Study guide

10.4 CT Systems, PACS, and Information Systems

Certified Nuclear Medicine Technologist CNMT

Track your progress as you read

Finished sections get a check mark.

1 Introduction & Exam Overview

1.1 Current CNMT Exam Facts 1.2 How to Use This Guide

2 Radiation Physics Fundamentals

2.1 Radioactive Materials, Isotopes, and Decay Modes 2.2 Decay Calculations and Counting Statistics 2.3 Interactions of Radiation with Matter 2.4 Radiation Detector Types and Principles

3 Biological Effects & ALARA

3.1 Biological Effects of Radiation Exposure 3.2 ALARA and Protection Calculations 3.3 Dose Limits, Surveys, and Personal Monitoring

4 NRC & Regulatory Compliance

4.1 NRC Areas, Spills, Waste, and Records 4.2 Written Directives and DOT Shipping 4.3 EPA, OSHA, HIPAA, FDA, and USP Standards 4.4 Radioactive Waste Disposal and Regulatory Records

5 Generators & Kit Preparation

5.1 Radionuclide Generators and Quality Control 5.2 Radiopharmaceutical Kit Preparation 5.3 Radiopharmaceutical Quality Control

6 Tc-99m Diagnostic Agents

6.1 Tc-99m Cardiac Imaging Agents 6.2 Tc-99m Bone, Renal, and GU Agents 6.3 Tc-99m GI and Hepatobiliary Agents 6.4 Tc-99m CNS, Infection, and Labeled Cells

7 Non-Tc Diagnostic & PET Agents

7.1 Iodine and Indium Diagnostic Agents 7.2 Gallium, Thallium, Xenon, and Miscellaneous Agents 7.3 PET Radiopharmaceuticals 7.4 Neurologic, Breast, and Neuroendocrine PET Agents

8 Therapeutics, Adjuncts & Administration

8.1 Therapeutic Radiopharmaceuticals 8.2 Interventional and Adjunct Pharmaceuticals 8.3 Blood Labeling and Administration Techniques 8.4 Blockade, Comfort, and Gastrointestinal Adjunct Agents

9 Non-Imaging Instrumentation

9.1 Well Counters, Probes, and Survey Meters 9.2 Dose Calibrator Quality Control 9.3 Auxiliary and Patient Care Equipment

10 Imaging Systems & Quality Control

10.1 Gamma Camera Quality Control 10.2 SPECT and SPECT/CT Systems 10.3 PET and PET/CT Systems and QC 10.4 CT Systems, PACS, and Information Systems 10.5 CT Radiation Dose, Safety Alerts, and Protocol Optimization

11 Pulmonary, Bone, Oncology & Infection Imaging

11.1 Pulmonary Imaging Procedures 11.2 Bone and Musculoskeletal Imaging 11.3 Oncology Imaging Procedures 11.4 Infection Imaging Procedures

12 Renal, Endocrine & Hematopoietic Imaging

12.1 Renal and Genitourinary Imaging 12.2 Endocrine Imaging (Thyroid, Parathyroid, Adrenal) 12.3 Hematopoietic and Bone Marrow Imaging 12.4 Thyroid Uptake Measurement and Radioiodine Survey Technique

13 Cardiovascular Procedures

13.1 Myocardial Perfusion Imaging 13.2 Cardiac Function, Viability, and Special Studies 13.3 Exercise and Pharmacologic Stress Testing

14 GI, CNS, Therapy & Hybrid CT Procedures

14.1 Gastrointestinal Procedures 14.2 Central Nervous System Procedures 14.3 Radionuclide Therapy Procedures 14.4 CT and Hybrid Imaging in Nuclear Medicine 14.5 CSF Studies: Cisternography, Leak Localization, and Shunt Patency

15 Patient Care & Clinical Workflow

15.1 Patient Care, Education, and Consent 15.2 Scheduling and Radiopharmaceutical Procurement 15.3 Acquisition, Positioning, and Image Quality 15.4 Image Processing and Quantitative Analysis 15.5 Non-Imaging Studies: Specimens, Counting, and Error Analysis

10.4 CT Systems, PACS, and Information Systems

Key Takeaways Explain This Topic AI

On hybrid cameras, CT may be low-dose for attenuation correction/localization or higher-technique diagnostic CT—know which protocol is ordered and justified

Co-registration of CT and emission data is required for accurate AC and fusion; motion between acquisitions breaks registration

CT QC includes noise, uniformity, spatial/contrast resolution, and artifact review; follow manufacturer and accreditation schedules

Matrix size, pixel size, and monitor quality affect what can be seen; soft-copy reading environments need appropriate luminance and ambient light control

PACS, RIS, and EMR integration demand correct patient ID, accession matching, and data integrity—wrong ID is a critical safety event

Last updated: August 2026

10.4 CT Systems, PACS, and Information Systems

Quick Answer: Hybrid CT = AC + localization (± diagnostic). Mind dose , registration , and CT QC . On the IT side: correct patient ID , PACS/RIS workflow, and monitors fit for the matrix you acquire.

Nuclear medicine technologists increasingly operate hybrid SPECT/CT and PET/CT systems. Domain IV expects fluency not only with emission detectors but also with CT roles , basic CT QC , digital image parameters , and information systems that move studies to interpretation.

