10.2 Physical & Chemical Drinking Water Testing: Turbidity, pH, Alkalinity, Hardness & Chlorine (DPD)

Colorado Water/Wastewater Operator Exam

Track your progress as you read

Finished sections get a check mark.

1 Chapter 1: State Laws, Regulations & CCWP Certification

1.1 CDPHE CCWP Program & Regulatory Framework (Regulation 100 / 5 CCR 1003-2) 1.2 Facility Classification & Operator in Responsible Charge (ORC) Requirements 1.3 Certification Levels, Education, Experience & Exam Eligibility (Class D-A, Grade 1-4, Class S/T) 1.4 License Renewal, Training Units (TUs) & Mandatory Regulatory Training 1.5 Colorado Primary Drinking Water Regulations (Regulation 11) & Compliance 1.6 Colorado Discharge Permit System (Regulations 61/62), DMRs & Biosolids Reporting

2 Chapter 2: Source Water & Drinking Water Treatment Processes

2.1 Raw Water Sources, High-Altitude Characteristics & Reservoir Quality 2.2 Coagulation Chemistry, Rapid Mixing & Chemical Dosing (Alum, Ferric, Polymers) 2.3 Flocculation & Sedimentation Basin Design, Operation & Settling Zones 2.4 Granular Media Filtration, Backwash Procedures & Turbidity Standards 2.5 Advanced Treatment: Membranes, Softening, GAC/PAC Adsorption & Aeration 2.6 Groundwater Sources, Well Hydraulics, Static and Pumping Levels & Specific Capacity 2.7 Source Water Characteristics, Contaminant Categories & Watershed Protection Plans 2.8 Pretreatment: Presedimentation, Aeration, Algae Control & Taste-and-Odor Management 2.9 Direct, In-Line, Pressure, Cartridge & Diatomaceous Earth Filtration 2.10 Lime-Soda Softening, Ion Exchange & Water Stabilization 2.11 Water Treatment Plant Residuals: Lagoons, Drying Beds, Land Application & Recycle

3 Chapter 3: Disinfection & Chemical Addition

3.1 Chlorine Chemistry: Hypochlorous Acid, Free/Combined Residual & Breakpoint Chlorination 3.2 Disinfection Kinetics, Baffle Factors, T10 & CT Compliance Calculations 3.3 Disinfection Byproducts (TTHMs, HAA5s) & Stage 1/2 DBP Rules 3.4 Alternative Disinfection: Chloramination, Ultraviolet (UV) Light & Chlorine Dioxide 3.5 Corrosion Control, Lead & Copper Rule Revisions (LCRR) & Fluoridation Practices

4 Chapter 4: Water Treatment Plant Equipment, Instrumentation & Maintenance

4.1 Raw Water Intakes, Screening Equipment & Raw Water Pumping 4.2 Chemical Feed Equipment: Feeders, Day Tanks, Carrier Water & Feed-Rate Calibration 4.3 Chlorination Equipment: Gas Systems, Vacuum Regulators, Hypochlorite Storage & Onsite Generation 4.4 Filter Underdrains, Support Gravel, Surface Wash, Air Scour & Filter-to-Waste 4.5 Plant Process Control, SCADA and PLC Systems, Alarm Management & Records of Operation 4.6 Preventive, Corrective and Predictive Maintenance, Work Orders & Asset Management

5 Chapter 5: Wastewater Treatment Principles & Biological Processes

5.1 Preliminary Treatment: Bar Screens, Shredders, Grit Chambers & Flow Equalization 5.2 Primary Clarification, Solids Separation & Scum Removal Mechanics 5.3 Activated Sludge Fundamentals: Aerobic Microbiology, DO Control & Biomass Growth 5.4 Return & Waste Activated Sludge Management (RAS/WAS Operations & Sludge Age) 5.5 Activated Sludge Troubleshooting: Bulking, Nocardia Foaming, Rising Sludge & Pin Floc 5.6 Wastewater Characteristics, Organic and Hydraulic Loading, Diurnal Flow & Septicity 5.7 Activated Sludge Arrangements: Conventional, Complete Mix, Step Feed, Contact Stabilization, Extended Aeration, Oxidation Ditch, SBR & MBR 5.8 Aeration Systems: Blowers, Diffusers, Surface Aerators, Oxygen Transfer & Altitude Effects

