Track your progress as you read
Finished sections get a check mark.
1 Introduction & Exam Overview
1.1 Current CNMT Exam Facts 1.2 How to Use This Guide
2 Radiation Physics Fundamentals
2.1 Radioactive Materials, Isotopes, and Decay Modes 2.2 Decay Calculations and Counting Statistics 2.3 Interactions of Radiation with Matter 2.4 Radiation Detector Types and Principles
3 Biological Effects & ALARA
3.1 Biological Effects of Radiation Exposure 3.2 ALARA and Protection Calculations 3.3 Dose Limits, Surveys, and Personal Monitoring
4 NRC & Regulatory Compliance
4.1 NRC Areas, Spills, Waste, and Records 4.2 Written Directives and DOT Shipping 4.3 EPA, OSHA, HIPAA, FDA, and USP Standards 4.4 Radioactive Waste Disposal and Regulatory Records
5 Generators & Kit Preparation
5.1 Radionuclide Generators and Quality Control 5.2 Radiopharmaceutical Kit Preparation 5.3 Radiopharmaceutical Quality Control
6 Tc-99m Diagnostic Agents
6.1 Tc-99m Cardiac Imaging Agents 6.2 Tc-99m Bone, Renal, and GU Agents 6.3 Tc-99m GI and Hepatobiliary Agents 6.4 Tc-99m CNS, Infection, and Labeled Cells
7 Non-Tc Diagnostic & PET Agents
7.1 Iodine and Indium Diagnostic Agents 7.2 Gallium, Thallium, Xenon, and Miscellaneous Agents 7.3 PET Radiopharmaceuticals 7.4 Neurologic, Breast, and Neuroendocrine PET Agents
8 Therapeutics, Adjuncts & Administration
8.1 Therapeutic Radiopharmaceuticals 8.2 Interventional and Adjunct Pharmaceuticals 8.3 Blood Labeling and Administration Techniques 8.4 Blockade, Comfort, and Gastrointestinal Adjunct Agents
9 Non-Imaging Instrumentation
9.1 Well Counters, Probes, and Survey Meters 9.2 Dose Calibrator Quality Control 9.3 Auxiliary and Patient Care Equipment
10 Imaging Systems & Quality Control
10.1 Gamma Camera Quality Control 10.2 SPECT and SPECT/CT Systems 10.3 PET and PET/CT Systems and QC 10.4 CT Systems, PACS, and Information Systems 10.5 CT Radiation Dose, Safety Alerts, and Protocol Optimization
11 Pulmonary, Bone, Oncology & Infection Imaging
11.1 Pulmonary Imaging Procedures 11.2 Bone and Musculoskeletal Imaging 11.3 Oncology Imaging Procedures 11.4 Infection Imaging Procedures
12 Renal, Endocrine & Hematopoietic Imaging
12.1 Renal and Genitourinary Imaging 12.2 Endocrine Imaging (Thyroid, Parathyroid, Adrenal) 12.3 Hematopoietic and Bone Marrow Imaging 12.4 Thyroid Uptake Measurement and Radioiodine Survey Technique
13 Cardiovascular Procedures
13.1 Myocardial Perfusion Imaging 13.2 Cardiac Function, Viability, and Special Studies 13.3 Exercise and Pharmacologic Stress Testing
14 GI, CNS, Therapy & Hybrid CT Procedures
14.1 Gastrointestinal Procedures 14.2 Central Nervous System Procedures 14.3 Radionuclide Therapy Procedures 14.4 CT and Hybrid Imaging in Nuclear Medicine 14.5 CSF Studies: Cisternography, Leak Localization, and Shunt Patency
15 Patient Care & Clinical Workflow
15.1 Patient Care, Education, and Consent 15.2 Scheduling and Radiopharmaceutical Procurement 15.3 Acquisition, Positioning, and Image Quality 15.4 Image Processing and Quantitative Analysis 15.5 Non-Imaging Studies: Specimens, Counting, and Error Analysis
10.3 PET and PET/CT Systems and QC
Key Takeaways Explain This Topic AI
PET detects near-simultaneous opposite annihilation photons (coincidence) to define lines of response (LORs); time-of-flight (TOF) uses arrival-time differences to localize events along the LOR
Daily blank scans and periodic normalization keep detector efficiencies matched; failed blank/normalization QC blocks accurate imaging
Well-counter (or equivalent dose calibrator/cross-calibration) pathways support quantitative SUV—activity concentration relative to injected dose and body size
CT provides attenuation correction and localization on PET/CT; metal, motion, and truncation distort AC maps
SUV pitfalls include wrong clock/dose times, residual syringe activity, hyperglycemia for FDG, and partial-volume underestimation of small lesions
Last updated: August 2026
10.3 PET and PET/CT Systems and QC
Quick Answer: PET = coincidence along LORs (± TOF ). Daily blank scan , normalization , and solid SUV cross-calibration. CT AC fails with metal , motion , and truncation ; SUV fails with timing , glucose , and partial volume .
Positron emission tomography (PET) images positron emitters (e.g., F-18, Ga-68, Rb-82). A positron annihilates with an electron, producing two 511 keV photons emitted roughly 180° apart . The scanner registers a true event when opposing detectors detect both photons within a short coincidence timing window.
