Study guide

10.2 Core Workflows and Decision Points

COA Study Guide Exam

Track your progress as you read

Finished sections get a check mark.

1 Chapter 1: COA Orientation and Exam Strategy

1.1 Current COA Exam Facts 1.2 Eligibility, Application, and Scheduling 1.3 Blueprint Domains and Weighting 1.4 Question Style and Score Report Thinking 1.5 Study Calendar and Practice Plan

2 Chapter 2: History and Documentation

2.1 History and Documentation Overview 2.2 Core Workflows and Decision Points 2.3 Scenario Practice for History and Documentation 2.4 Common Traps in History and Documentation 2.5 Practice Drills and Readiness Markers

3 Chapter 3: Visual Assessment

3.1 Visual Assessment Overview 3.2 Core Workflows and Decision Points 3.3 Scenario Practice for Visual Assessment 3.4 Common Traps in Visual Assessment 3.5 Practice Drills and Readiness Markers

4 Chapter 4: Visual Field Testing

4.1 Visual Field Testing Overview 4.2 Core Workflows and Decision Points 4.3 Scenario Practice for Visual Field Testing 4.4 Common Traps in Visual Field Testing 4.5 Practice Drills and Readiness Markers

5 Chapter 5: Pupil Assessment

5.1 Pupil Assessment Overview 5.2 Core Workflows and Decision Points 5.3 Scenario Practice for Pupil Assessment 5.4 Common Traps in Pupil Assessment 5.5 Practice Drills and Readiness Markers

6 Chapter 6: Tonometry

6.1 Tonometry Overview 6.2 Core Workflows and Decision Points 6.3 Scenario Practice for Tonometry 6.4 Common Traps in Tonometry 6.5 Practice Drills and Readiness Markers

7 Chapter 7: Keratometry

7.1 Keratometry Overview 7.2 Core Workflows and Decision Points 7.3 Scenario Practice for Keratometry 7.4 Common Traps in Keratometry 7.5 Practice Drills and Readiness Markers

8 Chapter 8: Lensometry

8.1 Lensometry Overview 8.2 Core Workflows and Decision Points 8.3 Scenario Practice for Lensometry 8.4 Common Traps in Lensometry 8.5 Practice Drills and Readiness Markers

9 Chapter 9: Biometry

9.1 Biometry Overview 9.2 Core Workflows and Decision Points 9.3 Scenario Practice for Biometry 9.4 Common Traps in Biometry 9.5 Practice Drills and Readiness Markers

10 Chapter 10: Diagnostic Ultrasound

10.1 Diagnostic Ultrasound Overview 10.2 Core Workflows and Decision Points 10.3 Scenario Practice for Diagnostic Ultrasound 10.4 Common Traps in Diagnostic Ultrasound 10.5 Practice Drills and Readiness Markers

11 Chapter 11: Supplemental Testing

11.1 Supplemental Testing Overview 11.2 Core Workflows and Decision Points 11.3 Scenario Practice for Supplemental Testing 11.4 Common Traps in Supplemental Testing 11.5 Practice Drills and Readiness Markers

12 Chapter 12: Microbiology

12.1 Microbiology Overview 12.2 Core Workflows and Decision Points 12.3 Scenario Practice for Microbiology 12.4 Common Traps in Microbiology 12.5 Practice Drills and Readiness Markers

13 Chapter 13: Pharmacology

13.1 Pharmacology Overview 13.2 Mydriatics, Cycloplegics, and Anesthetics 13.3 Glaucoma, Anti-Infective, and Anti-Inflammatory Agents 13.4 Stains, Instillation Technique, and Safety Traps 13.5 Practice Drills and Readiness Markers

14 Chapter 14: Surgical Assisting

14.1 Surgical Assisting Overview 14.2 Core Workflows and Decision Points 14.3 Scenario Practice for Surgical Assisting 14.4 Common Traps in Surgical Assisting 14.5 Practice Drills and Readiness Markers

15 Chapter 15: Ophthalmic Patient Services and Education

15.1 Ophthalmic Patient Services and Education Overview 15.2 Core Workflows and Decision Points 15.3 Scenario Practice for Ophthalmic Patient Services and Education 15.4 Common Traps in Ophthalmic Patient Services and Education 15.5 Practice Drills and Readiness Markers