6 Chapter 6: Advanced Wastewater Treatment & Solids Management

6.1 Biological Nutrient Removal: Nitrification, Denitrification & Anoxic Zones 6.2 Biological & Chemical Phosphorus Removal (EBPR & Coagulant Precipitation) 6.3 Sludge Thickening, Aerobic & Anaerobic Digestion (Mesophilic/Thermophilic) 6.4 Dewatering Equipment (Centrifuges, Belt Filter Presses) & Biosolids Standards (Reg 64) 6.5 Pond & Lagoon Operations, Trickling Filters & Rotating Biological Contactors (RBCs) 6.6 Secondary Clarification: Overflow and Solids Loading Rates, RAS Rate Setting & Clarifier Diagnostics 6.7 Tertiary Filtration, Effluent Polishing & Wastewater Disinfection with Chlorine, Dechlorination and UV 6.8 Effluent Disposal, Rapid Infiltration Basins, Colorado Water Reuse Law & Regulation 84

7 Chapter 7: Water Distribution Systems

7.1 Distribution Hydraulics, Pressure Zones, Static Head & Friction Headloss 7.2 Finished Water Storage Tanks, Standpipes, Mixing & Water Age Control 7.3 Distribution Piping, Valves (Gate, Butterfly, PRV), Hydrants & Main Flushing 7.4 Cross-Connection Control, Hazard Assessment & Backflow Preventers (RPZ, DCVA, PVB, Air Gap) 7.5 Water Main Installation, Trenching, Repair, Pressure Testing & Disinfection (AWWA C651) 7.6 Network Layout, Pressure Zones, Demand Patterns, Fire Flow & the Hydraulic Grade Line 7.7 Fire Hydrants, Dry Barrel Operation, Fire Flow Testing & Unidirectional Flushing 7.8 Booster Pump Stations, Hydropneumatic Systems, Controls & Surge Protection 7.9 Customer and Master Metering, Meter Sizing, Accuracy Testing, AMR & AMI 7.10 System Mapping, GIS, As-Builts, Operational Analysis & the AWWA Water Audit 7.11 Service Line Installation, Wet Tapping, Curb Stops & the Lead Service Line Inventory 7.12 Leak Detection, District Metered Areas, Minimum Night Flow & Pressure Management 7.13 Storage Facility Inspection, Sanitary Defects, Water Age Control, Cleaning & AWWA C652 Disinfection 7.14 Utility Locating, Colorado 811 Damage Prevention, the APWA Color Code & New Construction Inspection 7.15 Well and Wellhead Maintenance, Sanitary Condition, Rehabilitation & AWWA C654 Disinfection 7.16 Booster Chlorination, Remote Chemical Feed, Chemical Storage, Security & Supply Management 7.17 Distribution Sampling Technique, Chlorine Residual Monitoring & Nitrification Control in Chloraminated Systems

8 Chapter 8: Wastewater Collection Systems

8.1 Gravity Collection Systems: Flow Velocities, Slopes, Pipe Materials & Manholes 8.2 Wastewater Lift Stations: Wet Wells, Submersible Pumps, Level Controls & Force Mains 8.3 Inflow and Infiltration (I&I) Sources, Flow Monitoring, Smoke Testing & CCTV Inspection 8.4 Collection Maintenance: Hydraulic Jetting, Mechanical Rodding & FOG/Root Mitigation 8.5 Sanitary Sewer Overflows (SSOs): Root Causes, Containment, Reporting & Mitigation 8.6 Lift Station Electrical Devices, Level Detection, Telemetry, Gas Detection & Capacity Adjustment 8.7 Lift Station Configurations, Preventive Maintenance, Wet Well Cleaning, Troubleshooting & Emergency Bypass Pumping 8.8 Collection System Rehabilitation: CIPP, Pipe Bursting, Sliplining, Grouting, Manhole Repair & Tap Defects 8.9 Excavation, Hydro Excavation, Bedding and Haunching, Grade Control & Sewer Acceptance Testing 8.10 Collection System Flow Monitoring, Area-Velocity Metering, Weirs and Flumes & Capacity Assessment

9 Chapter 9: Pumps, Motors & Utility Mechanics

9.1 Centrifugal Pumps: Volutes, Impellers, Packing Seals & Mechanical Seals 9.2 Positive Displacement Pumps (Peristaltic, Diaphragm, Progressive Cavity) & Dosing Systems 9.3 Pump Performance Curves, System Head Curves, Cavitation & NPSH Calculations 9.4 Electric Motors (Single/Three Phase), Controls, Variable Frequency Drives (VFDs) & Power 9.5 Flow Meters (Venturi, Magnetic, Ultrasonic), Online Analyzers, PLCs & SCADA Telemetry 9.6 Air Compressors, Conveyors, Hoists and Cranes, Service Piping & Standby Power Generation 9.7 Valve Types and Duties, Air Release and Air-Vacuum Valves, Water Hammer & Surge Control

