Study guide

10.2 SPECT and SPECT/CT Systems

Certified Nuclear Medicine Technologist CNMT

Track your progress as you read

Finished sections get a check mark.

1 Introduction & Exam Overview

1.1 Current CNMT Exam Facts 1.2 How to Use This Guide

2 Radiation Physics Fundamentals

2.1 Radioactive Materials, Isotopes, and Decay Modes 2.2 Decay Calculations and Counting Statistics 2.3 Interactions of Radiation with Matter 2.4 Radiation Detector Types and Principles

3 Biological Effects & ALARA

3.1 Biological Effects of Radiation Exposure 3.2 ALARA and Protection Calculations 3.3 Dose Limits, Surveys, and Personal Monitoring

4 NRC & Regulatory Compliance

4.1 NRC Areas, Spills, Waste, and Records 4.2 Written Directives and DOT Shipping 4.3 EPA, OSHA, HIPAA, FDA, and USP Standards 4.4 Radioactive Waste Disposal and Regulatory Records

5 Generators & Kit Preparation

5.1 Radionuclide Generators and Quality Control 5.2 Radiopharmaceutical Kit Preparation 5.3 Radiopharmaceutical Quality Control

6 Tc-99m Diagnostic Agents

6.1 Tc-99m Cardiac Imaging Agents 6.2 Tc-99m Bone, Renal, and GU Agents 6.3 Tc-99m GI and Hepatobiliary Agents 6.4 Tc-99m CNS, Infection, and Labeled Cells

7 Non-Tc Diagnostic & PET Agents

7.1 Iodine and Indium Diagnostic Agents 7.2 Gallium, Thallium, Xenon, and Miscellaneous Agents 7.3 PET Radiopharmaceuticals 7.4 Neurologic, Breast, and Neuroendocrine PET Agents

8 Therapeutics, Adjuncts & Administration

8.1 Therapeutic Radiopharmaceuticals 8.2 Interventional and Adjunct Pharmaceuticals 8.3 Blood Labeling and Administration Techniques 8.4 Blockade, Comfort, and Gastrointestinal Adjunct Agents

9 Non-Imaging Instrumentation

9.1 Well Counters, Probes, and Survey Meters 9.2 Dose Calibrator Quality Control 9.3 Auxiliary and Patient Care Equipment

10 Imaging Systems & Quality Control

10.1 Gamma Camera Quality Control 10.2 SPECT and SPECT/CT Systems 10.3 PET and PET/CT Systems and QC 10.4 CT Systems, PACS, and Information Systems 10.5 CT Radiation Dose, Safety Alerts, and Protocol Optimization

11 Pulmonary, Bone, Oncology & Infection Imaging

11.1 Pulmonary Imaging Procedures 11.2 Bone and Musculoskeletal Imaging 11.3 Oncology Imaging Procedures 11.4 Infection Imaging Procedures

12 Renal, Endocrine & Hematopoietic Imaging

12.1 Renal and Genitourinary Imaging 12.2 Endocrine Imaging (Thyroid, Parathyroid, Adrenal) 12.3 Hematopoietic and Bone Marrow Imaging 12.4 Thyroid Uptake Measurement and Radioiodine Survey Technique

13 Cardiovascular Procedures

13.1 Myocardial Perfusion Imaging 13.2 Cardiac Function, Viability, and Special Studies 13.3 Exercise and Pharmacologic Stress Testing

14 GI, CNS, Therapy & Hybrid CT Procedures

14.1 Gastrointestinal Procedures 14.2 Central Nervous System Procedures 14.3 Radionuclide Therapy Procedures 14.4 CT and Hybrid Imaging in Nuclear Medicine 14.5 CSF Studies: Cisternography, Leak Localization, and Shunt Patency

15 Patient Care & Clinical Workflow

15.1 Patient Care, Education, and Consent 15.2 Scheduling and Radiopharmaceutical Procurement 15.3 Acquisition, Positioning, and Image Quality 15.4 Image Processing and Quantitative Analysis 15.5 Non-Imaging Studies: Specimens, Counting, and Error Analysis

10.2 SPECT and SPECT/CT Systems

Key Takeaways Explain This Topic AI

SPECT acquires multiple projections around the patient (step-and-shoot or continuous rotation) and reconstructs tomographic slices

Center-of-rotation (COR) QC is essential—COR error produces ring or circular artifacts and misregistration of activity

SPECT demands stricter field uniformity than planar imaging because nonuniformity reconstructs into rings and false lesions

Jaszczak (and similar) phantoms evaluate 3-D uniformity, resolution spheres/rods, and contrast; pixel size calibration ties matrix to real mm

Attenuation correction may use Chang’s method (calculated map) or CT-based maps on hybrid SPECT/CT; patient motion and COR errors remain major artifact sources

Last updated: August 2026

10.2 SPECT and SPECT/CT Systems

Quick Answer: SPECT needs COR QC , excellent uniformity , correct pixel size , and solid orbits. Prefer CT-based attenuation correction on hybrid systems when available; watch for COR rings , motion , and misregistration.

Single-photon emission computed tomography (SPECT) rotates one or more gamma-camera heads about the patient, collecting projection images used to reconstruct transverse, coronal, and sagittal slices. Many modern systems are SPECT/CT hybrids that add a CT for attenuation correction, anatomic localization, and sometimes diagnostic CT protocols.