16 Chapter 16: Optics and Spectacles

16.1 Optics and Spectacles Overview 16.2 Core Workflows and Decision Points 16.3 Scenario Practice for Optics and Spectacles 16.4 Common Traps in Optics and Spectacles 16.5 Practice Drills and Readiness Markers

17 Chapter 17: Contact Lenses

17.1 Contact Lens Types, Parameters, and Oxygen 17.2 Fitting Workflows: K Readings, Tear Lens, and the SAM/FAP Rule 17.3 Insertion, Removal, Care Systems, and Wear Schedules 17.4 Complications: Hypoxia, Infiltrates, and Corneal Ulcers 17.5 Specialty Lenses, Multifocals, and Readiness Drills

18 Chapter 18: Equipment Maintenance and Repair

18.1 Equipment Maintenance and Repair Overview 18.2 Core Workflows and Decision Points 18.3 Scenario Practice for Equipment Maintenance and Repair 18.4 Common Traps in Equipment Maintenance and Repair 18.5 Practice Drills and Readiness Markers

19 Chapter 19: Medical Ethics, Legal, and Regulatory Issues

19.1 Medical Ethics, Legal, and Regulatory Issues Overview 19.2 Core Workflows and Decision Points 19.3 Scenario Practice for Medical Ethics, Legal, and Regulatory Issues 19.4 Common Traps in Medical Ethics, Legal, and Regulatory Issues 19.5 Practice Drills and Readiness Markers

20 Chapter 20: General Medical Knowledge

20.1 General Medical Knowledge Overview 20.2 Medical Terminology, Anatomy, and Physiology 20.3 Systemic Disease with Ocular Manifestations 20.4 Pharmacology and Medical Emergencies 20.5 Microbiology, Infection Control, and Readiness Drills

21 Chapter 21: Refraction

21.1 Refraction Overview 21.2 Core Workflows and Decision Points 21.3 Scenario Practice for Refraction 21.4 Common Traps in Refraction 21.5 Practice Drills and Readiness Markers

22 Chapter 22: Ophthalmic Imaging

22.1 Ophthalmic Imaging Overview 22.2 Core Workflows and Decision Points 22.3 Scenario Practice for Ophthalmic Imaging 22.4 Common Traps in Ophthalmic Imaging 22.5 Practice Drills and Readiness Markers

23 Chapter 23: Ocular Motility Testing

23.1 Ocular Motility Testing Overview 23.2 Core Workflows and Decision Points 23.3 Scenario Practice for Ocular Motility Testing 23.4 Common Traps in Ocular Motility Testing 23.5 Practice Drills and Readiness Markers

24 Chapter 24: Final Review and Test Day

24.1 Timed Practice Strategy 24.2 Last-Week Review Map 24.3 Exam-Day Checklist 24.4 After the Exam and Next Steps

10.2 Core Workflows and Decision Points

Key Takeaways Explain This Topic AI

Contact (applanation) A-scan touches the cornea and risks compression artifact; immersion A-scan uses a saline shell and avoids corneal indentation.

A good A-scan trace shows a tall, steeply rising retinal spike and the probe aligned along the true visual axis (perpendicular to the macula).

The SRK regression formula is P = A − 2.5L − 0.9K, combining the lens A-constant, axial length L, and keratometry K.

B-scan workflow uses topographic, longitudinal, and transverse probe orientations with marked dynamic scanning to localize and characterize lesions.

Last updated: June 2026

10.2 Core Workflows and Decision Points

Good ultrasound data depends on technique, not just the machine. The COA must know the workflow for each mode and the decision points where errors creep in.

A-scan biometry: contact vs immersion

Contact (applanation) technique places the probe tip directly on the anesthetized cornea — the tear film acts as the coupling agent, so no gel is needed. It is fast but the operator can press too hard and compress the cornea , falsely shortening the axial length and over-plus-ing the IOL power. Immersion technique rests a saline-filled scleral shell on the eye so the probe floats above the cornea and never touches it, eliminating compression. Immersion is generally more accurate and is preferred for high-precision cases.

  • Step Contact A-scan Immersion A-scan
  • Anesthesia Topical drop required Topical drop required
  • Coupling Tear film Saline in shell
  • Corneal touch Yes — compression risk No — no compression
  • Speed Faster Slower, more setup
  • Accuracy Operator-dependent Generally higher
  • Recognizing a valid trace

A reliable A-scan shows distinct, properly spaced spikes: an initial corneal spike, anterior and posterior lens spikes, then a tall, steeply rising retinal spike . The probe must align along the true visual axis, perpendicular to the macula; an off-axis reading hits the optic-nerve region and falsely lengthens the measurement. The retinal spike should rise steeply from baseline because the beam is striking the retina at 90 degrees; a short, sloped, or rounded retinal spike means the probe is misaligned and the reading should be rejected.