10 Chapter 10: Laboratory Analysis & Water Quality Monitoring

10.1 Regulatory Sampling Procedures, Chain of Custody, Holding Times & Preservation Techniques 10.2 Physical & Chemical Drinking Water Testing: Turbidity, pH, Alkalinity, Hardness & Chlorine (DPD) 10.3 Wastewater Laboratory Analysis: BOD5, COD, Total Suspended Solids (TSS) & Settleometry 10.4 Microbiological Testing: Total Coliform, E. coli (Colilert, Membrane Filtration) & QA/QC 10.5 Bench-Scale Jar Testing, Coagulant Optimization & Streaming Current Monitoring 10.6 Laboratory Equipment, Calibration and Verification, QA/QC Program & Certified Laboratory Requirements 10.7 Nutrient Series, Metals Sampling, Whole Effluent Toxicity & Electrode Measurements (DO, pH, ORP, H2S)

11 Chapter 11: Operator Mathematics & Applied Calculations

11.1 Core Conversions, Geometric Area, Pipe Volume & Continuous Flow Rate (MGD, GPM, cfs) 11.2 Chemical Dosage, The Standard "Pounds Formula", Active Chemical & Solution Purity 11.3 Hydraulic Retention Time (HRT), Surface Overflow Rate (SOR) & Weir Overflow Rates 11.4 Process Control Calculations: F/M Ratio, MCRT/Solids Retention Time (SRT) & Sludge Volume Index (SVI) 11.5 Distribution Pressure, Static vs Dynamic Head, Water Horsepower & Wire-to-Water Efficiency 11.6 CT Calculations: Baffling Factors, T10, Temperature and pH Corrections & Multi-Segment Compliance

12 Chapter 12: Safety, Emergency Operations & Compliance

12.1 Permit-Required Confined Space (PRCS) Entry, Atmospheric Testing (O2/LEL/H2S/CO) & Rescue 12.2 Control of Hazardous Energy: Lockout/Tagout (LOTO), Arc Flash & Electrical Safety 12.3 Chemical Safety: Chlorine Gas Handling, Ton Cylinders, Sulfur Dioxide, Ammonia & SDS/GHS 12.4 Trenching & Excavation: Soil Classification, Sloping, Shoring, Shielding & Competent Person Rules 12.5 Emergency Action Plans, Security Vulnerability, Spill Containment & CDPHE Tier 1/2/3 Public Notice 12.6 Work Zone Traffic Control, PPE, Fall Protection, Fire Extinguishers, First Aid & Bloodborne Pathogens 12.7 Standard Operating Procedures, Operating Records, Regulatory Reporting, Customer Complaints & Supervisory Duties

10.2 Physical & Chemical Drinking Water Testing: Turbidity, pH, Alkalinity, Hardness & Chlorine (DPD)

Key Takeaways Explain This Topic AI

Nephelometric turbidity meters measure 90-degree scattered light in NTU; primary calibration requires formazin or AMCO Clear polymer standards, while optical cuvettes require silicone oil indexing.

Electrometric pH meters utilize a combination glass electrode with Nernstian temperature compensation, requiring multi-point calibrations with an acceptable slope between 95% and 105%.

Alkalinity titrations using 0.0200 N H2SO4 determine phenolphthalein alkalinity (pH 8.3) and total alkalinity (pH 4.5), fractionating hydroxide, carbonate, and bicarbonate buffering components.

Total hardness is quantified by EDTA complexometric titration at pH 10.0 with Eriochrome Black T indicator, converting between mg/L as CaCO3 and grains per gallon (1 gpg = 17.12 mg/L).

The DPD colorimetric method differentiates free available chlorine (instant magenta dye) from combined chloramines (magenta development catalyzed by potassium iodide addition).

Last updated: August 2026

Physical & Chemical Drinking Water Testing: Turbidity, pH, Alkalinity, Hardness & Chlorine (DPD)

Drinking water treatment plants rely on precise bench-top and continuous on-line laboratory instruments to verify regulatory compliance under CDPHE Regulation 11 (Colorado Primary Drinking Water Regulations) and optimize chemical coagulation, softening, corrosion control, and disinfection.

