10.1 Diagnostic Ultrasound Overview

COA Study Guide Exam

Track your progress as you read

Finished sections get a check mark.

1 Chapter 1: COA Orientation and Exam Strategy

1.1 Current COA Exam Facts 1.2 Eligibility, Application, and Scheduling 1.3 Blueprint Domains and Weighting 1.4 Question Style and Score Report Thinking 1.5 Study Calendar and Practice Plan

2 Chapter 2: History and Documentation

2.1 History and Documentation Overview 2.2 Core Workflows and Decision Points 2.3 Scenario Practice for History and Documentation 2.4 Common Traps in History and Documentation 2.5 Practice Drills and Readiness Markers

3 Chapter 3: Visual Assessment

3.1 Visual Assessment Overview 3.2 Core Workflows and Decision Points 3.3 Scenario Practice for Visual Assessment 3.4 Common Traps in Visual Assessment 3.5 Practice Drills and Readiness Markers

4 Chapter 4: Visual Field Testing

4.1 Visual Field Testing Overview 4.2 Core Workflows and Decision Points 4.3 Scenario Practice for Visual Field Testing 4.4 Common Traps in Visual Field Testing 4.5 Practice Drills and Readiness Markers

5 Chapter 5: Pupil Assessment

5.1 Pupil Assessment Overview 5.2 Core Workflows and Decision Points 5.3 Scenario Practice for Pupil Assessment 5.4 Common Traps in Pupil Assessment 5.5 Practice Drills and Readiness Markers

6 Chapter 6: Tonometry

6.1 Tonometry Overview 6.2 Core Workflows and Decision Points 6.3 Scenario Practice for Tonometry 6.4 Common Traps in Tonometry 6.5 Practice Drills and Readiness Markers

7 Chapter 7: Keratometry

7.1 Keratometry Overview 7.2 Core Workflows and Decision Points 7.3 Scenario Practice for Keratometry 7.4 Common Traps in Keratometry 7.5 Practice Drills and Readiness Markers

8 Chapter 8: Lensometry

8.1 Lensometry Overview 8.2 Core Workflows and Decision Points 8.3 Scenario Practice for Lensometry 8.4 Common Traps in Lensometry 8.5 Practice Drills and Readiness Markers

9 Chapter 9: Biometry

9.1 Biometry Overview 9.2 Core Workflows and Decision Points 9.3 Scenario Practice for Biometry 9.4 Common Traps in Biometry 9.5 Practice Drills and Readiness Markers

10 Chapter 10: Diagnostic Ultrasound

10.1 Diagnostic Ultrasound Overview 10.2 Core Workflows and Decision Points 10.3 Scenario Practice for Diagnostic Ultrasound 10.4 Common Traps in Diagnostic Ultrasound 10.5 Practice Drills and Readiness Markers

11 Chapter 11: Supplemental Testing

11.1 Supplemental Testing Overview 11.2 Core Workflows and Decision Points 11.3 Scenario Practice for Supplemental Testing 11.4 Common Traps in Supplemental Testing 11.5 Practice Drills and Readiness Markers

12 Chapter 12: Microbiology

12.1 Microbiology Overview 12.2 Core Workflows and Decision Points 12.3 Scenario Practice for Microbiology 12.4 Common Traps in Microbiology 12.5 Practice Drills and Readiness Markers

13 Chapter 13: Pharmacology

13.1 Pharmacology Overview 13.2 Mydriatics, Cycloplegics, and Anesthetics 13.3 Glaucoma, Anti-Infective, and Anti-Inflammatory Agents 13.4 Stains, Instillation Technique, and Safety Traps 13.5 Practice Drills and Readiness Markers

14 Chapter 14: Surgical Assisting

14.1 Surgical Assisting Overview 14.2 Core Workflows and Decision Points 14.3 Scenario Practice for Surgical Assisting 14.4 Common Traps in Surgical Assisting 14.5 Practice Drills and Readiness Markers

15 Chapter 15: Ophthalmic Patient Services and Education

15.1 Ophthalmic Patient Services and Education Overview 15.2 Core Workflows and Decision Points 15.3 Scenario Practice for Ophthalmic Patient Services and Education 15.4 Common Traps in Ophthalmic Patient Services and Education 15.5 Practice Drills and Readiness Markers