CT for Attenuation Correction vs Diagnostic CT

  • Role Typical intent Technique / dose concept
  • AC / localization CT μ-map + anatomic landmarks Often lower dose; may be non-diagnostic by policy

Diagnostic CT Primary CT interpretation (with appropriate supervision/credentialing) Higher technique, possible IV contrast, full diagnostic protocols

Never assume a localization CT is a full diagnostic study. Conversely, when a diagnostic CT is ordered with PET/CT or SPECT/CT, technique factors, breath-hold, and contrast timing follow CT—not emission—rules. Document which CT mode was performed.

Co-Registration

Emission and CT volumes must share spatial alignment. Systems use calibrated bed positions and software fusion. Patient motion , different arm positions, or table sag between CT and SPECT/PET shifts the μ-map relative to activity → false defects, hot rims, or wrong organ localization. If motion is obvious, repeat the affected portion when feasible and annotate the study.

CT Quality Control (Exam-Level)

Hybrid CT subsystems need QC analogous to radiology CT, scaled to departmental and manufacturer requirements.

Test concept What it checks

Noise Pixel standard deviation in a uniform water phantom ROI—too high noise may mean technique or detector issues

  • Uniformity Mean CT numbers across phantom positions should be consistent (and near expected HU for water/air)
  • Spatial resolution Ability to see high-contrast line pairs or beads
  • Contrast resolution / low-contrast detectability Seeing subtle density differences (often phantom inserts)
  • Artifact review Rings, streaks, bad detector channels

Failed CT QC can invalidate attenuation maps even if the gamma or PET detectors pass floods/blanks. After tube replacement or major service, complete required CT calibrations before hybrid clinical use.

CT Safety Alerts and Pre-Scan Notification

CT adds x-ray ionizing radiation and, for diagnostic protocols, possible IV contrast risks.

  • Safety focus Technologist actions
  • Pregnancy / screening Department screening questions before CT

Cumulative / appropriate use Correct protocol selection; avoid unnecessary diagnostic-dose CT when only AC is ordered

  • Contrast Allergy/renal history per policy when contrast used; emergency response readiness
  • Implants / devices Note metal; some devices need CT consideration

Pre-scan notification / alerts Respect dose alert/notification thresholds on modern CT consoles; do not casually override without justification and documentation per policy

Hybrid rooms still enforce controlled area practices: doors, warning lights, and ensuring non-essential persons leave during CT exposure.

Matrix Sizes, Pixels, and Monitors

Concept Why it matters

Matrix size (e.g., 64², 128², 256², 512²) Larger matrices can support finer sampling but need enough counts (NM) or dose (CT)

Pixel size FOV / matrix; determines mm per pixel

Display monitors Primary interpretation monitors need adequate resolution, luminance, and calibration; casual consumer screens can hide pathology

Window/level CT and NM grayscale mapping; wrong window hides disease

Technologists should acquire at protocol matrix/zoom and not “save counts” by unauthorized matrix changes that violate quantitative or interpretive standards.

PACS, RIS, and Networking

  • System Role
  • RIS (Radiology Information System) Orders, scheduling, accession numbers, reports workflow
  • PACS (Picture Archiving and Communication System) Stores and distributes images (DICOM)
  • EMR / HIS Enterprise medical record; may receive reports and some images
  • Modality worklist Pulls correct patient demographics to the camera/console

DICOM is the standard for medical image communication. Studies move from modality → PACS (sometimes via a gateway). Network outages require downtime procedures: do not invent temporary IDs that never get merged.

Data Integrity and Patient Identification

Wrong-patient imaging is a critical safety event in hybrid and pure NM workflows.

  • Practice Purpose
  • Two identifiers (e.g., name + DOB/MRN) Verify identity at every step
  • Modality worklist selection Avoid manual mistyping when possible
  • Accession / order match Right procedure, laterality, and radiopharmaceutical
  • Image check before push Correct patient, orientation, markers
  • QC of demographics Fix errors before PACS finalization per policy

If images are sent under the wrong name, follow institutional error disclosure and correction procedures immediately—do not silently “overwrite” without documentation.

Computers and Acquisition Consoles

The acquisition computer stores raw projections (SPECT), list-mode or sinogram data (PET), CT series, and reconstructed volumes. Protect data integrity by:

Completing studies under the correct worklist entry before reconstruction.

Avoiding unapproved software “experiments” on clinical databases.

Knowing downtime procedures (local archive path, paper orders, later reconciliation).

Applying HIPAA minimum-necessary access: lock consoles, no shared passwords, no PHI on portable media without encryption per policy.

Secondary captures and screenshots used for teaching must be de-identified. Quantitative screenshots (SUV tables) should match the series sent to PACS so the report and images agree.

Network troubleshooting (exam level): if the worklist is empty, check modality connectivity and RIS interface before manually entering demographics—manual entry is the highest-risk path for ID errors. If PACS send fails, do not delete local data until successful archive is confirmed.

Memory aid: Hybrid excellence = right patient , right CT mode , registered volumes , passing CT+NM QC , clean PACS send.