1. Turbidity Measurement via Nephelometry

Turbidity is an optical property of water causing light to be scattered and absorbed rather than transmitted in straight lines. In drinking water, suspended colloidal particles (clays, silts, organic matter, and microorganisms) shield pathogens from chemical disinfectants. Turbidity is quantified using a nephelometer and expressed in Nephelometric Turbidity Units (NTU) . [ Tungsten Lamp / LED (860 nm) ]

  • |
  • v (Incident Light Beam)
  • +-------------------+
  • | Sample Cell |
  • | (Indexed Glass) |
  • +-------------------+
  • / \
  • (Transmitted Light Trap) <----- -----> [ 90-Degree Scattered Light Detector ]
  • |
  • v
  • [ Nephelometer Digital Display ]
  • [ (Output in NTU) ]
  • Analytical Principles of Nephelometry

90-Degree Light Scatter: Standard Methods 2130B and EPA Method 180.1 specify that the primary photodetector must be positioned at a 90 ∘ ± 30 ∘ 90^\circ \pm 30^\circ 9 0 ∘ ± 3 0 ∘ angle relative to the incident light path. Scattered light intensity is directly proportional to particulate concentration.

Light Source Specifications: Incandescent tungsten filament lamp operated at a color temperature of 2200 – 3000 K 2200\text{--}3000\text{ K} 2200 – 3000 K , or an infrared LED light source ( 860 nm 860\text{ nm} 860 nm ) complying with ISO 7027 (minimizes dissolved color interferences).

Calibration Standards & Sample Handling

Primary Standards: Formazin is the international primary reference standard. A 4000 NTU stock solution is synthesized by combining hydrazine sulfate ( N 2 H 4 ⋅ H 2 SO 4 \text{N}_2\text{H}_4\cdot\text{H}_2\text{SO}_4 N 2 ​ H 4 ​ ⋅ H 2 ​ SO 4 ​ ) and hexamethylenetetramine ( ( CH 2 ) 6 N 4 (\text{CH}_2)_6\text{N}_4 ( CH 2 ​ ) 6 ​ N 4 ​ ) in ultrapure water, allowing 24 hours for polymer suspension maturation. AMCO Clear (styrene divinylbenzene copolymer beads) is also an EPA-approved primary standard that requires no mixing or dilution.

Secondary Standards: Sealed gel or silicone secondary standards are used strictly for daily calibration verification and drift checks. Secondary standards must never be used to calibrate or re-span a turbidimeter .

Optical Sample Cuvette Preparation: Cuvettes must be constructed of optical borosilicate glass. Clean cuvettes with non-abrasive laboratory detergent, rinse extensively with deionized water, and wipe with lint-free wipes (Kimwipes). Apply a single drop of high-purity silicone indexing oil to the exterior glass and buff with a velvet cloth to fill microscopic glass scratches that would otherwise refract incident light. Always align the cuvette index mark with the instrument indexing arrow.

Colorado Regulatory Turbidity Benchmarks

Combined Filter Effluent (CFE): Must be ≤ 0.30 NTU \le 0.30\text{ NTU} ≤ 0.30 NTU in at least 95% of monthly measurements, and must never exceed 1.0 NTU 1.0\text{ NTU} 1.0 NTU at any time.

Individual Filter Effluent (IFE) Triggers: Continuous recording every 15 minutes. An IFE exceeding 0.50 NTU 0.50\text{ NTU} 0.50 NTU after 4 hours of continuous operation, or exceeding 1.0 NTU 1.0\text{ NTU} 1.0 NTU in two consecutive 15-minute readings, or exceeding 2.0 NTU 2.0\text{ NTU} 2.0 NTU in two consecutive readings within 15 minutes of startup triggers mandatory filter profiling or a comprehensive filter self-assessment.

2. Electrometric pH Measurement & Electrode Calibration

pH represents the negative logarithm of hydrogen ion activity ( pH = − log ⁡ 10 [ H + ] \text{pH} = -\log_{10}[\text{H}^+] pH = − lo g 10 ​ [ H + ] ). In water treatment, pH governs coagulant hydrolysis, chemical precipitation, chlorine speciation, and lead/copper pipe passivation.

The Combination pH Glass Electrode

Electrometric pH systems utilize a combination glass electrode comprising two electrochemical half-cells integrated into a single probe:

Measuring Half-Cell: Contains a pH-sensitive, hydrated lithium glass bulb enclosing an internal reference element ( Ag/AgCl \text{Ag/AgCl} Ag/AgCl ) in a buffer of constant pH (typically pH 7.00). An electrical potential develops across the glass membrane proportional to the difference in hydrogen ion activity between the internal solution and the external sample.