16 Chapter 16: Optics and Spectacles

16.1 Optics and Spectacles Overview 16.2 Core Workflows and Decision Points 16.3 Scenario Practice for Optics and Spectacles 16.4 Common Traps in Optics and Spectacles 16.5 Practice Drills and Readiness Markers

17 Chapter 17: Contact Lenses

17.1 Contact Lens Types, Parameters, and Oxygen 17.2 Fitting Workflows: K Readings, Tear Lens, and the SAM/FAP Rule 17.3 Insertion, Removal, Care Systems, and Wear Schedules 17.4 Complications: Hypoxia, Infiltrates, and Corneal Ulcers 17.5 Specialty Lenses, Multifocals, and Readiness Drills

18 Chapter 18: Equipment Maintenance and Repair

18.1 Equipment Maintenance and Repair Overview 18.2 Core Workflows and Decision Points 18.3 Scenario Practice for Equipment Maintenance and Repair 18.4 Common Traps in Equipment Maintenance and Repair 18.5 Practice Drills and Readiness Markers

19 Chapter 19: Medical Ethics, Legal, and Regulatory Issues

19.1 Medical Ethics, Legal, and Regulatory Issues Overview 19.2 Core Workflows and Decision Points 19.3 Scenario Practice for Medical Ethics, Legal, and Regulatory Issues 19.4 Common Traps in Medical Ethics, Legal, and Regulatory Issues 19.5 Practice Drills and Readiness Markers

20 Chapter 20: General Medical Knowledge

20.1 General Medical Knowledge Overview 20.2 Medical Terminology, Anatomy, and Physiology 20.3 Systemic Disease with Ocular Manifestations 20.4 Pharmacology and Medical Emergencies 20.5 Microbiology, Infection Control, and Readiness Drills

21 Chapter 21: Refraction

21.1 Refraction Overview 21.2 Core Workflows and Decision Points 21.3 Scenario Practice for Refraction 21.4 Common Traps in Refraction 21.5 Practice Drills and Readiness Markers

22 Chapter 22: Ophthalmic Imaging

22.1 Ophthalmic Imaging Overview 22.2 Core Workflows and Decision Points 22.3 Scenario Practice for Ophthalmic Imaging 22.4 Common Traps in Ophthalmic Imaging 22.5 Practice Drills and Readiness Markers

23 Chapter 23: Ocular Motility Testing

23.1 Ocular Motility Testing Overview 23.2 Core Workflows and Decision Points 23.3 Scenario Practice for Ocular Motility Testing 23.4 Common Traps in Ocular Motility Testing 23.5 Practice Drills and Readiness Markers

24 Chapter 24: Final Review and Test Day

24.1 Timed Practice Strategy 24.2 Last-Week Review Map 24.3 Exam-Day Checklist 24.4 After the Exam and Next Steps

10.1 Diagnostic Ultrasound Overview

Key Takeaways Explain This Topic AI

Ophthalmic ultrasound uses high-frequency sound (A-scan ~8 MHz, B-scan ~10 MHz) to measure and image structures light cannot reach.

A-scan biometry produces axial length for intraocular lens (IOL) power; B-scan produces a two-dimensional cross-section of the globe and orbit.

Sound travels at 1641 m/s through cornea and lens, 1532 m/s through aqueous and vitreous, averaging 1555 m/s in a normal phakic eye.

On the COA exam, Diagnostic Ultrasound is a small slice of a 200-question, 180-minute test, but biometry numbers and probe safety are high-yield.

Last updated: June 2026

10.1 Diagnostic Ultrasound Overview

Diagnostic ultrasound is the use of high-frequency sound waves to measure and image ocular structures. It is indispensable when the view is blocked — a dense cataract, hyphema (blood in the anterior chamber), or vitreous hemorrhage — because sound passes through opaque media that light cannot. The certified ophthalmic assistant (COA) is expected to understand the two main modes, the velocities of sound in the eye, and how patient preparation affects accuracy.