SPECT Acquisition Modes

Mode How it works Teaching points

Step-and-shoot Detector stops at each angle, acquires, then steps to next angle Classic; less angular blur during count collection

Continuous (continuous rotation) Detector moves while acquiring Faster exams possible; motion during dwell can slightly blur projections

Common orbits are circular or body-contour / elliptical noncircular paths that keep detectors close to the patient (better resolution). Dual-head 180° opposed or 90° cardiac configurations shorten acquisition time. Parameters include matrix (e.g., 64×64 or 128×128), angular sampling (e.g., 3°–6° steps), time per stop, and zoom.

Pixel size / calibration: reconstruction assumes a known mm-per-pixel. Periodic pixel calibration (or equivalent factory/service checks) keeps quantitative distances and filter cutoffs meaningful. Wrong zoom or calibration warps anatomy and ROI sizes.

Why SPECT Uniformity Requirements Are Stricter

In planar imaging, mild nonuniformity may look like mottling. In SPECT, the same nonuniformity is seen from many angles and reconstructs into rings or circular artifacts , which can mimic lesions or defects (classic example: ring artifact from a cold PMT). Therefore SPECT-capable cameras need tight daily uniformity QC and up-to-date correction maps before tomography.

Center of Rotation (COR) QC

Center of rotation is the alignment between the mechanical axis of rotation and the electronic/matrix center used in reconstruction.

  • Concept Detail
  • How tested Point source (or multi-point) imaged through a full orbit; software computes COR offsets
  • Frequency Per manufacturer—often weekly or after head collision/service (know your SOP)
  • Failure pattern Ring artifacts , smeared point sources, misaligned activity on opposite projections
  • Action Recalibrate COR; do not run clinical SPECT with failed COR

COR errors are a high-yield CNMT topic: if you see concentric rings on reconstructed floods or patient slices without a physical ring source, think COR or uniformity—not the patient’s anatomy.

Three-Dimensional Performance: Jaszczak Phantom

The Jaszczak phantom (cylindrical phantom with cold/hot rods or spheres and a uniform section) evaluates SPECT tomographic uniformity , resolution , and contrast in three dimensions—beyond planar bar phantoms.

  • Phantom region What you assess
  • Uniform cylinder section 3-D uniformity; look for rings
  • Rod / sector inserts Tomographic spatial resolution
  • Spheres Contrast and partial-volume behavior

Periodic SPECT phantom testing (schedule per accreditation, manufacturer, and department) documents that reconstruction filters, COR, and detectors perform as a system.

Attenuation Correction

Photon attenuation in the body underestimates deep activity (e.g., inferior wall of the heart, deep abdominal structures).

Method Idea Pros / cons

Chang’s method Assumes uniform attenuation coefficient (soft tissue μ) and applies a calculated correction, often after first-pass reconstruction No CT needed; fails where density varies (lungs, bone, arms down vs up)

CT-based AC (SPECT/CT) CT provides patient-specific attenuation map scaled to emission energy More accurate when co-registered; metal, truncation, and motion can still err

Hybrid workflow sketch: position patient → scout/CT → SPECT orbit (or protocol order per system) → generate μ-map → reconstruct emission with AC (and scatter correction if used) → fuse display. Co-registration of CT and SPECT is mandatory; patient motion between CT and SPECT misplaces the μ-map and creates false defects or hot spots.

Common SPECT Artifacts

  • Artifact Clue Mitigation
  • COR error rings Concentric rings, especially on uniform phantom or liver COR QC / recalibration
  • Nonuniformity rings Rings tied to flood defects Fix uniformity; service

Patient motion Blur, misregistration, inconsistent projections (sinogram breaks) Immobilize, shorter protocols, motion correction if available, repeat

  • Truncation (CT FOV) Arms or body outside CT FOV on SPECT/CT Reposition arms; larger FOV CT if available
  • Metal CT streaks → wrong μ-map → emission bias Note metal; review non-AC images
  • SPECT/CT Practical Points

Low-dose CT often suffices for attenuation correction and localization; diagnostic CT needs appropriate technique factors, contrast protocols, and often a radiologist’s CT scope of practice.

Confirm correct patient, laterality, and study type before long dual-modality exams.

After collisions or service, repeat COR , uniformity , and any manufacturer-required hybrid calibrations before clinical use.

Reconstruction Awareness for Technologists

You may not design filters, but you must recognize that filtered back-projection (FBP) and iterative methods (e.g., OSEM) behave differently with noise and attenuation. Over-smoothing hides small lesions; under-smoothing creates noisy “pseudo-lesions.” Always use protocol-approved reconstruction parameters for quantitative or comparison studies (e.g., serial bone SPECT). Review sinograms or rotating cine projections when available—discontinuities flag motion better than a single static slice.

Scatter: many SPECT protocols apply energy-window or model-based scatter correction together with AC. Turning corrections on/off inconsistently between baseline and follow-up invalidates visual comparison.

Multi-head tips: dual- or triple-head systems need each head’s COR and uniformity within limits; a single bad head can ruin the entire orbit. After a head collision, treat the system as out of service until COR and floods pass.

Memory aid: SPECT quality = U niformity + C OR + O rbit closeness + R egistration (for hybrid) → “UCOR.”