Reference Half-Cell: Contains a silver/silver chloride ( Ag/AgCl \text{Ag/AgCl} Ag/AgCl ) reference wire submerged in a saturated electrolyte solution ( 3.0 M KCl 3.0\text{ M } \text{KCl} 3.0 M KCl ). The liquid junction (porous ceramic plug or Teflon wick) maintains continuous electrical contact with the sample.

The Nernst Equation & Temperature Compensation

The electrical potential ( E E E ) generated across the electrode membrane is governed by the Nernst Equation : E = E 0 − ( 2.303 ⋅ R ⋅ T n ⋅ F ) ⋅ pH E = E_0 - \left(\frac{2.303 \cdot R \cdot T}{n \cdot F}\right) \cdot \text{pH} E = E 0 ​ − ( n ⋅ F 2.303 ⋅ R ⋅ T ​ ) ⋅ pH

At 25.0 ∘ C 25.0^\circ\text{C} 25. 0 ∘ C (298.15 K) , the theoretical Nernstian slope is − 59.16 mV per pH unit -59.16\text{ mV per pH unit} − 59.16 mV per pH unit . Because this voltage slope varies with absolute temperature ( T T T ), laboratory meters must incorporate an Automatic Temperature Compensation (ATC) probe to adjust the millivolt-to-pH conversion slope. Theoretical Millivolt Output vs. pH at 25°C:

  • pH 4.01: +177.48 mV
  • pH 7.00: 0.00 mV (Isoelectric Point)
  • pH 10.01: -177.48 mV
  • Calibration Protocols & Electrode Slope Check

Multi-Point Calibration: Calibrate daily using at least two, preferably three, NIST-traceable standard buffer solutions (pH 4.01, 7.00, and 10.01) that bracket the anticipated sample pH.

Electrode Slope Verification: Modern meters calculate the actual electrode slope percentage relative to theoretical Nernstian output: Electrode Slope (%) = ( Δ V measured Δ V theoretical ) × 100 \text{Electrode Slope (\%)} = \left(\frac{\Delta V_{\text{measured}}}{\Delta V_{\text{theoretical}}}\right) \times 100 Electrode Slope (%) = ( Δ V theoretical ​ Δ V measured ​ ​ ) × 100

An acceptable electrode slope must fall between 95% and 105% ( 56.2 to 62.1 mV/pH unit 56.2\text{ to } 62.1\text{ mV/pH unit} 56.2 to 62.1 mV/pH unit ). A slope below 95% indicates electrode fouling, depleted electrolyte, or glass bulb dehydration, requiring electrode cleaning with 0.1 M HCl 0.1\text{ M } \text{HCl} 0.1 M HCl or electrode replacement.

Electrode Storage: Store electrodes in 3.0 M KCl 3.0\text{ M } \text{KCl} 3.0 M KCl or pH 4.0 buffer. Never store electrodes in deionized/distilled water , as osmotic pressure leaches the internal electrolyte through the ceramic junction.

Loading diagram... Alkalinity Titration & Chemical Speciation Decision Tree

3. Alkalinity Titration & Chemical Speciation

Alkalinity is the quantitative capacity of aqueous media to neutralize strong acids, serving as the chemical buffer against pH depression during alum coagulation, chlorination, and nitrification. Alkalinity is expressed in mg/L as CaCO 3 \text{mg/L as }\text{CaCO}_3 mg/L as CaCO 3 ​ and is primarily composed of three inorganic chemical species: hydroxide ( OH − \text{OH}^- OH − ), carbonate ( CO 3 2 − \text{CO}_3^{2-} CO 3 2 − ​ ), and bicarbonate ( HCO 3 − \text{HCO}_3^- HCO 3 − ​ ).

Analytical Titration Procedure

A 100 mL 100\text{ mL} 100 mL sample is titrated electrometrically (or with color indicators) using standard 0.0200 N H 2 SO 4 0.0200\text{ N } \text{H}_2\text{SO}_4 0.0200 N H 2 ​ SO 4 ​ :

Phenolphthalein Alkalinity ( P P P ): Titrate the sample to pH 8.3 (or until the pink color of phenolphthalein indicator turns colorless). Titration to pH 8.3 neutralizes all hydroxide ions and converts carbonate ions to bicarbonate: OH − + H + ⟶ H 2 O \text{OH}^- + \text{H}^+ \longrightarrow \text{H}_2\text{O} OH − + H + ⟶ H 2 ​ O CO 3 2 − + H + ⟶ HCO 3 − \text{CO}_3^{2-} + \text{H}^+ \longrightarrow \text{HCO}_3^- CO 3 2 − ​ + H + ⟶ HCO 3 − ​