The operator takes several readings and accepts only those that are consistent with one another and with the fellow eye, and that show the high, sharp retinal echo.

Two error directions to internalize

The two technique errors push the axial length in opposite directions, and confusing them is a classic mistake:

  • Error Effect on axial length Effect on IOL power
  • Corneal compression (too much pressure) Falsely short IOL power calculated too high
  • Off-axis aim (toward optic nerve) Falsely long IOL power calculated too low

Immersion removes the compression direction entirely; careful fixation removes the off-axis direction.

Feeding the IOL formula

Axial length feeds an IOL-power formula. The classic regression formula is the SRK formula :

P = A − 2.5L − 0.9K

where P is the implant power, A is the manufacturer's lens A-constant , L is axial length in millimeters, and K is the average keratometry reading in diopters. Each lens model has its own A-constant. Modern theoretical formulas (SRK/T, Holladay, Hoffer Q, Barrett) refine this, and SRK/T performs best for eyes roughly 24–28 mm long, but the COA should recognize the SRK variables because axial length and keratometry are the data the assistant supplies.

A worked IOL example

Suppose a lens has an A-constant of 118.0, the axial length L is 23.50 mm, and the average keratometry K is 44.00 D. Plugging into SRK: P = 118.0 − (2.5 × 23.50) − (0.9 × 44.00) = 118.0 − 58.75 − 39.6 = 19.65 D , rounded to the nearest available implant power. This illustrates why the COA must deliver an accurate L and K: a 0.4 mm error in L shifts the (2.5 × L) term by 1.0 D, and a keratometry error of about 1.1 D shifts the (0.9 × K) term by 1.0 D. Both inputs the assistant captures directly drive the result.

B-scan orientation

B-scan localizes findings using three probe orientations: transverse (probe marker horizontal, shows lateral extent of a lesion), longitudinal (marker aimed at a clock hour, shows radial extent toward the disc), and axial (through the cornea and lens, centering the optic nerve). Dynamic scanning — asking the patient to move the eye while watching for aftermovement — distinguishes a mobile vitreous membrane from a stiff retinal detachment. Coupling gel is applied to the closed lid or directly to the globe, and the marker dot is kept oriented so the examiner can map the echo back to a clock position.

A consistent convention — marker toward the examiner's reference point — keeps the recorded clock hour meaningful when the surgeon reviews the images later.

Reflectivity: reading what bounces back

A-scan also measures internal reflectivity , the height of echoes inside a lesion, which helps characterize tumors. A choroidal melanoma classically shows low-to-medium internal reflectivity with a hollow internal pattern, while a choroidal hemangioma shows high internal reflectivity . The COA does not diagnose, but recognizing that spike height carries diagnostic meaning explains why steady probe alignment and correct gain settings matter; pushing the gain too high washes out these distinctions and a too-low gain hides real echoes.

How this appears on the exam

Expect questions on why immersion beats contact, what a compression error does to the result, which spike is the retina, what the SRK letters stand for, and how probe orientation maps a lesion. Choose answers that protect measurement accuracy and align with the correct probe technique rather than the fastest option.

Keratometry: the other half of the equation

The COA almost never supplies axial length alone. The IOL formula also needs keratometry (K) , the corneal curvature in diopters, measured by a manual or automated keratometer (or by topography). Because the K value is multiplied by 0.9 in the SRK formula, a keratometry mistake is nearly as damaging as an axial-length mistake. Common keratometry pitfalls the assistant must avoid: an irregular or dry tear film distorting the mires, a recently worn contact lens warping the cornea, and failing to instill artificial tears on a dry eye before reading.

The lesson is that biometry is a paired task — clean axial length and clean keratometry — and both must be verified before the surgeon plans the implant.

Documenting and handing off

After acceptable readings, the assistant records the axial length, keratometry, the velocity setting used, the technique (contact or immersion), and which eye, then flags any large inter-eye asymmetry for the surgeon. Good documentation lets the surgeon trust the numbers and lets a later reviewer reproduce the calculation. The workflow is not finished when the spike looks good; it is finished when the data are validated, labeled by eye, and clearly handed off.