Coincidence, LOR, and Time-of-Flight
Concept Meaning
Coincidence detection Two 511 keV interactions accepted as a pair if within the timing window and energy windows
Line of response (LOR) The line connecting the two detection points—annihilation is assumed somewhere along that line
True / scatter / random coincidences Trues = one annihilation; scatter = deflected photon; randoms = unrelated photons within the window
Time-of-flight (TOF) Measures small arrival-time difference to localize the event along the LOR, improving SNR especially in large patients
PET does not use lead parallel-hole collimators for primary spatial encoding; electronic collimation via coincidence defines geometry. Physical septa appear in some historical 2-D systems (below).
2-D vs 3-D Acquisition Concepts
- Mode Idea Trade-off
- 2-D Interplane septa reduce out-of-plane scatter and randoms Lower sensitivity
3-D Septa retracted/absent; accept more oblique LORs Higher sensitivity; more scatter/randoms—needs good corrections
Most modern whole-body PET/CT systems operate in 3-D with sophisticated scatter and randoms correction. Exam items may still contrast the concepts.
- Core PET QC
- Daily Blank Scan
A blank scan (or equivalent daily detector check using a transmission rod/source or manufacturer daily protocol) verifies that detectors respond uniformly enough for clinical use. It is analogous in spirit to a gamma-camera flood: catch dead blocks, unstable electronics, or temperature issues before patients.
- If blank fails Action
- Localized cold block Service; do not image through a dead detector region
- Global count loss Check sources, HV, cooling, coincidence timing
- Pass Document; proceed with remaining daily checks
- Normalization
Normalization measures relative efficiency of each LOR or detector crystal pair so reconstruction can compensate for efficiency differences. It is performed on a schedule (and after major service) using manufacturer protocols with a known source geometry. Skipping normalization after crystal or module replacement yields patterned artifacts and bad quantitation.
Cross-Calibration / Well Counter for SUV
Standardized uptake value (SUV) compares lesion activity concentration to injected activity normalized by body weight (or lean body mass/BSA variants):
- SUV = (activity concentration in ROI) / (injected activity / body mass)
- (with consistent units and decay correction)
The PET scanner’s quantitative scale must match the dose calibrator used to assay the syringe. Departments perform cross-calibration (sometimes involving a well counter or phantom with known activity) so that 1 Bq/mL in the phantom reads correctly on PET. Without it, SUVs drift between cameras or after software updates.
CT Attenuation Correction on PET/CT
CT images are converted to a 511 keV attenuation map for PET reconstruction. Benefits: fast AC, anatomic fusion, and localization. Requirements: same patient position and good registration between CT and PET bed positions.
PET Artifacts and Pitfalls
- Problem Effect on images / SUV
- Metal implants CT streak artifacts → incorrect μ → over/under-corrected emission
- Patient motion Misregistration; false defects or hot rims; wrong SUV
- Truncation Body outside CT FOV → incomplete μ-map
- Residual activity / wrong injection time Wrong decay correction → SUV error
- Extravasation Dose not fully in patient; SUV underestimated
- High plasma glucose (FDG) Competes with FDG; tumor SUV may fall
- Partial-volume effect Small lesions look colder than true concentration
Timing discipline: record assay time , injection time , and scan start ; assay residual syringe activity when protocol requires. Clock mismatch between hot lab and scanner is a classic quantitative error.
SUV Basics for the Exam
- Factor Direction of effect (typical)
- Longer uptake time (FDG) Tumor SUV often rises over early intervals
- Larger patient / wrong weight entry SUV changes if weight wrong
- Hotter reconstruction / sharper filter Can raise max SUV
- Larger ROI / mean vs max Different numeric SUV
Report which SUV metric (SUVmax, SUVmean) and reconstruction the site uses. Compare serial studies only when protocols match.
Randoms, Dead Time, and High-Activity Studies
At high count rates (e.g., Rb-82 MPI, high-activity dynamic studies), random coincidences and dead time rise. Systems estimate randoms (delayed window or singles-based methods) and apply dead-time correction, but extreme rates can still degrade quantitation. Follow manufacturer activity guidelines for injected dose and start times.
Daily / periodic PET QC summary (teaching):
- Test Typical role
- Blank / daily detector check Catch failed blocks before patients
- Normalization Relative LOR/crystal efficiencies
- Well-counter / dose-calibrator cross-cal Absolute scale for SUV
- Phantom (uniform / ACR-type) Image quality, SUV recovery, artifact survey
Worked SUV pitfall: Injected assay 10.0 mCi at 08:00; residual syringe 1.0 mCi at 08:05 not subtracted; scan at 09:00 with decay correction assuming 10.0 mCi in the patient. True net injected activity is lower → SUV systematically high . Always apply residual and correct clocks.
TOF note: TOF does not remove the need for AC or scatter correction; it improves effective sensitivity/SNR by constraining where along the LOR the event likely occurred.
Memory aid: PET QC B lank, N ormalization, C ross-cal → “BNC” before trusting SUV.