Total (Methyl Orange) Alkalinity ( T T T ): Continue titrating past pH 8.3 down to pH 4.5 (bromocresol green-methyl red indicator endpoint turning from blue-green to pinkish-gray). Titration from pH 8.3 to 4.5 converts all remaining bicarbonate ions to carbonic acid / dissolved carbon dioxide: HCO 3 − + H + ⟶ H 2 CO 3 ⇌ CO 2 ↑ + H 2 O \text{HCO}_3^- + \text{H}^+ \longrightarrow \text{H}_2\text{CO}_3 \rightleftharpoons \text{CO}_2\uparrow + \text{H}_2\text{O} HCO 3 − ​ + H + ⟶ H 2 ​ CO 3 ​ ⇌ CO 2 ​ ↑ + H 2 ​ O

Alkalinity Calculation Formula Alkalinity (mg/L as CaCO 3 ) = A × N × 50,000 Sample Volume (mL) \text{Alkalinity (mg/L as }\text{CaCO}_3) = \frac{A \times N \times 50{,}000}{\text{Sample Volume (mL)}} Alkalinity (mg/L as CaCO 3 ​ ) = Sample Volume (mL) A × N × 50 , 000 ​

Where:

  • A A A = Volume of standard acid titrant used ( m L mL m L )
  • N N N = Normality of sulfuric acid ( 0.0200 N 0.0200\text{ N} 0.0200 N )
  • 50,000 50{,}000 50 , 000 = Milligram equivalent weight factor for CaCO 3 \text{CaCO}_3 CaCO 3 ​

For a standard 100 mL 100\text{ mL} 100 mL sample volume and 0.0200 N 0.0200\text{ N} 0.0200 N acid titrant, the equation simplifies to: Alkalinity = A × 10 \text{Alkalinity} = A \times 10 Alkalinity = A × 10 .

Alkalinity Speciation Table (Standard Methods 2320B)

Titration Result Hydroxide Alkalinity as CaCO 3 \text{CaCO}_3 CaCO 3 ​ Carbonate Alkalinity as CaCO 3 \text{CaCO}_3 CaCO 3 ​ Bicarbonate Alkalinity as CaCO 3 \text{CaCO}_3 CaCO 3 ​

  • P = 0 P = 0 P = 0 (pH ≤ 8.3 \le 8.3 ≤ 8.3 ) 0 0 0 0 0 0 T T T
  • P < 1 2 T P < \frac{1}{2}T P < 2 1 ​ T 0 0 0 2 P 2P 2 P T − 2 P T - 2P T − 2 P
  • P = 1 2 T P = \frac{1}{2}T P = 2 1 ​ T 0 0 0 2 P = T 2P = T 2 P = T 0 0 0
  • P > 1 2 T P > \frac{1}{2}T P > 2 1 ​ T 2 P − T 2P - T 2 P − T 2 ( T − P ) 2(T - P) 2 ( T − P ) 0 0 0
  • P = T P = T P = T T T T 0 0 0 0 0 0
  • Worked Example 9.2.1: Determining Alkalinity Fractions

A 100 mL 100\text{ mL} 100 mL water sample has an initial pH of 9.2. Titration with 0.0200 N H 2 SO 4 0.0200\text{ N } \text{H}_2\text{SO}_4 0.0200 N H 2 ​ SO 4 ​ requires 3.6 mL 3.6\text{ mL} 3.6 mL to reach the phenolphthalein endpoint (pH 8.3) and a cumulative total of 14.2 mL 14.2\text{ mL} 14.2 mL to reach the methyl red endpoint (pH 4.5). Calculate the Phenolphthalein Alkalinity ( P P P ), Total Alkalinity ( T T T ), and the individual concentrations of hydroxide, carbonate, and bicarbonate alkalinity in mg/L as CaCO 3 \text{mg/L as }\text{CaCO}_3 mg/L as CaCO 3 ​ .