The two core modes

A-scan (amplitude scan) sends a single sound beam down the visual axis and displays returning echoes as vertical spikes on a one-dimensional graph. Each spike marks an acoustic interface (cornea, anterior lens, posterior lens, retina). The distance between the first corneal spike and the retinal spike is the axial length — the single most important measurement for choosing an intraocular lens (IOL) before cataract surgery. A-scan probes operate near 8 MHz .

B-scan (brightness scan) sweeps the beam to build a two-dimensional cross-sectional picture of the globe and orbit, displaying echo strength as dots of varying brightness. B-scan answers "what does it look like" — retinal detachment, tumor, foreign body, optic-disc drusen — while A-scan answers "how long is it." B-scan probes operate near 10 MHz ; higher frequency means better resolution but less penetration.

Speed of sound in ocular media

Ultrasound machines convert echo travel time into distance, so the assumed velocity must be correct. Memorize these values:

  • Tissue Sound velocity
  • Cornea and crystalline lens (solid) 1641 m/s
  • Aqueous and vitreous (fluid) 1532 m/s
  • Average normal phakic eye 1555 m/s
  • Aphakic eye (no lens) 1532 m/s
  • Silicone-oil-filled eye ~1040 m/s (much slower)

If the wrong velocity is set — for example, leaving the phakic setting on a silicone-oil eye — the axial length and IOL power will be badly wrong. The assistant selects the velocity that matches the patient's actual lens status. A dense cataract changes lens density, and pseudophakic eyes contain an implant of known material, so most machines offer dedicated phakic, aphakic, pseudophakic (PMMA, silicone, or acrylic), and dense-cataract settings. Choosing the right mode is part of the assistant's job, not an afterthought.

What each spike represents

Reading the A-scan trace is a core skill. From left to right, the expected echoes are: the corneal spike (the entry surface), the anterior lens spike, the posterior lens spike, and finally the tall retinal spike followed by lower scleral and orbital echoes. The axial length is the distance from the cornea to the retina. A trace missing the lens spikes suggests an aphakic eye; a trace with extra spikes between lens and retina suggests vitreous opacity such as hemorrhage or a detached membrane.

Indications you should recognize

  • Clinical situation Why ultrasound is ordered
  • Cataract surgery planning A-scan axial length feeds the IOL power calculation
  • Dense cataract or vitreous hemorrhage B-scan images the retina the examiner cannot see
  • Suspected retinal detachment B-scan confirms and maps the detachment
  • Suspected intraocular tumor (melanoma) A-scan and B-scan assess internal reflectivity and size
  • Suspected foreign body after trauma B-scan localizes a high-reflectivity object
  • Why precision matters

A worked example shows the stakes: a corneal compression (indentation) error of just 0.4 mm during contact A-scan produces roughly a 1.00 diopter error in the calculated IOL power. Because the eye is only about 23–24 mm long on average, even sub-millimeter errors are clinically large. That is the difference between a patient seeing well and needing glasses or a lens exchange. The COA learns to take readings with the lightest possible corneal touch, or to use the immersion technique that avoids touching the cornea at all, then to repeat the measurement and compare it against the fellow eye before trusting any single number.

Patient preparation and comfort

Good data also depends on the patient. The assistant explains that a numbing drop will be used, that the probe or shell may feel like a light touch, and that the test is painless and brief. For A-scan and B-scan alike, topical anesthetic is instilled; the patient is asked to fixate steadily so the beam stays on the visual axis. Contact lenses are removed first because they alter the corneal surface and coupling. A calm, well-positioned, fixating patient produces far cleaner echoes than a squeezing, anxious one.

How this appears on the exam

COA questions in this area favor concrete facts: which mode gives axial length, what velocity applies to a given eye, what causes a falsely short axial length, what each spike on the trace represents, and how to prepare and reassure the patient. Treat distractors that confuse A-scan and B-scan roles, or that misstate a velocity, as traps. The exam rewards knowing the numbers and the indication, not vague descriptions of what ultrasound is. A final orientation point: ultrasound is one of several biometry pathways.

Optical biometry (partial coherence interferometry, the IOLMaster and similar devices) measures axial length with a laser and is now first-line for most clear-media eyes, but it fails through dense cataracts and media opacities — exactly when ultrasound A-scan remains essential. Knowing when optical biometry cannot be used, and that ultrasound is the fallback, is itself an exam-relevant judgment.