Large adult 35-44 cm
  • Thigh cuff 45-52 cm

Critical rule: wrong cuff size means a wrong reading. A cuff too small reads falsely HIGH (it cannot compress the artery fully); a cuff too large reads falsely LOW .

  • Manual Technique, Korotkoff Sounds, and Errors
  • Step-by-Step Manual BP

Prepare: rest the patient 5 minutes, seated, feet flat, arm supported at heart level , sleeve not rolled into a tight band.

Place the cuff: center the bladder over the brachial artery, lower edge 1 inch above the elbow crease, snug enough that only two fingers slide under.

Find the brachial pulse in the antecubital space and rest the stethoscope diaphragm there.

Inflate with the valve closed to about 30 mmHg above the expected systolic (commonly 160-180).

Deflate slowly, releasing 2-4 mmHg per second , and listen.

Record both numbers, the arm used, and the patient's position.

Korotkoff Sounds

  • Phase Sound Meaning
  • Phase I First clear tapping Systolic pressure
  • Phase IV Muffling Nearing diastolic
  • Phase V Sound disappears Diastolic pressure

The first tap is systolic; the silence is diastolic.

Common Errors and Their Effect

  • Error Effect Fix
  • Cuff too small Falsely high Match cuff to arm circumference
  • Cuff too loose Falsely high Two-finger snugness test
  • Arm below heart level Falsely high Support arm at heart level
  • Arm above heart level Falsely low Support arm at heart level
  • Measured after activity Falsely high Rest patient 5 minutes
  • Deflating too fast Missed systolic Release 2-4 mmHg/second
  • Re-inflating mid-reading Inaccurate Fully deflate, wait 1-2 min, restart
  • Restricted Arms and Special Situations

Never take BP on an arm with an IV line , a dialysis access (AV fistula or shunt) , a recent fracture or injury, or on the side of a mastectomy . Document which arm you used and stay consistent. Orthostatic (postural) BP is measured lying, sitting, then standing to detect drops that cause dizziness; a fall of more than 20 mmHg systolic on standing is reported.

When to Report

Report immediately if systolic is above 160 or below 90 , diastolic is above 100 or below 60 , the reading changed sharply from baseline, or the patient reports dizziness, a pounding headache, or chest pain.

  • Electronic Cuffs, the Two-Step Method, and Worked Scenarios
  • Electronic vs. Manual

Many facilities use an electronic (automatic) BP machine that inflates and reads on its own. Some states let candidates test on the electronic skill, but the classic NNAAP version uses the manual sphygmomanometer and stethoscope, so master listening for Korotkoff sounds . Even with an electronic machine the fundamentals do not change: correct cuff size, arm at heart level, a rested patient, and the same arm each time. If an electronic reading looks wildly off, the trained response is to repeat it manually rather than chart a number you do not trust.

The Two-Step (Palpated) Method

To avoid the auscultatory gap , a window where Korotkoff sounds briefly vanish, experienced staff first palpate the radial pulse while inflating: the pressure at which the pulse disappears estimates systolic. They then inflate 30 mmHg above that point before listening. This keeps you from under-inflating and recording a falsely low systolic, and it is why the procedure says inflate to about 30 above the expected systolic rather than a fixed number for everyone.

Worked Scenario

A CNA gets 168/96 on a resident who is normally 124/78. Before reporting an alarming spike, check the controllable variables: was the resident rested 5 minutes, was the arm supported at heart level, was the cuff the right size and snug, and was talking avoided during the reading? Suppose the resident had just hurried back from the dining room and the arm was hanging at the side. The CNA reseats the resident, supports the arm at heart level, waits 5 minutes, and re-measures, getting 130/82. The correct documentation notes both, and the nurse is informed of the rested value.

Orthostatic Readings and Safety

For orthostatic (postural) BP , measure lying, then sitting, then standing, waiting 1 to 3 minutes between positions, and stay with the patient because a large drop can cause fainting. A systolic fall greater than 20 mmHg or a diastolic fall greater than 10 mmHg on standing, especially with dizziness, is reported, and the patient is kept safe from a fall during the test.