Step 1: Calculate P P P and T T T alkalinity: P = 3.6 mL × 10 = 36.0 mg/L as CaCO 3 P = 3.6\text{ mL} \times 10 = 36.0\text{ mg/L as }\text{CaCO}_3 P = 3.6 mL × 10 = 36.0 mg/L as CaCO 3 ​ T = 14.2 mL × 10 = 142.0 mg/L as CaCO 3 T = 14.2\text{ mL} \times 10 = 142.0\text{ mg/L as }\text{CaCO}_3 T = 14.2 mL × 10 = 142.0 mg/L as CaCO 3 ​

Step 2: Compare P P P and T T T : 1 2 T = 142.0 2 = 71.0 mg/L \frac{1}{2}T = \frac{142.0}{2} = 71.0\text{ mg/L} 2 1 ​ T = 2 142.0 ​ = 71.0 mg/L

Since P ( 36.0 ) < 1 2 T ( 71.0 ) P (36.0) < \frac{1}{2}T (71.0) P ( 36.0 ) < 2 1 ​ T ( 71.0 ) , this falls into Case 2 ( P < 1 2 T P < \frac{1}{2}T P < 2 1 ​ T ).

Step 3: Calculate species distribution:

Hydroxide Alkalinity = 0 mg/L as CaCO 3 0\text{ mg/L as }\text{CaCO}_3 0 mg/L as CaCO 3 ​

Carbonate Alkalinity = 2 P = 2 × 36.0 = 72.0 mg/L as CaCO 3 2P = 2 \times 36.0 = 72.0\text{ mg/L as }\text{CaCO}_3 2 P = 2 × 36.0 = 72.0 mg/L as CaCO 3 ​

Bicarbonate Alkalinity = T − 2 P = 142.0 − 72.0 = 70.0 mg/L as CaCO 3 T - 2P = 142.0 - 72.0 = 70.0\text{ mg/L as }\text{CaCO}_3 T − 2 P = 142.0 − 72.0 = 70.0 mg/L as CaCO 3 ​

4. Total and Calcium Hardness via EDTA Titrimetry

Water Hardness is defined as the total concentration of polyvalent metallic cations in solution, predominantly Calcium ( Ca 2 + \text{Ca}^{2+} Ca 2 + ) and Magnesium ( Mg 2 + \text{Mg}^{2+} Mg 2 + ). Hardness causes scale formation in boilers and distribution piping while consuming soap.

Hardness Classifications (USGS / EPA)

Soft: 0 – 60 mg/L as CaCO 3 0\text{--}60\text{ mg/L as }\text{CaCO}_3 0 – 60 mg/L as CaCO 3 ​ ( 0 – 3.5 gpg 0\text{--}3.5\text{ gpg} 0 – 3.5 gpg )

Moderately Hard: 61 – 120 mg/L as CaCO 3 61\text{--}120\text{ mg/L as }\text{CaCO}_3 61 – 120 mg/L as CaCO 3 ​ ( 3.5 – 7.0 gpg 3.5\text{--}7.0\text{ gpg} 3.5 – 7.0 gpg )

Hard: 121 – 180 mg/L as CaCO 3 121\text{--}180\text{ mg/L as }\text{CaCO}_3 121 – 180 mg/L as CaCO 3 ​ ( 7.0 – 10.5 gpg 7.0\text{--}10.5\text{ gpg} 7.0 – 10.5 gpg )

Very Hard: > 180 mg/L as CaCO 3 > 180\text{ mg/L as }\text{CaCO}_3 > 180 mg/L as CaCO 3 ​ ( > 10.5 gpg > 10.5\text{ gpg} > 10.5 gpg )

Conversion Factor: 1 grain per gallon (gpg) = 17.12 mg/L as CaCO 3 1\text{ grain per gallon (gpg)} = 17.12\text{ mg/L as }\text{CaCO}_3 1 grain per gallon (gpg) = 17.12 mg/L as CaCO 3 ​

The EDTA Complexometric Titration Method

Ethylenediaminetetraacetic acid ( EDTA ) chelates divalent metal ions to form stable, soluble chelate complexes:

Total Hardness: Buffer a 50.0 mL 50.0\text{ mL} 50.0 mL sample to pH 10.0 ± 0.1 \pm 0.1 ± 0.1 using an ammonium chloride/ammonium hydroxide buffer ( NH 4 Cl / NH 4 OH \text{NH}_4\text{Cl}/\text{NH}_4\text{OH} NH 4 ​ Cl / NH 4 ​ OH ). Add Eriochrome Black T (EBT) or Calmagite indicator. In the presence of Ca 2 + \text{Ca}^{2+} Ca 2 + and Mg 2 + \text{Mg}^{2+} Mg 2 + , the indicator forms a wine-red complex. Titrate with standard 0.0100 M ( 0.0200 N ) 0.0100\text{ M } (0.0200\text{ N}) 0.0100 M ( 0.0200 N ) EDTA . The EDTA displaces the indicator, turning the solution to a sharp, distinct sky blue at the endpoint.

Calcium Hardness: Buffer a separate sample to pH 12.0 to 13.0 by adding 1.0 N NaOH 1.0\text{ N } \text{NaOH} 1.0 N NaOH . At this elevated pH, magnesium precipitates quantitatively as solid magnesium hydroxide ( Mg(OH) 2 ↓ \text{Mg(OH)}_2\downarrow Mg(OH) 2 ​ ↓ ). Add Hydroxy Naphthol Blue (or Murexide) indicator and titrate with EDTA from pink/red to pure royal blue .

Magnesium Hardness Determination: Calculated mathematically by subtraction: Magnesium Hardness (mg/L as CaCO 3 ) = Total Hardness − Calcium Hardness \text{Magnesium Hardness (mg/L as }\text{CaCO}_3) = \text{Total Hardness} - \text{Calcium Hardness} Magnesium Hardness (mg/L as CaCO 3 ​ ) = Total Hardness − Calcium Hardness Total Hardness (mg/L as CaCO 3 ) = mL EDTA Titrant × 1,000 Sample Volume (mL) \text{Total Hardness (mg/L as }\text{CaCO}_3) = \frac{\text{mL EDTA Titrant} \times 1{,}000}{\text{Sample Volume (mL)}} Total Hardness (mg/L as CaCO 3 ​ ) = Sample Volume (mL) mL EDTA Titrant × 1 , 000 ​

(When utilizing 0.0100 M 0.0100\text{ M} 0.0100 M EDTA standard titrant where 1.00 mL EDTA = 1.00 mg CaCO 3 1.00\text{ mL EDTA} = 1.00\text{ mg }\text{CaCO}_3 1.00 mL EDTA = 1.00 mg CaCO 3 ​ .)

5. Chlorine Residual Analysis: DPD Colorimetric & FAS Titration

Maintaining a legal disinfectant residual is mandatory across distribution systems ( 0.20 mg/L 0.20\text{ mg/L} 0.20 mg/L free chlorine minimum under Colorado Reg 11).

DPD Colorimetric Method (Standard Methods 4500-Cl G)

Free Available Chlorine (HOCl + OCl⁻): N , N -diethyl- p -phenylenediamine N,N\text{-diethyl-}p\text{-phenylenediamine} N , N -diethyl- p -phenylenediamine ( DPD ) reagent buffered to pH 6.2–6.5 is added to the sample. Free chlorine instantly oxidizes DPD to form a magenta/pink Würster dye. Photometric absorbance is read at 515 nm 515\text{ nm} 515 nm within 1 minute .

Combined Chlorine (Chloramines): Excess potassium iodide ( KI \text{KI} KI ) crystals or DPD-3 reagent is introduced. Chloramines oxidize iodide ( I − \text{I}^- I − ) to iodine ( I 2 \text{I}_2 I 2 ​ ), which in turn oxidizes additional DPD to produce total magenta color representing Total Chlorine Residual .

Combined Residual Calculation: Combined Chlorine = Total Chlorine − Free Chlorine \text{Combined Chlorine} = \text{Total Chlorine} - \text{Free Chlorine} Combined Chlorine = Total Chlorine − Free Chlorine .

Analytical Interferences & Matrix Precautions

Oxidized Manganese ( MnO 2 \text{MnO}_2 MnO 2 ​ ): Produces a false-positive free chlorine reading. Corrected by running a sodium arsenite ( NaAsO 2 \text{NaAsO}_2 NaAsO 2 ​ ) or thioacetamide blank.

High Chlorine Bleaching: High chlorine concentrations ( > 4.0 – 5.0 mg/L > 4.0\text{--}5.0\text{ mg/L} > 4.0 – 5.0 mg/L ) oxidize DPD past the colored dye to a colorless imine compound, yielding a false-low reading. If a transient flash of pink vanishes instantly, dilute the sample with chlorine-demand-free water and re-test.

DPD FAS Titrimetric Method (SM 4500-Cl F): Standard Ferrous Ammonium Sulfate ( 0.00282 N FAS 0.00282\text{ N } \text{FAS} 0.00282 N FAS ) titrates the pink DPD dye to a colorless endpoint ( 1.00 mL FAS = 0.100 mg Cl 2 1.00\text{ mL FAS} = 0.100\text{ mg Cl}_2 1.00 mL FAS = 0.100 mg Cl 2 ​ ), eliminating turbidity and colorimeter optical bias.